The Rocket Indwelling Pleural Catheter (IPC) Insertion Kit (Model # R51400-16-PL and R51400-MI-PL) is indicated for intermittent, long-term drainage of symptomatic, recurrent, pleural effusion, including malignant pleural effusion and other recurrent effusions that do not respond to medical management of underlying disease. The device is indicated for: - The palliation of dyspnea due to pleural effusion - Providing pleurodesis (resolution of the pleural effusion) This device should only be used by or under the supervision of trained personnel and in conjunction with current clinical practice guidelines.
Device Story
Rocket IPC Insertion Kit provides fenestrated silicone catheter for intermittent, long-term drainage of pleural effusions; includes drainage line, accessories for insertion/dressing, and optional lidocaine/DuraPrep. Catheter features polyester cuff for tissue attachment and one-way silicone valve to prevent backflow of air/fluid into pleural space. Used in clinical settings by trained personnel (surgeons/physicians). Catheter connects to vacuum bottles or wall suction via drainage line. Output is pleural fluid drainage; clinical benefit includes symptom relief (palliation of dyspnea) and pleurodesis. Kit provided sterile, single-use.
Clinical Evidence
Bench testing only. No clinical data. Predicate performance data (resistance to deformation, force at break, leakage) and biocompatibility (ISO 10993-1) are leveraged. Lidocaine HCl tested post-EtO sterilization per USP monograph; DuraPrep validated for EtO cycles.
Indicated for patients with symptomatic, recurrent pleural effusion (including malignant) unresponsive to medical management. Used for palliation of dyspnea and to provide pleurodesis. For use by or under supervision of trained personnel.
Regulatory Classification
Identification
A patient care suction apparatus is a device used with an intrathoracic catheter to withdraw fluid from the chest during the recovery period following surgery.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's logo, which consists of a stylized caduceus with three intertwined snakes and a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the logo in a circular fashion.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 17, 2017
Rocket Medical PLC Tracy Charlton Regulatory Affairs Manager 2-4 Sedling Rd, Wear Industrial Estate Washington, NE38 9BZ, GB
Re: K163321
Trade/Device Name: Rocket Indwelling Pleural Catheter (IPC) Insertion Kit Regulation Number: 21 CFR 870.5050 Regulation Name: Patient Care Suction Apparatus Regulatory Class: Class II Product Code: DWM Dated: July 20, 2017 Received: July 21, 2017
Dear Tracy Charlton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Tara A. Ryan -S
for Lori Wiggins Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known)
#### K163321
Device Name
Rocket Indwelling Pleural Catheter (IPC) Insertion Kit
Indications for Use (Describe)
The Rocket Indwelling Pleural Catheter (IPC) Insertion Kit (Model # R51400-16-PL and R51400-MI-PL) is indicated for intermittent, long-term drainage of symptomatic, recurrent, pleural effusion, including malignant pleural effusion and other recurrent effusions that do not respond to medical management of underlying disease.
The device is indicated for:
- · The palliation of dyspnea due to pleural effusion
- · Providing pleurodesis (resolution of the pleural effusion)
This device should only be used by or under the supervision of trained personnel and in conjunction with current clinical practice guidelines.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Section 5 – 510(k) Summary
| Submitter: | Rocket Medical PLC<br>2-4 Sedling Road, Wear Industrial Estate<br>Washington, Tyne and Wear<br>NE38 9BZ UK |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Tracy Charlton, Regulatory Affairs Manager<br>Phone: 00 44 191 419 4488 Ext 222<br>Fax: 00 44 191 416 5693<br>tracy@rocketmedical.com |
| Date Prepared: | August 1, 2017 |
| Trade Name: | Rocket Indwelling Pleural Catheter (IPC) Insertion Kit |
| Classification: | Class II<br>Patient Care Suction Apparatus<br>21 CFR 870.5050 |
| Product Code: | DWM |
| Predicate<br>Device(s): | The subject device is equivalent to the following devices:<br>Predicate: Rocket Indwelling Pleural Catheter (IPC) Insertion Kit;<br>O<br>(Rocket Medical PLC, K123033)<br>Reference Predicate: Rocket Indwelling Peritoneal Catheter (IPC)<br>O<br>Insertion Kit; (Rocket Medical PLC, K162457). This device kit<br>components and accessories are identical to the subject device.<br>Therefore, performance is identical to the subject device. |
