The Embla N7000 is intended for use by a physician or trained technician for the acquisition of EEG and polysomnography (PSG) signals and transmission of these signals to a PC during neurophysiologic or sleep examinations. The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments, The use of the Embla N7000 system does not involve any patient monitoring or diagnosis.
Device Story
Embla N7000 is a full polysomnography and EEG acquisition system. It collects EEG, bipolar, and respiratory signals, plus auxiliary inputs (e.g., CPAP, CO2 machines). Device amplifies and digitizes signals, transmitting them via Ethernet to a PC for external analysis. Used in hospitals, sleep labs, and clinics by physicians or technicians. System does not perform patient monitoring or diagnosis; it serves as a data acquisition front-end. Benefits include streamlined cabling and flexible, rugged design for clinical/research environments. Healthcare providers use the transmitted data on a PC for subsequent review and clinical decision-making.
Clinical Evidence
Bench testing only. Validation included design review, internal verification, and practical testing in a simulated sleep lab setting. Hazard analysis performed and verified. External testing by SEMKO confirmed compliance with EMC requirements and electrical safety standards (UL2601).
Technological Characteristics
System features 32 referential EEG channels, 8 bipolar channels, and respiratory signal inputs. Uses internal microprocessor and Flash memory. Connectivity via standard Ethernet with signal-transformer isolation. Powered by internal mains isolation transformer. Complies with EN60601-1, UL-2601, and EN60601-1-2. Software level of concern is minor.
Indications for Use
Indicated for patients undergoing neurophysiologic or sleep examinations requiring acquisition of EEG and polysomnography signals. Intended for use by physicians or trained technicians in clinical or research settings. No specific patient population exclusions or contraindications noted.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the word "Medcare" in bold black font. Above the word, there is a handwritten string of characters, "K024322". The handwritten string is oriented diagonally in the upper left corner of the image.
OCT 2 7 2003
### 510(k) Summary
December 19th, 2002
#### Submitter
Medcare Flaga Vesturhlid 7 105 Reykjavik Iceland Europe
Tel: 011 354 510 2000 Fax: 011 354 510 2010
Contact Person: Berglind Hallgrímsdóttir
e-mail: berglind@medcare.is, guality@medcare.is
#### Device
Trade Name: Common Name: Classification Product Code Classification Panel Embla N7000 EEG Amplifier Electroencephalograph, Class II GWQ Neurology
#### Substantial Equivalence
Artisan™ Acquisition System For EEG and polysmnography.
Airsep Corporation
290 Creekside Drive Buffalo, NY 14228
#### 510(k) Number K992283
#### Description of the Device
The Embla N7000 system is a full polysomnography system with expanded EEG capability. It is used to perform online sleep studies in the sleep lab, hospital or clinical environment under the supervision of a clinician or sleep technician. It does not do the recording, monitoring or analysis of this data.
The Embla N7000 system integrates digital technology and precision engineering into a flexible, rugged, full polysomnography system. The system itself, featuring an Ethernet network connection, is simple to assemble and the cables have been streamlined to provide a comfortable and reliable system. Meeting the most demanding clinical and research needs by offering 32 referential EEG channels, 8 bipolar channels plus an extensive set of respiratory signals, the system also includes auxiliary inputs for additional devices such as CPAP and CO2 machines.
{1}------------------------------------------------
# Medcare
#### Intended Use of the device
The Embla N7000 is intended for use by a physician or trained technician for the acquisition of EEG and polysomnography (PSG) signals and transmission of these signals to a PC during neurophysiologic or sleep examinations. The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments,
The use of the Embla N7000 system does not involve any patient monitoring or diagnosis.
#### Technological Charaterics
Both Artisan and Embla N7000 are used for amplifying EEG and sleep signals in clinical and research laboratories. Both are designed for collection and transmission of sleep data to a personal computer. Both systems are only acquisition systems from amplifying and digitizing EEG and other polygraph channels and their transmission to a PC.
The amplification and data acquisition functions are similar for both systems. The main difference is the extended hardware possibilities of re-referencing in the Embla N7000. Also the Embla N7000 is controlled by software that is stored in internal Flash memory and runs on internal microprocessors. Artisans programs are also stored in internal Flash memory and run on a PLD (Programmable Logic Device). The difference between a microprocessor and a PLD is mainly how the commands are processed in each device. The software for PLDs sets up the configuration of the PLDs electronics to make it perform certain logical operations. The software for microprocessors performs the same logical operations by a sequence of instructions executed in the processors CPU. The software development was done according to the firmware development process at Medcare Flaga and it was verified that the software requirements are met. The software level of concern for both products was minor.
The patient mains-isolation is equivalent for the two devices. The N7000 has internal mains isolation transformer to power the device. The Artisan is powered from external mains transformer that provides the same isolation. In both cases the mains isolation fulfills the reinforced requirements of over 4000V manufacturing test voltage, set by the EN60601-1 and UL-2601.
The isolation from external devices is done in similar way in both devices. In Artisan, a special external box containing AD converters and opto-couplers are used to isolate the signals, but in N7000 the similar converters and opto-couplers are built in the device.
