K171968 · Smiths Medical Asd, Inc. · BSP · Jan 25, 2018 · Anesthesiology
Device Facts
Record ID
K171968
Device Name
EchoGlo Needle
Applicant
Smiths Medical Asd, Inc.
Product Code
BSP · Anesthesiology
Decision Date
Jan 25, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EchoGlo™ Peripheral Nerve Block Needle: Single Shot This device is intended for use in the administration of regional anesthetic blocks. The EchoGlo™ Peripheral Nerve Block Needle: Continuous This device is intended for use in the administration of regional anesthetic blocks.
Device Story
EchoGlo Peripheral Nerve Block (PNB) needles are sterile, single-use disposable devices used for regional anesthesia administration under ultrasound guidance. The device consists of a polycarbonate luer hub, stainless steel cannula, and polyethylene sheath. The distal end of the cannula features a helical-shaped groove design and visible bands to enhance ultrasound visibility and monitor insertion depth in 1cm increments. The needles are available with either a bevel tip or a Tuohy tip. For single-shot procedures, the needle is supplied with a 24-inch PVC extension set. Used in clinical settings like the ICU or OR by physicians, the device facilitates needle placement via ultrasound visualization, aiding in the delivery of anesthetic agents. The enhanced echogenicity allows for improved needle tracking, potentially increasing procedural accuracy and patient safety during nerve block placement.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included biocompatibility (cytotoxicity, sensitization, systemic toxicity, pyrogenicity, hemocompatibility), mechanical testing (gauging, leakage, assembly, separation force, unscrewing torque, stress cracking), and ultrasound visibility verification.
Technological Characteristics
Materials: Stainless steel cannula, polycarbonate luer hub, polyethylene sheath, PVC extension tubing. Sensing: Passive ultrasound reflection via helical-grooved cannula and visible depth markings. Dimensions: 18G-22G, 35-150mm lengths. Connectivity: None. Sterilization: Ethylene Oxide (EtO). Standards: ISO 594-1/2 (luer), ISO 7864 (needles), ISO 9626 (tubing), ISO 10993 (biocompatibility).
Indications for Use
Indicated for administration of regional anesthetic blocks in adults.
Regulatory Classification
Identification
An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym along with the full name of the agency on the right. The FDA part is in blue, with the acronym in a square and the words "U.S. FOOD & DRUG ADMINISTRATION" next to it.
January 25, 2018
Smiths Medical ASD, Inc. Michael Johnson Sr. Regulatory Affairs Specialist 6000 Nathan Lane North Minneapolis, Minnesota 55442
Re: K171968
Trade/Device Name: EchoGlo Periphreal Nerve Block Needle: Single Shot and EchoGlo Periphreal Nerve Block Needle: Continuous Catheter Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: Class II Product Code: BSP Dated: December 20, 2017 Received: December 22, 2017
Dear Michael Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171968
Device Name
EchoGloTM Peripheral Nerve Block Needle: Single Shot EchoGlo™ Peripheral Nerve Block Needle: Continuous
Indications for Use (Describe)
The EchoGlo™ Peripheral Nerve Block Needle: Single Shot This device is intended for use in the administration of regional anesthetic blocks.
The EchoGlo™ Peripheral Nerve Block Needle: Continuous This device is intended for use in the administration of regional anesthetic blocks.