| Device<br>Description: | Model R51400-16-PL IPC Insertion Kit consists of the IPC, drainage line<br>and accessories for insertion, attachment and dressing. In addition, it<br>contains currently licensed 1% lidocaine (5 mL x 3) and DuraPrep (6 mL<br>x 2). Model R51400-MI-PL is the Mini Kit with fewer accessories and no<br>drugs. It is for use by surgeons where the patients are already draped and<br>insertion components are readily available. The Rocket IPC Insertion Kits<br>are provided sterile and are for single use only.<br>The Rocket Indwelling Pleural Catheter (IPC) is a fenestrated silicone<br>catheter, with a barium sulfate stripe through its length, intended for the<br>drainage of pleural effusion. There is a polyester cuff for attachment to the<br>patient and a silicone one-way valve to prevent air and fluid from<br>migrating back into the pleural space. This valve can only be operated<br>with the specifically designed drainage line and drainage bottles supplied<br>by Rocket Medical. |
{5}------------------------------------------------
| Indications for<br>Use: | The Rocket Indwelling Pleural Catheter (IPC) Insertion Kit (Model #<br>R51400-16-PL and R51400-MI-PL) is indicated for intermittent, long-<br>term drainage of symptomatic, recurrent, pleural effusion, including<br>malignant pleural effusion and other recurrent effusions that do not<br>respond to medical management of underlying disease.<br><br>The device is indicated for:<br>• The palliation of dyspnea due to pleural effusion<br>• Providing pleurodesis (resolution of the pleural effusion)<br> | | | |
|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | This device should only be used by or under the supervision of trained<br>personnel and in conjunction with current clinical practice guidelines. | | | |
| Substantial<br>Equivalence: | Parameter | Subject Device<br>Rocket Indwelling<br>Pleural Catheter<br>(IPC) Insertion Kit | Predicate Device<br>Rocket Indwelling<br>Pleural Catheter<br>(IPC) Insertion Kit | Similarities and<br>Differences |
| | 510(k) Number | K163321 | K123033 | n/a |
| | Classification | Class II | Same as subject<br>device | identical |
| | Product Code | DWM | Same as subject<br>device | identical |
| | Regulation | 21 CFR 870.5050 | Same as subject<br>device | identical |
| | Indications for<br>Use | The Rocket<br>Indwelling Pleural<br>Catheter (IPC)<br>Insertion Kit (Model<br># R51400-16-PL and<br>R51400-MI-PL) is<br>indicated for<br>intermittent, long-<br>term drainage of<br>symptomatic,<br>recurrent, pleural<br>effusion, including<br>malignant pleural<br>effusion and other<br>recurrent effusions<br>that do not respond<br>to medical<br>management of<br>underlying disease.<br>This device should<br>only be used by or<br>under the<br>supervision of<br>trained personnel<br>and in conjunction<br>with current clinical<br>practice guidelines. | The Rocket IPC<br>Insertion Kit and<br>Rocket IPC<br>Dressing Pack and<br>Bottle Set are<br>indicated for<br>intermittent, long-<br>term drainage of<br>symptomatic,<br>recurrent, pleural<br>effusion, including<br>malignant pleural<br>effusion and other<br>recurrent effusions<br>that do not respond<br>to medical<br>management of<br>underlying disease.<br>This device should<br>only be used by or<br>under the<br>supervision of<br>trained personnel<br>and in conjunction<br>with current clinical<br>practice guidelines. | Same as subject<br>device except:<br><br>The bottle sets and<br>dressing packs<br>accessories were<br>already cleared in<br>the predicate<br>(K123033);<br>therefore, they are<br>not part of the<br>subject device<br>convenience kit.<br>However, the bottle<br>sets and dressing<br>packs are used with<br>the subject device;<br>therefore, there is<br>no impact on<br>substantial<br>equivalence. |
| | The device is<br>indicated for:<br>· The palliation of<br>dyspnea due to<br>pleural effusion<br>· Providing<br>pleurodesis<br>(resolution of the<br>pleural effusion) | The devices are<br>indicated for:<br>1. The palliation of<br>dyspnea due to<br>pleural effusion<br>2. Providing<br>pleurodesis<br>(resolution to the<br>pleural effusion).<br>The Rocket IPC<br>Bottle Sets are<br>indicated for use<br>only with the<br>Rocket Indwelling<br>Pleural Catheter for<br>intermittent<br>drainage. The<br>Rocket Dressing<br>Packs are indicated<br>for dressing of a<br>catheter and exit<br>site. | | |
