← Product Code [BSP](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSP) · K093209

# FOCUS ( TM) ECHOGENIC NERVE STIMULATING NEEDLE (K093209)

_Cook, Inc. · BSP · Feb 26, 2010 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSP/K093209

## Device Facts

- **Applicant:** Cook, Inc.
- **Product Code:** [BSP](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSP.md)
- **Decision Date:** Feb 26, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5150
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The Focus™ Echogenic Nerve Stimulating Needle is intended for locating and stimulating peripheral nerves and nerve plexuses for nerve block anesthesia techniques using a Peripheral Nerve Stimulator and/or ultrasound guidance.

## Device Story

Focus Echogenic Nerve Stimulating Needle; sterile, single-use, FEP-coated stainless steel needle; B-bevel tip. Features black 10 mm marker bands for placement; distal 10 mm dimpled for ultrasound visibility. Insulated wire connects cannula to external peripheral nerve stimulator. Used by clinicians for nerve block anesthesia; facilitates nerve localization via electrical stimulation and ultrasound imaging. Benefits include enhanced needle visualization and precise nerve targeting for anesthesia delivery.

## Clinical Evidence

Bench testing only. Verification performed: air/fluid leakage, bond strength (needle-to-hub, wire-to-connector), tensile strength, cannula break strength, penetration force, sterilization, and biocompatibility. Animal model evaluation confirmed electrical current applied via the needle successfully stimulates nerves.

## Technological Characteristics

Stainless steel cannula with FEP (fluorinated ethylene propylene) polymer coating; plastic hub; insulated wire for nerve stimulator connection. Distal 10 mm dimpled for ultrasound echogenicity. Available in 22 and 23 gauge. Sterile, single-use.

## Regulatory Identification

An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.

## Predicate Devices

- EchoStim Facet Tip Needle ([K063380](/device/K063380.md))
- Temena Hybrid Nerve Location Needle ([K080603](/device/K080603.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Cook Incorporated FocusTM Echogenic Nerve Stimulating Needle Traditional 510(k) 09 October 2009

K093209
6
page 1 of 3

#### 5. 510(k) Summary

# FEB 2 6 2010

Cook Incorporated Focus™ Echogenic Nerve Stimulating Needle 510(k) Summary 21 CFR 807.92

## 1. Submitter Information:

Applicant:

Cook Incorporated

Address:

Phone Number: Fax Number:

Contact: Contact Address:

750 Daniels Way P.O. Box 489 Bloomington, IN 47402

(800) 468-1379 (812) 332-0281

Sean Werner, PhD Cook Incorporated 750 Daniels Way P.O. Box 489 Bloomington, IN 47402

Contact Phone Number: Contact Fax Number:

812-339-2235 x2685 812-332-0281

#### 2. Device Information:

Trade name: Common name: Classification:

Regulation:

Product Code:

Focus™ Echogenic Nerve Stimulating Needle Needle, Conduction, Anesthetic (W/Wo introducer) Class II

21 CFR § 868.5150 Anesthesia Conduction Needle

BSP

Company Confidential

{1}------------------------------------------------

# 3. Predicate Devices:

Cook Incorporated's Focus Echogenic Nerve Stimulating Needle (hereafter referred to as the Focus Needle) is substantially equivalent to the EchoStim Facet Tip Needle manufactured by Hakko Medical Co., Ltd. of Japan and distributed in the U.S. by Havel's Inc. under 510(k) clearance K063380. The Focus Echogenic Nerve Stimulating Needle is also substantially equivalent to the Temena Hybrid Nerve Location Needle manufactured by TE ME NA under 510(k) clearance K080603.

#### 4. Device Description:

The Focus Needle is a sterile, single use FEP-coated stainless steel needle with a B Bevel tip. The needle contains black marker bands every 10 mm to aid in placement and echogenic dimpling on the distal 10 mm to enhance visibility under ultrasound guidance. An insulated wire is attached to the stainless steel cannula and can be connected to a peripheral nerve stimulator unit. The device will be available in the following gauge sizes: 22 and 23.

## 5. Intended Use:

The Focus™ Echogenic Nerve Stimulating Needle is intended for locating and stimulating peripheral nerves and nerve plexuses for nerve block anesthesia techniques using a Peripheral Nerve Stimulator and/or ultrasound guidance.

{2}------------------------------------------------

## 6. Technological Characteristics:

The Focus Needle consists of a stainless steel cannula coated with a fluorinated ethylene propylene polymer to provide resistance to loss of electrical current along the shaft of the needle, a plastic hub with an attachment point for a connection tube, and an insulated wire attached to the cannula to allow connection to a nerve stimulator unit. Additionally, the distal 10 mm of the needle are dimpled using a pattern designed to enhance visibility by ultrasound. The technological characteristics of the Focus Needle and the two predicate devices, the EchoStim Facet Tip Needle and the Temena Hybrid Nerve Location Needle are substantially equivalent in that all devices are coated with an electrically resistant polymer to allow for electric nerve stimulation, and all devices can be visualized with ultrasound technology. No new technological aspects are being introduced with the proposed device.

To demonstrate reliable design and performance of the Focus Needle, the following verification testing was performed:

- . Air leakage under vacuum,
- . Fluid leakage under pressure,
- Force to remove protector. .
- . Needle to hub bond strength,
- Wire to electrical connector bond, .
- . Wire tensile strength.
- . Wire to hub tensile strength,
- Cannula break strength, .
- . Penetration force,
- Sterilization testing, .
- Biocompatibility testing. .

Additionally, an evaluation in an animal model was carried out to verify that electrical current applied through the Focus Needle by a Peripheral Nerve Stimulator unit can stimulate a nerve.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Dr. Sean Werner Regulatory Science Associate Cook, Incorporated 750 Daniels Way PO Box 489 Bloomington, Indiana 47402

FEB 2 6 2010

Re: K093209

Trade/Device Name: Focus™ Echogenic Nerve Stimulating Needle Regulation Number: 21CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: II Product Code: BSP Dated: February 19, 2010 Received: February 22, 2010

Dear Dr. Werner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 Dr. Werner

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Rh. for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Cook Incorporated FocusTM Echogenic Nerve Stimulating Needle Traditional 510(k) 09 October 2009

## 4. Indications for Use Statement

510(k) Number (if known): K093209

Device Name: Focus™ Echogenic Nerve Stimulating Needle

#### Indications for Use:

The Focus™ Echogenic Nerve Stimulating Needle is intended for locating and stimulating peripheral nerves and nerve plexuses for nerve block anesthesia techniques using a Peripheral Nerve Stimulator and/or ultrasound guidance.

Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

L. Schulten

sion Slan-Off Division of Anesthestology. General Hospital Infection Control. Dental Devices

810(k) Number: K093209

Company Confidential

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSP/K093209](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSP/K093209)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
