K081037 · Itamar Medical · MNR · Jun 9, 2008 · Anesthesiology
Device Facts
Record ID
K081037
Device Name
WATCH-PAT 200I (WP200I)
Applicant
Itamar Medical
Product Code
MNR · Anesthesiology
Decision Date
Jun 9, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.2375
Device Class
Class 2
Indications for Use
The WATCH PAT™ 200i ("WP200i") is a non-invasive home care device for use with patients suspected to have sleep related breathing disorders. The WP200i is a diagnostic aid for the detection of sleep-related breathing disorders and rapid eye movement ("REM") sleep stages. The WP200i generates a peripheral arterial tonometry ("PAT"), respiratory disturbance index ("PRDI"), and PAT REM sleep stage identification ("PREM"). The WP200i's PREM provides supplemental information to its PRDI. The WP200i's PREM is not intended to be used as the sole or primary basis for diagnosing any sleep related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted. The WP200i is not indicated for children less than 17 years old. The WP200i is contraindicated for patients with latex allergy.
Device Story
Watch-PAT 200i is a wrist-worn, non-invasive home diagnostic aid for sleep-related breathing disorders. Inputs include: PAT signal from a finger probe; blood oxygen saturation from an embedded pulse oximeter; motion data from an embedded actigraph. Device uses optopneumatic sensing to measure vasomotor activity in the distal finger. Electronics/controller process these signals to generate PAT, respiratory disturbance index (PRDI), and REM sleep stage identification (PREM). Used in home settings; operated by patients. Healthcare providers review the generated indices to assist in clinical assessment of sleep disorders. PREM is supplemental and not intended as a sole diagnostic basis. Benefits include non-invasive, home-based screening for sleep-related breathing disorders.
Clinical Evidence
Bench testing only. Testing included electrical and electromagnetic safety, software verification and validation, electronics design verification, and performance testing to demonstrate accuracy of captured signals and device reproducibility.
Technological Characteristics
Wrist-worn device with finger-mounted optopneumatic PAT probe, pulse oximeter, and actigraph. Measures vasomotor activity via plethysmography. Compact controller with low power consumption. Software-based signal processing. Connectivity/interoperability details not specified. Sterilization method not specified.
Indications for Use
Indicated for patients >= 17 years old suspected of sleep-related breathing disorders. Contraindicated for patients with latex allergy.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
KOY1037
# 510(k) SUMMARY
JUN - 9 2008
#
| Applicant's Name: | Itamar Medical Itd.<br>2 Ha'eshel st.<br>Caesarea 38900, Israel<br>Tel: +972 4 617 7000<br>Fax: +972 4 627 5598 |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jonathan Kahan, Esq.<br>Hogan & Hartson, L.L.P.<br>Columbia Square<br>555 Thirteenth Street, NW<br>Washington, DC 20004-1109<br>Tel: (202)637-5794<br>Fax: (202)637-5910 |
| | and<br>Dorit Winitz, Ph. D<br>Biomedical Strategy (2004) Ltd.<br>Moshe Aviv Tower, 34th Floor,<br>7 Jabotinsky Street<br>Ramat Gan 52520, Israel<br>Tel: +972-3-612-3281<br>Fax: +972-3-612-3282<br>dorit@ebms.co.il |
| Date Prepared: | March, 2008 |
| Trade Name: | Watch-PAT 200i ("WP200i") |
| Common Name: | Ventilatory Effort Recorder |
. 1
{1}------------------------------------------------
## Classification:
21 CFR 868.2375 Class:II MNR (Ventilatory Effort Recorder).
## Classification Name: Breathing Frequency Monitor
## Medical Specialty: Anesthesiology
#### Predicate Devices:
Watch-Pat100S ("WP100S") (Itamar Ltd), cleared under K042916; product code MNR (ventilatory effort recorder).
