K042501 · Hitachi Medical Corp. · DQA · Oct 12, 2004 · Cardiovascular
Device Facts
Record ID
K042501
Device Name
ETG-4000 OPTICAL TOPOGRAPHY SYSTEM
Applicant
Hitachi Medical Corp.
Product Code
DQA · Cardiovascular
Decision Date
Oct 12, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
The intended use of the ETG-4000 is the measurement of relative levels of cerebral deoxyhemoglobin and oxyhemoglobin.
Device Story
ETG-4000 is a non-invasive optical topography system; measures relative changes in cerebral oxy-hemoglobin, deoxy-hemoglobin, and total hemoglobin concentrations. Device beams frequency-modulated near-infrared light (670-1300nm) into scalp/skull; light reflects through cerebral cortex and is collected by optical fiber array. System processes absorption data to generate 2D and 3D dynamic images representing metabolic and circulatory status of cerebral cortex. Used in clinical settings; operated by trained personnel. Output provides time-course graphic representations and images of brain surface activity; assists clinicians in assessing cerebral hemodynamic status.
Clinical Evidence
Bench testing only. No clinical data provided. The device was validated in accordance with design controls and applicable safety standards to demonstrate substantial equivalence to the predicate.
Technological Characteristics
Non-invasive optical topography system; uses near-infrared light (670-1300nm) for sensing. Employs an array of optical fiber tips for scalp contact. Measures light absorption to calculate hemoglobin concentration changes. No moving parts. Complies with applicable safety standards.
Indications for Use
Indicated for the measurement of relative levels of cerebral deoxyhemoglobin and oxyhemoglobin in patients.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
K182242 — VO200 NeurOs Cerebral Oximetry System · Mespere Lifesciences, Inc. · Dec 26, 2018
K073036 — FORE-SIGHT, MC-2000 SERIES · Cas Medical Systems, Inc. · Feb 25, 2008
K123700 — FORE-SIGHT ELITE ABSOLUTE TISSUE OXIMETER · Cas Medical Systems, Inc. · Apr 11, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
OCT 1 2 2004 Attachment 2
# HITACI
#### HITACHI MEDICAL CORPORATION 1-1-14 Uchi-Kanda, Chiyoda-ku Tokyo 101-0047, Japan
### 510(k) Summary
KO42501
#### Submitter Information
#### Hitachi Medical Corporation Submitter: 1-1-14 Uchi-Kanda, Chiyoda-ku Tokyo 101-0047, Japan
Douglas J. Thistlethwaite Contact: ph: (330) 425-1313 fax: (330) 425-1410
Date: September 13, 2004
#### Device Name
Device Name: Optical Topography System Trade/Proprietary Name: ETG-4000 Optical Topography System Common Name: Oximeter Classification Name: Oximeter Classification Number: Sec. 870.2700
#### Predicate Device
Predicate Device: Hitachi ETG-100, K011320
{1}------------------------------------------------
# Device Description
#### Function
The ETG-4000 is a device that measures relative changes in tissue concentration of oxy-The E I G-4000 is a device that measures relative changuin to the level of blood) in
hemoglobin and deoxy-hemoglobin and total hemoglobin (678-1200 m) that sam hemoglobil and deoxy-nemogroom and total near-infrared light (670-1300nm) that can the surface area of the cerebra. Sed is absorbed by the hemoglobin in the blood.
The ETG-4000 displays the changes of overall Hemoglobin concentration in time- course I he ETO-4000 uisplays the changes of overally images (motion images and still images) based on the data from multiple point measurements.
ouse on a
This is a non-invasive test that is done by contacting an array of small optical fiber tips on the surface of the scalp.
The ETG-4000 beams frequency modulated near-infrared light into the surface of the brain I he £ I 0-4000 beams frequency modula it is a light passes through the scalp, skull and upper layer of illiough several or the optical noves read by the hemoglobin in the blood and is reflected back and is collected by optical fibers.
