The Denver® PLEURX Pleural Catheter Kit and Denver® PLEURX Home Drainage Kit are indicated for intermittent,longterm drainage of symptomatic, recurrent, malignant pleural effusion. The devices are indicated for the palliation of dyspnea due to malignant pleural effusion and for providing pleurodesis.
Device Story
The Denver® PLEURX Pleural Catheter Kit is a sterile, single-use system for managing malignant pleural effusions. It consists of a flexible, fenestrated silicone catheter with a cuff and valve, placed in the pleural space via the Seldinger technique; the cuff is tunneled subcutaneously. The Home Drainage Kit serves as an accessory for patient-managed drainage, utilizing vacuum bottles to collect pleural fluid. The system allows for intermittent, ambulatory drainage, providing relief from dyspnea and facilitating pleurodesis. The device is used in clinical and home settings; operation involves healthcare providers for placement and patients or caregivers for drainage. By enabling controlled, intermittent removal of fluid, the device improves patient quality of life and daily activity capacity.
Clinical Evidence
Clinical evaluation under IDE #930085. Subjects showed significant improvement in dyspnea (Friedman test) and quality of life (t-test) compared to baseline. Approximately 50% of subjects achieved pleurodesis; no subjects developed symptomatic recurrence of effusion during the study.
Technological Characteristics
Materials: Silicone catheter with radiopaque stripe. Design: Fenestrated catheter with cuff and valve. Energy source: Vacuum-based drainage. Connectivity: None. Sterilization: Single-use, sterile. Software: None.
Indications for Use
Indicated for patients with symptomatic, recurrent, malignant pleural effusion requiring intermittent, long-term drainage for palliation of dyspnea and to provide pleurodesis.
Regulatory Classification
Identification
A patient care suction apparatus is a device used with an intrathoracic catheter to withdraw fluid from the chest during the recovery period following surgery.
Predicate Devices
Argyle® Silicone Thoracic Catheter
Denver® Pleural Effusion Shunt with Externalized Pump Chamber
Submission Summary (Full Text)
{0}------------------------------------------------
K971753
Summary of Safety and Effectiveness Denver® PLEURX Pleural Catheter Kit Denver® PLEURX Home Drainage Kit
JUN 2 7 1997
SUBMITTER INFORMATION
Denver Biomaterials, Inc. 14998 W. 6th Avenue, Bldg. E700 Golden, Colorado 80401 USA
Lynne Leonard Director, Regulatory Affairs/Quality Assurance
DEVICE COMMON NAME
Patient Care Suction Apparatus
DEVICE CLASSIFICATION NAME
21 CFR 870.5050
IDENTIFICATION OF SUBSTANTIALLY EQUIVALENT DEVICES
The devices are substantially equivalent to the Argyle® Silicone Thoracic Catheter in intended use. They differ in that the Argyle® Silicone Thoracic Catheter is intended for continuous drainage, while the PLEURX Catheter is intended for intermittent, ambulatory drainage.
The devices are substantially equivalent to the Denver® Pleural Effusion Shunt with Externalized Pump Chamber in materials and dimensional specifications. They differ in that the Denver® Pleural Effusion Shunt with Externalized Pump Chamber transfers pleural fluid to the peritoneal cavity, while the PLEURX Catheter transfers fluid to a vacuum source.
Differences between the devices and their predicate devices should not affect their safety or effectiveness.
DEVICE DESCRIPTION:
The Denver® PLEURX Pleural Catheter Kit (#50-7000) is a sterile device, for single use only, and is not to be resterilized. The kit contains the PLTURX Catheter; a specifically matched drainage line that can be configured for either wall suction or a vacuum bottle; and the components needed to place the catheter within the pleural space and to dress the catheter exit site after placement .
{1}------------------------------------------------
The PLEURX Catheter contained in the kit is a flexible, fenestrated silicone catheter with a cuff and a valve. The fenestrated portion of the catheter is inserted into the pleural space utilizing a peel-away introducer (Seldinger technique). The portion of the catheter containing the cuff is tunneled subcutaneously, and the remaining portion is left external to the A radiopaque stripe runs the entire length of the body . . catheter.
