PORTEX® products and components with the NRFit™ Connector are intended for the injection or regional anesthetics or narcotics. The PORTEX® LOR Syringe with NRFit™ Connector is intended for use with the NRFit™ compatible components to verify needle tip placement in the epidural space by the Loss of Resistance technique using air or saline. The NRFit™ Connector is designed to be used with regional anesthesia systems only. The PORTEX® EpiFuse Catheter with NRFit™ Connector is intended for use with the catheter and compatible components for the injection or infusion of regional anesthetics or narcotics. The key is a tool for re-opening the catheter connector. The PORTEX® Epidural Flat Filter with NRFit™ Connector is designed for use when administering injections and/or infusions of regional anesthetics or narcotics to a patient via compatible system components. The NRFit™ Connector is designed to be used with regional anesthesia systems only. The PORTEX® Filter Needle with NRFit™ Connector and PORTEX® Filter Straw with NRFit™ Connector are intended to draw up medication when using the PORTEX® LOR Syringe.
Device Story
Portfolio of regional anesthesia components including LOR syringes, catheters, flat filters, filter needles, and filter straws; all featuring ISO 80369-6 compliant NRFit™ connectors. Used in ICU/OR settings by clinicians. LOR syringes (glass/plastic) assist in locating epidural space via air/saline resistance. Catheters facilitate anesthetic/narcotic infusion; include key tool for connector access. Filters minimize particulate contamination. Filter needles/straws assist in medication draw-up. NRFit™ connectors are color-coded yellow and non-compatible with standard luer connectors to prevent misconnections and accidental infusion of non-neuraxial medications. Devices are single-use, sterile, and disposable. Output is the delivery or preparation of regional anesthetic agents; clinical decision-making relies on clinician verification of needle placement and appropriate drug administration. Benefits include reduced risk of medication misconnection errors.
Clinical Evidence
Bench testing only. Performance evaluated against ISO 80369-6 for connector integrity (leakage, axial load, unscrewing, overriding, stress cracking) and non-interconnectability. Biocompatibility testing performed per ISO 10993 series (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, chemical characterization). Packaging and sterilization validated per ISO 11607 and ISO 11135.
Indicated for patients requiring injection or infusion of regional anesthetics or narcotics via epidural or regional anesthesia systems. Includes use for verifying epidural needle tip placement via Loss of Resistance technique.
Regulatory Classification
Identification
An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.
{0}------------------------------------------------
June 1, 2018
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Smiths Medical ASD, Inc. Sunita Teekasingh Senior Principal Regulatory Affairs Specialist 6000 Nathan Lane North Minneapolis, Minnesota 55442
### Re: K172410
Trade/Device Name: PORTEX® LOR Syringe with NRFit™ Connector PORTEX® EpiFuse Catheter with NRFit™ Connector PORTEX® Epidural Flat Filter with NRFit™ Connector PORTEX® Filter Needle with NRFit™ Connector PORTEX® Filter Straw with NRFit™ Connector Regulation Number: 21 CFR 868.5140 Regulation Name: Anesthesia Conduction Kit Regulatory Class: Class II Product Code: CAZ Dated: Mav 3. 2018 Received: May 4, 2018
Dear Sunita Teekasingh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Food and Drug Administration
## Indications for Use
Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K172410
Device Name
PORTEX® LOR Syringe with NRFit™ Connector PORTEX® EpiFuse Catheter with NRFit™ Connector PORTEX® Epidural Flat Filter with NRFit™ Connector PORTEX® Filter Needle with NRFit™ Connector and PORTEX® Filter Straw with NRFit™ Connector Indications for Use (Describe)
PORTEX® products and components with the NRFit™ Connector are intended for the injection or regional anesthetics or narcotics.
The PORTEX® LOR Syringe with NRFit™ Connector is intended for use with the NRFit™ compatible components to verify needle tip placement in the epidural space by the Loss of Resistance technique using air or saline. The NRFit™ Connector is designed to be used with regional anesthesia systems only.
