GE DATEX-OHMEDA AVANCE ANESTHESIA SYSTEM

K071142 · Datex-Ohmeda · BSZ · Jun 6, 2007 · Anesthesiology

Device Facts

Record IDK071142
Device NameGE DATEX-OHMEDA AVANCE ANESTHESIA SYSTEM
ApplicantDatex-Ohmeda
Product CodeBSZ · Anesthesiology
Decision DateJun 6, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.5160
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The GE Datex-Ohmeda Avance Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The device is intended for volume or pressure control ventilation. The Avance is not suitable for use in a MRI environment.

Device Story

The GE Datex-Ohmeda Avance is an anesthesia workstation providing inhalation anesthesia and ventilatory support. It utilizes electronic gas mixing, controlled via keypad and rotary controller, with electronic flow meters. The system includes the 7900 Anesthesia Ventilator, a microprocessor-based, pneumatically driven unit that monitors inspired oxygen, airway pressure, and exhaled volume. It compensates for compression losses, fresh gas contribution, and system leaks. The device supports multiple ventilation modes (Volume, Pressure Control, PCV-VG, SIMV, Pressure Support). It accepts Tec 4-7 vaporizers and features integrated respiratory gas monitoring via the M-Gas Module. Used in clinical settings by trained medical professionals, the system displays parameters on a main unit. It includes safety features to prevent hypoxic mixtures and gas supply failures. Output data assists clinicians in managing patient ventilation and anesthesia delivery during surgical procedures.

Clinical Evidence

Bench testing only. The device was validated through rigorous testing to support compliance with recognized voluntary standards, including EN 740, EN 60601-1, and related safety/performance standards for anesthesia workstations and medical electrical equipment.

Technological Characteristics

Anesthesia workstation; electronic gas mixing; microprocessor-controlled pneumatic ventilation. Features Selectatec manifold for vaporizers (Tec 4-7). Connectivity via RS-232 serial port. Complies with EN 740, EN 60601-1, EN 60601-1-2, IEC 60601-1-4, EN 475, EN 850, EN 980, EN 1041, EN 1089-3, ISO 5356-1, and EN 1820. Integrated respiratory gas monitoring via M-Gas Module.

Indications for Use

Indicated for general inhalation anesthesia and ventilatory support for a wide range of patients requiring volume or pressure control ventilation. Contraindicated for use in MRI environments.

