Reusable, Finger Clip Nonin Medical CO-Met Oximetry Sensor
K172625 · Nonin Medical, Inc. · DQA · May 11, 2018 · Cardiovascular
Device Facts
Record ID
K172625
Device Name
Reusable, Finger Clip Nonin Medical CO-Met Oximetry Sensor
Applicant
Nonin Medical, Inc.
Product Code
DQA · Cardiovascular
Decision Date
May 11, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Multi-Sensing Oximetry System when used with Model 8330AA sensor, is intended for noninvasive measuring of functional oxygen saturation of arterial hemoglobin (%SpO2), carboxyhemoglobin saturation (%COHb), and pulse rate. The measurements may be multiple spot-checks to observe change and/or continuous monitoring. This device is not meant for sole use in clinical decision making; it must be used in conjunction with additional methods of assessing clinical signs and symptoms. • For %SpO2 and pulse rate, the multi-sense system is intended for use in professional healthcare facilities, mobile, and EMS settings in adult and pediatric patients (> 66 lbs/30 kg). • For %COHb, the multi-sense system is intended for use in professional healthcare facilities in adult and pediatric patients (> 66 lbs/30 kg). This device should not be used to screen, diagnose or make treatment decisions related to suspected carbon monoxide poisoning.
Device Story
Multi-Sensing Oximetry System with Model 8330AA sensor uses noninvasive optical sensing to measure functional oxygen saturation (%SpO2), carboxyhemoglobin saturation (%COHb), and pulse rate. Device inputs are optical signals from finger clip sensor; signal processor transforms inputs into real-time data output. Used in professional healthcare, mobile, and EMS settings by trained personnel. Output displayed on monitor as numerical values and pleth waveform; assists clinicians in patient assessment alongside other clinical signs. Benefits include continuous or spot-check monitoring of physiological parameters. System includes Bluetooth, serial RS-232, and nurse call/analog output interfaces.
Clinical Evidence
Clinical testing conducted per ISO 80601-2-61 on healthy adult subjects (male/female, non-smoking, various skin tones). COHb accuracy measured over 0-15% range with 80-100% SaO2. SpO2 accuracy measured over 70-100% range with 0-15% COHb. Results demonstrated compliance with acceptance criteria and substantial equivalence to predicate.
Technological Characteristics
Reusable finger clip sensor; optical sensing principle. Powered by 7.2V Li-ion battery or 100-240 VAC adapter. Connectivity: Bluetooth, Serial RS-232, Nurse Call/Analog. Enclosure IP32. Standards: IEC 60601-1, IEC 60601-1-2, ISO 80601-2-61, IEC 62304.
Indications for Use
Indicated for noninvasive measurement of %SpO2, %COHb, and pulse rate in adult and pediatric patients (> 66 lbs/30 kg). %SpO2 and pulse rate indicated for professional healthcare, mobile, and EMS settings. %COHb indicated for professional healthcare facilities. Contraindicated for screening, diagnosing, or making treatment decisions for suspected carbon monoxide poisoning.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Model X-100C System with Model 8300AA Sensor (K160231)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 11, 2018
Nonin Medical Inc. Kim Aves Principal Regulatory Affairs Specialist 13700 1st Avenue North Plymouth, Minnesota 55441
Re: K172625
Trade/Device Name: Multi-Sensing Oximetry System with Model 8330AA Sensor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: April 3, 2018 Received: April 9, 2018
Dear Kim Aves:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tina Kiang
s
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172625
Device Name
Multi-Sensing Oximetry System with Model 8330AA Sensor
#### Indications for Use (Describe)
The Multi-Sensing Oximetry System when used with Model 8330AA sensor, is intended for noninvasive measuring of functional oxygen saturation of arterial hemoglobin (%SpO2), carboxyhemoglobin saturation (%COHb), and pulse rate. The measurements may be multiple spot-checks to observe change and/or continuous monitoring. This device is not meant for sole use in clinical decision making; it must be used in conjunction with additional methods of assessing clinical signs and symptoms.
· For %SpO2 and pulse rate, the multi-sense system is intended for use in professional healthcare facilities, mobile, and EMS settings in adult and pediatric patients (> 66 lbs/30 kg).