The PORTEX® EpiFuse Catheter with NRFit™ Connector is intended for use with the catheter and compatible components for the injection or infusion of regional anesthetics or narcotics. The key is a tool for re-opening the catheter connector.
The PORTEX® Epidural Flat Filter with NRFit™ Connector is designed for use when administering injections and/or infusions of regional anesthetics or narcotics to a patient via compatible system components. The NRFit™ Connector is designed to be used with regional anesthesia systems only.
The PORTEX® Filter Needle with NRFit™ Connector and PORTEX® Filter Straw with NRFit™ Connector are intended to draw up medication when using the PORTEX® LOR Syringe.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) K172410 Applicant's Name Smiths Medical ASD, Inc. and Address 6000 Nathan Lane North Minneapolis, MN 55442 USA Contact Person Sunita Teekasingh RN, BSN, CCRN Senior Principal Regulatory Affairs Specialist Vascular Access and Infusion Regulatory Affairs Interim Manger Smiths Medical ASD. Inc. 6000 Nathan Lane North Minneapolis, MN 55442 sunny.teekasingh@smiths-medical.com Phone: (763) 383-3336 May 31, 2018 Date Regulation No. 21 CFR 880.5140 Regulation Name Anesthesia conduction kit, and Anesthesia conduction needle Primary Product CAZ Codes PORTEX® LOR Syringe with NRFit™ Connector Trade Name PORTEX® EpiFuse Catheter with NRFit™ Connector PORTEX® Epidural Flat Filter with NRFit™ Connector PORTEX® Filter Needle with NRFit™ Connector PORTEX® Filter Straw with NRFit™ Connector
### 1. ADMINISTRATIVE INFORMATION
### 2. REASON FOR SUBMISSION
The purpose of this submission is to make a modification to the currently marketed Smiths Medical PORTEX® LOR Syringe with NRFit™ Connector, PORTEX® EpiFuse Catheter with NRFit™ Connector, PORTEX® Epidural Flat Filter with NRFit™ Connector, PORTEX® Filter Needle with NRFit™ Connector, PORTEX® Filter Straw with NRFit™ Connector are being updated to include an ISO 80369-6 compliant connector for neuraxial applications.
## 3. DEVICE INFORMATION
| | Predicate Device | Subject Device |
|------------------|----------------------------------|---------------------------------------------------|
| Trade Name | CorrectInject Syringe | PORTEX® LOR Syringe with<br>NRFit™ connector |
| Regulation No. | 21CFR868.5140 | 21CFR868.5140 |
| Regulation Name | Anesthesia conduction kit | Anesthesia conduction kit |
| Regulatory Class | II | II |
| Product Code | CAZ | CAZ |
| 510(k) | K110053 | K172410 |
| | Predicate Device | Subject Device |
| Trade Name | CorrectInject Catheter Connector | PORTEX® EpiFuse Catheter with<br>NRFit™ Connector |
| Regulation No. | 21CFR868.5140 | 21CFR868.5140 |
| Regulation Name | Anesthesia conduction kit | Anesthesia conduction kit |
{4}------------------------------------------------
bringing technology to life
| Regulatory Class | II | II |
|------------------|------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Product Code | CAZ | CAZ |
| 510(k) | K110053 | K172410 |
| | Predicate Device | Subject Device |
| Trade Name | CorrectInject Filter | PORTEX® Filter Needle with NRFit Connector and PORTEX® Filter Straw with NRFit™ Connector |
| Regulation No. | 21CFR868.5140 | 21CFR868.5140 |
| Regulation Name | Anesthesia conduction kit | Anesthesia conduction kit |
| Regulatory Class | II | II |
| Product Code | CAZ | CAZ |
| 510(k) | K110053 | K172410 |
| | Predicate Device | Subject Device |
| Trade Name | CorrectInject Filter Needle,<br>CorrectInject Filter Straw | PORTEX® NRFit™ Filter Needle, |
| Regulation No. | 21CFR868.5140 | 21CFR868.5140 |
| Regulation Name | Anesthesia conduction kit | Anesthesia conduction kit |
| Regulatory Class | II | II |
| Product Code | CAZ | CAZ |
| 510(k) | K110053 | K172410 |
## 4. DEVICE DESCRIPTION
The PORTEX® Regional Anesthesia Portfolio with NRFit™, consisting of five (5) categories of product including PORTEX® LOR Syringe with NRFit™ Connector, PORTEX® EpiFuse Catheter with NRFit™ Connector, PORTEX® Epidural Flat Filter with NRFit™ Connector and the PORTEX® Filter Straw with NRFit™ Connector.