Intended population is adults.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Confidential
{3}------------------------------------------------
## 1. ADMINISTRATIVE INFORMATION
| 510(k) | K171968 |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant's Name<br>and Address | Smiths Medical ASD, Inc.<br>6000 Nathan Lane North<br>Minneapolis, MN 55442 USA |
| Contact Person | Michael R. Johnson<br>Sr. Regulatory Affairs Specialist<br>Phone: 763-383-3341<br>Fax: 763-383-3679<br>Email: mike.johnson@smiths-medical.com |
| Date | December 14, 2017 |
| Regulation No. | 21 CFR § 868.5150 |
| Regulation Name | Anesthesia conduction needle |
| Primary Product<br>Code | BSP |
| Trade Name | EchoGlo™ Peripheral Nerve Block Needle: Single Shot<br>EchoGlo™ Peripheral Nerve Block Needle: Continuous |
## 2. DEVICE INFORMATION
| | Predicate Device | Subject Device |
|------------------|------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Trade Name | SONO Tap<br>Tuohy Sono | EchoGlo™ Peripheral Nerve Block<br>Needle: Single Shot<br>EchoGlo™ Peripheral Nerve Block<br>Needle: Continuous |
| Regulation No. | 21 CFR § 868.5150 | 21 CFR § 868.5150 |
| Regulation Name | Anesthesia conduction needle | Anesthesia conduction needle |
| Regulatory Class | II | II |
| Product Code | BSP | BSP |
| 510(k) | K113207 | K171968 |
## 3. DEVICE DESCRIPTION
The EchoGlo™ Peripheral Nerve Block (PNB) Needle consist of a polycarbonate luer hub, stainless steel cannula and a polyethylene sheath. The needle cannula will be supplied with a bevel tip or a Tuohy tip. The EchoGlo™ PNB Needle will be used for peripheral nerve block under ultrasound guidance. The distal end of the cannula has been engraved with two bands to increase ultrasound visibility by optimally reflecting ultrasound waves. These needles are provided as sterile, nonpyrogenic, single-use disposable devices. The needles may be used for single shot peripheral nerve block or for continuous infusion nerve block. The EchoGlo™ PNB Needle used for a single shot procedure will be supplied in a pouch along with a 24 inch extension set. The extension set will consist of polyvinyl chloride (PVC) tubing and have a Female luer at one end of the tubing and a Male luer at the opposite end of the tubing.
{4}------------------------------------------------
## 4. INDICATIONS FOR USE
The EchoGlo™ Peripheral Nerve Block Needle: Single Shot
This device is intended for use in the administration of regional anesthetic blocks.
The EchoGlo™ Peripheral Nerve Block Needle: Continuous
This device is intended for use in the administration of regional anesthetic blocks.
Intended population is adults.
## 5. SUBSTANTIAL EQUIVALENCE DISCUSSION
### Indications for Use
The Echoglo indications for use is similar to the predicate devices include enhanced for ultrasound visibility, similar transient delivery and the option to facilitate placement of a catheter.
A. Both the subject and predicate devices are enhanced for ultrasound visibility. The EchoGlo™ needles are enhanced for ultrasound visibility via grooved echogenic feature. The EchoGlo™ needles also have visible bands which identify the needle length and needle insertion depth in 1cm increments after echogenic feature, visible to the eye without ultrasound imaging.
B. Both the subject and predicate devices offer similar transient use.
C. Both the subject and predicate devices offer the option to facilitate placement of a catheter. Testing demonstrating equivalency for placement.
#### Summary of Technology Characteristics
The Smiths Medical EchoGlo Peripheral Nerve Block Needles and predicate devices are both designed for enhanced ultrasound visibility, delivery of regional anesthesia, and to facilitate placement of a catheter. The Smiths Medical EchoGlo™ Peripheral Nerve Block Needles has similar technological characteristics as the predicate device. Both subject and predicate device incorporate 1cm incremented visible depth markings to monitor insertion depth through ultrasound imaging. The subject device and predicate devices are prescribe by physician only, for single use and provided Ethylene Oxide (EO) sterile to the healthcare facility and/or end user.
The difference between the predicate is the Smiths EchoGlo has echogenic visible bands with helical markers while the SonoTAP and Tuohy Sono include diamond shape markers. Additionally, the Echoglo offers a Bevel tip in 20G-22G, and a Tuohy tip in 18G.