· For %COHb, the multi-sense system is intended for use in professional healthcare facilities in adult and pediatric patients (> 66 lbs/30 kg). This device should not be used to screen, diagnose or make treatment decisions related to suspected carbon monoxide poisoning.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☑</span> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☐</span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. "510(k) Summary" as required by section 807.92(c)
| Submitter: | Nonin Medical, Inc.<br>13700 1st Ave. North<br>Plymouth, MN 55441-5443 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Phone: | 763 577 3196 |
| Fax: | 763 553 7807 |
| Contact Person: | Kim E. Aves<br>Principal Regulatory Affairs Specialist |
| Date Prepared: | 10 May, 2018 |
| Trade Names: | Multi-Sensing Oximetry System with Model 8330AA Sensor |
| Common Name: | Oximeter |
| Classification Name: | Oximeter |
| Regulation Number: | Class II, 21 CFR 870.2700 (Oximeter) |
| Product Code, Panel: | DQA, Anesthesiology |
| Predicate Device(s): | Model X-100C System with Model 8300AA Sensor (K160231) |
| | The Model 8300AA reusable finger clip sensor is intended for noninvasive<br>measuring of functional oxygen saturation of arterial hemoglobin<br>(%SpO2), carboxyhemoglobin saturation (%COHb), and pulse rate of<br>adult and pediatric patients (> 66 lbs / 30 kg). The measurements may be<br>multiple spot-checks to observe change and/or continuous monitoring. It<br>is intended for use in Emergency Medical Service (EMS), hospitals,<br>medical facilities, and mobile environments. This device is not meant for<br>sole use in clinical decision making; it must be used in conjunction with<br>additional methods of assessing clinical signs and symptoms. |
| Device Description: | Nonin Medical's Multi-Sensing Oximetry system, when used with Model<br>8330AA reusable finger clip sensor, is based on Nonin's SenSmart™<br>technology platform, displays (%COHb), pulse oxygen saturation (%SpO2),<br>and pulse rate (PR) data transmitted from one channel of data through a<br>direct connection of a signal processor to the monitor. The model<br>8330AA sensor provides real time data output of COHb, SpO2, and PR. |
| Intended Use: | Multi-Sensing Oximetry System with 8330AA Sensor |
| | The Multi-Sensing Oximetry System when used with Model 8330AA<br>sensor, is intended for noninvasive measuring of functional oxygen<br>saturation of arterial hemoglobin ( $%SpO2$ ), carboxyhemoglobin<br>saturation ( $%COHb$ ), and pulse rate. The measurements may be multiple<br>spot-checks to observe change and/or continuous monitoring. This<br>device is not meant for sole use in clinical decision making; it must be<br>used in conjunction with additional methods of assessing clinical signs<br>and symptoms. |
| | • For $%SpO2$ and pulse rate, the multi-sense system is intended for use<br>in professional healthcare facilities, mobile, and EMS settings in adult and<br>pediatric patients (> 66 lbs/30 kg). |
| | • For $%COHb$ , the multi-sense system is intended for use in professional<br>healthcare facilities in adult and pediatric patients (> 66 lbs/30 kg). This<br>device should not be used to screen, diagnose or make treatment<br>decisions related to suspected carbon monoxide poisoning. |
| Testing: | The Nonin Model 8330AA sensor when used with Multi-Sensing<br>Oximetry System is supported by electrical safety testing,<br>electromagnetic compatibility, device performance, and clinical testing<br>to ensure appropriate functionality and to demonstrate substantial<br>equivalence to the predicate devices. |
| Functional Testing: | The results of the testing demonstrate equivalency with the predicate<br>devices and compliance to current recognized standards. Table 1<br>summarizes test results for the proposed devices, which met the relevant |
{4}------------------------------------------------
| Table 1 | | |