PORTEX® products and components with NRFit™ connector are intended for the injection or infusion of regional anesthetics or narcotics.
The NRFit ™ connector conform to ISO 80369-6. Small bore connector for liquids and gases in healthcare applications -- Part 6: Connectors for neuraxial applications. The connectors are not compatible with standard luer connectors which are intended to reduce the risk of misconnection that may result in the infusion of medications not intended for neuraxial or regional anesthetic use.
The PORTEX® Regional Anesthesia Portfolio with NRFit™ connector are color-coded yellow to indicate medication intended for neuraxial or regional anesthetic delivery.
This Premarket Notification Traditional 510(k) includes various configurations of PORTEX® Regional Anesthesia Portfolio devices with NRFit™ Connector. A description of each configuration is provided in the table below.
{5}------------------------------------------------
| Device Type | Description |
|-----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PORTEX® LOR<br>Syringe with<br>NRFit™ Connector | The Loss of Resistance (LOR) syringes are a range of sterile, single use syringes<br>used as an aid in locating the epidural space using the Loss of Resistance technique.<br>PORTEX® LOR syringes are available in glass or plastic, and 5, 7, and 10 mL sizes. |
| PORTEX® EpiFuse<br>Catheter with<br>NRFit™<br>Connector: | The Catheter Connector is designed to provide a secure interface between the<br>catheter and the delivery system. The key is a tool for re-opening the catheter<br>connector. The Catheter Connector includes a Thread Assist Guide which provides<br>support for the catheter threading. |
| PORTEX® Epidural<br>Flat Filter with<br>NRFit™ Connector: | The Epidural Flat Filter is a microporous filter used to minimize particulate (foreign<br>material) contamination of fluids and drugs. |
| PORTEX®<br>PORTEX® Filter<br>Needle with NRFit™<br>Connector and<br>PORTEX® Filter<br>Straw with NRFit™<br>Connector: | The PORTEX® NRFit™ Filter Needle and Filter Straw is used to draw up<br>medication when using an NRFit™ Syringe. |
## 5. INDICATIONS FOR USE
PORTEX® products and components with the NRFit™ Connector are intended for the injection or infusion of regional anesthetics or narcotics.
| Device Category | Indications for Use |
|-----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PORTEX® LOR<br>Syringe with<br>NRFit™ Connector: | The PORTEX® LOR Syringe with NRFit™ Connector is intended for use with<br>NRFit™ compatible components to verify needle tip placement in the epidural space<br>by the Loss of Resistance technique using air or saline. The NRFit™ connector is<br>designed to be used with regional anesthesia systems only. |
| PORTEX® EpiFuse<br>Catheter with<br>NRFit™ Connector | The PORTEX® EpiFuse Catheter with NRFit™ Connector is intended for use with<br>the catheter and compatible components for the injection or infusion of regional<br>anesthetics or narcotics. The key is a tool for re-opening the catheter connector. |
| PORTEX® Epidural<br>Flat Filter with<br>NRFit™ Connector: | PORTEX® Epidural Flat Filter with NRFit™ Connector is designed for use when<br>administering injections and/or infusions of regional anesthetics or narcotics to a<br>patient via compatible system components. The NRFit™ Connector is designed to<br>be used with regional anesthesia systems only. |
| PORTEX® Filter<br>Needle with<br>NRFit™ Connector<br>and PORTEX®<br>Filter Straw with<br>NRFit™ Connector | The PORTEX® Filter Needle with NRFit™ Connector<br>and PORTEX® Filter Straw with NRFit™ Connector are intended to draw up<br>medication when using the PORTEX® LOR Syringe. |
## 6. SUBSTANTIAL EQUIVALENCE DISCUSSION
The Smiths Medical PORTEX® LOR Syringe with NRFit™ Connector, PORTEX® EpiFuse Catheter with NRFit™ Connector, PORTEX® Epidural Flat Filter with NRFit™ Connector, PORTEX® Filter Needle with NRFit™ Connector and PORTEX® Filter Straw with NRFit™ Connector have the same technological characteristics as the predicate devices with the exception of the NRFit TM Connectors.