| Kit Components<br>and Accessories | IPC Catheter,<br>drainage line and<br>accessories for<br>insertion, attachment<br>and dressing<br>including currently<br>licensed Lidocaine<br>HCl, USP (5mL x 3)<br>and DurePrep (6 mL<br>x 2) and four update<br>kit accessories | IPC Catheter,<br>drainage line and<br>accessories for<br>insertion,<br>attachment and<br>dressing | The subject device<br>and predicate and<br>reference predicate<br>device insertion kits<br>have the identical<br>main components:<br>indwelling catheter,<br>drainage line and<br>accessories for<br>insertion,<br>attachment and<br>dressing. The<br>subject device kit<br>components<br>including currently<br>licensed Lidocaine<br>HCl, USP (5mL x<br>3) and DuraPrep (6<br>mL x 2) and four<br>updated kit<br>accessories are<br>identical to the<br>reference predicate<br>(K162457). | |
| Catheter and<br>Drainage Line<br>Technological<br>Features | Catheter: Tubing,<br>cuff, one-way valve<br>and cap;<br>Drainage line:<br>tubing, one-way<br>valve and locking<br>section | Same as subject<br>device | Since the subject<br>device and the<br>predicate device<br>catheters and<br>drainage line are<br>identical, they have<br>identical<br>technological<br>features. | |
| | | | | |
| Catheter and<br>Drainage Line<br>Principle of<br>Operation | Catheter is attached<br>via the one-way<br>valve to vacuum<br>bottles or wall<br>suction with<br>drainage line for<br>drainage of the<br>pleural space; the<br>one-way value on<br>catheter or drainage<br>line prevents air and<br>fluid from migrating<br>back into the pleural<br>space | Same as subject<br>device | Since the subject<br>device and the<br>predicate catheters<br>and drainage line<br>are identical, they<br>have identical<br>principles of<br>operation. | |
| Catheter and<br>Drainage Line<br>Materials | Catheter: Silicone<br>tubing with 30%<br>barium sulfate stripe,<br>polyester cuff,<br>silicone adhesive,<br>ABS one-way valve<br>and cap;<br>Drainage line: PVC<br>tubing and ABS one-<br>way valve and<br>locking section | Same as subject<br>device | Since the subject<br>device and the<br>predicate catheters<br>and drainage line<br>are identical, they<br>have identical<br>materials | |
| Kits | Model # R51400-16-<br>PL and R51400-MI-<br>PL | Model # R51400-<br>16-00 and R51400-<br>MI-PL | Unique model<br>numbers for each<br>device. | |
| Performance per<br>EN 1617:1997<br>EN 1618:1997 | Resistance to<br>deformation; force at<br>break- connections;<br>force at break-<br>drainage catheters<br>and all other parts;<br>freedom from<br>leakage | Same as subject<br>device | Since the main<br>components of the<br>devices are<br>identical, the<br>performance testing<br>performed on the<br>predicate is<br>applicable to the<br>subject device. | |
| Biocompatibility<br>per ISO 10993-1 | Cytotoxicity.<br>sensitization,<br>irritation, system<br>toxicity, subchronic<br>toxicity.<br>genotoxicity,<br>implantation,<br>exhaustive<br>extraction, summary<br>report and biological<br>risk assessment | Same as subject<br>device | Since the main<br>components of the<br>devices are<br>identical, the<br>biocompatibility<br>testing performed<br>on the predicate is<br>applicable to the<br>subject device. | |
| | | | | |
| | Packaging<br>components and<br>materials | APET tray, blue<br>field wrap, uncoated<br>Tyvek 1073B/PET<br>pouch, folding<br>carton | Same as subject<br>device | The subject device<br>and predicate use<br>the same packaging<br>materials and<br>process. |
| | Sterility | SAL 106 | Same as subject<br>device | The subject device<br>and predicate are<br>sterilized using the |
| | Sterilization<br>Method per ISO<br>11135-1 and<br>ISO 10993-7 | EtO and<br>EtO residuals | Same as subject<br>device | sterilized using the<br>same method, the<br>same sterility level<br>and the same<br>residual testing. |
| | Shelf life | The shelf life will<br>not exceed the shelf<br>life of any of its kit<br>components when<br>packaged | 3 years | Since the subject<br>device is identical<br>to the predicate<br>except for the<br>Lidocaine HCl,<br>USP and DuraPrep,<br>the stability testing<br>on the predicate is<br>applicable to the<br>subject device. The<br>shelf life of the<br>Lidocaine HCl,<br>USP and DuraPrep<br>are determined and<br>monitored by the<br>manufacturers. |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