Regulatory Classification

Identification

A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K071142 Image /page/0/Picture/1 description: The image shows the General Electric (GE) logo, which consists of the letters 'G' and 'E' intertwined within a circular border. The logo is presented in black and white, with the letters and the inner part of the circle filled in black, while the border and the background are white. The design is stylized and recognizable, representing the brand identity of General Electric. ## JUN -6 2007 PO Box 7550 Madison, WI 53707-7550 USA GE Healthcare T 608 221 1551 F 608 223 2476 May 23, 2007 Date: 510(k) Summary of Safety and Effectiveness Information Subject: for the GE Datex-Ohmeda Avance Proprietary: GE Datex-Ohmeda Avance Gas Machine, Anesthesia Common: Classification: Anesthesiology, 73 BSZ, 21 CFR 868.5160 The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 1992. The GE Datex-Ohmeda Avance is substantially equivalent to the following currently marketed devices: Datex-Ohmeda S/5 Avance - Class II - 21CFR868.5160, which has been the subject of a cleared 510(k) with FDA log number K040743 Datex-Ohmeda Aisys - Class II - 21CFR868.5160, which has been the subject of a cleared 510(k) with FDA log number K061609 The GE Datex-Ohmeda Avance is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. It represents one of the systems in a long line of products based on the Datex-Ohmeda Excel, Aestiva, Aespire, and Aisys Anesthesia Systems. It is to be used only by trained and qualified medical professionals. The Datex-Ohmeda Avance Anesthesia System supplies set flows of medical gases to the breathing system using electronic gas mixing. Gas flows are selected by the user using the keypad and rotary controller on the main display unit and then displayed as electronic flow meters on the system display unit. The Avance is equipped with a pneumatic back-up O2 delivery system and traditional flow tube, as well. A large selection of frames, gases, and vaporizers are available to give the user control of the system configuration. The Avance is also available in wall-mount and pendant models. It is available with two or three gases, up to two vaporizer positions and up to three cylinder connections. All models have 02. The Avance comes with up to two optional gases (air, N2O). Image /page/0/Picture/16 description: The image shows the handwritten text "1 of 3". The number 1 is on the left side of the image. The words "of" are in the middle of the image, and the number 3 is on the right side of the image. General Electric Company Datex-Ohmeda, Inc. {2}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a sequence of handwritten digits and a letter. The sequence starts with the letter 'K', followed by the numbers '0', '7', '1', '1', '4', and '2'. The numbers are written in a simple, slightly slanted style, and the image has a clear, legible quality. The Avance systems accept Tec 4, Tec 5, Tec 6, and Tec 7 vaporizers on a Selectatec manifold. Safety features and devices within the Avance are designed to decrease the risk of hypoxic mixtures, agent mixtures and complete power or sudden gas supply failures. The Avance system is available with optional integrated respiratory gas monitoring. When supplied as an option, the integrated respiratory gas monitoring is provided via the Datex-Ohmeda M-Gas Module (M-CAiO and M-CAiOV software revision 3.2 and above K# 001814) which is physically integrated into the Avance, receives electronic power from the Avance and communicates measured values to the Avance for display on the system display unit. The Datex-Ohmeda 7900 Anesthesia Ventilator is used in the Avance Anesthesia System. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh gas contribution and small leakage in the breathing absorber, bellows and system. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple and secure setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is regulated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices. Ventilator modes for the device include Volume Mode, Pressure Control Mode, Pressure Support with Apnea Backup Mode (Optional), Synchronized Intermittent Mandatory Ventilation (SIMV) Mode (Optional), Pressure Controlled Ventilation with Volume Guarantee (PCV-VG), and Volume Control Ventilation Mode for Cardiac Bypass Mode, Ventilator parameters and measurements are displayed on the system display unit. Several options enable the mounting of the Datex-Ohmeda S/5 Anesthesia Monitor (most recently cleared via K030812). An additional option allows the S/5 AM to be linked to the power supply of the Avance such that when the Avance is turned on, the S/5 AM is also turned on. Additional configurations allow for the mounting of various patient monitors on the top shelf of the Avance. 2 of 3 {3}------------------------------------------------ K071142 The GE Datex-Ohmeda Avance was designed to comply with the applicable portions of the following voluntary standards: | Standard | Title | |------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | EN 740 : 1998 | Anesthesia Workstations and their components | | EN 60601-1:1990 | Medical Electrical Equipment Part 1: General<br>Requirements for Safety Incorporates Corrigendum<br>July 1994; Includes Amendments A1:1993, A11:1993,<br>A12:1993, A2:1995, A13:1996, IEC 601-1:1998 +<br>A2:1995 + Corrigendum 1995, Modified | | EN 60601-1-1:2000 | Medical Electrical Equipment - Medical Electrical<br>Systems | | EN 60601-1-2:2001 | Medical Electrical Equipment - Electromagnetic<br>Compatibility | | IEC 60601-1:<br>4:2000 | Safety of Programmable Electronic Medical Systems | | EN 475:1995 | Electrically Generated Alarm Signals | | EN 850:1997 | Small Medical Gas Cylinders - Pin Indexed | | EN 980:1997 | Graphical Symbols | | EN 1041:1998 | Information to be supplied with medical devices | | EN 1089-3:1997 | Color coding for medical gases | | ISO5356-1:1996 | Conical Connectors | | EN 1820:1997 | Reservoir Bags | The GE Datex-Ohmeda Avance and the currently marketed device are substantially equivalent in design concepts, technologies and materials. The GE Datex-Ohmeda Avance has been validated through rigorous testing that, in part, supports the compliance of GE Datex-Ohmeda Avance to the standards listed above. - Contact: Dan Kosednar Manager, Regulatory Affairs Phone: 608-221-1551 x 3581 Fax: 608-223-2496 {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN -6 2007 Mr. Dan Kosednar Manager, Regulatory Affairs Datex-Ohmeda, Incorporated Life Support Solutions PO Box 7550 Madison, Wisconsin 53707 Re: K071142 Trade/Device Name: GE Datex-Ohmeda Avance Anesthesia System Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: II Product Code: BSZ Dated: May 23, 2007 Received: May 24, 2007 Dear Mr. Kosednar: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ ## Page 2 - Mr. Kosednar Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Sylitta Y. Micheltons. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health ## Enclosure {6}------------------------------------------------ ## Indications for Use 510(k) Number (if known): KO7 | VY A Device Name: GE Datex-Ohmeda Avance Anesthesia System Indications For Use: The GE Datex-Ohmeda Avance Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The device is intended for volume or pressure control ventilation. The Avance is not suitable for use in a MRI environment. Prescription Use _XXX_ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) O. R. L. ിvision Sign-Off) ിഴിടിഠന വിട്ടീസ് Division of Anesthesiology, General Hospital, Infection Control, Dental Devices 510(k) Number:_ Page 1 of __ 1_
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...