The main difference is the NRFit™ connectors, which conform to ISO 80369-6, Small bore connectors for liquids and gases in healthcare applications -- Part 6: Connectors for neuraxial applications. The connectors are not compatible with standard luer connectors which are intended to reduce the risk of misconnection that may result in the infusion of medications not intended for
{6}------------------------------------------------
neuraxial or regional anesthetic use.
The Smiths Medical PORTEX® Regional Anesthesia Portfolio with NRFit™ and predicate devices are both designed for the injection or infusion of regional anesthetics or narcotics.
Subject and predicate are made of similar materials, similar size ranges, chemical composition, and have the same design features excluding the NRFit TM connector design.
Potential risks introduced by the differences are addressed through biocompatibility and bench testing and validation and verification data. The differences are not critical to the intended therapeutic use of the device and do not raise different questions of safety and effectiveness of the device when used as labeled.
A comparative analysis is provided in the following Tables;
| Characteristic | Predicate Device<br>K110053 | Subject Device | Discussion |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | Smiths Medical | Smiths Medical | N/A |
| FDA Product<br>Code & CFR | CAZ<br>21 CFR 868.5140 | CAZ<br>21 CFR 868.5140 | Same |
| Regulation Name | Anesthesia conduction kit | Anesthesia conduction kit | Same |
| Regulatory Class | II | II | Same |
| Trade Name | CorrectInject Syringe | PORTEX® LOR Syringe with<br>NRFit™ Connector | N/A |
| Common Name | Syringe | Loss of Resistance Syringe | Similar. Subject is a more<br>specific application type of<br>syringe. |
| Indications for<br>Use | The CorrectInject Syringe<br>is intended for use with<br>CorrectInject compatible<br>components for the injection<br>of medications. | The PORTEX® L.O.R. Syringe<br>with NRFit™ Connector is<br>intended for use with NRFit™<br>compatible components to<br>verify needle tip placement in<br>the epidural space by the Loss<br>of Resistance technique using<br>air or saline. NRFit™<br>connectors are designed to be<br>used with regional anesthesia<br>systems only. | Similar. Both are syringes<br>that are used with<br>compatible components for<br>injection.<br>Predicate is intended for<br>CorrectInject compatible<br>components for the<br>injection of medications.<br>Subject is intended for use<br>with NRFit™ compatible<br>components to verify needle<br>tip placement in the epidural<br>space by the Loss of<br>Resistance technique, and<br>designed to be used with<br>regional anesthesia systems<br>only. This difference does<br>not raise different questions<br>of safety and effectiveness. |
| Hospital Location<br>Use | ICU/OR | ICU/OR | Same |
#### Table 1: PORTEX® LOR Syringes with NRFit™ Connector
{7}------------------------------------------------
## smiths medical
bringing technology to life
| Characteristic | Predicate Device<br>K110053 | Subject Device | Discussion |
|----------------|--------------------------------------|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Connector | Neuraxial safety system | ISO 80369-6 NRFitTM | ISO 80369-6 NRFitTM connector intended to reduce risk of misconnections. Both met the requirements of the respective standards and these standards are recognized by FDA. |
| Tip Design | Lock and slip tip | Lock and slip tip | Same |
| Packaging | Tyvek pouch | Tyvek pouch | Same |
| Sterility | Sterile, EO | Sterile, EO | Same |
| Use | Single Use Disposable | Single Use Disposable | Same |
| mL | 3, 5, 10, 20 | 5, 7 and 10 | Subject capacity is different than predicate's mL range due to the specific application of the LOR syringe. |
