← Product Code [BSO](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSO) · K063234

# ON-Q INTRODUCERS (K063234)

_I-Flow Corp. · BSO · Nov 21, 2006 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSO/K063234

## Device Facts

- **Applicant:** I-Flow Corp.
- **Product Code:** [BSO](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSO.md)
- **Decision Date:** Nov 21, 2006
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 868.5120
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology

## Indications for Use

The ON-Q Introducers are intended for the percutaneous introduction and placement of catheters into or around surgical wound sites and/or close proximity to nerves. Introducers with a luer hub may be used to aspirate or inject a bolus of fluid or medication prior to placing the catheters.

## Device Story

ON-Q Introducers facilitate percutaneous catheter placement near surgical wounds or nerves. System comprises two components: stainless steel trocar (11-17 GA, 3.25-12 inches) with sharp/beveled or blunt/rounded tip; and peelable plastic T-handle sheath. Blunt-tip versions (Tunnelers) may be reusable; sharp-tip versions (Introducer Needles) are single-use. Luer hub models allow fluid aspiration or medication bolus injection prior to catheter placement. Used by clinicians in surgical or clinical settings to assist in regional anesthesia or wound site pain management. Device provides mechanical access path for catheter insertion; benefits patient by enabling precise delivery of therapeutic agents.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Stainless steel trocar (ASTM ISO 9626:1991); peelable plastic sheath. Conical 6% luer fittings (ISO 594-1:1986, ISO 594-2:1998). Sterile, single-use (except some reusable Tunnelers). Design meets ISO 10555-1:1995 and ISO 10555-5:1996 requirements for intravascular catheters.

## Regulatory Identification

An anesthesia conduction catheter is a flexible tubular device used to inject local anesthetics into a patient and to provide continuous regional anesthesia.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K063234

ON-Q Introducers Section 5 - Summary of Safety and Effectiveness NOW 2 1 2006

## 510(K) - SUMMARY OF SAFETY AND EFFECTIVENESS

| Submitter:                | I-Flow Corporation<br>20202 Windrow Drive<br>Lake Forest, CA. 962630                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                  | Shane Noehre<br>Director, Regulatory Affairs<br>I-Flow Corporation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Trade Names:              | ON-Q Introducer, ON-Q Needle, ON-Q Tunneler, ON-Q Sheath                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Common Name:              | Catheter Introducer Needle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Device Description:       | The ON-Q Introducers consist of the following two components:<br>The first component is a stainless steel shaft (trocar) with a handle or                     luer hub. The trocar diameter ranges from 11 to 17 GA and the length                     ranges from 3.25 to 12 inches. The trocar may have a sharp, beveled                     tip or a blunt, rounded tip. The second component is a peelable (split T-Handle) plastic sheath                     that is designed to fit over the trocar.              The model configurations are designated as follows:  Models with a blunt tip are called ON-Q Tunnelers. The sheaths may                     be sold separately and are single use only. Some versions of the                     ON-Q Tunnelers may be reusable. Models with a sharp tip are called ON-Q Introducers Needles which                     are single use only.              The ON-Q Introducers are intended for the percutaneous introduction and             placement of catheters into or around surgical wound sites and/or close             proximity to nerves. Introducers with a luer hub may be used to aspirate             or inject a bolus of fluid or medication prior to placing the catheters. |
| Standards:                | The ON-Q Introducers meet the requirements of the following FDA             recognized consensus standards for the device design and performance             requirements:  ISO 10555-5:1996, Sterile, Single-Use Intravascular Catheters -                     Over Needle Peripheral Catheters ISO 10555-1:1995, Sterile, Single-Use Intravascular Catheters -                     General Requirements ISO 594-1:1986, Conical Fittings with 6% (Luer) Taper - Part 1 ISO 594-2:1998, Conical Fittings with 6% (Luer) Taper - Part 2 ISO 9626:1991, Stainless Steel Needle Tubing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Technology<br>Comparison: | The ON-Q Introducers use the same technology as legally marketed<br>predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Conclusion:               | The ON-Q Introducers are substantially equivalent to existing introducers<br>that are currently marketed.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Shane Noehre Director, Regulatory Affairs I-Flow Corporation 20202 Windrow Drive Lake Forest, California 92630

NOV 2 1 2006

Re: K063234

Trade/Device Name: On-Q Introducers Regulation Number: 868.5120 Regulation Name: Anesthesia Conduction Catheter Regulatory Class: II Product Code: BSO Dated: October 24, 2006 Received: October 25, 2006

Dear Mr. Noehre:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Mr. Noehre

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Cuy Cholionn for

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

K063234 510(k) Number (if known):

ON-Q Introducers Device Name:

Indications For Use:

- 1. The ON-Q Introducers are intended for the percutaneous introduction and placement of catheters into or around surgical wound sites and/or close proximity to nerves. Introducers with a luer hub may be used to aspirate or inject a bolus of fluid or medication prior to placing the catheters.
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Reds C. Alopez L. Amm 11/21/2006

for Anesthesiology, General Hospital,
Control, Dental Devices

K063234

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSO/K063234](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSO/K063234)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