| Table 5-1 | | | |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Predicate Device<br>K113207 | Subject Device<br>K171968 | Discussion |
| Company | Pajunk GMBH | Smiths Medical | NA |
| FDA Product Code<br>& CFE | BSP<br>21 CFR 868.5150. | BSP<br>21 CFR 868.5150. | Same |
| Trade Name | SONO Tap<br>Tuohy Sono | EchoGlo™ Peripheral<br>Nerve Block Needle: | NA |
| Characteristic | Predicate Device<br>K113207 | Subject Device<br>K171968 | Discussion |
| | | Single Shot | |
| | | EchoGlo™ Peripheral<br>Nerve Block Needle:<br>Continuous | |
| Product | Needle, Conduction,<br>Anesthetic (w/wo<br>introducer) | Single Shot and<br>Continuous Nerve Block<br>Needle | NA |
| Common /<br>Classification<br>Name | Anesthesia conduction<br>needle | Anesthesia conduction<br>needle | Same |
| Indications for Use | The cannulas/ needles<br>for anesthesia and<br>analgesia enhanced for<br>ultrasound visibility -<br>Tuohy Sono, Sono<br>TAP, Quincke Sono,<br>Chiba Sono, Sono and<br>Crawford Sono - are<br>intended for the<br>transient delivery of<br>anesthetics to provide<br>regional anesthesia<br>and analgesia or to<br>facilitate placement of<br>a catheter | The EchoGlo™ Peripheral<br>Nerve Block Needle:<br>Single Shot<br>This device is intended for<br>use in the administration<br>of regional anesthetic<br>blocks.<br>The EchoGlo™ Peripheral<br>Nerve Block Needle:<br>Continuous<br>This device is intended for<br>use in the administration<br>of regional anesthetic<br>blocks. | Similar. Both are<br>enhanced for<br>ultrasound<br>visibility. Also,<br>both are intended<br>for the delivery of<br>regional<br>anesthesia, or to<br>facilitate<br>placement of a<br>catheter. |
| Hospital Location<br>Use | ICU/OR | ICU/OR | Same |
| Visibility feature | Etched cannula/needle<br>to enhance ultrasound<br>visibility. The needle<br>cannula is equipped<br>with a specific<br>diamond shaped<br>reflector pattern that is<br>designed to reflect<br>ultrasound waves. | Etched cannula/needle to<br>enhance ultrasound<br>visibility. A special<br>helical-shaped groove<br>design is cut into the<br>needle cannula. The<br>grooves are designed to<br>optimally reflect<br>ultrasound waves to make<br>the needle more visible<br>under ultrasound. | Similar. Both are<br>etched needles that<br>are designed to<br>enhance visibility<br>under ultrasound.<br>Difference is the<br>etched pattern. |
| Proximal End<br>Connection | Luer, Per ISO 594-1,<br>and ISO 594-2 | Luer, Per ISO 594-1, and<br>ISO 594-2 | Same |
| Single Use | Yes | Yes | Same |
| Sterilization<br>Method | EtO | EtO | Same |
| Tip Design | Tuohy, Bevel | Tuohy, Bevel | Same |
| Needle Gauge | Tuohy SONO : 16G -<br>26G | Bevel : 20G-22G<br>Tuohy: 18G | Similar. Subject<br>diameter is |
| Characteristic | Predicate Device<br>K113207 | Subject Device<br>K171968 | Discussion |
| | Quincke SONO: 16G -<br>26G<br>Sono TAP, Chiba<br>SONO: 16G - 26G<br>Crawford SONO: 16G<br>- 26G | | between<br>predicate's gauge<br>range |
| Needle Length | Tuohy SONO: 20mm-<br>180mm<br>Sono TAP, Quincke<br>SONO: 20mm-180mm<br>Chiba SONO: 20mm-<br>180mm<br>Crawford SONO:<br>20mm-180mm | Single Shot: 35-150mm<br>Continuous: 50-150mm | Similar. Subject<br>length is within<br>predicate's length<br>range |
| Cannula Material | Stainless steel | Stainless steel | Same |
| Needle Hub<br>Material | Polycarbonate | Polycarbonate | Same |
| Sheath Material | Polyethylene | Polyethylene | Same |
| Extension Set<br>Tubing Material | Polyvinyl<br>chloride (PVC) | Polyvinyl<br>chloride (PVC) | Same |
| Extension Set Male<br>Luer Material | Polyacrylic | PVC | Similar |
| Extension Set<br>Female Luer<br>Material | PVC | PVC | Same |
A comparative analysis of the technological characteristics is provided in Table 5-1.