| Material | Polypropylene and poly -<br>isoprene | Polypropylene,<br>Isobutylene/isoprene rubber tip | Similar. Subject uses isoprene and predicate uses poly isoprene.<br>Plastic syringe contains Polypropylene in both Subject and predicate.<br>Glass syringe is different. Subject made of glass, predicate made of polypropylene. |
{8}------------------------------------------------
## smiths medical
bringing technology to life
| 510(k) Summary |
|------------------------------------------------------------|
| K172410; PORTEX® Regional Anesthesia Portfolio with NRFit™ |
| Characteristics | Predicate<br>Device | Subject | SE Rationale | Risk Impact |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Syringe Volume<br>capacity (ml) | Volume/size<br>of in<br>subject=5, 7,<br>10 ml<br>ml in<br>predicate = 3,<br>5, 10, 20 ml | Subject has a<br>7 ml size,<br>which is<br>within range<br>of the 5 and<br>10ml<br>predicate<br>device sizes. | Subject capacity is<br>different than<br>predicate's mL range<br>due to the specific<br>application of the LOR<br>syringe. | No risk: The<br>LOR is not used<br>clinically for a<br>precise injection<br>of fluids. The<br>physician will<br>add an<br>approximate<br>amount per their<br>preference to<br>perform the LOR<br>technique with.<br>As such, the<br>volume markings<br>are for reference. |
| Material in glass<br>syringe | Glass | Subject uses<br>glass with a<br>metal - nickel<br>and chrome<br>plated brass.<br>ISO 80369-6<br>tip | The body of the syringes<br>is the same glass<br>material. The NRFit™<br>LOR syringe has a metal<br>tip that meets ISO<br>80369-6 dimensions.<br>Biocompatibility testing<br>was performed on the<br>NRFit™ Glass LOR<br>with passing results. | No risk impact -<br>Materials were<br>verified through<br>biocompatibility<br>testing. |
| Connector | CorrectInject | NRFit™<br>connector in<br>subject | The difference is the<br>subject uses an ISO<br>80369-6 NRFit™<br>connector. This is an<br>industry standard design<br>as opposed to the<br>CorrectInject proprietary<br>standard. | Risk reduction to<br>avoid<br>misconnections. |
| Indications For<br>Use | The predicate<br>is intended<br>for<br>CorrectInject<br>compatible<br>components<br>for the<br>injection of<br>medications. | Subject and<br>predicate are<br>both syringes<br>that are used<br>with<br>compatible<br>components<br>for injection.<br>Subject is<br>intended for<br>NRFit™<br>compatible<br>components<br>to verify<br>needle tip | The subject IFU<br>contains additional<br>information to clarify<br>clinical use. | No risk.<br>Additional<br>information to<br>clarify use. |
#### Table 2: Syringe – Characteristic Comparison
{9}------------------------------------------------
## smiths medical
bringing technology to life
| Characteristics | Predicate<br>Device | Subject | SE Rationale | Risk Impact |
|-----------------|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|-------------|
| | | placement in<br>the epidural<br>space by the<br>Loss of<br>Resistance<br>technique<br>using air or<br>saline.<br>NRFit™<br>connectors<br>are designed<br>to be used<br>with regional<br>anesthesia<br>systems only. | | |
### Table 3: PORTEX® EpiFuse Catheter with NRFit™ Connector
| Characteristic | Predicate Device<br>K110053 | Subject Device | Discussion |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | Smiths Medical | Smiths Medical | N/A |
| FDA Product<br>Code & CFR | CAZ<br>21 CFR 868.5140 | CAZ<br>21 CFR 868.5140 | Same |
| Regulation<br>Name | Anesthesia conduction kit | Anesthesia conduction kit | Same |
| Regulatory Class | II | II | Same |
| Trade Name | CorrectInject Catheter<br>Connector | PORTEX® EpiFuse Catheter<br>with NRFit™ Connector | N/A |
| Common Name | Catheter Connector | Catheter Connector | Same |
