The AG100s is intended to monitor and control the cuff pressures of endotracheal tube (ETT) or tracheostomy tube (TT) and to evacuate secretions from the subglottic space above the cuff during mechanical ventilation. 1. When used with a standard ETT or TT, the device automatically maintains the cuff pressure as pre-set by the user. 2. When used with ETT or TT with suction line, the device automatically maintains the cuff pressure as pre-set by the user and performs intermittent evacuation of subglottic secretions from above the tube's cuff. 3. When used with ETT with suction and venting lines (e.g., AG ETT), the device automatically maintains the cuff pressure as pre-set by the user, or automatically adjusts the cuff pressure based on monitoring of the Carbon Dioxide (CO2) concentration above the cuff. In addition, it performs evacuation, or rinsing and evacuation of subglottic secretions from above the tube's cuff.
Device Story
AG100s is a control unit for managing endotracheal/tracheostomy tube cuff pressures and subglottic secretions. System inputs include CO2 concentration levels (via dedicated tube lumen) and user-defined pressure setpoints. The device uses a CO2 analyzer, peristaltic pump, and vacuum regulator to maintain cuff pressure, evacuate secretions, and perform rinsing. Outputs include automated cuff pressure adjustments and secretion removal. Used in clinical settings during mechanical ventilation; operated by healthcare providers. Benefits include reduced aspiration risk, maintained ventilation, and minimized tracheal tissue damage from excessive cuff pressure. The device connects to airway tubes via a connection kit/harness; includes a secretions canister and antibacterial filter.
Clinical Evidence
Bench testing only. Performance data included comparative cuff pressure control and suction performance testing when connected to ETTs and TTs, software system verification and validation, and usability validation.
Technological Characteristics
System includes control unit, connection kit, secretions canister, and antibacterial filter. Features CO2 analyzer module, peristaltic pump for rinsing, and vacuum regulator for suction. Cuff pressure control range: 10-50 mmHg. Suction pressure range: -20 to -120 mmHg. Power: 100-240 V with backup battery. Closed system design. Materials are biocompatible per ISO 10993-1. Software-controlled.
Indications for Use
Indicated for adult patients requiring mechanical ventilation via endotracheal tube (ETT) or tracheostomy tube (TT) for airway management, cuff pressure control, and subglottic secretion evacuation.
Regulatory Classification
Identification
An inflatable tracheal tube cuff is a device used to provide an airtight seal between a tracheal tube and a patient's trachea.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 29, 2022
Hospitech Respiration Ltd. % Bosmat Friedman Regulatory Consultant ProMedoss, Inc. 3521 Hatwynn Rd. Charlotte, North Carolina 28269
Re: K221477
Trade/Device Name: AG100s Regulation Number: 21 CFR 868.5750 Regulation Name: Inflatable Tracheal Tube Cuff Regulatory Class: Class II Product Code: BSK Dated: May 18, 2022 Received: May 23, 2022
Dear Bosmat Friedman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Ethan L. Nyberg -S
for James Lee, PhD Division Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K221477
Device Name AG100s
Indications for Use (Describe)
The AG100s is intended to monitor and control the cuff pressures of endotracheal tube (ETT) or tracheostomy tube (TT) and to evacuate secretions from the subglottic space above the cuff during mechanical ventilation.
1. When used with a standard ETT or TT, the device automatically maintains the cuff pressure as pre-set by the user. 2. When used with ETT or TT with suction line, the device automatically maintains the cuff pressure as pre-set by the user and performs intermittent evacuation of subglottic secretions from above the tube's cuff.
3. When used with ETT with suction and venting lines (e.g., AG ETT), the device automatically maintains the cuff pressure as pre-set by the user, or automatically adjusts the cuff pressure based on monitoring of the Carbon Dioxide (CO2) concentration above the cuff. In addition, it performs evacuation, or rinsing and evacuation of subglottic secretions from above the tube's cuff.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(K) SUMMARY
[as required by section 807.92(c)] AG100s 510(k) Number K221477
#### 5.1 SUBMITTER
#### Applicant's Name:
Hospitech Respiration Ltd. 15 Atir Yeda Kfar-Saba, 4464312, Israel TEL: 972-3-919-1648 FAX: 972-3-919-1647
#### Contact Person:
Bosmat Friedman Regulatory Affairs Consultant 3521 Hatwynn Rd. Charlotte, NC 28269 Phone: 980-308-1636 bosmat.f(@promedoss.com
#### Date Prepared:
September 14, 2022
#### 5.2 DEVICE
Trade Name: AG100s
Classification Code: Classification Name: Cuff, Tracheal Tube, Inflatable Product Code: BSK Regulation No: 868.5750 Class: 2 Review Panel: Anesthesiology
#### 5.3 PREDICATE DEVICES
Predicate Device 1 (Primary)
- AG100s, by Hospitech Respiration Ltd., Product code BSK cleared Under: ● K180991.
