The Somté PSG is intended for use in the recording, displaying, analysis, printing and storage of human biological parameters such as heart and muscle activity, eye movement, breathing and body movements to assist in the diagnosis of various sleep disorders or sleep related respiratory or cardiac disorders. The Somté PSG is designed for ambulatory and mobile operation and can be used in either the patient's home, the hospital or other environments, thus enabling patients to be investigated under as realistic conditions as possible. The Somté PSG is only to be used under the direction of a physician.
Device Story
Somté PSG System records, displays, and analyzes physiological signals including EEG, EOG, ECG, respiratory effort, airflow, and body position. Device consists of a Patient Input Box (PIB) and a Recorder connected by cable; powered by 2 AA batteries. Used in clinical or home settings by physicians or trained staff to aid in diagnosing sleep disorders. Data is stored on CF cards (up to 2GB) and transferred to a PC for review, analysis, and report generation via Somté ECG Analysis software. Remote surveillance capabilities allow monitoring of lead quality and recording parameters via broadband. Provides objective physiological data to assist clinicians in diagnostic decision-making, enabling investigation under realistic patient conditions.
Clinical Evidence
Bench testing only. Performance validation included internal and external sleep authority review (Woody Clinical Validation Report, Vampire Heart Rate Verification Report). Compliance testing performed for hardware, software, and firmware against design requirements. Safety and electromagnetic compatibility testing conducted per IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
Portable, battery-powered (2 AA) system with Patient Input Box and Recorder. Features monochrome graphic LCD with 3 push buttons. Supports up to 19 channels; 16-bit A/D resolution. Connectivity via IrDA and Bluetooth. Storage on CF card (up to 2GB). Sampling rates 1-1024 Hz. Software-based analysis for ECG and sleep parameters.
Indications for Use
Indicated for patients requiring diagnostic investigation of sleep disorders or sleep-related respiratory or cardiac disorders. Used in hospital or home environments under physician direction.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
K072201
page 1 of 5
# 4 510(k) SUMMARY
APR 1 4 2008
In accordance with 21 CFR 807.92, the following information constitutes the Compumedics Pty Ltd summary for the SOMTÉ PSG SYSTEM.
| SUBMITTER'S NAME: | Compumedics Pty Ltd |
|---------------------|--------------------------------------------------------------------|
| ADDRESS: | 30-40 Flockhart Street,<br>Abbotsford, Victoria, 3067<br>Australia |
| CONTACT PERSON: | Gordon Ferguson |
| TELEPHONE NUMBER: | +61 3 8420 7300 |
| FAX NUMBER: | +61 3 8420 7399 |
| DATE OF SUBMISSION: | 3May07 |
## 1. Identification of device
| Proprietary Name: |
|------------------------|
| Common Name: |
| Classification Status: |
| Product Codes: |
Compumedics Somté PSG System SOMTE PSG SYSTEM Class II per regulations 868.2375 MINR
### 2. Predicate devices
Compumedics Limited believes the SOMTÉ PSG SYSTEM is substantially equivalent to:
| Device name: | Compumedics Sleep Monitoring System |
|----------------|-------------------------------------|
| Supplier: | Compumedics Limited |
| 510(k) number: | K955841. |
| Device name: | Compumedics Somté Recording System |
| Supplier: | Compumedics Limited |
| 510(k) number: | K021176. |
#### Description of the Device 3.
The Somté PSG System comprises hardware and software which provides up to 27 separate parameters for recording, review and analysis of collected and stored physiological parameters, including EEG, EOG, ECG and respiratory signals, which are then used as an aid in the diagnosis of respiratory and/or cardiac related sleep disorders by qualified physicians.
The Somté PSG device consists of two components permanently connected by a cable: the Patient Input Box (PIB) and the Recorder. The Somté PSG device is powered by 2 AA Batteries - Alkaline or NiMh.
{1}------------------------------------------------
# 4. Intended use
The Somté PSG is intended for use by or under the direction of physicians in private practices or hospital environments to assist in the diagnosis of various sleep disorders or sleep related respiratory or cardiac disorders.
: :