{5}------------------------------------------------
# smiths medical
bringing technology to life
{6}------------------------------------------------
# smiths medical
bringing technology to life
{7}------------------------------------------------
## 6. SUMMARY OF NON-CLINICAL TESTING
The Smiths Medical EchoGlo™ Peripheral Nerve Block Needle was evaluated via non-clinical performance testing to demonstrate the devices are substantially equivalent the predicate devices. All testing met pre-established specifications, and successfully demonstrated that the EchoGlo™ Peripheral Nerve Block Needle performed as intended. A summary of the Non-Clinical performance testing is in Table 5-2. Biological and Bench testing was conducted to the Standards in Table 5-3.
| Category | Evaluation | Standards |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | Cytotoxicity | ISO 10993-7:2009: Biological<br>evaluation of medical devices—<br>Part 5: Tests for in vitro<br>cytotoxicity |
| | Sensitization | ISO 10993-10:2010: Biological<br>evaluation of medical devices—<br>Part 10: Tests for irritation and skin<br>sensitization |
| | Acute systemic toxicity | ISO 10993-11:2006: Biological<br>evaluation of medical devices—<br>Part 11: Tests for systemic toxicity |
| | Material-Mediated Pyrogenicity | USP<151>Pyrogen Test |
| | Particulate Matter in Injections | USP<788>:2012 Particulate Matter<br>in Injections |
| | Exaggerated Leachables | ISO 10993-17:2002: Biological<br>evaluation of medical devices—<br>Part 17: Establishment of allowable<br>limits for leachable substances |
| | Hemocompatibility | ISO 10993-4:2002/(R)2013 &<br>A1:2006/(R)2013: Biological<br>evaluation of medical devices - Part<br>4: Selection of tests for interaction<br>with blood |
| Packaging | Existing packaging material for currently<br>marketed devices will be used for the<br>packaging of the EchoGlo™ Peripheral<br>Nerve Block Needle. Packaging integrity<br>test were performed to ensure packaging<br>meet specifications. For details reference<br>Section 14.3 | ASTM D4169:2016: Standard<br>practice for performance testing of<br>shipping containers and systems. |
| | | ASTM F1886:2013: Determining<br>integrity of seals for flexible<br>packaging by visual inspection. |
| | | ASTM F2825:2015: Standard<br>practice of climatic stressing of<br>packaging systems for single parcel<br>delivery. |
| Category | Evaluation | Standards |
| | | test method for seal strength of<br>flexible barrier materials. |
| | | ASTM F2096:2011: Standard test<br>method for detecting gross leaks in<br>medical packaging by internal<br>pressurization. |
| | | ISO 11607-1:2006/(R):2010:<br>Packaging for terminally sterilized<br>medical devices Part 1:<br>Requirements for materials, sterile<br>barrier systems, and packaging<br>systems. |
| Device Performance | Gauging (T=0, T=6 months accelerated<br>aging (AA))<br>Leakage, Air (T=0, T=6 months AA)<br>Leakage, Liquid (T=0, T=6 months AA)<br>Ease of Assembly (T=0, T=6 months AA)<br>Separation Force (T=0, T=6 months AA)<br>Resistance to Overriding (T=0, T=6 months<br>AA)<br>Unscrewing Torque (T=0, T=6 months AA)<br>Stress Cracking (T=0, T=6 months AA) | ISO 594-1: Conical fittings with a<br>6% (luer) taper for syringes,<br>needles, and certain other medical<br>equipment - Part 1: General<br>Requirements<br>ISO 594-2 second ed. 1998-09-01:<br>Conical Fittings with a 6% taper<br>(luer) for syringes, needles, and<br>certain other medical equipment -<br>Part 2: Lock fittings |
| | Needle Inspections<br>Bond - Needle and hub (T=0, T=6 months<br>AA)<br>Limits of Acidity/Alkalinity<br>Patency of Lumen<br>Limits for Extractable materials | ISO 7864 Fourth Edition: 2016-08-<br>01: Sterile hypodermic needles for<br>single use - Requirements and test<br>methods |
| | Needle Inspections<br>Needle Stiffness<br>Resistance to Breakage<br>Resistance to Corrosion | ISO 9626 Second Edition 2016-08-<br>01: Stainless steel needle tubing for<br>the manufacture of medical devices<br>- Requirements and test methods |