The Denver® PLEURX Home Drainage Kit (#50-7500), an accessory to the Denver® PLEURX Pleural Catheter Kit, is a sterile, disposable device designed for single use only, and is not to be The kit contains drainage lines, vacuum bottles, resterilized. and the components needed to drain the effusion and to redress the catheter exit site.
### INDICATED USE
The devices are indicated for intermittent, long-term drainage of symptomatic, recurrent, malignant pleural effusion. They are indicated for the palliation of dyspnea due to malignant pleural effusion, and for providing pleurodesis.
BIOCOMPATABILITY TESTING:
The biological safety of the Denver® PLEURX Pleural Catheter Kit has been assured through the selection of materials which are currently used in legally marketed devices.
CLINICAL TESTING:
The Denver® PLEURX Pleural Catheter Kit and Denver® PLEURX Home Drainage Kit were evaluated clinically under IDE #930085.
- Effectiveness of PLEURX Catheter for Relief of Dyspnea 1.
When compared to baseline, subjects showed a significant improvement in their rating of dyspnea after treatment with the PLEURX Catheter, as determined by the Friedman test.
#### Effectiveness of PLEURX Catheter for Improving the 2. Quality of Life
Subjects experienced a significant improvement in quality of life after treatment with the PLEURX Catheter, as determined by the Although the subjects had serious malignant disease, t-test. they experienced an improved ability to carry out daily activities when their pleural effusions were controlled.
#### Effectiveness of PLEURX Catheter for Resulting in 3. Pleurodesis
Approximately half of the subjects treated with the PLEURX Catheter achieved pleurodesis while in the study. No subjects in the PLEURX Catheter arm developed a documented symptomatic recurrence of their effusion.
{2}------------------------------------------------
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Lynne Leonard Director, Regulatory Affairs and Quality Assurance Denver Biomaterials, Inc. 14998 W. 6th Avenue, Bldg. E 700 Golden, Colorado 80401
JUN 27 1397
K971753 Re: Denver® PLEURX Pleural Catheter Kit and Denver® PLEURX Home Drainage Kit Regulatory Class: II (two) Product Code: 74 DWM Dated: May 8, 1997 Received: May 12, 1997
Dear Ms. Leonard:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510 (k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Please note: If you purchase your device components in Amendments. bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510 (k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to additional controls. Existing major regulations affecting your device
{3}------------------------------------------------
# Page 2 - Ms. Lynne Leonard
can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, FDA will verify such assumptions. Failure to comply with the CMP regulation may result in requlatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
FDA notes that your device will contain sutures for which you have provided evidence that the suture characteristics are not altered by the sterilization process used for the device. However, you should be aware of the following additional information regarding the inclusion of a suture as a component of your device:
- The labeling, packaging and method of sterilization of 1. the suture cannot be changed without prior notification, review and clearance by FDA.
- 2. The supplier of the sutures used in your device can not be changed without prior notification, review and clearance by FDA.
In addition, we have determined that your device kit contains lidocaine which is subject to regulation as a drug.
Our substantially equivalent determination does not apply to the drug component of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug component. For information on applicable Agency requirements for marketing this drug, we suggest you contact:
> Director, Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0063
{4}------------------------------------------------
Page 3 - Ms. Lynne Leonard
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Fart 801 and additionally 809 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act, may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Atora J. Callaba
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
| 510 (K) Number: | K 971753 |
|----------------------|-----------------------------------------------------------------------------|
| Device Names: | Denver® PLEURX Pleural Catheter Kit,<br>Denver® PLEURX Home Drainage Kit |
| Indications for Use: | The Denver® PLEURX Pleural Catheter Kit<br>and Denver® PLEURX Home Drainage |
and Denver® PLEURX Home Drainage Kit are indicated for intermittent,longterm drainage of symptomatic, recurrent, malignant pleural effusion. The devices are indicated for the palliation of dyspnea due to malignant pleural effusion and for providing pleurodesis.
Concurrence of CDRH, Office of Device Evaluation (ODE)
ui. Daguinal m.D.
6/26/27
Prescription Use: Yes OR
(Per 21 CFR 801.109)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Over-the-Counter Use: No
----------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.