There have been no technological modifications made to the subject device Technological catheter or drainage line as it relates to the predicate device (they are and identical), therefore the performance of the subject device catheter and Performance drainage line is identical to the performance of the predicate device. Comparison: Currently licensed Lidocaine HCl, USP and DuraPrep have been added as accessories to the Pleural IPC convenience kit, but they do not affect the technology, performance or principle of operation of the catheter and drainage line. The Lidocaine HCl, USP and DuraPrep were cleared as kit components in the Reference predicate (K162457). The Lidocaine HCL, USP was tested post EtO sterilization per the Lidocaine HCl, USP monograph and confirmed it met all Lidocaine HCl, USP monograph requirements as labeled. The DuraPrep is validated for 2 cycles of EtO sterilization by the manufacturer.
{9}------------------------------------------------
| Component | Subject Device | Regulatory<br>Status | Predicate<br>Device | Reference<br>Predicate |
|-------------------------------------|--------------------------------------|----------------------|---------------------|------------------------------|
| Tunneler | Metal | K152105 | Plastic | |
| Suture Forceps/<br>Dressing Forceps | Blue tipped to<br>signify single use | 510(k)<br>Exempt | Not blue<br>tipped | Same as<br>subject<br>device |
| Needle Holder | Blue tipped to<br>signify single use | 510(k)<br>Exempt | Not blue<br>tipped | |
| Suture | Dissolvable | K981935 | Silk | |
| Pinch Clamp and<br>Slide Clamp | Removed from<br>kit | N/A | Present in kit | |
Comparison of subject device updated kit accessories to predicate and Reference predicate:
Summary of Because the subject device catheter and drainage line are identical to the Performance predicate device, and no manufacturing or packaging materials or processes Testing: have changed and the sterilization method has not changed, no new performance, packaging, biocompatibility or shelf life testing was performed for this submission in order to establish substantial equivalence. Performance testing on the predicate included resistance to deformation, force at break-connections, force at break- drainage catheters and all other parts, and freedom from leakage. All testing met acceptance criteria. Biocompatibility testing on the predicate included cytotoxicity, sensitization, irritation, system toxicity, subchronic toxicity, genotoxicity, implantation, exhaustive extraction, and biological risk assessment. All testing met acceptance criteria. The shelf life will not exceed the shelf life of any of its kit components. The Lidocaine HCl, USP and DuraPrep were cleared as kit components in the Reference predicate (K162457). Lidocaine HCl, USP was tested post sterilization per the Lidocaine HCl, USP monograph and confirmed it met all Lidocaine HCl, USP requirements as labeled. DuraPrep is validated by the manufacturer for two EtO sterilization cycles and therefore required no additional testing.
{10}------------------------------------------------
| Conclusion: | Rocket Medical considers the Rocket IPC Indwelling Pleural Insertion Set<br>with Lidocaine and DuraPrep to be substantially equivalent to the predicate<br>and Reference predicate devices listed above. This conclusion is based upon<br>the fact that the subject device catheter and drainage line technological<br>features, principle of operation, materials, biocompatibility, performance,<br>packaging and sterilization are identical to the predicate. Also, all kit<br>components including currently licensed Lidocaine HCl, USP and DuraPrep<br>and four updated kit accessories (two cosmetic and two material changes<br>identified in the table above: Comparison of updated kit accessories to<br>predicate and Reference predicate) are identical and have the same shelf life<br>as the Reference predicate. The Lidocaine HCl, USP was tested post EtO<br>sterilization per the Lidocaine HCl, USP monograph and met all USP<br>requirements as labeled. The difference in the Indication for Use statements<br>of the subject and predicate as described above (exclusion of Bottle Sets and<br>Dressing Packs for subject device) have not changed the intended use or<br>substantial equivalence of the devices. |
|-------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.