The ETG-4000 measures relative changes in tissue concentration of oxy-hemoglobin and deoxy-The LTG-4000 incasures relative changes in multiple points on the head simultaneously by utilizing the changes of light absorption.
The ETG-4000 provides data that show the activity status of the cerebral cortex by displaying the The LTG++000 provides data that mich in the brain surface, and the metabolic and circulatory status of the cerebral cortex in a time-course graphic representation of oxy-hemoglobin, deoxy-Status of the coronal corent in bit, 2D dynamic images and 3D dynamic images.
# Device Intended Use
The intended use of the ETG-4000 is the measurement of relative levels of cerebral deoxyhemoglobin and oxyhemoglobin.
{2}------------------------------------------------
### Device Technological Characteristics
The characteristics of the ETG-4000 Optical Topography System compare substantially to the The characteristics of the LTG-400 Option Topography of and functional methodology.
ETG-100 predicate device in materials used, technology applied, and functional methodoney E I G-100 predicate device in materials acoup residently - intended use, or application Differences do not affect safety and effect for that is substantially equivalent to the cleared memods. The LTO-4000 operates and enhancement of the technology of the predicate.
# Safety
The ETG-4000 is a non-invasive device with no moving parts. It has been designed to comply with all applicable safety standards.
# Conclusions
The ETG-4000 system has been developed and validated in accordance with design controls and The LTG-4000 System has been developed at the system is safe and effective for the indicated appricable standards: Testing nas pe vat there are no new safety issues associated with this system as compared with the predicate device.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
OCT 1 2 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Hitachi Medical Corporation c/o Mr. Douglas J. Thistlethwaite Manager of Regulatory Affairs Hitachi Medical Systems America, Inc. 1959 Summit Commerce Park Twinsburg, Ohio 44087
Re: K042501
K042301
Trade/Device Name: ETG-4000 Optical Topography System Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: September 13, 2004 Received: September 14, 2004
Dear Mr. Thistlethwaite:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 910(x) premained is substantially equivalent (for the indications referenced above and nave ucterimined the actived predicate devices marketed in interstate for use stated in the encrosule, to regally mancess to the Medical Device Amendments, or to commerce prior to May 28, 1770, the encentines with the provisions of the Federal Food, Drug, devices that have been reclassified in accreanse who a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a previsions of the Act. T and Cosment Act (Act) that do not require subject to the general controls provisions of the Act. The You may, therefore, market the device, seeject to the many for annual registration, listing of general controls provisions of the free labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) me existing major regulations affecting your device can may be subject to such additional controller "Entrological" - " " " " " " be found in the Code of Pouchal Regeraing your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that IDA s issualled of a basedana.
that FDA has made a determination that your device complies with other requirements of the Act that FDA has made a decemination administered by other Federal agencies. You must or any Federal statures and regulations daminding, but not limited to: registration and listing (21 comply with an the Act 3 requirements, more and manufacturing practice requirements as set CrK Fart 807), labeling (21 CFR Part 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Douglas J. Thistlethwaite
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will allow you to begin harketing your antial equivalence of your device to a legally
premarket notification. The FDA finding of substantial equivale of the porm premarket notification. The PDA miding of Subscribed of the Pressure and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
rs the latest of the more on 2011 5 - Alama lease note the roculation entit If you desire specific advice for your do no no no the regulation entitled, the regulation entitled, contact the Office of Compliance at (210) = 10 = 1 = 1 = 1 = 8 (1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = "Misbranding by relefence to premanter to be it from the Division of Small other general information on your responsible in to toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance of Architect (demansing http://www Manufacturers, micrnational and October http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam L. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K042501
> ETG-4000 Optical Topography System Device Name:
Indications for Use:
The intended use of the ETG-4000 is the measurement of relative levels of cerebral deoxyhemoglobin and oxyhemoglobin.
X Prescription Use
AND/OR
Over-the-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
uriam C. Provost
Division of General, Restorative, and Neurological Devices
8042501 510(k) Number_
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.