K160450 · Care Fusion · DWM · Oct 31, 2016 · Cardiovascular
Device Facts
Record ID
K160450
Device Name
Pleurx Pleural Catheter System
Applicant
Care Fusion
Product Code
DWM · Cardiovascular
Decision Date
Oct 31, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PleurX Pleural Catheter Kits are indicated for intermittent, long term drainage of symptomatic, recurrent, pleural effusion, including malignant pleural effusion and other recurrent effusions that do not respond to medical management of the underlying disease. The devices are indicated for 1) the palliation of dyspnea due to pleural effusion and 2) providing pleurodesis (resolution of the pleural effusion). The PleurX Pleural Catheter Kits are indicated for adults only. The PleurX Lockable Drainage Line is used to drain fluid using standard wall suction, water seal drainage system, vacuum bottle, or other appropriate method. The PleurX Access Valve attaches only to the PleurX Catheter. The PleurX Valve Kit is intended to repair the PleurX Catheter and replace the PleurX Valve. The PleurX Valve Kit is indicated for adults only. The PleurX LP Catheter Mini Kit is indicated for intermittent, long term drainage of symptomatic, recurrent, pleural effusion, including malignant pleural effusion and other recurrent effusions that do not respond to medical management of the underlying disease. The devices are indicated for the palliation of dyspnea due to pleural effusion and providing pleurodesis (resolution of the pleural effusion). The PleurX Low Profile Catheter is indicated for adults only.
Device Story
Pleurx Pleural Catheter System facilitates intermittent, long-term drainage of pleural effusions in home settings. System includes indwelling percutaneously tunneled catheter with fenestrations, radiopaque markings, and cuff; access valve; and drainage kits. Device allows fluid removal via wall suction, water seal, or vacuum bottle. Designed for adult patients to manage dyspnea and achieve pleurodesis. System components include Pleurx Catheter, Pleurx Drainage Kits, Pleurx Lockable Drainage Line, and Pleurx Access Valve/Valve Kit for repairs. Healthcare providers perform initial placement; patients or caregivers manage intermittent drainage. Output is pleural fluid drainage, providing symptomatic relief and potential resolution of effusion.
Clinical Evidence
No clinical tests were conducted for this submission; bench testing only.
Technological Characteristics
Materials: Silicone, Barium Sulfate, Polyester, Polyisoprene, Polypropylene, Colorant, ABS, Ink, Adhesive. Features: Percutaneously tunneled indwelling catheter, fenestrations, radiopaque markings, cuff, and valved closed system. Standards: ISO 10993 (biocompatibility), EN 1617/1618 (drainage catheters), ISO 11607 (packaging), ISO 11135 (EO sterilization), ASTM F1980 (aging), ASTM F2052/F2119/F2182/F2213 (MRI safety).
Indications for Use
Indicated for adults with symptomatic, recurrent pleural effusion (including malignant) unresponsive to medical management. Used for palliation of dyspnea and to provide pleurodesis.
Regulatory Classification
Identification
A patient care suction apparatus is a device used with an intrathoracic catheter to withdraw fluid from the chest during the recovery period following surgery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 31, 2016
CareFusion Tamara Brey Advisor, Regulatory Management 75 North Fairway Drive Vernon Hills, Illinois 60061
Re: K160450
Trade/Device Name: Pleurx Pleural Catheter System Regulation Number: 21 CFR 870.5050 Regulation Name: Patient Care Suction Apparatus Regulatory Class: Class II Product Code: DWM Dated: September 26, 2016 Received: September 28, 2016
Dear Tamara Brey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160450
Device Name Pleurx Pleural Catheter System
#### Indications for Use (Describe)
The Pleural Catheter Kits are indicated for intermittent, long term drainage of symptomatic, recurrent, pleural effusion, including malignant pleural effusion and other recurrent effusions that do not respond to medical management of the underlying disease. The devices are indicated for 1) the palliation of dyspnea effusion and 2) providing pleurodesis (resolution of the pleural effusion). The Pleural Catheter Kits are indicated for adults only.
