FUNCTIONAL ANAESTHETIC DISCOGRAPHY (FAD) SYSTEM, INCLUDING FAD CATHETER KIT AND FAD INTRODUCER NEEDLE
Applicant
Innospine, Inc.
Product Code
BSP · Anesthesiology
Decision Date
Apr 15, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Functional Anaesthetic Discography (FAD) Catheter System is intended for use in delivering either a single dose or continuous administration of a radiopague contrast, local anaesthetics, and/or saline solution to the intradiscal space. The FAD Introducer Needle is intended for use to access the area adjacent to the intradiscal space for the purpose of facilitating placement of the FAD Catheter and Guidewire into the intradiscal space.
Device Story
The FAD System is a medical device kit comprising a micro-catheter and an introducer needle used to access the intradiscal space. The system allows for the delivery of radiopaque contrast, local anesthetics, or saline solutions. The catheter features a flexible polymer shaft with an expandable anchor mechanism, controlled via a dedicated lumen, to secure the catheter in place during fluid administration. Two radiopaque markers facilitate radiographic positioning. The introducer needle, made of stainless steel with a stylet, provides the initial access path for the guidewire and catheter. The device is intended for use by clinicians in a clinical setting. By enabling precise placement and anchoring of the catheter, the system facilitates targeted delivery of diagnostic or therapeutic agents to the intradiscal space, potentially aiding in the diagnosis or management of spinal conditions.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
System consists of a flexible polymer micro-catheter with an expandable anchor mechanism and a stainless steel introducer needle with stylet. Includes radiopaque markers for positioning. Accessories include guidewire, syringes, and Touhy-Borst adapters. Non-electronic, mechanical device.
Indications for Use
Indicated for patients requiring delivery of radiopaque contrast, local anesthetics, or saline to the intradiscal space. Used to access the intradiscal space via an introducer needle for catheter and guidewire placement.
Regulatory Classification
Identification
An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.
{0}------------------------------------------------
APR 1 5 2005
INNOSPINE, INC. 510(k) Submission
## 510(k) SUMMARY InnoSpine, Inc. Functional Anaesthetic Discography (FAD) System 510(k) Notification K043500
#### GENERAL INFORMATION
#### Applicant:
InnoSpine, Inc. 1766 Sand Hill Road, Unit 404 Palo Alto, CA 94304 Phone: 650-996-2110 FAX: 650-249-0218
#### Contact Person:
Laraine Pangelina Regulatory Consultant Experien Group, LLC 155 Moffett Park Drive, Suite A-101 Sunnyvale, CA 94089-1330 Phone: 408-400-0856 FAX: 408-400-0865
#### Date Prepared:
December 17, 2004
#### DEVICE INFORMATION
The Functional Anaesthetic Discography (FAD) System consists of the FAD Catheter Kit and the FAD Introducer Needle.
#### Classification:
Class II, Anesthesia conduction catheter, 21 CFR §868.5120 Class II, Anesthesia conduction needle, 21 CFR §868.5150
#### Trade Name:
Functional Anaesthetic Discography (FAD) System:
- FAD Catheter Kit .
- FAD Introducer Needle .
CONFIDENTIAL
{1}------------------------------------------------
## Generic/Common Name:
Anesthesia conduction catheter Anesthesia conduction needle
#### PREDICATE DEVICES
#### FAD Catheter Kit:
- Spinal Specialties Discography Kit, K960082, cleared 3/21/96 .
- Epimed Versa-Kath Epidural Catheter, K023140, cleared 8/15/03 .
- Braun Perifix Catheter, K042488, cleared 9/30/04 .
## FAD Introducer Needle:
- Braun Perifix Safety Epidural Needle, K013610, cleared 1/25/02 .
- Epimed Quincke Spinal Needle, K022029, cleared 7/30/02 .
- Pajunk Touhy Needles, K040965, cleared 9/7/04 .
#### INTENDED USE
The Functional Anaesthetic Discography (FAD) System is intended for use in delivering The I unchonal I rindothers Broosean inistration of radiopaque contrast, local anaesthetics, and/or saline solution to the intradiscal space. The FAD Introducer Needle is intended for and of same to access the intradiscal space for the purpose of facilitating use to access the area agatheter and Guidewire into the intradiscal space.