# Technological characteristics, comparison to predicate device. 5.
# COMPARATIVE TABLE
| Characteristic | Compumedics<br>Somté PSG<br>Recording System | Compumedics<br>P-Series<br>Sleep Monitoring<br>System<br>(predicate device) | Compumedics<br>Somté Recording<br>System<br>(predicate device) |
|----------------------------------------------------------------------|------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Intended Use | Sleep/Respiratory<br>recording and<br>analysis,<br>Ambulatory ECG<br>recording and<br>analysis | Sleep/Respiratory<br>recording and<br>analysis | Sleep/Respiratory<br>recording and<br>analysis,<br>Ambulatory ECG<br>recording and<br>analysis |
| 510(k) number | | K955841 | K021176 |
| Environmental | 5 to 45 deg C<br>20 to 90 % relative<br>humidity<br>less than 3000m<br>altitude | 0 to 45 deg C<br>30 to 95% relative<br>humidity<br>less than 3000m<br>altitude | 0 to 45 deg C<br>30 to 95% relative<br>humidity<br>less than 3000m<br>altitude |
| Local User Interface | Monochrome graphic<br>LCD, 3 push buttons | Monochrome graphic<br>LCD, 3 push buttons | Monochrome graphic<br>LCD, 3 push buttons |
| Physical<br>dimensions | Recorder:113x65x30<br>mm<br>PIB: 53 x 133 x 25<br>mm | Recorder:<br>240x120x58mm<br>PIB: 127x66x27mm | Recorder:116x65x30<br>mm<br>PIB: 78 x 45 x 22<br>mm |
| Weight | 270 grams (including<br>batteries) | 1570 grams<br>(including battery) | 250 grams (including<br>batteries) |
| Number of Patients<br>Can Record<br>Simultaneously | 1 per Unit<br>2 Units per PC | 1 per Unit<br>2 Units per PC | 1 per Unit |
| Portable Design | Yes | Yes | Yes |
| Patient worn device | Yes | Yes | Yes |
| Main Recording unit | Yes | Yes | Yes |
| Patient Interface<br>Box | Yes | Yes | Yes |
| Use for in-lab<br>monitored studies | Yes | Yes | No |
| Use for ambulatory<br>studies at home | Yes | Yes | Yes |
| General<br>Functionalities: | | | |
| Data Collection | Yes | Yes | |
| Displays Raw Data<br>during Recording | Yes, PC during in-lab<br>use, or LCD on unit | Yes, PC during in-lab<br>use, or LCD on unit | Yes, on LCD on unit |
| Built in Display for<br>waveform preview | Yes - Integrated LCD<br>including full<br>disclosure waveform<br>preview | Yes - Integrated<br>LCD including full<br>disclosure waveform<br>preview | Yes - LCD |
| Characteristic | Compumedics<br>Somté PSG<br>Recording System | Compumedics<br>P-Series<br>Sleep Monitoring<br>System<br>(predicate device) | Compumedics<br>Somté Recording<br>System<br>(predicate device) |
| Data display on PC<br>for interpretation | Yes, during or after<br>recording | Yes, during or after<br>recording | Yes, after recording<br>only |
| Data Analysis | Optional | Optional | Optional |
| Report Generation | Optional | Optional | Optional |
| Capable of Data<br>Transfer for<br>Analysis and Report<br>Generation | Yes | Yes | Yes |
| Data Inputs | Generic and custom<br>sensors. Data inputs<br>as listed below | Generic and custom<br>sensors. Data inputs<br>as listed below | Generic and custom<br>sensors. Data inputs<br>as listed below |
| Signal Conditioning | Yes | Yes | Yes |
| Data Analysis<br>(Computer,<br>Computer-Assisted<br>or manual). | Optional | Optional | Optional |
| Comprehensive<br>Report Generation | Optional | Optional | Optional |
| Remote Sleep<br>Surveillance | Yes, via broadband | Yes, via dial-up | No |
| Remote Capability<br>to Monitor Lead<br>Quality | Yes | Yes | No |
| Remote Capability<br>to Monitor<br>Recording<br>Parameters | Yes | Yes | No |
| Look<br>Forward/Watch<br>Back Display | Yes | Yes | No |
| Data Recorded: | | | |
| # Channels of Data<br>Recorded | Up to 19 | Up to 18 | Up to 13 |
| Respiratory Effort<br>Channels | 2 Inductive<br>plethysmography | 2 Inductive<br>plethysmography | 2 Inductive<br>plethysmography |
| Airflow | 1 Nasal Cannula | 1 Nasal Cannula | 1 Nasal Cannula |
| Pressure | 1 Nasal Cannula | 1 Nasal Cannula | 1 Nasal Cannula |
| Snore<br>Body Position | From Cannula<br>1 built in | From Microphone<br>1 | From Cannula<br>1 |
| SaO2 | Yes | Yes | Yes |
| Heart Rate | Yes | Yes | Yes |
| Pulse Wave | Yes | No | Yes |
| Oximeter status<br>(signal quality) | Yes | Yes | Yes |
| ECG Channels | Up to 2 | 1 | Up to 2 |
| EEG Channels | 2 or 3 | Up to 2 | Up to 2 |
| Chin EMG | Up to 2 | 1 | Up to 2 |
| Diaphragmatic EMG | Up to 1 | optional | Up to 2 |
| EOG | 2 | 2 | Up to 2 |
| Leg Movement | Yes 2 Piezo Electric | Yes 2 Piezo Electric | Yes 2 Piezo Electric |
| Characteristic | Compumedics<br>Somté PSG<br>Recording System | Compumedics<br>P-Series<br>Sleep Monitoring<br>System<br>(predicate device) | Compumedics<br>Somté Recording<br>System<br>(predicate device) |