| | Bond Pull - Extension set (T=0, T=6<br>months AA)<br>Demonstration<br>Penetration Force<br>Ultrasound Visibility | N/A |
| Human Factors | Evaluate user interface | FDA guidance, Applying Human<br>Factors and Usability Engineering<br>to Medical Devices: Guidance for<br>Industry and Food and Drug<br>Administration Staff (3 February<br>2016) |
Table 5-2 - Non-Clinical Performance Testing
{8}------------------------------------------------
# smiths medical
bringing technology to life
| Recognition<br>Number | Standard Number | Assessment |
|-----------------------|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 14-452 | ISO 11135 2nd Edition 2014 | Sterilization Of Health-Care Products -<br>Ethylene Oxide - Requirements For The<br>Development, Validation And Routine |
| Recognition<br>Number | Standard Number | Assessment |
| 2-156 | ISO 10993-1:2009 R 2013 | Control Of A Sterilization Process For<br>Medical Devices. (Sterility) |
| 6-11 | ISO 594-1 | Biological Evaluation Of Medical<br>Devices - Part 1: Evaluation And Testing<br>Within A Risk Management Process |
| 6-129 | ISO 594-2 Second Ed. 1998-09-01 | Conical Fittings With A 6% (Luer)<br>Taper For Syringes, Needles And<br>Certain Other Medical Equipment - Part<br>1: General Requirements |
| 14-457 | 11607-1:2006/(R)2010 | Conical Fittings With A 6% (Luer)<br>Taper For Syringes, Needles And<br>Certain Other Medical Equipment - Part<br>2: Lock Fittings |
| 2-234 | ISO 10993-4:2002/(R)2013 &<br>A1:2006/(R)2013 | Packaging For Terminally Sterilized<br>Medical Devices - Part 1: Requirements<br>For Materials, Sterile Barrier Systems<br>And Packaging Systems [Including:<br>Amendment 1 (2014)] |
| 14-278 | 10993-7:2008(R)2012 | Biological evaluation of medical devices<br>- Part 4: Selection of tests for interaction<br>with blood |
| 2-118 | AAMI ANSI ISO 10993-<br>11:2006/(R)2010 | Biological Evaluation Of Medical<br>Devices - Part 7: Ethylene Oxide<br>Sterilization Residuals |
| 14-227 | AAMI ANSI ISO 11737-1:2006<br>(R)2011 | Biological evaluation of medical devices<br>Part 11: Tests for systemic toxicity |
| 14-360 | ANSI/AAMI ST72:2011 | Sterilization Of Health Care Products -<br>Microbiological Methods - Part 1:<br>Determination Of The Population Of<br>Microorganisms On Product |
| 6-379 | ISO 7864 Fourth Edition 2016-08-01 | Bacterial Endotoxins - Test Methods,<br>Routine Monitoring, And Alternatives<br>To Batch Testing |
| 6-380 | ISO 9626 Second Edition 2016-08-01 | Sterile Hypodermic Needles For Single<br>Use - Requirements And Test Methods |
| 5-108 | ASTMD 4169-16 2014 | Stainless Steel Needle Tubing For The<br>Manufacture Of Medical Devices -<br>Requirements And Test Methods |
| N/A | USP<151> | Standard Practice for Performance<br>Testing of Shipping Containers and<br>Systems |
| N/A | USP<788>:2012 | Pyrogen Test |
{9}------------------------------------------------
# smiths medical
bringing technology to life
# 7. SUBSTANTIAL EQUIVALENCE CONCLUSION
The Smiths Medical EchoGlo™ Peripheral Nerve Block Needle is substantially equivalent to the
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for Smiths Medical. The logo consists of the words "smiths medical" in blue font. Below the logo is the tagline "bringing technology to life" in a smaller, lighter blue font.
predicate device based on comparisons of the device classifications, intended use, and technological characteristics. Verification and validation tests confirmed the suitability of the subject devices for its intended uses. The results of biocompatibility testing and performance confirmed the substantial equivalence of the EchoGlo™ Peripheral Nerve Block Needle. The test results do not raise different safety or performance questions, and confirmed that the EchoGlo™ Peripheral Nerve Block Needle is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.