Predicate Device 2 (Secondary)
- IntelliCuff, manufactured by Hamilton Medical AG., cleared under K150893; . Product Code: BSK.
#### 5.4 DEVICE DESCRIPTION
The AG100s system is comprised of the following main components:
{5}------------------------------------------------
- The AG100s control unit о
- The AnapnoGuard connection kit/harness (AG Connection Kit) O connecting a cuffed airway to the AG100s control unit.
Additional device components include cart, secretions canister (Trap Bottle), rinsing fluid (saline) bag and antibacterial air filter.
When the AG100s, is connected to AG ETT, it monitors leaks between endotracheal cuff and the trachea by measuring the Carbon Dioxide levels in the subglottic area above the cuff through a dedicated lumen in the endotracheal tube. Detection of a high level of Carbon Dioxide is an objective indicator for a leak (improper sealing of the trachea by the AG ETT). The system continuously monitors and adjusts the cuff pressure to prevent a leak at minimum possible pressure (all within pressure limits preset by the user). Preventing a leak reduces the likelihood of aspiration of secretions from the upper airways into the lungs and increases the likelihood for no loss of ventilation and delivery of anesthetic and nebulized drugs into the lungs. Keeping the cuff pressure as low as possible reduces the mechanical pressure of the cuff on the tracheal tissue throughout the intubation period. The system also performs evacuation of secretions from above the endotracheal tube's cuff through a dedicated lumen at the dorsal side of the endotracheal tube.
The AG100s also performs cuff pressure monitoring and control in standard FDA cleared TTs and ETTs; depending on the tube type, the system may also perform secretions removal.
#### 5.5 INDICATIONS FOR USE
The AG100s is intended to monitor and control the cuff pressures of endotracheal tube (ETT) or tracheostomy tube (TT) and to evacuate secretions from the subglottic space above the cuff during mechanical ventilation.
- When used with a standard ETT or TT, the device automatically maintains 1. the cuff pressure as pre-set by the user.
- When used with ETT or TT with suction line, the device automatically 2. maintains the cuff pressure as pre-set by the user and performs intermittent evacuation of subglottic secretions from above the tube's cuff.
- 3. When used with ETT with suction and venting lines (e.g., AG ETT), the device automatically maintains the cuff pressure as pre-set by the user, or automatically adjusts the cuff pressure based on monitoring of the Carbon Dioxide (CO2) concentration above the cuff. In addition, it performs evacuation, or rinsing and evacuation of subglottic secretions from above the tube's cuff.
#### 5.6 SUBSTANTIAL EQUIVALENCE
The AG100s is substantially equivalent to the predicate device based on the following:
#### Intended Use
The intended use of the proposed device remains unchanged. The indication for use has been modified to include the ability to connect to TT as well as to provide
{6}------------------------------------------------
clarifications regarding device functionality when connected to various airway tubes.
## Technology
The proposed device modifications introduce the ability to connect the system to TT through minor GUI modifications. All changes have been validated demonstrating that the device functions as intended. Testing demonstrates that the modified AG100s functions in an equivalent manner as the previously cleared AG100s.
#### Discussion
The modified AG100s has the same intended use as the previously cleared AG100s, and a slightly modified indications for use statement (for clarification purposes). The main technological difference between the devices is the ability to connect the AG100s to a TT as well as ETT.
Verification and Validation testing demonstrated that the revised device is substantially equivalent to the previously cleared AG100s. Consequently, the AG100s is as safe and effective as its predicates without raising any new safety and/or effectiveness concerns.