| Indications for<br>Use | The CorrectInject Catheter<br>Connector is intended for use<br>with an epidural anesthesia<br>catheter and CorrectInject<br>compatible components for<br>the injection of local or<br>regional anesthetics,<br>narcotics or other<br>medications indicated for<br>injection into the epidural<br>space. | The PORTEX® EpiFuse<br>Catheter with NRFit TM<br>Connector is intended for use<br>with a catheter and compatible<br>components for the injection or<br>infusion of regional anesthetics<br>or narcotics. The key is a tool<br>for re-opening the catheter<br>connector. | Similar. Both used with a<br>catheter. Both are indicated<br>for regional anesthetics or<br>narcotics, and both to be<br>used with compatible<br>components. This<br>difference does not raise<br>different questions of<br>safety and effectiveness. |
| Hospital<br>Location Use | ICU/OR | ICU/OR | Same |
| Connector | CorrectInject Luer | ISO 80369-6 NRFit™ | ISO 80369-6 NRFit™<br>connector intended to<br>reduce risk of<br>misconnections. Both met<br>the requirements of the<br>respective standards and |
{10}------------------------------------------------
## smiths medical
bringing technology to life
| Characteristic | Predicate Device<br>K110053 | Subject Device | Discussion |
|-------------------------------------|--------------------------------------------------------------|-----------------------------------------------------------|-------------------------------------------|
| | | | these standards are<br>recognized by FDA. |
| Packaging | Tyvek pouch | Tyvek pouch | Same |
| Sterility | Sterile, EO | Sterile, EO | Same |
| Use | Single Use Disposable | Single Use Disposable | Same |
| Compatible<br>Needle Length<br>(mm) | 16-19 | 16-19 | Same |
| Catheter<br>Connector -<br>Material | Polypropylene body with<br>a thermoplastic elastomer<br>tube | Polypropylene body with a<br>thermoplastic elastomer tube | Same |
#### Table 4: Catheter Connector – Characteristic Comparison
| Characteri<br>stic | Predicate | Subject<br>Device | SE Rationale | Risk Impact |
|-------------------------|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| Catheter<br>connector : | The predicate<br>uses a luer tip<br>syringe to re-<br>open the<br>catheter<br>connector | Subject uses a<br>key as a tool<br>for re-opening<br>the catheter<br>connector | Both devices require the<br>use of a secondary tool to<br>open the connector The<br>NRFitTM connector<br>requires a specially<br>designed tool as the slip<br>NRFitTM connector has a<br>shroud that surrounds the<br>cone | Risk reduction as<br>Luer syringes<br>cannot open the<br>NRFitTM catheter<br>connector. |
| Connector | CorrectInject<br>Luer | NRFitTM<br>connector in<br>subject | Subject uses an ISO 80369-<br>6 NRFitTM connector,<br>which is designed to reduce<br>risk of misconnections. | Risk reduction to<br>avoid<br>misconnections. |
### Table 5: PORTEX® Epidural Flat Filter with NRFit™ Connector
| Characteristic | Predicate Device<br>K110053 | Subject Device | Discussion |
|---------------------------|-----------------------------|-------------------------------------------------------|------------|
| Company | Smiths Medical | Smiths Medical | N/A |
| FDA Product<br>Code & CFR | CAZ<br>21 CFR 868.5140 | CAZ<br>21 CFR 868.5140 | Same |
| Regulation<br>Name | Anesthesia conduction kit | Anesthesia conduction kit | Same |
| Regulatory Class | II | II | Same |
| Trade Name | CorrectInject Filter | PORTEX® Epidural Flat Filter<br>with NRFit™ Connector | N/A |
| Common Name | Flat Filter | Flat Filter | Same |
{11}------------------------------------------------
## smiths medical
bringing technology to life
| Characteristic | Predicate Device | Subject Device | Discussion |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | K110053 | | |
