GE DATEX-OHMEDA AESTIVA/5 MRI ANESTHESIA SYSTEM

K050055 · Datex-Ohmeda, Inc. · BSZ · Feb 4, 2005 · Anesthesiology

Device Facts

Record IDK050055
Device NameGE DATEX-OHMEDA AESTIVA/5 MRI ANESTHESIA SYSTEM
ApplicantDatex-Ohmeda, Inc.
Product CodeBSZ · Anesthesiology
Decision DateFeb 4, 2005
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.5160
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Aestiva/5 MRI Anesthesia System provides the functional feature set offered by the conventional Aestiva/5 to the clinician with the added ability to be used in the MR environment. Among those standard Aestiva/5 features is the Datex-Ohmeda user interface, all the ventilation parameters of the SmartVent along with the Aestiva breathing circuit. The Aestiva/5 MRI Anesthesia System performed to specifications when tested directly next to 1.5 and 3.0 Tesla shielded MRI devices in a field strength that did not exceed 300 gauss.

Device Story

Anesthesia delivery system designed for use in magnetic resonance (MR) environments. Device incorporates non-ferrous materials to prevent attraction to cryogenic magnets. Provides standard anesthesia ventilation parameters and user interface features. Operated by clinicians in clinical settings near MRI scanners. Safety features mitigate risks of hypoxic mixtures, agent mixtures, and power or gas supply failures. Benefits include enabling anesthesia delivery during MRI procedures without compromising device function or patient safety in high-field magnetic environments.

Clinical Evidence

Bench testing only. Device performance validated against voluntary standards including EN 740, IEC 60601-1, IEC 60601-1-2, ISO 5358, and ASTM F1208-94. Testing confirmed operational specifications when placed directly next to 1.5 and 3.0 Tesla shielded MRI devices in field strengths up to 300 gauss.

Technological Characteristics

Constructed primarily of non-ferrous materials. Includes Datex-Ohmeda user interface, SmartVent ventilation system, and Aestiva breathing circuit. Complies with EN 740, IEC 60601-1, IEC 60601-1-2, ISO 5358, and ASTM F1208-94. Designed for operation in 1.5T and 3.0T MRI environments.

Indications for Use

Indicated for use by clinicians to provide anesthesia in the MR environment (1.5T and 3.0T shielded MRI systems).

Regulatory Classification

Identification

A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K050055 December 28th, 2004 Subject: 510(k) Summary of Safety and Effectiveness Information for the GE Datex-Ohmeda Aestiva/5 MRI Anesthesia System Datex-Ohmeda Aestiva/5 MRI Anesthesia System Proprietary: Common: Gas Machine, Anesthesia Anesthesiology, 73CBK, 21CFR868.5160 Classification: The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 1992. The Aestiva SmartVent MRI Anesthesia System is substantially equivalent to the following currently marketed device: 1. Datex-Ohmeda Aestiva SmartVent MRI Anesthesia System - Class II -21CFR868.5160 The Aestiva/5 MRI provides the functional feature set offered by the conventional Aestiva 3000 (K973896) to the clinician with the added ability to be used in the MR environment (as cleared in K993410). Among those standard Aestiva 3000 features is the Datex-Ohmeda user interface, all the ventilation parameters of the SmartVent (Including those cleared in K023366) along with the Aestiva breathing circuit. The Aestiva/5 MRI is constructed of primarily non-ferrous materials to help prevent attraction to the cryogenic magnets in the MRI systems. The Aestiva/5 MRI performed to specifications when tested directly next to an MRI device of the field strength listed in the product labeling. Safety features and devices within the Aestiva/5 MRI decrease the risk of hypoxic mixtures, agent mixtures and complete power or sudden gas supply failures. The Aestiva SmartVent MRI was designed to comply with the applicable portions of the following voluntary standards: - 1. EN 740 Anesthetic Work Stations - 2. EN 60601-1, IEC 601-1: 1988 Medical Electrical Equipment - 3. EN 60601-1-2, IEC 601-1-2: 1998 Medical Electrical Equipment -Electromagnetic Compatibility - 4. ISO 5358 Anesthetic Gas Machines - 5. ASTM F1208-94 Anesthesia Breathing Circuit Standard In addition, the FDA Document, A Primer on Medical Device Interactions wi part of the Resonance Imaging Systems, was used to help determine testing and labeling requirements. The Datex-Ohmeda Aestiva/5 MRI and the currently marketed devices are subst equivalent in design concepts, technologies and materials. The Datex-Ohmeda MRI has been walidated through rigorous testing that, in part, support the comp Aestivation 12 the above mentioned standards. **Datex-Ohmeda, Inc.** P.O. Box 7550 Madison, WI 53707-7550 www.datex-ohmeda.com Now GE family {2}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread, symbolizing protection and service. The image is in black and white. Public Health Service FEB - 4 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Dan Kosednar Manager, Regulatory Planning and Submissions, CARE Datex-Ohmeda, Incorporated P.O Box 7550 Madison, Wisconsin 53707-7550 Re: K050055 Trade/Device Name: GE Datex-Ohmeda Aestiva/5 MRI Anesthesia System Regulation Number: 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: II Product Code: BSZ Dated: January 10, 2005 Received: January 11, 2005 Dear Mr. Kosednar: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Kosednar Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. CluLs Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use KOSOOSS 510(k) Number (if known): Device Name: GE Datex-Ohmeda Aestiva/5 MRI Anesthesia System Indications For Use: The Aestiva/5 MRI Anesthesia System provides the functional feature set offered by the conventional Aestiva/5 to the clinician with the added ability to be used in the MR environment. Among those standard Aestiva/5 features is the Datex-Ohmeda user interface, all the ventilation parameters of the SmartVent along with the Aestiva breathing circuit. The Aestiva/5 MRI Anesthesia System performed to specifications when tested directly next to 1.5 and 3.0 Tesla shielded MRI devices in a field strength that did not exceed 300 gauss. Prescription Use XXX (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of ____________________________________________________________________________________________________________________________________________________________________ Division Sign-Off Division of Anesthaslology. General F Intection Control, Dental Devi 510(k) Number:
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...