The Pleur X Lockable Drainage Line is used to drain fluid using standard wall suction, water seal drainage system, vacuum bottle, or other appropriate method.
The Pleur X Access Valve attaches only to the PleurX Valve Kit is intended to repair the Pleur X Catheter and replace the PleurX Valve. The PleurX Valve Kit is indicated for adults only.
The Pleur X LP Catheter Mini Kit is indicated for intermittent, long term drainage of symptomatic, recurrent, pleural effusion, including malignant pleural effusion and other recurrent effusions that do not respond to medical management of the underlying disease. The devices are indicated for the palliation of dyspnea effusion and providing pleurodesis (resolution of the pleural effusion). The PleurX Low Profile Catheter is indicated for adults only.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY K160450
A summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92.
| SUBMITTER INFORMATION | |
|---------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | CareFusion |
| Address | 75 North Fairway Drive, Vernon Hills, IL 60061 USA |
| Phone number | (847) 362-9485 |
| Fax number | (312) 949-9245 |
| Establishment Registration<br>Number | 1423507 |
| Name of contact person | Tamara Brey |
| Date prepared | February 15, 2016 |
| DESCRIPTION OF DEVICE | |
| Trade or proprietary name | Pleurx Pleural Catheter System |
| Common or usual name | Pleural Drainage Catheter |
| Classification name | Patient Care Suction Apparatus |
| Classification panel | Anesthesiology |
| Regulation | Class II per 21CFR §870.5050, Procode DWM |
| Product Code(s) | Multiple |
| Legally marketed device(s) to<br>which equivalence is claimed | CareFusion Pleurx Catheter System: K141965, and<br>Bard Aspira Pleural Drainage System: K110409 |
| Reason for 510(k) submission | This 510(k) submission is for a change in material formulation and<br>concentration, and minor design modifications to the devices cleared<br>under K141965. |
| Device description | The Pleurx Pleural Catheter System provides patients with a convenient<br>method to relieve pleural effusion symptoms at home. The primary<br>components of the Pleurx Catheter System are the Pleurx Pleural<br>Catheter and the Pleurx Drainage Kits. |
| Intended use of the device | The PleurX Pleural Catheter Kits are indicated for intermittent, long term<br>drainage of symptomatic, recurrent, pleural effusion, including malignant<br>pleural effusion and other recurrent effusions that do not respond to<br>medical management of the underlying disease. The devices are indicated<br>for 1) the palliation of dyspnea due to pleural effusion and 2) providing<br>pleurodesis (resolution of the pleural effusion). The PleurX Pleural<br>Catheter Kits are indicated for adults only.<br>The PleurX Lockable Drainage Line is used to drain fluid using standard<br>wall suction, water seal drainage system, vacuum bottle, or other<br>appropriate method.<br>The PleurX Access Valve attaches only to the PleurX Catheter. The<br>PleurX Valve Kit is intended to repair the PleurX Catheter. |
| | Catheter and replace the PleurX Valve. The PleurX Valve Kit is indicated<br>for adults only. |
| | The PleurX LP Catheter Mini Kit is indicated for intermittent, long term<br>drainage of symptomatic, recurrent, pleural effusion, including malignant<br>pleural effusion and other recurrent effusions that do not respond to<br>medical management of the underlying disease. The devices are indicated<br>for the palliation of dyspnea due to pleural effusion and providing<br>pleurodesis (resolution of the pleural effusion). The PleurX Low Profile<br>Catheter is indicated for adults only. |
{4}------------------------------------------------
## SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE
| Characteristic | New Device | Predicate |
|---------------------------------|---------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Catheter Description | Internal: fenestrations,<br>radiopaque markings & cuff<br>External: valve | Same as predicate: CareFusion Pleurx<br>Catheter Systems: K141965 |
| Method | Percutaneously tunneled -<br>indwelling | Same as predicate: CareFusion Pleurx<br>Catheter Systems: K141965 |
| Access Valve Description | Unique access tip, valved closed<br>system to inject and withdraw<br>fluids. | Same as predicate: Bard Aspira Drainage<br>System: K110409 |
| Materials of Construction | Silicone, Barium Sulfate,<br>Polyester, Polyisoprene,<br>Polypropylene, Colorant, ABS,<br>Ink, Adhesive | Same as predicate: CareFusion Pleurx<br>Catheter Systems: K141965 |
| CONCLUSION OF DEVICE COMPARISON | | |
The technological characteristics of the proposed devices are substantially equivalent to the predicates.