|-------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Test | Reference | Results |
| Strain Relief Pull Strength | Internal | Pass |
| Operating Temp & Humidity | IEC 60601-1:2005 Sub-Clause 7.9.3.1,<br>EN1789:2007 Sub-clause 6.3.2<br>IEC 60601-1-11:2010 Sub-clause 4.2.2. | Pass |
| Thermal Shock | IEC 60601-1-11:2010 Sub-clause 4.2.3 | Pass |
| Drop & Swing<br>(Free Fall) | IEC 60601-1:2005 Sub-clauses 15.3.4.1 and 15.3.4.2<br>ISO 80601-2-61:2011 Sub-Clause 201.15.3.5.101.2<br>IEC 60601-1-11:2010 Sub-clause 10.1.3<br>IEC 60601-1-12:2014 Sub-clause 10.1.3<br>EN 1789:2007 Sub-clause 6.4.2 | Pass |
| ESU Interference | ANSI/AAMI EC 13:2002/(R) 2007 | Pass |
| Defib Proof | IEC 60601-1:2005 Clause 8.5.5 | Pass |
| Storage & Transport Tempo & Humidity | IEC 60601-1-11:2010 Sub-clause 4.2.1 & 4.2.2<br>IEC 60601-1-12 Sub-clause 4.2.1 & 4.2.2<br>IEC60601-1:2005/Al:2012 Sub-clause 7.9.3.1 | Pass |
| Dielectric Strength, Patient Leakage<br>Current, Pre-conditioning | IEC 60601-1:2005 Sub-clause 5.7, 6.3 8.7 & 8.8.3 | Pass |
| Excessive Temp | IEC 60601-1:2005 Sub-clause 11.1.2 | Pass |
| Hyperbaric Pressure | Internal | Pass |
requirements of the applicable recognized standards.
{5}------------------------------------------------
| Sensor Image SW V&V | FDA 938:2002<br>IEC 62304:2008 | Pass |
|---------------------|-----------------------------------|------|
| CoHb Clinical Study | ISO 14155<br>ISO 80601-2-61 (REF) | Pass |
| SpO2 Clinical Study | ISO 14155<br>ISO 80601-2-61 (REF) | Pass |
| Pulse Rate Testing | ISO 80601-2-61 (REF) | Pass |
#### Accuracy (clinical) testing:
#### COHb Accuracy Testing
COHb accuracy testing was conducted at an independent research laboratory on healthy, male and female, non-smoking, light to darkskinned subjects that were 18 years of age and older.
The measured carboxyhemoglobin value (%COHb) of the sensors was compared to simultaneous arterial blood samples as assessed by cooximetry. The accuracy of the sensors in comparison to the co-oximeter samples was measured over the COHb range of 0-15% with 95 - 100% SaO2. Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61, Medical Electrical Equipment-Particular requirements for basic safety and essential performance of pulse oximeter equipment.
#### COHb Accuracy Testing with Elevated HHb (Deoxyhemoglobin)
COHb accuracy testing was conducted at an independent research laboratory on healthy, male and female, non-smoking, light to darkskinned subjects that were 18 years of age and older.
The measured carboxyhemoglobin value (%COHb) of the sensors was compared to simultaneous arterial blood samples as assessed by cooximetry. The accuracy of the sensors in comparison to the co-oximeter samples was measured over the COHb range of 0-15% with 80 - 100% SaO2. Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61, Medical Electrical Equipment-Particular requirements for basic safety and essential performance of pulse oximeter equipment.
#### SpO2 Accuracy Testing in Presence of COHb
During no-motion conditions at an independent research laboratory, SpO2 accuracy testing in the presence of COHb was conducted during induced hypoxia studies on healthy, male and female, non-smoking, lightto dark-skinned subjects that were 18 years of age and older.
{6}------------------------------------------------
| The measured arterial hemoglobin saturation value (SpO₂) of the sensors |
|-------------------------------------------------------------------------|
| was compared to arterial hemoglobin oxygen (SaO₂) value, determined |
| from blood samples with a laboratory co-oximeter. |
The accuracy of the sensors in comparison to the co-oximeter samples was measured over the SpO₂ range of 70-100%, range 0-3% COHb and 80 – 100%, range 0 – 15% COHb. Testing was also conducted over the SpO₂ range of 70-100%.
Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61, Medical Electrical Equipment— Particular requirements for basic safety and essential performance of pulse oximeter equipment.
