K031041 · Bard Peripheral Vascular · JCT · Jun 19, 2003 · General, Plastic Surgery
Device Facts
Record ID
K031041
Device Name
FLUENCY TRACHEOBRONCHIAL STENT GRAFT
Applicant
Bard Peripheral Vascular
Product Code
JCT · General, Plastic Surgery
Decision Date
Jun 19, 2003
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.3720
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The FLUENCY™ Tracheobronchial Stent Graft is indicated for use in the treatment of tracheobronchial strictures produced by malignant neoplasms or in benign strictures after all alternative therapies have been exhausted.
Device Story
Self-expanding Nitinol stent encapsulated with ePTFE; pre-loaded on flexible pull-back delivery system. Designed to maintain patency of tracheobronchial tree. Radiopaque Tantalum markers facilitate placement. Used by physicians in clinical settings to treat strictures. Stent provides mechanical support to airway; benefits patient by restoring airway patency.
Clinical Evidence
Bench testing only.
Technological Characteristics
Self-expanding Nitinol stent; ePTFE encapsulation; Tantalum radiopaque markers. Flexible pull-back delivery system. Single-use. Available in various lengths and diameters.
Indications for Use
Indicated for patients with tracheobronchial strictures caused by malignant neoplasms or benign strictures where alternative therapies have been exhausted.
Regulatory Classification
Identification
The tracheal prosthesis is a rigid, flexible, or expandable tubular device made of a silicone, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the trachea or trachealbronchial tree. It may be unbranched or contain one or two branches. The metal tracheal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
Predicate Devices
WALLGRAFT™ Tracheobronchial Endoprosthesis
Bard LUMINEXX™ 7F Biliary Stent
Bard memotherm® Covered Esophageal Stent
VIABAHN™ Endoprosthesis
Related Devices
K050832 — FLUENCY PLUS TRACHEOBRONCHIAL STENT GRAFT · Bard Peripheral Vascular, Inc. · Jul 7, 2005
K992510 — WALLSTENT TRACHEOBRONCHIAL ENDOPROSTHESIS AND UNISTEP PLUS DELIVERY SYSTEM · Boston Scientific Corp · Nov 18, 1999
K961296 — WALLSTENT TRACHEOBRONCHIAL ENDOPROSTHESIS · Boston Scientific Scimed, Inc. · Jul 10, 1996
K121048 — ULTRAFLEX TRACHEOBRONCHIAL STENT SYSTEM- STERILE UNCOVERED · Boston Scientific Corp · Aug 3, 2012
K961507 — WALSTENT TRACHEOBRONCHIAL ENDOPROSTHESIS · Boston Scientific Scimed, Inc. · Jul 10, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
JUN 19 2003
Kø3/ø41 (P. 1 of 2)
## 510(k) Summary of Safety and Effectiveness Information VI.
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a summary of the safety and effectiveness information upon which substantial equivalence determination is based follows:
- Submitter Information: A.
| Applicant: | Bard Peripheral Vascular, a division of C.R. Bard, Inc.<br>1625 West 3rd Street<br>Tempe, Arizona 85280 |
|------------|---------------------------------------------------------------------------------------------------------|
| Phone: | 480-303-2752 |
| Fax: | 480-449-2546 |
| Contact: | Aymee R. Berry, Associate Manager, Regulatory Affairs |
- B. Device Name:
| Trade Name: | FLUENCY™ Tracheobronchial Stent Graft |
|----------------|---------------------------------------|
| Common or | |
| Usual Name: | Tracheal Prosthesis |
| Classification | |
| Name: | Prosthesis, tracheal |
- C. Predicate Device Name(s): WALLGRAFT™® Tracheobronchial Endoprosthesis
Boston Scientific
Bard LUMINEXX™ 7F Biliary Stent C.R. Bard. Inc.
Bard memotherm® Covered Esophageal Stent C.R. Bard, Inc.
VIABAHN™ Endoprosthesis W.L. Gore & Associates, Inc.
- D. Device Description:
The FLUENCY™ Tracheobronchial Stent Graft includes a self-expanding Nitinol Stent encapsulated with ePTFE pre-loaded on a flexible "pull-back" delivery system. It is a single use prosthetic device designed to maintain the patency of the tracheobronchial tree in patients with tracheobronchial strictures produced by malignant neoplasms or in benign strictures after all alternatives have been exhausted. Highly radiopaque Tantalum markers on the stent graft ends facilitate stent graft placement. The FLUENCY™ Tracheobronchial Stent Graft is available in various lengths and
{1}------------------------------------------------
\\$\phi3/1041 (P.2 of 2)
diameters. It is preloaded into various size delivery catheters, depending on the size of the stent graft.
- E. Statement of Intended Use:
The FLUENCY™ Tracheobronchial Stent Graft is indicated for use in the treatment of tracheobronchial strictures produced by malignant neoplasms or in benign strictures after all alternative therapies have been exhausted.
- ட் Substantial Equivalence:
A variety of tests, assessments, and comparisons demonstrate that the FLUENCY™ Tracheobronchial Stent Graft is substantially equivalent to the following predicates in terms of composition, design, intended use, and performance attributes as noted below:
- The FLUENCY™ Stent Graft is substantially equivalent to the . WALLGRAFT™ Endoprosthesis in indication for use, method of deployment, and in-vitro testing.
- The FLUENCY™ Stent Graft is substantially equivalent to the ● VIABAHN™ in both stent graft materials and indication for use.
In addition, safety and effectiveness of the FLUENCY™ Stent Graft is further supported by comparability to other legally marketed devices as follows:
- The FLUENCY™ Stent Graft and delivery system is comparable to the . LUMINEXX™ 7F Biliary Stent Graft (Nitinol stent) and delivery system as previously demonstrated.
- The FLUENCY™ Stent Graft is comparable to the memotherm® . Esophageal in covering and stent material.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a seal for the U.S. Department of Health & Human Services. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Public Health Service
JAN 1 6 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Bard Peripheral Vascular, Inc. % Mr. Joshua Smale Regulatory Affairs Specialist 1625 West 3rd Street P.O. Box 1740 Tempe, Arizona 85280
Re: K031041
Trade/Device Name: FLUENCY™ Tracheobronchial Stent Graft Regulation Number: 21 CFR 878.3720 Regulation Name: Tracheal prosthesis Regulatory Class: II Product Code: JCT Dated: March 31, 2003 Received: April 7, 2003
Dear Mr. Smale:
This letter corrects our substantially equivalent letter of June 19, 2003.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Joshua Smale
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark N. Melkerson
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K031041
Device Name: FLUENCY® Tracheobronchial Stent Graft
Indications for Use:
The FLUENCY® Tracheobronchial Stent Graft is indicated for use in the treatment of tracheobronchial strictures produced by malignant neoplasms.
Prescription Use X (Part21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) storative. officies Listices and Of General, free Jrative,
Image /page/4/Picture/14 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, giving them a distinct and modern appearance. The font style is consistent throughout the word, creating a unified and recognizable logo or brand name.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.