K072420 · Oceanic Medical Products, Inc. · BSZ · Jan 22, 2008 · Anesthesiology
Device Facts
Record ID
K072420
Device Name
MAGELLAN-2200, MODEL-3 ANESTHESIA MACHINE
Applicant
Oceanic Medical Products, Inc.
Product Code
BSZ · Anesthesiology
Decision Date
Jan 22, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5160
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Magellan-2200, Model-3 Anesthesia Machine is intended for spontaneous or manually assisted ventilation of patients during anesthesia. It may be used for the delivery of gases and anesthetic vapor. It is capable of monitor/alarm functions for oxygen concentration, breathing pressure and source oxygen pressure.
Device Story
Magellan-2200, Model-3 is an anesthesia gas machine; modification of Model-1. Device delivers gases and anesthetic vapor for spontaneous or manually assisted patient ventilation. Components include inlet gas manifold (oxygen/air), pressure gauges, low-pressure alarm, oxygen analyzer/monitor, oxygen flow selector (nasal cannula/mask), oxygen flush button, flowmeters, agent-specific vaporizer, common gas outlet, CO2 absorber with one-way valves, and reservoir bag/pressure relief arm. Used in clinical settings by healthcare providers to manage patient ventilation and anesthesia delivery. Output includes gas/vapor delivery and real-time monitoring of oxygen concentration and breathing pressure; alerts provider to pressure/concentration changes to support clinical decision-making and patient safety during anesthesia.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Gas machine for anesthesia/analgesia. Components: inlet gas manifold, pressure gauges, low-pressure alarm, oxygen analyzer/monitor, oxygen flow selector, oxygen flush button, flowmeters, agent-specific vaporizer, common gas outlet, CO2 absorber with one-way valves, reservoir bag, pressure relief/scavenger arm. Mechanical/pneumatic operation.
Indications for Use
Indicated for spontaneous or manually assisted ventilation of patients during anesthesia, including delivery of gases and anesthetic vapor, with monitoring/alarm functions for oxygen concentration, breathing pressure, and source oxygen pressure.
Regulatory Classification
Identification
A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).
Cardinal Medical Specialties OBA-1 Anesthesia Machine (K000859)
Related Devices
K060193 — MAGELLAN-2200 MODEL -2 ANESTHESIA MACHINE, MODEL 2 · Oceanic Medical Products, Inc. · Apr 10, 2006
K965041 — OHMEDA APAC (ADVANCED PORTABLE ANESTHESIA CARE) SYSTEM · Ohmeda Medical · Jun 17, 1997
K151954 — A7 Anesthesia System · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Oct 27, 2015
K031400 — FABIUS TIRO ANESTHESIA SYSTEM · Draeger Medical, Inc. · Jul 15, 2003
K973896 — OHMEDA EXCEL 3000 ANESTHESIA SYSTEM · Ohmeda Medical · Apr 21, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) SUMMARY
# K072420 21 : 1 2
# Oceanic Medical Products, Inc.'s Magellan-2200, Model-3 Anesthesia Machine
#### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
William M. Gates Vice President, Innovation and Development Oceanic Medical Products, Inc. 8005 Shannon Industrial Park Lane Atchison, Kansas 66002 Telephone: (913) 874-2000 Facsimile: (913) 874-2005
JAN 22 2008
# OR
Howard M. Holstein, Esq. Hogan and Hartson, LLP 555 Thirteenth Street, NW Washington, DC, 20004 Telephone: (202) 637-5813 Facsimile: (202) 637-5910
Date Prepared: December 26, 2007
#### Name of Device and Name/Address of Sponsor
Magellan-2200, Model-3 Anesthesia Machine
Oceanic Medical Products, Inc. 8005 Shannon Industrial Park Lane Atchison, Kansas 66002 Telephone: (913) 874-2000 Facsimile: (913) 874-2005
#### Common or Usual Name
Anesthesia Gas Machine
#### Classification Name
Gas Machine, Anesthesia
{1}------------------------------------------------
## Product Code
BSZ
### Regulation Number
21 C.F.R. § 868.5160
### Predicate Devices
Oceanic Medical Products, Inc.'s Magellan-2200, Model-1 Anesthesia Machine (K010613) Cardinal Medical Specialties OBA-1 Anesthesia Machine (K000859)
# Purpose of the Special 510(k) Notice
The Magellan-2200, Model-3 Anesthesia Machine is a modification to the Magellan-2200, Model-1 Anesthesia Machine.
## Intended Use
The Magellan-2200, Model-3 Anesthesia Machine is intended for spontaneous or manually assisted ventilation of patients during anesthesia. It may be used for the delivery of gases and anesthetic vapor. It is capable of monitor/alarm functions for oxygen concentration, breathing pressure and source oxygen pressure.
#### Technological Characteristics
The Model-3 is a gas machine for anesthesia or analgesia. The device consists of the following components: an inlet gas manifold for oxygen and air; pressure gauges for oxygen and air; a low-pressure alarm for inlet oxygen surveillance; an oxygen analyzer and monitor; an oxygen flow selector for nasal cannula or oxygen mask for pre- and post-anesthesia use; an oxygen flush button; oxygen and air flowmeters; an agent-specific vaporizer; a common gas outlet; a carbon dioxide absorber with one way valves; and a reservoir bag/pressure gauge/pressure relief/scavenger outlet arm.
#### Substantial Equivalence
The Model-3 has the same intended use and similar indications, principles of operation, and technological characteristics as the predicate devices. Thus, the Model-3 is substantially equivalent to its predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and head. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 22 2008
Oceanic Medical Products, Incorporated C/O Mr. Howard M. Holstein Regulatory Counsel Hogan & Hartson L.L.P. 555 Thirteenth Street, NW Washington, DC 20004
Re: K072420
Trade/Device Name: Magellan-2200, Model-3 Anesthesia Machine Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: II Product Code: BSZ Dated: December 26, 2007 Received: December 26, 2007
Dear Mr. Holstein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Holstein
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sutte Jr. Michaudavis.
Chip Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Attachment 8
#### Indications for Use Statement
Device Name: Magellan-2200, Model-3 Anesthesia Machine
Indications for Use:
The Magellan-2200, Model-3 Anesthesia Machine is intended/indicated for spontaneous or manually assisted ventilation of patients during anesthesia. It may be used for the delivery of gases and anesthetic vapor. It is capable of monitor/alarm functions for oxygen concentration, breathing pressure and source oxygen pressure.
Prescription Use X (Per 21 C.F.R. 801.109) AND/OR
Over-The-Counter Use (Per 21 C.F.R. 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
inh Nhh
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K072420
\\\DC - 022875/000001 - 2588040 v1
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.