The Belun Sleep System BLS-100 is a wearable device intended to record, analyze, display, export, and store biophysical parameters to aid in evaluating moderate to severe sleep-related breathing disorders suspected of sleep apnea. The device is intended for use in clinical and home settings under the direction of a Healthcare Professional (HCP).
Device Story
Wearable ring-based system for home sleep apnea testing (HSAT). Device consists of Belun Ring (worn on proximal phalanx of index finger) and Belun Sleep AI (standalone PC software). Ring contains red/infrared LEDs for PPG, and accelerometer for movement detection. Records oxygen saturation, pulse period, PPG, and actigraphy to on-board memory. Data exported via USB/Bluetooth to PC software. AI processes signals to generate apnea-hypopnea index (bAHI) and sleep staging (bSTAGES: Wake, REM, NREM). Used by patients at home; results reviewed by HCP to aid in sleep apnea evaluation. Benefits include non-invasive, convenient sleep monitoring for clinical decision-making.
Clinical Evidence
Prospective clinical study of 106 adult patients suspected of OSA. Compared Belun Sleep System output to manual PSG scoring (AASM v2.4). At AHI cutoff 15: 87.7% accuracy, 89.8% sensitivity, 86.0% specificity. At AHI cutoff 30: 92.5% accuracy, 84.0% sensitivity, 95.1% specificity. Sleep stage classification (Wake/REM/NREM) accuracy ranged from 82.7% to 90.8%. Mean difference between device and PSG total sleep time was 21.8 minutes (SD 41.6).
Technological Characteristics
Wearable ring with biocompatible enclosure. Sensors: optical PPG (red/IR) and accelerometer. Energy: 3.7V Lithium battery. Connectivity: USB/Bluetooth for data export. Software: Standalone PC-based AI analysis. Standards: IEC 60601-1, IEC 60601-1-11, ISO 80601-2-61, ISO 10993 (biocompatibility), IEC 62366-1 (usability).
Indications for Use
Indicated for adult patients (22+ years) suspected of moderate to severe sleep-related breathing disorders/sleep apnea. Intended for use in clinical and home settings under HCP direction.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
February 23, 2023
Belun Technology Company Limited Leung Lap Wai Lydia CEO Unit 218, 2/F, Core Building 2, No. 1 Science Park West Avenue, Hong Kong Science Park Sha Tin, Hong Kong China
Re: K222579
Trade/Device Name: Belun Sleep System BLS-100 Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: MNR Dated: January 10, 2023 Received: January 13, 2023
Dear Leung Lap Wai Lydia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Rachana Visaria -S
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Opthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222579
Device Name Belun Sleep System BLS-100
### Indications for Use (Describe)
The Belun Sleep System BLS-100 is a wearable device intended to record, analyze, display, export, and store biophysical parameters to aid in evaluating moderate to severe sleep-related breathing disorders suspected of sleep apnea. The device is intended for use in clinical and home settings under the direction of a Healthcare Professional (HCP).
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) is being submitted in accordance with the requirements of 21 CFR 807.92 on 23-Feb-2023.
#### l. SUBMITTER
Belun Technology Company Limited Address: Unit 218, 2 Floor, Core Building 2, 1 Science Park West Avenue, Hong Kong Science Park, Shatin, Hong Kong Contact Person: Leung Lap Wai Lydia Phone: +852 706 5640
#### II. PROPOSED DEVICE
Trade/Device Name: Belun Sleep System BLS-100 Model: BLS-100 Classification Name: Ventilatory Effort Recorder Regulation Number: 21 CFR 868.2375 Regulatory Class: II Product Code: MNR
#### PREDICATE DEVICE III.
Device Name: NightOwl (K220028) Classification Name: Ventilatory Effect Recorder Regulation Number: 21 CFR 868.2375 Regulatory Class: II Product Code: MNR
#### IV. DEVICE DESCRIPTION
The Belun Sleep System BLS-100 is prescribed by a Health Care Professional (HCP) for the patient to use in the home as a 'home sleep apnea test' (HSAT).
The Belun Sleep System BLS-100 comprises a sensor that is worn on the proximal phalanx of index finger (Belun Ring) over-night whilst the subject is sleeping and a stand-alone analysis software (Belun Sleep AI).
The Belun Ring has a small biocompatible enclosure. The sensor has 2 LEDs, one in the red spectrum and the other in the infrared spectrum, and an accelerometer. The sensor is placed on the proximal phalanx of the index finger, with the sensor window applied against the palmar side of the proximal phalanx of the index finger.
{4}------------------------------------------------
The sensor measures the reflected red/infrared signals to record the photoplethysmograph (PPG) signal. The accelerometer is used to detect movement.
The data recorded by the Belun Ring is stored in device on-board memory. The data is exported when the Belun Ring is returned to the prescribing HCP via USB or Bluetooth and passed to the Belun Sleep AI Software, which is standalone PC software. The Belun Sleep Al loads and processes the signal from the exported data and generates the apnea-hypopnea index (bAHI) and sleep staging identification (bSTAGES).