| Indications for<br>Use | The CorrectInject anesthesia<br>conduction filter is a<br>microporous filter used while<br>administering to a patient<br>injections of local<br>anesthetics to minimize<br>particulate (foreign material)<br>contamination of the injected<br>fluid when used with<br>CorrectInject compatible<br>components. | The PORTEX® Epidural Flat<br>Filter with NRFit™ connectors<br>is designed for use when<br>administering injections and/or<br>infusions of regional<br>anesthetics or narcotics to a<br>patient via compatible system<br>components. NRFit™<br>connectors are designed to be<br>used with regional anesthesia<br>systems only. | Similar. Both are indicated<br>for filter application for<br>injection of anesthetics.<br>This difference does not<br>raise different questions of<br>safety and effectiveness. |
| Hospital<br>Location Use | ICU/OR | ICU/OR | Same |
| Connector | CorrectInject Luer | ISO 80369-6 NRFit™ | ISO 80369-6 NRFit™<br>connector intended to<br>reduce risk of<br>misconnections. Both met<br>the requirements of the<br>respective standards and<br>these standards are<br>recognized by FDA. |
| Filter Pore Size<br>(micron) | 0.02 | 0.02 | Same |
| Packaging | Tyvek pouch | Tyvek pouch | Same |
| Sterility | Sterile, EO | Sterile, EO | Same |
| Use | Single Use Disposable | Single Use Disposable | Same |
| Material | Polyether sulfone (PES) filter<br>enclosed in a modified<br>acrylic housing with a<br>polypropylene rotating collar.<br>The collar is colored yellow<br>using a Clariant colorant. | Polyether sulfone (PES) filter<br>enclosed in a modified acrylic<br>housing cover, with a<br>polypropylene rotating hub.<br>The hub is colored yellow using<br>a Clariant colorant. | Same |
#### Table 6: Flat Filter - Characteristic Comparison
| Characteristic | Predicate | Subject<br>Device | SE Rationale | Risk Impact |
|----------------|----------------|---------------------|-----------------------------|-----------------------------------------------|
| Connector | Luer connector | NRFit™<br>connector | Different<br>connector type | Risk reduction to<br>avoid<br>misconnections. |
#### Table 7: PORTEX® Filter Needle with NRFit™ Connector and Filter Straw with NRFit™ Connector
| Characteristic | Predicate Device<br>K110053 | Subject Device | Discussion |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | Smiths Medical | Smiths Medical | N/A |
| FDA Product<br>Code & CFR | CAZ<br>21 CFR 868.5140 | CAZ<br>21 CFR 868.5140 | Same |
| Characteristic | Predicate Device<br>K110053 | Subject Device | Discussion |
| Regulation<br>Name | Anesthesia conduction kit | Anesthesia conduction kit | Same |
| Regulatory Class | II | II | Same |
| Trade Name | CorrectInject<br>Filter Needle<br>Filter Straw | PORTEX®<br>NRFit™ Filter Needle with<br>NRFit™ Connector and<br>PORTEX®<br>Filter Straw with NRFit™<br>Connector | N/A |
| Common Name | Anesthesia Conduction Kit | Anesthesia Conduction Kit | Same |
| Indications for<br>Use – Filter<br>Straw | The CorrectInject filter straw<br>is intended to adapt a Luer<br>draw-up device to the<br>CorrectInject syringe. | The PORTEX® NRFit™ Filter<br>Needle and Filter Straw is<br>intended to draw up medication<br>when using the PORTEX®<br>Regional Anaesthesia Portfolio<br>with NRFit™ Syringe. | Similar.<br>Both connect to a syringe,<br>and both draw up<br>medication. This difference<br>does not raise different<br>questions of safety and<br>effectiveness. |
| Indications for<br>Use – Filter<br>Needle | The CorrectInject filter<br>needle is intended to draw<br>up medication when using<br>the CorrectInject Syringe. | The PORTEX® NRFit™ Filter<br>Needle and Filter Straw is<br>intended to draw up medication<br>when using the PORTEX®<br>Regional Anaesthesia Portfolio<br>with NRFit™ Syringe. | Similar.<br>Both connect to a syringe,<br>and both draw up<br>medication. This difference<br>does not raise different<br>questions of safety and<br>effectiveness. |
| Hospital<br>Location Use | ICU/OR | ICU/OR | Same |
