DISCYPHOR CATHETER SYSTEM, MODEL D01A; DISCYPHOR INTRODUCER NEEDLE, MODEL D01B
Applicant
Kyphon, Inc.
Product Code
BSP · Anesthesiology
Decision Date
Apr 13, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Kyphon Discyphor Catheter System for the Functional Anaesthetic Discography Procedure, and its components, are intended for use in delivering either a single dose or continuous administration of radiopaque contrast, local anaesthetics, and/or saline solution to the intradiscal space. The Discyphor Introducer Needle is intended for use to access the area adjacent to the intradiscal space for the purpose of facilitating sequential placement of the Discyphor Spinal Needle, Discyphor Catheter and Discyphor Guidewire into the intradiscal space. The Kyphon Discyphor Introducer Needle is intended to be used only with the Kyphon Discyphor Catheter System. The Kyphon Discyphor Spinal Needle is intended to access the nucleus of an intervertebral disc for the purpose of performing provocative discography and facilitating placement of the Discyphor Catheter and Guidewire into the intradiscal space. The Discyphor Spinal Needle can be used to deliver contrast, antibiotic, and/or saline into an intervertebral disc. The Kyphon Discyphor Spinal Needle is intended to be used only with the Kyphon Discyphor Catheter System.
Device Story
System facilitates Functional Anaesthetic Discography (F.A.D.) procedure; provides access to intervertebral disc nucleus. Components include stainless steel introducer needle, spinal needle, micro-catheter with distal balloon, and guidewire. Physician uses introducer needle to access area adjacent to disc; spinal needle accesses nucleus for provocative discography; catheter/guidewire placed into intradiscal space. System delivers contrast, local anesthetics, antibiotics, or saline. Guidewire is radiopaque for fluoroscopic visualization. Stopcock and syringes manage balloon pressure. Device used in clinical setting by physicians. Benefits include targeted delivery of diagnostic/therapeutic agents to intradiscal space.
Clinical Evidence
Bench testing only. In vitro testing evaluated bending stiffness, peak axial strength, fracture, and kink resistance of the modified guidewire. Biocompatibility testing performed on the catheter system and guidewire assembly. Laboratory physician preference testing conducted.
Technological Characteristics
System components: stainless steel needles (introducer/spinal), micro-catheter with distal balloon, Nitinol/Elgiloy guidewire. Guidewire diameter 0.009". Radiopaque under fluoroscopy. Polymer hubs on needles. Manual operation via syringes and stopcock. No software or electronic components.
Indications for Use
Indicated for patients undergoing provocative discography or requiring delivery of contrast, local anesthetics, antibiotics, or saline to the intradiscal space of an intervertebral disc.
Regulatory Classification
Identification
An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.
Predicate Devices
Functional Anaesthetic Discography (F.A.D.) Catheter System (K061210)
{0}------------------------------------------------
Kyphon Inc. Page 1 of 3
Special Premarket Notification, K063071, Amendment 2 Discyphor™ Catheter System Discyphor™ Introducer Needle and Discyphor™ Spinal Needle
APR 1 3 2007
## Premarket Notification [510(k)] Summary
| Trade Name: | DiscyphorTM Catheter System<br>DiscyphorTM Introducer Needle<br>DiscyphorTM Spinal Needle |
|------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Anesthesia conduction catheter<br>Anesthesia conduction needle |
| Classification<br>/Name: | Class II, Anesthesia conduction needle, 21 CFR 868.5150 |
| Device Code: | BSP |
| Manufacturer's<br>Name:<br>Address: | Kyphon Inc.<br>1221 Crossman Avenue<br>Sunnyvale, CA 94089 |
| Corresponding<br>Official:<br>Title:<br>Address:<br>Phone: | Cindy Domecus<br>Clinical Research and Regulatory Affairs Consultant<br>1221 Crossman Avenue<br>Sunnyvale, CA 94089<br>(408) 548-5421 |
| Predicate<br>Device(s): | K061210, Functional Anaesthetic Discography (F.A.D.)<br>Catheter System, cleared on June 27, 2006.<br>K043500, Functional Anaesthetic Discography (F.A.D.) |
Ary statement regarding "substantial equivalence" made in this submission only relates to whether the product addressed in this submission may be lawfully marketed without premarket approval or reclassification and is not intended to be interpreted as an admission or any other type of evidence in any patent proceeding, including patent infringement litigation or proceeding before any Patent Office. The present submission and statements therein therefore should not be construed as affecting or relating to the scope of any patent application, or to whether the product addressed in this submission, or its use, may be considered indistinct, from a a atestability perspective, from any of the other devices referenced in this filing.
{1}------------------------------------------------
Kyphon Inc. Page 2 of 3
Intended Use: The Kyphon Discyphor Catheter System for the Functional Anaesthetic Discography Procedure, and its components, are intended for use in delivering either a single dose or continuous administration of radiopaque contrast, local anaesthetics, and/or saline solution to the intradiscal space.
> The Discyphor Introducer Needle is intended for use to access the area adjacent to the intradiscal space for the purpose of facilitating sequential placement of the Discyphor Spinal Needle, Discyphor Catheter and Discyphor Guidewire into the intradiscal space. The Kyphon Discyphor Introducer Needle is intended to be used only with the Kyphon Discyphor Catheter System.