Signals from the Belun Ring device include the following:
- . Oxygen saturation
- Beat-to-beat pulse period
- Photoplethysmography
- . Actigraphy
#### V. INDICATIONS FOR USE
The Belun Sleep System BLS-100 is a wearable device intended to record, analyze, display, export, and store biophysical parameters to aid in evaluating moderate to severe sleep-related breathing disorders of adult patients suspected of sleep apnea. The device is intended for use in clinical and home settings under the direction of a Healthcare Professional (HCP).
{5}------------------------------------------------
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE AND THE REFERENCE DEVICE VI.
| Table 1 Performance Specification Comparison Table between the Belun Sleep System BLS-100 and Predicate Device | | | |
|----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Proposed Device<br>Belun Sleep System BLS-100 | Predicate Device<br>NightOwl (K220028) | Comparison |
| Regulation No. | 21 C.F.R. 868.2375 | 21 C.F.R. 868.2375 | Identical to predicate device |
| Classification | ll | ll | Identical to predicate device |
| Classification Name | Breathing frequency monitor | Breathing frequency monitor | Identical to predicate device |
| Product Code | MNR | MNR | Identical to predicate device |
| Intended Use | The Belun Sleep System BLS-100 is a<br>wearable device intended to record,<br>analyze, display, export, and store<br>biophysical parameters to aid in<br>evaluating moderate to severe sleep-<br>related breathing disorders of adult<br>patients suspected of sleep apnea.<br>The device is intended for use in<br>clinical and home settings under the<br>direction of a Healthcare Professional<br>(НСР). | The NightOwl is a wearable device<br>intended for use in the recording,<br>analysis, displaying, exporting, and<br>storage of biophysical parameters to<br>aid in the evaluation of sleep-related<br>breathing disorders of adult patients<br>suspected of sleep apnea. The device<br>is intended for the clinical and home<br>setting use under the direction of a<br>Healthcare Professional (HCP). | Substantially equivalent to<br>predicate<br>The proposed system and the<br>predicate have the similar<br>intended use. |
| Table 1 Performance Specification Comparison Table between the Belun Sleep System BLS-100 and Predicate Device | | | |
| Characteristic | Proposed Device<br>Belun Sleep System BLS-100 | Predicate Device<br>NightOwl (K220028) | Comparison |
| Use Environment | Recording in the home environment with the<br>report interpretation performed in the<br>clinical setting. | Recording in the home environment with the<br>report interpretation performed in the<br>clinical setting. | Identical to predicate device |
| Target population | 22 years old and older | 22 years old and older | Identical to predicate device |
| Sensor software | Firmware is limited to control the recording<br>and communications processes. No<br>presentation of test results to the patient.<br>Data analyzed and presented in a separate<br>software suite. | Firmware is limited to control the recording<br>and communications processes. No<br>presentation of test results to the patient.<br>Data analyzed and presented in a separate<br>software suite. | Identical to predicate device |
| Analysis software<br>location | Analysis performed off the recording device,<br>exclusively stand-alone by the Belun software | Analysis performed off the recording device,<br>exclusively cloud-based by the NightOwl<br>software | Substantially equivalent to<br>predicate<br>The software is verified and validated.<br>The difference of the analysis software<br>location does not raise different questions of<br>safety and effectiveness. |
| Table 1 Performance Specification Comparison Table between the Belun Sleep System BLS-100 and Predicate Device | | | |
| Characteristic | Proposed Device<br>Belun Sleep System BLS-100 | Predicate Device<br>NightOwl (K220028) | Comparison |
| Analysis software<br>algorithm - AHI | bAHI calculation tuned to the AASM's '1B Rule' for the scoring of hypopnea | pAHI calculation tuned to the AASM's '1A Rule' for the scoring of hypopnea AND<br>pAHI calculation tuned to the AASM's<br>'1B Rule' for the scoring of hypopnea | Substantially equivalent to<br>predicate<br>The clinical evaluation has confirmed that<br>the Belun Sleep System deep-learning<br>algorithms calculating the Belun Apnea<br>Hypopnea Index (bAHI) and Belun Sleep |
| Analysis software<br>algorithm - Sleep<br>stages | bSTAGES: WAKE, NREM and REM<br>bTST calculation (time summation of REM and<br>NREM stages) | Total Sleep Time (TST) calculation<br>Total REM Time calculation | Stage (bSTAGES) generate comparable<br>output to human manual scoring of an<br>Apnea Hypopnea Index (AHI) from<br>Polysomnography (PSG) studies, using<br>American Academy of Sleep Medicine<br>(AASM) scoring guidelines for adult<br>patients, with accuracy, sensitivity and<br>specificity similar to the predicate and<br>reference devices. |
| Sensors | Optical plethysmography sensor,<br>accelerometer | Optical plethysmography sensor,<br>accelerometer | Identical to predicate device |
| Battery | 3.7V Lithium Battery | Battery powered by coin cell | Substantially equivalent to<br>predicate<br>The battery safety is verified and<br>validated. The difference in battery type<br>does not raise different questions of safety<br>and effectiveness |