| Aux AC | 1 (or DC) | 1 | No |
| Aux DC | 1 (or AC) | option | No |
| Impedance Check | Yes | Yes | No |
| Power Source | 2 AA Batteries –<br>Alkaline or NiMh | 7.2V NiMh | 2 AA Batteries –<br>Alkaline or NiMh |
| Maximum recording<br>duration | 30 hours | 15 hours | 30 hours |
| Storage Media | CF Card up to 2Gb | CF Card up to<br>128Mb | CF Card up to 2Gb |
| Channel Sampling<br>Rates | 1-1024 Hz channel<br>dependent | 1-256 Hz channel<br>dependent | 1-1024 Hz channel<br>dependent |
| Channel Storage<br>Rates | 1-256 Hz channel<br>dependent | 1-256 Hz channel<br>dependent | 1-256 Hz channel<br>dependent |
| A/D Vertical<br>Resolution (in Bits) | 16 | 8 or 12 | 16 |
| Average study size<br>per 8 hour study | 100Mb | 30Mb | 42Mb |
| Ability to calibrate<br>DC channels | Yes | Yes | Yes |
| Communications to<br>PC | IrDA & Bluetooth | Comms port | Not applicable |
| When used for<br>ECG: | | | |
| Number of ECG<br>channels | 1 or 2 | Not applicable | 1 or 2 |
| Sensors | Standard ECG<br>electrodes | Not applicable | Standard ECG<br>electrodes |
| Sample rate | 1024 | Not applicable | 1024 |
| Recording rate | 256 | Not applicable | 256 |
| Frequency<br>response | 0.048Hz to 102Hz<br>(-3dB) | Not applicable | 0.05Hz to 100Hz<br>(-3dB) |
| Full scale range | 10mV p-p | Not applicable | 10mV p-p |
| ECG analysis<br>software | Somté ECG Analysis<br>V1.02 | Not applicable | Somté ECG Analysis<br>V1.02 |
{2}------------------------------------------------
:
.
.
{3}------------------------------------------------
:
.
・
-
{4}------------------------------------------------
#### 6. Discussion of performance testing.
An extensive collection of tests has been conducted and successfully completed, including:
- Safety Tests to conform to IEC 60601-1, to ensure that there is no A potential for detrimental effects on patients, other persons, animals or the surroundings. (refer to IEC60601-1 rpt-46275.1-6Dec06-SomtePSG-IEC60601-1 rpt). - refer to Attachment F.
- A Electromagnetic Compatability Tests to IEC 60601-1-2 to ensure no intolerable magnetic disturbances are introduced into its electromagnetic environment. (refer to IEC60601-1-2 rpt-M060825-21Sep06-SomtePSG-IEC60601-1-2 rpt). - refer to Attachment F.
- Performance Validation was conducted, results were recorded and an A internal and external Sleep authority generated a Somté PSG Performance Validation Report (refer to AC550-01 Woody Clinical Validation Report, AD052-01 Vampire Heart Rate Verification Report and AC551-01 Report to EN60601-1-4 for Woody-Somte PSG). ). refer to Attachment D.
- ﺮ ﺍﻟﻤﺮﺍﺟﻊ Compliance tested to hardware, software and firmware test specifications to ensure conformance to all design requirements. ). refer to Attachments C & D.
### 7. Conclusion
Based on extensive performance testing and a comparison to the predicate device, it is the conclusion of Compumedics Limited that the SOMTE PSG SYSTEM is substantially equivalent to devices already on the market (cleared by the 510(k) process) and presents no new concerns about safety and effectiveness.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Gordon Ferguson Compumedics Pty Limited 30-40 Flockhart Street Abbotsford, Victoria 3067 AUSTRALIA
APR 1 4 2008
Rc: K072201
Trade/Device Name: Somté PSG System Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: MNR Dated: April 10, 2008 Received: April 11, 2008
Dear Mr. Ferguson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Ferguson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sutte Y. Michael Davis
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# 3 STATEMENT OF INDICATIONS FOR USE
510(k) Number: K072201
Device Name: SOMTÉ PSG SYSTEM
Indications for Use:
The Somté PSG is intended for use in the recording, displaying, analysis, printing and storage of human biological parameters such as heart and muscle activity, eye movement, breathing and body movements to assist in the diagnosis of various sleep disorders or sleep related respiratory or cardiac disorders.
The Somté PSG is designed for ambulatory and mobile operation and can be used in either the patient's home, the hospital or other environments, thus enabling patients to be investigated under as realistic conditions as possible. The Somté PSG is only to be used under the direction of a
Prescription Use
(Per 21 CFR 801 Subpart D)
physician.
AND/OR
Over the Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M. N.
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.