#### PRODUCT DESCRIPTION
The Functional Anaesthetic Discography (FAD) System consists of the FAD Catheter Kit The FAD Introducer Needle. The FAD Catheter Kit includes one FAD Catheter, one alle the I AD Introducer Needle, and other purchased accessories, including one Ouldewine, one 1 AD introducer Nocarener, two 1 cc syringes and two Touhy-Borst adapters.
The FAD Catheter is a micro-catheter with a flexible polymer shaft. There is a polymer TIC I AD Catheter 15 a miero cannote Catheter, which is expanded once the catheter is in allow near the unsul up of the ring over catheter placement during the delivery of fluids place for the purpose of manualing pror to removal of the catheter. Two mo the mitradibed spares in polymer shaft of the catheter, one for the purpose of expanding the anchor, the other for the purpose of tracking over the guidewire and injecting the desired fluid. There are two radiopaque markers, one proximal to the anchor and one the desired from. "There as w for radiographic positioning of the anchor. There is also a disur to the ancher ver which the catheter is tracked to the desired location.
The FAD Introducer Needle is provided in the FAD Catheter Kit and will also be sold The TAD Introducer Neducer Needle is stainless steel and has a stainless steel stylet. The
#### CONFIDENTIAL
{2}------------------------------------------------
FAD Introducer Needle provides access to the intradiscal space. The Guidewire and Catheter are delivered through the FAD Introducer Needle.
## SUBSTANTIAL EQUIVALENCE:
The Functional Anaesthetic Discography (FAD) System consists of the FAD Catheter Kit and the FAD Introducer Needle.
## FAD Catheter Kit:
FAD Catheter Kit has substantially equivalent indications for use, and technological characteristics as the predicate devices which include:
- Spinal Specialties Discography Kit, K960082, cleared 3/21/96 .
- Spimal Specialites Discognify, 1023140, cleared 8/15/03 .
- Braun Perifix Catheter, K042488, cleared 9/30/04 .
The indications for use for the predicate devices are substantially equivalent to the proposed I he indications for use for the predicals derress in the technological characteristics
indications for use for the FAD System. Any differences in the technological of AD indications for use for the PAD Dystem vissues of safety or efficacy. Thus, the FAD Catheter Kit is substantially equivalent to the predicate devices.
## FAD Introducer Needle:
F AD Introducer Needle has substantially equivalent indications for use, and The I AD Introducer Prostics as the predicate devices which include:
- Braun Perifix Safety Epidural Needle, K013610, cleared 1/25/02 .
- Draum Formix Spinal Needle, K022029, cleared 7/30/02 .
- Pajunk Touhy Needles, K040965, cleared 9/7/04 .
The indications for use for the predicate devices are substantially equivalent to the proposed The indications for use for the predicate do recordinate. Any differences in the technological indications for use for the PAD introducer receinty new issues of safety or efficacy. Thus, characteristics between the devices do not rail ranswitzent to the predicate devices.
## TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION
Any differences in technological characteristics between the FAD System and the predicate Any differences in technological characteristics occircuit. The safety of the FAD System was devices do not false ally new issues of sales, or sales, or docompatibility testing, evaluated through design vermeation the FAD Catheter Kit and the FAD introducer I he collective results have demonstation that the respective predicate devices with regard to safety.
### SUMMARY
The FAD Catheter Kit and the FAD Introducer Needle are substantially equivalent to the predicate devices.
### CONFIDENTIAL
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 5 2005
InnoSpine, Incorporated C/O Ms. Laraine Pangelina Regulatory Consultant Experien Group, LLC 11240 Magdalena Road Los Altos Hills, California 94024
Re: K043500
Trade/Device Name: Functional Anaesthetic Discography (FAD) Catheter System Regulation Number: 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: II Product Code: BSP Dated: March 21, 2004 Received: March 24, 2005
Dear Ms. Pangelina:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Pangelina
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. or must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K043500
Device Name: Functional Anaesthetic Discography (FAD) Catheter System
Indications for Use:
The Functional Anaesthetic Discography (FAD) Catheter System is intended for use in delivering either a single dose or continuous administration of a radiopague contrast, local anaesthetics, and/or saline solution to the intradiscal space. The FAD Introducer Needle is intended for use to access the area adjacent to the intradiscal space for the purpose of facilitating placement of the FAD Catheter and Guidewire into the intradiscal space.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Clu Cs
y General Hospital,
fection Com...onic Devices
Number K043500
Page 1 of 1
(Posted November 13, 2003)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.