K080983 · Ventus Medical, Inc. · MNR · Aug 7, 2008 · Anesthesiology
Device Facts
Record ID
K080983
Device Name
PROVENT NASAL CANNULA
Applicant
Ventus Medical, Inc.
Product Code
MNR · Anesthesiology
Decision Date
Aug 7, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2375
Device Class
Class 2
Indications for Use
The Provent Nasal Cannula is indicated for the transmission of respiratory airflow signals between the Provent™ Professional Sleep Apnea Therapy device and the pressure transducers used with physiologic recorders during sleep studies.
Device Story
Provent Nasal Cannula transmits respiratory airflow pressure signals from Provent Professional Sleep Apnea Therapy (PSAT) device to external pressure transducers of physiologic recorders; used during sleep evaluations. Device consists of bifurcated tubing, luer lock connector, and anti-microbial filter; worn with proximal ends in nares and tubing routed over ears/under chin. Functions as passive conduit for pressure signals; enables clinicians to monitor patient airflow during sleep studies while patient uses PSAT therapy. Single-use, disposable design.
Clinical Evidence
Bench testing only.
Technological Characteristics
Bifurcated tubing, luer lock connector, anti-microbial filter. Passive pressure signal transmission via hollow tubing. Single-use, disposable. No electronic components or software.
Indications for Use
Indicated for transmission of respiratory airflow signals between Provent Professional Sleep Apnea Therapy device and pressure transducers used with physiologic recorders during sleep studies.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
Predicate Devices
Salter Labs Bi-NAPS Nasal Airflow and Snore Transducer (K051313)
{0}------------------------------------------------
Korogrz3
:AUG - 7 2008
## 510(K) SUMMARY
| 510(k) Applicant: | Ventus Medical, Inc.<br>1301 Shoreway Rd., Suite 340<br>Belmont, CA 94002<br>(650) 632-4199 (phone)<br>(650) 632-4198 (fax) |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------|
|-------------------|-----------------------------------------------------------------------------------------------------------------------------|
Cindy Domecus. R.A.C. (US & EU) Contact: Domecus Consulting Services LLC (650) 343-4813
Date Summary Prepared: July 17, 2008
Provent™ Nasal Cannula Name of Device:
Nasal Cannula Common Name:
Classification Name: Breathing Frequency Monitor (21 CFR 868.2375, Product Code MNR)
Predicate Device: Predicate devices included in the 510(k) Notification are as follows: K051313. Salter Labs Bi-NAPS Nasal Airflow and Snore Transducer K984431. Braebon Medical Ultima Airflow Sensor
K971501, Bio-logic Airflow Pressure Transducer K053063, Pro-Flow Multipurpose Cannula
Device Description:
The Provent Nasal Cannula was developed to transmit pressure signals from the Provent Professional Sleep Apnea Therapy device to a pressure transducer of a physiologic recorder during sleep evaluations. The Provent Nasal Cannula is specifically designed to attach the Provent Professional Sleep Apnea Therapy (Provent PSAT, K071560) device to standard pressure transducers.
The Provent Nasal Cannula is essentially identical to standard nasal cannulae except for its custom connection to the Provent PSAT. The Provent Nasal Cannula is comprised of the same design elements as a standard nasal cannula:
- 1. a bifurcated tube with a luer lock connector,
- 2. an anti-microbial filter at one end, and
- 3. a component to hold the tubes in the correct location near the nostrils at the other end.
4-1
{1}------------------------------------------------
The Provent nasal cannula is worn in the same way as the standard nasal cannulae, with the proximal ends held in place in the path of nasal airflow and the tubing routed over the ears and under the chin.
Indications for Use:
The Provent Nasal Cannula is indicated for the transmission of respiratory airflow signals between the Provent Professional Sleep Apnea Therapy device and the pressure transducers used with physiologic recorders during sleep studies.
Comparison to Predicate Devices:
The table below shows a comparison the Provent Nasal Cannula to its predicates.