### PERFORMANCE DATA
### SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE
### Performance Test Summary
| Characteristic | Standard/Test/FDA Guidance |
|------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | AAMI/ANSI/ISO 10993-1:2009 Biological evaluation of Medical Devices Part 1:<br>Evaluation and Testing |
| Biocompatibility | AAMI/ANSI/ISO 10993-5: 2009 Biological Evaluation of Medical Devices – Part 5<br>Tests for In Vitro Cytotoxicity |
| Biocompatibility | AAMI/ANSI/ISO 10993-6: 2007 (R) 2010 Biological Evaluation of Medical Devices –<br>Part 6 Tests for Local Effects After Implantation |
| Residuals | AAMI/ANSI/ISO 10993-7:2008 Biological evaluation of Medical Devices Part 7:<br>Ethylene Oxide Sterilization Residuals |
| Biocompatibility | AAMI/ANSI/ISO 10993-10: 2010 Biological Evaluation of Medical Devices – Part 10<br>Tests for Irritation and Skin Sensitivity |
| Biocompatibility | ISO 10993-17:2002 Biological evaluation of medical devices --Establishment of<br>allowable limits for leachable substances |
| Biocompatibility | ISO 10993-18:2005 Biological evaluation of medical devices -- Chemical<br>characterization of materials |
| Performance | EN 1617:1997 Sterile Drainage Catheters and Accessory Devices for Single Use |
{5}------------------------------------------------
| Performance | EN 1618:1997 Catheters Other Than Intravascular Catheters - Test Methods for<br>Common Properties |
|--------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance | ANSI/AAMI/ISO 11607-1:2006/(R)2010 Packaging for Terminally Sterilized Medical<br>Devices- Part 1:Requirements For Materials, Sterile Barrier Systems And Packaging<br>Systems [Including: Amendment 1 (2014)] |
| Performance | ANSI/AAMI/ISO 11607-2:2006/(R)2010 Packaging for Terminally Sterilized Medical<br>Devices- Part 2: Validation Requirements For Forming, Sealing And Assembly<br>Processes [Including: Amendment 1 (2014)] |
| Performance | ASTM F1980-07 Accelerated Aging of Sterile Barrier Systems |
| Performance | ISO 11138-1, 2nd Ed Sterilization of healthcare products - Biological Indicators - Part<br>1: General Requirements |
| Performance | ISO 11737-1 2nd Ed Sterilization of Medical Devices – Microbiological Methods Part<br>1 |
| Performance | ISO 11135 2nd Ed Sterilization of Health-Care Products -Ethylene Oxide –<br>Requirements for the Development, Validation and Routine Control of a Sterilization<br>Process for Medical Devices |
| Performance | AAMI TIR28:2009(R)2013 Product Adoption and Process Equivalency for Ethylene<br>Oxide Sterilization |
| Performance | ISO 14630 Non-active Surgical Implants - General Requirements |
| Performance | ASTM F2052 Measurement of Magnetically Induced Displacement Force on Medical<br>Devices in the MR Environment |
| Performance | ASTM F2119 Evaluation of MR Image Artifacts from Passive Implants |
| Performance | ASTM F2182 A Measurement of Radiofrequency Induced Heating on or Near<br>Passive Implants During MRI |
| Performance | ASTM F2213 Measurement of Magnetically Induced Torque on Medical Devices in<br>the Magnetic Resonance Environment |
| SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL<br>EQUIVALENCE AND/OR OF CLINICAL INFORMATION | |
N/A – No clinical tests were conducted for this submission
CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA
The results of the non-clinical tests show that the CareFusion Pleural Catheter System is as safe, as effective, and performs as well as the legally marketed predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.