K102594 · Baxano, Inc. · BSO · Jan 19, 2011 · Anesthesiology
Device Facts
Record ID
K102594
Device Name
IO-FLEX CATHETER
Applicant
Baxano, Inc.
Product Code
BSO · Anesthesiology
Decision Date
Jan 19, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5120
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The iO-Flex Catheter is for continuous or intermittent preoperative, perioperative or post-operative delivery of local anesthetics and analgesics in the epidural space or near a nerve for regional anesthesia and pain management. The catheter may also be used perioperatively during open spinal procedures to irrigate, aspirate or inject a bolus of fluid or medication. The iO-Flex Catheter is designed for use with the iO-Flex Wire.
Device Story
iO-Flex Catheter; flexible catheter body with Y-adapter featuring rotating hemostasis valve (RHV); central port for guidewire insertion over Baxano iO-Flex Wire; side port for syringe connection; external depth markers on shaft. Used by clinicians during open spinal procedures or for regional anesthesia/pain management. Enables irrigation, aspiration, or bolus injection of fluids/medications at surgical site. Benefits patient through targeted delivery of anesthetics/analgesics and surgical site management.
Clinical Evidence
No clinical data. Bench testing included dimensional, flexibility/kink resistance, leak, guidewire compatibility, ink adhesion, flow rate, and tensile strength assessments. Biocompatibility testing performed per ISO-10993-1 (cytotoxicity, irritation, acute systemic toxicity, sensitization, pyrogenicity). Cadaver model study assessed ease of introduction, positioning, guidewire compatibility, irrigation, injection, aspiration, and introduction of hemostatic matrices.
Technological Characteristics
Catheter shaft with Y-adapter and rotating hemostasis valve. External depth markers. Biocompatible materials tested per ISO-10993-1. External communicating device for limited exposure (<24 hours). Sterilization via double ethylene oxide.
Indications for Use
Indicated for patients requiring regional anesthesia or pain management via continuous or intermittent delivery of local anesthetics/analgesics in the epidural space or near a nerve, and for irrigation, aspiration, or fluid/medication injection during open spinal procedures.
Regulatory Classification
Identification
An anesthesia conduction catheter is a flexible tubular device used to inject local anesthetics into a patient and to provide continuous regional anesthesia.
{1}------------------------------------------------
K102594
Image /page/0/Picture/1 description: The image shows the word "BAXANO" in a bold, sans-serif font. A stylized graphic element intersects the "X", resembling a curved line with a pointed end. The overall design is simple and modern, with a focus on the typography and the unique graphic element.
JAN 1 9 2011
# 510(K) SUMMARY (21 CFR 807.92) iO-Flex™ Catheter
510(k) Owner:
Baxano, Inc. 655 River Oaks Parkway San Jose, CA 95134 Tel: 408-514-2200 Fax: 408-514-2201
Contact Person:
Edward J. Sinclair Tel: 408-514-2220 E-mail: esinclair@baxano.com
Date Prepared: November 12, 2010
Trade Name: iO-Flex™ Catheter
Anesthesia Conduction Catheter Common Name:
Classification: Class II (21 CFR 868.5120)
BSO
Product Code:
I-Flow Corporation, ON-Q Introducers (K063234) Predicate Devices: B.Braun, Regional Anesthesia Catheters (K042488) Axiom Medical, Inc., MultiPurpose Wound Drain Catheter (K993592)
The iO-Flex Catheter is comprised of two primary components: (1) Device Description: Y-Adapter with a rotating hemostasis valve (RHV) and (2) a flexible catheter body. The Y-Adapter consists of two ports: (1) a central guidewire port with rotating hemostasis valve and (2) a side port for connection of a standard syringe for the irrigation, injection and / or aspiration of fluids at the surgical site. The central guidewire port with RHV allows for insertion of the iO-Flex Catheter over the Baxano iO-Flex Wire. The flexible catheter shaft contains external markers for depth adjustment by the clinician.