Testing conclusion: The proposed Model 8330AA Sensor meets all acceptance criteria. Based on test results, analysis and comparison to the legally marketed predicates, the Model 8330AA when used with Nonin Medical's Multi-Sensing Oximetry System is substantially equivalent to the predicate device for its intended use.
| Summary of<br>Substantial<br>Equivalence: | The Model 8330AA Sensor has the following similarities to the predicate devices:<br>• Similar indications for use - different wording, same indications<br>• Similar intended use environments<br>• Identical patient population<br>• Same primary mode of operation<br>• Performs equivalently to the similar specifications<br>A comparison table is on the following pages. |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | The proposed system is substantially equivalent to the predicate |
system.
{7}------------------------------------------------
Page 5 of 8
# K172625 Comparison of Proposed Multi-Sensing Oximetry System with Model 8330AA Sensor to Predicate, Model X-100C CO-Met Oximetry System with Model 8300AA
| CATEGORY | Identical/<br>Similar/<br>Different | Model X-100C CO-Met Oximetry System<br>Multi-Sensing Oximetry System<br>With Model 8330AA Sensor (Proposed) | Model X-100C System<br>With Model 8300AA Sensor (Predicate) |
|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | | | |
| System | Similar – The<br>indications<br>are differently<br>worded for<br>clarification of<br>intended use<br>environments.<br>The subject<br>device also<br>does not have<br>MetHb<br>functionality. | The Multi-Sensing Oximetry System with the Model 8330AA<br>reusable finger clip sensor is intended for noninvasive measuring<br>of functional oxygen saturation of arterial hemoglobin (%SpO2),<br>carboxyhemoglobin saturation (%COHb), and pulse rate. The<br>measurements may be multiple spot-checks to observe change<br>and/or continuous monitoring. This device is not meant for sole<br>use in clinical decision making; it must be used in conjunction<br>with additional methods of assessing clinical signs and<br>symptoms.<br>• For %SpO2 and pulse rate, the 8330AA is intended for use in<br>professional healthcare facilities, mobile, and EMS settings in<br>adult and pediatric patients (> 66 lbs/30 kg).<br>• For %COHb, the 8330AA is intended for use in professional<br>healthcare facilities settings in adult and pediatric patients (> 66<br>lbs/30 kg). This device should not be used to screen, diagnose or<br>make treatment decisions related to suspected carbon monoxide<br>poisoning.<br><br>Rxonly CAUTION: Federal law (USA) restricts this device to sale<br>by or on the order of a licensed practitioner. | Nonin's X-100C System is indicated for noninvasive spot-<br>checking and/or continuous monitoring of<br>Carboxyhemoglobin (COHb), Methemoglobin (MetHb),<br>functional oxygen saturation of arterial hemoglobin (SpO2)<br>and pulse rate of adult and pediatric patients using the<br>8300AA sensor.<br>The X-100C System is indicated for use by trained personnel<br>in clinical and non-clinical settings, including Emergency<br>Medical Service (EMS), hospitals, medical facilities, mobile<br>environments, and home healthcare.<br><br>Rxonly CAUTION: Federal law (USA) restricts this device to<br>sale by or on the order of a licensed practitioner. |
| Oxygen Saturation<br>Range (%SpO2) | Identical | 0-100% | 0-100% |