> The Kyphon Discyphor Spinal Needle is intended to access the nucleus of an intervertebral disc for the purpose of performing provocative discography and facilitating placement of the Discyphor Catheter and Guidewire into the intradiscal space. The Discyphor Spinal Needle can be used to deliver contrast, antibiotic, and/or saline into an intervertebral disc. The Kyphon Discyphor Spinal Needle is intended to be used only with the Kyphon Discyphor Catheter System.
The Discyphor" Catheter System is comprised of a Discyphor" Catheter with Device Description: guidewire funnel, a Discyphor" Guidewire, a Discyphor" Introducer Needle, a Discyphor" Spinal Needle, a stopcock, syringes, catheter connectors and labels. The Discyphor Catheter is a micro-catheter with a flexible shaft and a balloon located near the distal tip. The metallic Discyphor Guidewire is .009" in diameter and is radiopaque when viewed using fluoroscopy. The stopcock retains pressure within the balloon when used in conjunction with the syringes. The catheter connectors are connected to the lumens of the Discyphor Catheter and provide access to the lumens. Labels are used to identify the solutions used during the procedure or to denote the level at which the Discyphor Catheter has been placed.
> The Discyphor Introducer Needle is designed specifically for use with the Discyphor Catheter System. The Discyphor Introducer Needle is stainless steel with a removable stainless steel stylet. The Discyphor Catheter is delivered through the Discyphor Introducer Needle and is tracked over the Discyphor Guidewire.
The Discyphor" Spinal Needle is a standard stainless steel needle with a removable stainless steel stylet. The hubs of the needle and stylet are comprised of polymer.
Substantial This 510(k) describes a material change to the Discyphor Guidewire. The Equivalence: Discyphor Guidewire is one component of the Discyphor Catheter System.
Any statement regarding "substantial equivalence" made in this submission only relates to whether the product addressed in this submission may be lawfully marketed without premarket approval or reclassification, and is not intended to be interpreted as an admission or any other type of evidence in any patent proceeding, including patent infringement litigation or proceeding before any Patent Office. The present submission and statements therein therefore should not be construed as affecting or relating to the scope of any patent application, or to whether the product addressed in this submission, or its use, may be considered indistinct, from a pateriability perspective, from any of the other devices referenced in this filing.
{2}------------------------------------------------
Kyphon Inc. Page 3 of 3
The guidewire materials for the Discyphor Guidewire include Nitinol and Elgiloy. In vitro testing was conducted on the materials to evaluate bending stiffness, peak axial strength, fracture and kink resistance. In addition laboratory physician preference testing was conducted. Biocompatibility testing conducted on the Discyphor Catheter System and Guidewire assembly demonstrates that the materials are acceptable for the intended use. The results of testing demonstrate that the guidewire materials are safe and effective for the intended use. The changes described in this submission do not represent a change to the intended use and they do not represent a change in the fundamental scientific technology of the device. Therefore, the devices included in this submission are substantially equivalent to the predicate devices already cleared under K061210.
Any statement regarding "substantial equivalence" made in this submission only relates to whether the product addressed in this submission may be lawfully marketed without premarket approval or reclassification, and is not intended to be interpreted as an admission or any other type of evidence in any patent proceeding, including patent The present submission and statements infringement litigation or proceeding before any Patent Office. therein therefore should not be construed as affecting or relating to the scope of any patent application, or to whether the product addressed in this submission, or its use, may be considered indistinct, from a patentability perspective, from any of the other devices referenced in this filing.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three wing-like shapes. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the bird-like figure. The logo is black and white.
## APR 1 3 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Cindy Domecus Clinical Research and Regulatory Affairs Consultant Kyphon, Incorporated 1221 Crossman Avenue Sunnyvale, California 94089
Re: K063071
Trade/Device Name: Discyphor™ Catheter System, Discyphor™ Introducer Needle, Discyphor™ Spinal Needle Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: II Product Code: BSP Dated: March 14, 2007 Received: March 15, 2007
Dear Ms. Domecus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Domecus
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Qives
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K063071
Device Name: Discyphor™ Catheter System Discyphor
Discyphor Discyphor
Di Introducer Needle Discyphor Spinal Needle
Indications for Use:
The Kyphon Discyphor Catheter System for the Functional Anaesthetic Discography™ Procedure, and its components, are intended for use in delivering either a single dose or continuous administration of radiopaque contrast, local anaesthetics, and/or saline solution to the intradiscal space.
The Kyphon Discyphor Introducer Needle is intended for use to access the area adjacent to the intradiscal space for the purpose of facilitating sequential placement of the Discyphor Spinal Needle, Discyphor Catheter and Discyphor Guidewire into the intradiscal space. The Kyphon Discyphor Introducer Needle is intended to be used only with the Kyphon Discyphor Catheter System.
The Kyphon Discyphor Spinal Needle is intended to access the nucleus of an intervertebral disc for the purpose of nerforming provocative discography and facilitating placement of the Discyphor Catheter and Guidewire into the intradiscal space. The Discyphor Spinal Needle can be used to deliver contrast, antibiotic, and/or saline into an intervertebral disc. The Kyphon Discyphor Spinal Needle is intended to be used only with the Kyphon Discyphor Catheter System.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
|--|
| of Anesthesia / General Hospital, |
|------------------------------------|
| Infection Control, Device Services |
| 510(k) Number: | K063071 |
|----------------|---------|
|----------------|---------|
| | Page 1 of 1 |
|--|-------------|
|--|-------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.