| Characteristics | Subject Device<br>AG100s System | Primary Predicate<br>AG100s (K180991) | Comments |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Hospitech Respiration, Ltd. | Hospitech Respiration, Ltd. | Same |
| 510(k) Number | TBD | K180991 | -- |
| Product Code(s) | BSK | BSK | Same |
| Regulation | 868.5750 | 868.5750 | Same |
| Class | 2 | 2 | Same |
| Indications for Use | The AG100s is intended to monitor and control the<br>cuff pressures of endotracheal tube (ETT) or<br>tracheostomy tube (TT) and to evacuate secretions<br>from the subglottic space above the cuff during<br>mechanical ventilation.<br>a. When used with a standard ETT or TT, the<br>device automatically maintains the cuff<br>pressure as pre-set by the user.<br>b. When used with ETT or TT with suction line,<br>the device automatically maintains the cuff<br>pressure as pre-set by the user and performs<br>intermittent evacuation of subglottic<br>secretions from above the tube's cuff.<br>c. When used with ETT with suction and<br>venting lines (e.g., AG ETT), the device<br>automatically maintains the cuff pressure as<br>pre-set by the user, or automatically adjusts<br>the cuff pressure based on monitoring of the<br>Carbon Dioxide (CO2) concentration above<br>the cuff. In addition, it performs evacuation,<br>or rinsing and evacuation of subglottic<br>secretions from above the tube's cuff. | AG100s is intended for<br>airway management by<br>oral/nasal intubation while<br>providing continuous<br>endotracheal cuff pressure<br>control using non-invasive<br>measurement and monitoring<br>of carbon dioxide<br>concentration in the subglottic<br>space and evacuation of<br>secretions from above the<br>endotracheal tube's cuff. | Same overall intended use.<br>Indications for use slightly<br>modified to include the ability to<br>connect to TT as well as provide<br>clarity regarding device<br>functionality with various tubes.<br>Changes have been validated via<br>performance testing. No new<br>question of safety and<br>effectiveness. |
| Indications for single<br>patient use | No | No | Same |
| Allows tube replacement<br>without disconnecting<br>patient from ventilator | Yes, both TT and ETT | Yes, ETT only | Similar, in predicate device TT<br>was not used. |
| Patient Population | Adults | Adults | Same |
| Biocompatibility | No new patient contacting materials | All materials that come in<br>contact with the patient body | Same |
#### Substantial Equivalency Table
{7}------------------------------------------------
| Characteristics | Subject Device<br>AG100s System | Primary Predicate<br>AG100s (K180991) | Comments |
|------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | or liquids are biocompatible<br>and compliant with ISO<br>10993-1 | |
| Latex Free | Yes | Yes | Same |
| Power Supply | 100-240 V with backup battery | 100-240 V with backup<br>battery | Same |
| CO2 analyzer module | CO2analyzer measures the CO2levels in the air<br>coming from the subglottic space above the ETT<br>cuff and cuff pressure inflate/deflate accordingly. | CO2analyzer measures the<br>CO2levels in the air coming<br>from the subglottic space<br>above the ETT cuff and cuff<br>pressure inflate/deflate<br>accordingly. | Same |
| Rinse module | Includes peristaltic pump and sensors with closed<br>loop control on the saline volume. | Includes peristaltic pump and<br>sensors with closed loop<br>control on the saline volume. | Same |
| Cuff pressure control<br>module | Regulate the cuff pressure according to a pre-<br>determined set point. | Regulate the cuff pressure<br>according to a pre-determined<br>set point. | Same |
| Cuff Pressure Control<br>Range | 10-50mmHg<br>(13.6-68 mmH2O) | 10-50mmHg<br>(13.6 - 68 mmH2O) | Same |
| Vacuum regulator module<br>(suction module) | Regulates the vacuum level during suction<br>procedure. | Regulates the vacuum level<br>during suction procedure. | Same |
| Suction Pressure Range | Subglottic suction: -20 up to -120 mmHg | Subglottic suction: -20 up to -<br>120 mmHg | Same |
| Mode of operation | Manual Intermittent | Manual Intermittent | Same |
| Closed System | Yes | Yes | Same |
| Manual control of vacuum | Yes | Yes | Same |
| Evacuation of secretions<br>from above the<br>endotracheal tube's cuff | Yes | Yes | Same |
| Flow Rate | 0 to 12 L/min | 0 to 12 L/min | Same |
| General suction | Not available | Yes | Different; the general suction<br>option was removed as it was<br>not regularly used by end user.<br>No impact on device safety or<br>effectiveness. |
#### 5.7 PERFORMANCE DATA
In order to support the proposed modifications, the following tests were conducted:
- Comparative AG100s Cuff pressure control + suction performance when । connected to ETTs & TTs
- Software System V&V Test l
- -AG100s Usability Validation
# 5.8 CONCLUSION
Hospitech Respiration has demonstrated that the AG100s is substantially equivalent to the predicate device. Differences between the AG100s and the predicate device do not raise new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.