| Connector | CorrectInject Luer | ISO 80369-6 NRFitTM | ISO 80369-6 NRFit™<br>connector intended to<br>reduce risk of<br>misconnections. Both met<br>the requirements of the<br>respective standards and<br>these standards are<br>recognized by FDA. |
| Filter Pore Size<br>(micron) | 5 | 5 | Same |
| Packaging | Tyvek pouch | Tyvek pouch | Same |
| Sterility | Sterile, EO | Sterile, EO | Same |
| Use | Single Use Disposable | Single Use Disposable | Same |
| Filter Needle<br>Material | Stainless steel needle with a<br>polycarbonate hub and a<br>polybutylene terephthalate<br>adaptor. | Stainless steel needle with an<br>ABS hub. | Similar. Both use stainless<br>steel needle. Difference is<br>hub materials with predicate<br>made of polycarbonate and<br>subject made of ABS. |
| Filter Straw<br>Material | Polycarbonate and an acrylic<br>copolymer attached to a<br>polybutylene terephthalate<br>(PBT) adaptor. | DEHP-free PVC straw tubing<br>with an ABS hub. | Similar. Predicate hub is<br>made of acrylic copolymer<br>attached to a BPT adaptor<br>and subject made from a<br>DEHP-free PVC with ABS<br>hub. |
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the logo for Smiths Medical. The text "smiths" is in a dark blue color, while the text "medical" is in a lighter blue color. The logo is simple and modern.
bringing technology to life
{13}------------------------------------------------
## smiths medical
bringing technology to life
| Characteristic | Predicate | Subject<br>Device | SE<br>Rationale | Risk Impact |
|----------------|----------------|----------------------|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material | Polycarbonate | ABS | Needle and filter<br>materials are<br>similar. | Risk mitigated -<br>The materials<br>incorporated in<br>the NRFitTM<br>Filter Needle are<br>suitable for the<br>devices' intended<br>use and do not<br>pose a biological<br>risk to clinical<br>subjects. |
| Connector | Luer connector | NRFitTM<br>connector | Different<br>connector type | Risk reduction to<br>avoid<br>misconnections. |
#### Table 8: Filter Needle – Characteristic Comparison
#### Table 9: Filter Straw – Characteristic Comparison
| Characteristic | Predicate | Subject<br>Device | SE<br>Rationale | Risk Impact |
|----------------|----------------------------------------------------------------------------------------------|--------------------------------------------------------------|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material | PVC straw and a<br>polycarbonate filter<br>housing with a PBT<br>CorrectInject<br>connector. | PVC straw<br>with an ABS<br>NRFitTM<br>hub/filter<br>housing | Different<br>material type | No risk impact –<br>The materials<br>incorporated in the<br>NRFitTM Filter<br>Straw are suitable<br>for the devices'<br>intended use and do<br>not pose a<br>biological risk to<br>clinical subjects. |
| Connector | Luer connector | NRFitTM<br>connector | Different<br>connector type | Risk reduction to<br>avoid<br>misconnections. |
{14}------------------------------------------------
## 7. SUMMARY OF NON-CLINICAL TESTING
The PORTEX® Regional Anesthesia Portfolio with NRFit™ Connector were evaluated via nonclinical performance testing to demonstrate the devices are as safe, as effective, and perform as well as or better than the predicate devices. All testing met pre-established specifications, and successfully demonstrated that the PORTEX® Regional Anesthesia Portfolio with NRFit™ connectors performed as intended. A summary of the evaluation is provided below in Table 10.
| Category | Evaluation | Test Criteria |
|---------------------------|-------------------------------------------------------------------------|--------------------------------------------------------------…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.