## Device Description:
The WATCH-PAT200i System (WP200i) is a non-invasive home care device for use with patients suspected to having sleep related breathing disorders. The WP200i is a diagnostic aid for the detection of sleep related breathing disorders and rapid eye movement ("REM") sleep stag, based on Peripheral Arterial Tonometry (PAT); a noninvasive technology.
The WP200i System is a compact version of the WP100S System, both consist of: (1) a finger PAT probe, which is used to detect the PAT signal; (2) an embedded pulse oximeter using a second probe that is attached to another finger for measuring blood oxygen saturation; (3) an embedded actigraph which is used to determine periods of sleep based on the motion of the wrist; (4) Electronics which include a controller that records the information supplied by the PAT finger probe, oximeter, and actigraph; (5) the device software; and (6) a power supply.
The device is worn on the wrist, and continuously measures the relative state of the vasomotor activity in the distal part of the finger, by a finger-mounted probe based on a plethysmographic method. The measured signal is acquired from a self contained, optopneumatic sensor.
#### Intended Use:
The WATCH PAT™ 200i ("WP200i") is a non-invasive home care device for use with patients suspected to have sleep related breathing disorders. The WP200i is a diagnostic aid for the detection of sleep-related breathing disorders and rapid eye movement ("REM") sleep stages. The WP200i generates a peripheral arterial tonometry ("PAT"), respiratory disturbance index ("PRDI"), and PAT REM sleep stage identification ("PREM"). The WP200i's PREM provides supplemental information to its PRDI. The WP200i's PREM is not intended to be used as the sole or primary basis for diagnosing any sleep related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted. The WP200i is not indicated for children less than 17 years old. The WP200i is contraindicated for patients with latex allergy.
{2}------------------------------------------------
## Performance Data & Substantial Equivalence
The WP100i System is substantially equivalent in all aspects, including tcchnological characteristics, mode of operation, performance characteristics, intended use, etc., to the commercially available WP100S System, cleared under K042916.
The principle changes between the devices include minor hardware modifications that were made to the controller components to enable more compact design and lower power consumption.
A series of safety and performance testing, including bench testing were performed to demonstrate that the modified WP200i System does not raise any new questions of safety and efficacy. These tests include:
- 트 Electrical and electromagnetic testing
- l Software verification and validation
- 트 Electronics design verification test
- 트 Performance testing demonstrating the accuracy of the captured signals and device's reproducibility
Based on these tests results, Itamar Medical Ltd. believes that the WP200i System is substantially equivalent to the cleared WP100S System without raising new safety and/or effectiveness issues.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The seal is simple and monochromatic.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUN - 9 2008
Itamar Medical Limited C/O Mr. Jonathan S. Kahan Regulatory Affairs Consultant Hogan & Hartson, L.L.P. 555 Thirteenth Street, NW Washington, DC 20004-1109
Re: K081037
Trade/Device Name: Watch-PATTM 200i Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: MNR Dated: May 13, 2008 Received: May 13, 2008
Dear Mr. Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 – Mr. Kahan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
#### Indications for Use Statement
510(k) Number (if known): K081037
Device Name: WATCH PATTM 200i
Indications for Use:
The WATCH PATTM 200i ("WP200i") is a non-invasive home care device for use with patients suspected to have sleep related breathing disorders. The WP200i is a diagnostic aid for the detection of sleep-related breathing disorders and rapid eye movement ("REM") sleep stages. The WP200i generates a peripheral arterial tonometry ("PAT"), respiratory disturbance index ("PRDI'), and PAT REM sleep stage identification ("PREM"). The WP200i's PREM provides supplemental information to its PRDI. The WP200i's PREM is not intended to be used as the sole or primary basis for diagnosing any sleep related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted.
The WP200i is not indicated for children less than 17 years old. The WP200i is contraindicated for patients with latex allergy.
Prescription Use X (Per 21 C.F.R. 801.109) AND/OR
Over-The-Counter Use (Per 21 C.F.R. 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) |
|---------------------|
|---------------------|
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
| 510(k) Number: | K08 1037 |
|----------------|----------|
|----------------|----------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.