{2}------------------------------------------------
| Intended Use: | The iO-Flex Catheter is for continuous or intermittent preoperative,<br>perioperative or post-operative delivery of local anesthetics and<br>analgesics in the epidural space or near a nerve for regional<br>anesthesia and pain management. The catheter may also be used<br>perioperatively during open spinal procedures to irrigate, aspirate or<br>inject a bolus of fluid or medication.<br><br>The iO-Flex Catheter is designed for use with the iO-Flex Wire. |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics: | The Baxano iO-Flex Catheter is similar in use and has similar<br>technological characteristics to the I-Flow Corporation ON-Q<br>Introducer, the B. Braun Inc. Regional Anesthesia Catheters and<br>Axiom Medical Inc. MultiPurpose Wound Drain Catheter. All<br>devices incorporate a catheter shaft and a mechanism similar to the<br>iO-Flex Catheter Y-adapter for the irrigation, aspiration or injection<br>of fluids or medications. |
| Non-Clinical<br>Performance Data: | Bench performance testing was conducted to demonstrate the<br>functional performance of the iO-Flex Catheter and to ensure that<br>the requirements of the product specification were met. iO-Flex<br>Catheters were subjected to double ethylene oxide sterilization,<br>environmental conditioning, and transportation testing prior to<br>undergoing packaging, bench performance and functional testing.<br>Bench performance testing included dimensional, flexibility / kink<br>resistance, leak, guide wire compatibility, ink adhesion, flow rate<br>and tensile strength assessments. Additionally, the packaging and<br>device characteristics were tested after accelerated aging to support<br>shelf-life labeling. The results of the bench testing demonstrated<br>that the iO-Flex Catheter satisfied all of the dimensional and<br>functional performance requirements, which are designed to ensure<br>that the device is safe and effective for its intended use.<br><br>To verify the biocompatibility of all components and materials for<br>the iO-Flex Catheter, biocompatibility testing was conducted<br>pursuant to FDA's Guidance Document (#G95-1), Use of<br>International Standard ISO-10993-1, "Biological Evaluation of<br>Medical Devices Part 1: Evaluation and Testing (1995)".<br>Biocompatibility testing was conducted in accordance with the GLP<br>regulations using polar and nonpolar solvents. According to the<br>FDA guidance and ISO standards, the iO-Flex Catheter, is an<br>"external communicating device" in contact with tissue and bone<br>for limited exposure (<24 hour) during spinal procedures. Based on<br>these characteristics, the following biocompatibility tests were<br>conducted: Cytotoxicity, Irritation / Acute Intracutaneous<br>Reactivity, Acute Systemic Toxicity, Sensitization / Maximization<br>and Pyrogenicity. Based on the test results provided, the iO-Flex<br>Catheter has been demonstrated to be biocompatible for its intended<br>use. |
Design validation testing was conducted to confirm that the iO-Flex Catheter design meets the requirements of the product specification.
{3}------------------------------------------------
Testing was conducted using a cadaver model. The following assessments were performed: Ease of introduction and positioning, guide wire compatibility, irrigation and injection of fluids as well as aspiration. The ability of the catheter to introduce commerciallyavailable hemostatic matrices was validated. The results demonstrate that the iO-Flex Catheter satisfies all of the functional performance requirements, which are designed to ensure that the device is safe and effective for its intended use.
The iO-Flex Catheter is substantially equivalent to the I-Flow Substantial Corporation ON-Q Introducers (K063224), B. Braun Regional Equivalence: Anesthesia Catheters (K042488), and the Axiom Medical MultiPurpose Wound Drain Catheter (K993592) in terms of its intended use / indications for use, as well as technological characteristics and principles of operation.
> The iO-Flex Catheter has the same indications for use and fundamental scientific technology as the predicate devices. Based upon the indications for use, technological characteristics and performance test results, iO-Flex Catheter does not raise any new questions of safety or effectiveness.
Baxano, Inc. has determined based on bench performance, Conclusions: biocompatibility and cadaver studies that the iO-Flex Catheter conforms to its design specifications and is substantially equivalent to the predicate devices.
Any statement regarding "substantial equivalence" made in this 510(k) submission and summary only relates to whelher the product addressed in this submission may be lawfully marketed without premarket approval or reclassification, and is not interpreted as un admission or any other type of evidence in any patent proceeding, including patent infringement, litigation or proceeding before any Patent Office. The present submission and statements therein therefore should not be construed as affecting or relating to the scope of any patent application, or to whether the product addressed in this submission, or its use, may be considered indistinct, from a patentability perspective, from any of the other devices referenced in this filing.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three wavy lines, which are positioned to the right of the text. The text is arranged in a circular pattern around the left side of the graphic. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Edward J. Sinclair Vice President, Clinical, Regulatory and Quality affairs Baxano, Incorporated 655 River Oaks Parkway San Jose. California 95134
## JAN 1-9 2011
Re: K102594
Trade/Device Name: Baxano iO-Flex Catheter Regulation Number: 21 CFR 868.5120 Regulation Name: Anesthesia Conduction Catheter Regulatory Class: II Product Code: BSO Dated: December 13, 2010 Received: December 14, 2010
Dear Mr. Sinclair:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2- Mr. Sinclair
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Rh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Baxano iO-Flex Catheter
Indications For Use:
The iO-Flex Catheter is for continuous or intermittent preoperative, perioperative or postoperative delivery of local anesthetics and analgesics in the epidural space or near a nerve for regional anesthesia and pain management. The catheter may also be used perioperatively during open spinal procedures to irrigate, aspirate or inject a bolus of fluid or medication.
The iO-Flex Catheter is designed for use with the iO-Flex Wire.
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital Division of Anesthouselongy.
Infection Control, Dental Devicesage 1 of 1
KIS 2594
510(k) Number: CONFIDENTIAL
Baxano, Inc.
Page 5-2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.