| | Provent<br>Nasal<br>Cannula | Salter Labs<br>BI-NAPS<br>Nasal Airflow<br>and Snore<br>Transducer | Biologic<br>Airflow<br>Pressure<br>Transducer | Braebon<br>Medical Ultima<br>Airflow<br>Sensor | Pro-Flow<br>Multipurpos<br>e Cannula |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The Provent<br>Nasal<br>Cannula is<br>indicated for<br>the<br>transmission<br>of respiratory<br>airflow<br>signals<br>between the<br>Provent<br>Professional<br>Sleep Apnea<br>Therapy<br>device and<br>the pressure<br>transducers<br>used with<br>physiologic<br>recorders<br>during sleep<br>studies. | The Salter<br>Labs Airflow<br>Pressure<br>Transducer is<br>a reusable<br>device<br>intended for<br>use during<br>sleep disorder<br>studies to<br>detect<br>respiratory<br>airflow and<br>snoring for<br>recording onto<br>a<br>polysomnogra<br>phy recorder<br>via nasal<br>pressure<br>changes. | The Bio-logic<br>Airflow<br>Pressure<br>Transducer is<br>indicated for<br>use during<br>sleep disorder<br>studies to<br>detect<br>respiratory<br>airflow for<br>recording on a<br>physiological<br>recorder. It is a<br>battery-<br>powered<br>device, with a<br>disposable<br>nasal cannula<br>which attaches<br>to the patient<br>and plugs into<br>the input of the<br>Airflow<br>Pressure<br>Transducer<br>device. | The Ulitma<br>Airflow<br>Pressure<br>Sensor is<br>intended for<br>use during<br>sleep disorder<br>studies as a<br>qualitative<br>measure of<br>respiratory<br>airflow ... A<br>disposable<br>nasal cannula<br>with a 0.2-<br>micron<br>hydrophobic<br>filter attaches<br>to the patient<br>and connects<br>to the input of<br>the Ultima<br>Airflow<br>Pressure<br>Sensor. | Intended for<br>use for<br>simultaneous detection<br>of<br>respiratory<br>airflow and<br>sampling, or<br>delivery, of<br>gases, such<br>as EtCO2 or<br>oxygen |
| Product<br>Code | MNR | MNR | MNR | BZQ | CCK |
| Purpose | Carry nasal<br>airflow<br>signals | Carry and<br>process nasal<br>airflow signals | Carry and<br>process nasal<br>airflow signals | Carry and<br>process nasal<br>airflow signals | Carry nasal<br>airflow<br>signals and<br>sample or<br>deliver gas |
{2}------------------------------------------------
| | Provent<br>Nasal<br>Cannula | Salter Labs<br>BI-NAPS<br>Nasal Airflow<br>and Snore<br>Transducer | Biologic<br>Airflow<br>Pressure<br>Transducer | Braebon<br>Medical Ultima<br>Airflow<br>Sensor | Pro-Flow<br>Multipurpos<br>e Cannula |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|------------------------------------------------------------------------|------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Interface<br>with nasal<br>air flow | Two open<br>tubing ends<br>which attach<br>to the<br>Provent<br>Professional<br>Sleep Apnea<br>Therapy<br>device,<br>which, in turn<br>enters the<br>nares | Two open<br>tubing ends on<br>nasal prongs<br>which enter<br>the nares | Two open<br>tubing ends on<br>nasal prongs<br>which enter the<br>nares | Two open<br>tubing ends on<br>nasal prongs<br>which enter the<br>nares | Two open<br>tubing ends<br>on nasal<br>prongs<br>which enter<br>the nares |
| Cannula<br>interface<br>with<br>pressure<br>transducer | Single Luer<br>fitting | Single Luer<br>fitting | Single Luer<br>fitting | Single Luer<br>fitting | One Luer<br>fitting for<br>each<br>purpose<br>(one for<br>airflow<br>monitoring<br>and one for<br>gas<br>delivery/sam<br>pling) |
| Method of<br>pressure<br>signal<br>transmissio<br>n | Hollow<br>tubing | Hollow tubing | Hollow tubing | Hollow tubing | Hollow<br>tubing |
| Reuse | Single use,<br>disposable | Cannula<br>portion is<br>single use,<br>disposable | Cannula<br>portion is single<br>use, disposable | Cannula<br>portion is single<br>use, disposable | Single use,<br>disposable |
Table 4-1. Comparison of the Provent Nasal Cannula with predicate devices
## Performance Data:
Bench data were submitted to support the 510(k) Notification. The bench testing demonstrated that the device is safe and effective for its intended use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ventus Medical, Incorporated C/O Ms. Cindy Domecus Principal Domecus Consulting Services LLC 1301 Shoreway Road, Suite 340 Belmont, California 94002
AUG - 7 2008
Re: K080983
Trade/Device Name: Provent™ Nasal Cannula Regulation Number: 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: MNR Dated: July 17, 2008 Received: July 21, 2008
Dear Ms. Domecus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Clare
Chiu S. Lin, Ph. D Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number (if known):
Device Name: Provent™ Nasal Cannula
Indications for Use:
The Provent Nasal Cannula is indicated for the transmission of respiratory airflow signals between the Provent™ Professional Sleep Apnea Therapy device and the pressure transducers used with physiologic recorders during sleep studies.
Prescription Use × Prescription Use
(Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K080983
3-1