| Characteristic | Proposed Device<br>Belun Sleep System BLS-100 | Predicate Device<br>NightOwl (K220028) | Comparison |
| Channels | Photoplethysmography (PPG), Pulse rate, Oximetry, Actigraphy | PAT, Pulse rate, Oximetry, Actigraphy | Substantially equivalent to predicate<br>The proposed system and the predicate |
| Wearable sensor<br>location | The photoplethysmography (PPG) sensor and accelerometer components are worn on the proximal phalanx of index finger. | The photoplethysmography (PPG) sensor and accelerometer components are worn on the fingertip. | use the similar sensors: optical plethysmography sensor and accelerometer. The differences between predicate are that the proposed device feeds PPG whose envelop is similar to the information provided by PAT to its AI algorithms. Also, the PPG sensor and accelerometer components are worn on the proximal phalanx of index finger instead of fingertip. The clinical evaluation has confirmed that the Belun Sleep System deep-learning algorithms calculating the Belun Apnea Hypopnea Index (bAHI) and Belun Sleep Stage (bSTAGES) generate comparable output to human manual scoring of an Apnea Hypopnea Index (AHI) from Polysomnography (PSG) studies. As such, the proposed system does not introduce any change to the safety and effectiveness of the predicate device.<br>Similar to Somnapatch (K183625) which uses PPG instead of PAT to aid in evaluating OSA. |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
#### VII. PERFORMANCE DATA
# Non-clinical Test
The proposed device Belun Ring is tested in accordance with the following standards and FDA guidance documents (leveraged from K211407), including:
- . IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- . IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- . IEC 60601-1:2005 + a1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- . IEC 62133:2012 Secondary cells and batteries containing alkaline or other nonacid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications
- . ISO 80601-2-61:2017 Medical electrical equipment -- Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
- ISO 10993-1:2018, Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process
- . ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
- . ISO 10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
- . ISO 10993-10:2010 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity
- . IEC 62366-1:2015 Medical devices — Part 1: Application of usability engineering to medical devices
- . BQB Bluetooth qualification
- FCC Part 15B & FCC Part 18 certification .
- . FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
- . FDA Guidance for the Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
- . FDA Guidance for Applying Human Factors and Usability Engineering to Medical Devices: 2016
{10}------------------------------------------------
The Belun Ring BLR-100X has been validated for low perfusion saturation and pulse rate in K211407. There are no changes in those features or algorithms for clearance of this device to be used as a home sleep apnea test (HSAT).
# Clinical Study
The Clinical Test consists of 2 parts:
- (1) SpO2 accuracy: This testing was performed in K211407 and leveraged for this submission.
- (2) Comparison to PSG sleep lab results: The purpose of this study is to evaluate the accuracy, sensitivity and specificity of the Apnea-Hypopnea Index (AHI) and sleep staging (wake, Rapid Eye Movement (REM) and Non-Rapid Eye Movement (NREM)) of the Belun Sleep System BLS-100 compared with overnight polysomnography (PSG) study according to the American Academy of Sleep Medicine (AASM) guidelines in a sleep laboratory with 106 patients suspected of obstructive sleep apnea (OSA). All sleep studies were manually scored based on the AASM scoring manual (version 2.4) by a senior sleep tech scorer and reviewed by a board-certified sleep physician. All investigators, sleep lab team, and scorers were blinded to the results until statistical analysis was performed. The accuracy, sensitivity and specificity at AHI cutoff 15 and 30 are summarized in the table below.
| Cutoff | Accuracy | Sensitivity | Specificity |
|--------|----------|-------------|-------------|
| 15 | 0.877 | 0.898 | 0.860 |
| 30 | 0.925 | 0.840 | 0.951 |
The accuracy, sensitivity, and specificity of 3-categorization sleep stages [wake, Rapid Eye Movement (REM) and Non-Rapid Eye Movement (NREM)] are summarized in the table below.
| Class | Number of<br>Epochs | Accuracy | Sensitivity | Specificity |
|-------|---------------------|----------|-------------|-------------|
| Wake | 85471 | 0.885 | 0.604 | 0.961 |
| REM | | 0.908 | 0.712 | 0.944 |
| NREM | | 0.827 | 0.904 | 0.695 |
The total sleep time (bTST) of the Belun Sleep System BLS-100 is the time summation of REM and NREM stages. The mean difference between bTST and PSG-TST was 21.8 minutes with a standard deviation of 41.6 minutes. The mean absolute difference between bTST and PSG-TST was 30.8 minutes.
#### VIII. CONCLUSIONS
In conclusion, the proposed device of Belun Sleep System BLS-100 has the same classification information, similar intended use, similar product design and specification as
{11}------------------------------------------------
the predicated device. According to the results of non-clinical test and clinical study, the proposed device is Substantially Equivalent (SE) to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.