Pajunk's anesthesia conduction needles - Tuohy, Quincke, Chiba, and Crawford - are intended for the transient delivery of anesthetics to provide regional anesthesia or to facilitate placement of an epidural catheter.
Device Story
Anesthesia conduction needles (Tuohy, Quincke, Chiba, Crawford) consist of stainless steel cannula, stainless steel stylet, and plastic luer hub; used for regional anesthesia. Operated by professional anesthetists in clinical settings. Device facilitates transient delivery of anesthetics or placement of epidural catheters for continuous infusion. Cannula stabilized by inner stylet during puncture; stylet withdrawn upon reaching anatomical site. Introducer needle used for initial skin puncture. Single-use, sterile, disposable. Enables targeted anesthetic delivery; provides pain relief via catheter placement.
Clinical Evidence
Bench testing only. Biocompatibility testing performed; no clinical data provided.
Technological Characteristics
Stainless steel cannula and stylet; plastic luer hub. Various tip geometries: pencil-point (Sprotte), bent/sharp (Tuohy), sharp (Quincke), elongated sharp (Chiba), short (Crawford). Sterile, single-use, disposable. No energy source or software.
Indications for Use
Indicated for patients requiring regional anesthesia via transient delivery of anesthetics or placement of an epidural catheter for pain relief. Used by professional anesthetists.
Regulatory Classification
Identification
An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.
{0}------------------------------------------------
Revised Pages 54-56, August 2004
K040965
## Section 6.0
#### SEP - 7 2004 510(k) Summary of Safety and Effectiveness
This section contains the 510(k) Summary of Safety and Effectiveness. (This document can be copied and submitted to interested parties as required by 21CFR 807.92).
# 510(k) Summary of Safety and Effectiveness
#### Submitter's Information:
Pajunk Medical Technology GmbH Karl-Hall-Str. 1 78187 Geisingen Germany
# USA Contact:
Lynette Howard (Pajunk's USA agent) Lyle Howard Corporation 203 Main Street, 166 Flemington, NJ 08822 USA P 908.788.4580 F 908.788.4590 e-mail: Lylehoward@aol.com
#### Device Name:
Trade Names: Tuohy needle; Quincke needle; Chiba needle; Crawford needle Common Names: Anesthesia conduction needles, epidural/spinal needles
Classification Name: Anesthesia conduction needles (Reference, 21 CFR, 868.5150, April 1 2003)
Product Class: Class II
Product Code: BSP, MIA
Preparation date of summary: March 2. 2004
{1}------------------------------------------------
# Predicate Devices:
- 1. Becton Dickinson spinal needles: K953790 & K954953.
- 2. Pajunk's anesthesia conduction needles Tuohy, Quincke, Chiba, and Crawford consist of the same plastic luer hub and metal cannula materials as Pajunk's Sprotte needles cleared under 510(k) numbers K911202, K911221, K911260, and K923003. The only difference between these needle types is the needle tip, which is either pencil-point-tip (Sprotte), bent and sharp needle tip (Tuohy), sharp needle tip (Quincke), elongated sharp needle tip (Chiba), or short needle tip (Crawford).
# Device Description:
Anesthesia conduction needles consist of a luer hub, a stainless steel cannula with various tip types, and a stainless steel stylet. These needles are provided as sterile, single use, disposable devices. They may be packaged individually or included in regional anesthesia trays (kits). Anesthesia conduction needles fit into an introducer needle. This is a simple hypodermic needle to make the initial puncture through the skin to aid in the placement of the anesthesia conduction needle. The later can facilitate the placement of an epidural catheter for continuous infusion of local anesthetics into the epidural space for longer pain relief.
The Pajunk anesthesia conduction needles - Tuohy, Quincke, Chiba, and Crawford - are single use, sterile, non-pyrogenic and latex-free medical devices for transient delivery of anesthetics during regional anesthesia. The cannula is stabilized during puncture with use of an inner stylet (mandrin). This stylet is withdrawn after the anesthesia conduction needle has reached its anatomical site for regional anesthesia. Then the anesthetics can be applied transiently (i.e. within minutes) by the professional anesthetist. Alternatively or additionally, an evidural catheter may be placed through the anesthesia conduction needle. The needle is withdrawn and the epidural catheter tip may remain in the epidural space for pain treatment.
# Intended Use:
Pajunk's anesthesia conduction needles - Tuohy, Quincke, Chiba, and Crawford - are intended for the transient delivery of anesthetics to provide regional anesthesia or to facilitate placement of an epidural catheter.
{2}------------------------------------------------
# Technology Characteristics:
The Pajunk anesthesia conduction needles - Tuohy, Quincke, Chiba, and Crawford - have the same technological characteristics as the predicate devices identified above. The Pajunk anesthesia conduction needles - Tuohy, Quincke, Chiba, and Crawford - are equivalent in design, physical dimensions, luer hub, metal and plastics materials, and packaging to Pajunk's Sprotte needles cleared under 510(k) numbers K911202, K911221, K911260, and K923003.
Biocompatibility testing of Pajunk's anesthesia conduction needles is located in Section 7 of this submission.
#### Conclusion:
The majority of anesthesia conduction needles are pre-amendment devices. This is due to the fact that anesthesia conduction needles have been available to anesthesiologists for about 100 years. The comparison between the predicate devices and the proposed devices demonstrates that the proposed devices are safe and effective, as well as substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 7 2004
Pajunk GMBH C/O Ms. Lynette L. Howard Submission Corespondent Lyle Howard Corporation 203 Main Street, PMB 166 Flemington, New Jersey 08822
Re: K040965
Trade/Device Name: Pajunk Anesthesia Conduction Needles Regulation Number: 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: II Product Code: BSP Dated: August 16, 2004 Received: August 23, 2004
Dear Ms. Howard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Howard
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Chiu Lin, Ph.D.
hiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use - Revised
510(k) Number (if known):
Device Name: __ Pajunk Anesthesia Conduction Needles
Indications For Use:
Pajunk's anesthesia conduction needles - Tuohy, Quincke, Chiba, and Crawford - are F ajunk 3 anesthesia conduction nodales - Fabriy, provide regional anesthesia or to facilitate placement of an epidural catheter.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Aun A. Sytion
(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
510(k) Number: R040965
Page 1 of 1 _
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.