The construction of the communication is also equivalent. Artisan uses fiber-optic cable for the isolation, using a proprietary communication standard for the point to point communications with the PC. The N7000 uses a signal-transformer to isolate the communications and communicates over standard Ethernet with the PC.
#### Test
The Embla N7000 has been tested and verified in various phases, the testing included design review, internal verification and validation as well as external testing/validation. The design was verified throughout the design process. Internal validation included bench testing and practical testing in a simulated sleep lab setting. Hazard analysis was done and appropriate measures were implemented and their effectiveness was verified. External test house, SEMKO, was used to confirm compliance to EMC requirements and standards for electrical safety including UL2601.
Testing performed and the results verify that the Embla N7000 is safe and effective for its intended use.
{2}------------------------------------------------
# Medcare
#### Compliance with guidelines and standards
Embla N/S7000 complies with the FDA "Electroencephalograph Devices Guidance for 510(k) Content, Draft Document, Version 1.0"
Embla N/S7000 is certified to carry the CE mark (CE 0413). The CE mark is a declaration that Embla N7000 is in compliance with the directive set forth by the European Union for medical devices.
Embla N/S7000 is manufactured by Medcare Flaga and the system conforms to the following standards and requlations:
Class IIa, Medical Device Directive (MDD), Annex V, "EC Directive" 93/42/EEC
EN60601-1:1990 with A1 and A12:1993, A2:1995 and A13:1996. (IEC 60601-1, 2nd ed., 1988 with A1,1991 and A2, 1995) : MEDICAL ELECTRICAL EQUIPMENT -PART 1: GENERAL REQUIREMENTS FOR SAFETY
EN60601-1-2:2001 MEDICAL ELECTRICAL EQUIPMENT -PART 1: GENERAL REQUIREMENTS FOR SAFETY - 2. COLLATERAL STANDARD: ELECTROMAGNETIC COMPATIBILITY -REQUIREMENTS AND TESTS
EN 61000-3-2:1995 and A14:2000. ELECTROMAGNETIC COMPATIBILITY (EMC) - PART 3-2: LIMITS - LIMITS FOR HARMONIC CURRENTS EMISSIONS (EQUIPMENT INPUT CURRENT UP TO AND INCLUDING16 A PER PHASE)
EN 61000-3-3:1995, ELECTROMAGNETIC COMPATIBILITY (EMC) - PART 3-3: LIMITS -LIMITATION OF VOLTAGE FLUCTUATIONS AND FLICKER IN LOW-VOLTAGE SUPPLY SYSTEMS FOR EQUIPMENT WITH RATED VOLTAGE UP TO 16 A
EN865:1997 PULSE OXIMETERS - PARTICULAR REQUIREMENTS
IEC 60601-2-25:1993 PARTICULAR REQUIREMENTS FOR THE SAFETY OF ELECTROCARDIOGRAPHS
60601-2-26:1994 PARTICULAR FOR IEC REQUIREMENTS THE SAFETY OF ELECTROENCEPHALOGRAPHS
IEC-60601-2-40:1998 PARTICULAR REQUIREMENTS FOR THE SAFETY OF ELECTROMYOGRAPHS AND EVOKED RESPONSE EQUIPMENT
EN60601-1-4 MEDICAL ELECTRICAL EQUIPMENT - PART 1: GENERAL REQUIREMENTS FOR SAFETY - 4. COLLATERAL STANDARD: PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS
UL 2601-1: MEDICAL ELECTRICAL EQUIPMENT, PART 1:GENERAL REQUIREMENTS FOR SAFETY, 2nd ed. 1997
CAN/CSA C22.2 No.601.1-M90 MEDICAL ELECTRICAL EQUIPMENT PART 1: GENERAL REQUIREMENTS FOR SAFETY
AS/NZS 3200.1.0: 1998 MEDICAL ELECTRICAL EQUIPMENT - GENERAL REQUIREMENTS FOR SAFETY - PARENT STANDARD
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines that suggest the wings and body of the bird.
Public Health Service
OCT 27 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Kumar Kulkarni Medcare Flaga c/o RTS, Incorporated 46 Loch Haven Lane Battle Creek, MI 49015
Re: K024322
Trade/Device Name: Embla N7000 Regulation Number: 21 CFR 868.2375 Regulation Name: Ventilatory Effort Recorder Regulatory Class: II Product Code: MNR Dated: July 27, 2003 Received: July 31, 2003
Dear Mr. Kulkarni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Mr. Kumar Kulkarni
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Joflis
Chiu Lin, Ph.D.
fer
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
(Optional Format 3-10-98)
Applicant: Medare Flaga
510(k) Number (if known): K024322
Device Name: Embla N7000
Indications For Use:
The Embla N7000 is intended for use by a physician or trained technician for the acquisition of EEG and polysomnography (PSG) signals and transmission of these signals to a PC during neurophysiologic or sleep examinations. The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments.
The use of the Embla N7000 system does not involve any patient monitoring or diagnosis.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Q. Patricku 10/24/03
(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
510(k) Numb
5
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.