| Carboxyhemoglobin<br>Saturation Range | Identical | 0-99% (%COHb) | 0-99% (%SpCO) |
| Methemoglobin<br>Saturation Range | Different –<br>Subject<br>devices does<br>not include<br>MetHb<br>measurement | N/A | 0-99.9% (%SpMet) |
| Pulse Rate Display<br>Range | Identical | 18-321 beats per minute (bpm) | 18-321 beats per minute (bpm) |
| K172625<br>Page 6 of 8 | | | |
| CATEGORY | Identical/<br>Similar/<br>Different | Multi-Sensing Oximetry System<br>With Model 8330AA Sensor (Proposed) | Model X-100C System<br>With Model 8300AA Sensor (Predicate) |
| Accuracy | | | |
| Oxygen Saturation<br>(%SpO₂) non-motion | Identical | Refer to Oxygen Saturation (SpO₂) with elevated<br>Carboxyhemoglobin and Oxygen Saturation (SpO₂) | Refer to Oxygen Saturation (SpO2) with elevated<br>Carboxyhemoglobin and Oxygen Saturation (SpO2) with<br>elevated Methemoglobin |
| Carboxyhemoglobin<br>Accuracy | Similar | Adults, Pediatrics<br>• ±3 Arms (COHb range of 0-15%, SaO₂ 95-100%)<br>• ±3 Arms (COHb range of 0-5%, SaO₂ 95-100%)<br>• ±2 Arms (COHb range of 5-10%, SaO₂ 95-100%)<br>• ±2 Arms (COHb range of 10-15%, SaO₂ 95-100%) | Adults, Pediatrics<br>• ±2 Arms (COHb range of 0-15%, SaO₂ 95-100%)<br>• ±3 Arms (COHb range of 0-5%, SaO₂ 95-100%)<br>• ±2 Arms (COHb range of 5-10%, SaO₂ 95-100%)<br>• ±2 Arms (COHb range of 10-15%, SaO₂ 95-100%) |
| Carboxyhemoglobin<br>Accuracy with<br>elevated HHb | Similar | Adults, Pediatrics<br>• ±3 Arms (COHb range of 0-15%, SaO₂ 80-100%)<br>• ±3 Arms (COHb range of 0-5%, SaO₂ 80-100%)<br>• ±2 Arms (COHb range of 5-10%, SaO₂ 80-100%)<br>• ±3 Arms (COHb range of 10-15%, SaO₂ 80-100%) | Adults, Pediatrics<br>• ±3 Arms (COHb range of 0-15%, SaO₂ 80-100%)<br>• ±3 Arms (COHb range of 0-5%, SaO₂ 80-100%)<br>• ±2 Arms (COHb range of 5-10%, SaO₂ 80-100%)<br>• ±3 Arms (COHb range of 10-15%, SaO₂ 80-100%) |
| Oxygen Saturation<br>(SpO₂) with elevated<br>Carboxyhemoglobin | Similar | Adults, Pediatrics<br>• ±2 Arms (COHb range of 0-3%, SaO₂ 70-100%)<br>• ±3 Arms (COHb range of 0-3%, SaO₂ 70-80%)<br>• ±2 Arms (COHb range of 0-3%, SaO₂ 80-90%)<br>• ±2 Arms (COHb range of 0-3%, SaO₂ 90-100%) | Adults, Pediatrics<br>• ±2 Arms (COHb range of 0-3%, SaO₂ 70-100%)<br>• ±2 Arms (COHb range of 0-3%, SaO₂ 70-80%)<br>• ±3 Arms (COHb range of 0-3%, SaO₂ 80-90%)<br>• ±2 Arms (COHb range of 0-3%, SaO₂ 90-100%) |
| Pulse Rate Accuracy | | | |
| Pulse Rate Accuracy<br>non-motion | Identical | Adults, Pediatrics: ±3 digits (40-250 bpm) | Adults, Pediatrics: ±3 digits (40-250 bpm) |
| Electrical | | | |
| Standalone<br>• AC Power<br>Requirements<br>• Power<br>consumption<br>• Fuses | Identical | 100-240 VAC, 50-60 Hz | 100-240 VAC, 50-60 Hz |
| Batteries<br>• Type<br>• Capacity<br>• Charging Time | Identical | 7.2 volt Li-ion battery pack, 2.4 Ah when charged<br>6 hours¹<br>2.5 hours | 7.2 volt Li-ion battery pack, 2.4 Ah when charged<br>6 hours¹<br>2.5 hours |
| K172625<br>Page 7 of 8 | | | |
| CATEGORY | Identical/<br>Similar/<br>Different | Multi-Sensing Oximetry System<br>With Model 8330AA Sensor (Proposed) | Model X-100C System<br>With Model 8300AA Sensor (Predicate) |
| Environmental | | | |
| Operating<br>Temperature | Identical | 32 °F to 104 °F (0 °C to 40 °C) | 32 °F to 104 °F (0 °C to 40 °C) |
| Storage/Transport<br>Temperature | Identical | -22 °F to 158 °F (-30 °C to 70 °C) | -22 °F to 158 °F (-30 °C to 70 °C) |
| Operating Humidity | Identical…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.