The Masimo Disposable Transflectance Forehead Sensors are indicated for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Masimo Disposable Transflectance Forehead Sensors are intended for use with adult and pediatric patients, weighing greater than 10 kg, who are well or poorly perfused in healthcare environments.
Device Story
Disposable transflectance forehead sensor; uses spectrophotometry and photoplethysmography to measure SpO2, pulse rate, perfusion index (PI), and pleth variability index (PVI). Sensor placed on patient forehead, secured with headband, and connected to Masimo SET or Rainbow SET technology-enabled pulse oximeter instruments. Provides continuous monitoring in healthcare environments. Healthcare providers use output to assess patient oxygenation and perfusion status; aids in clinical decision-making regarding respiratory and circulatory management. Benefits include noninvasive, continuous monitoring for patients >10 kg.
Clinical Evidence
Clinical testing per ISO 80601-2-61 conducted on healthy adult volunteers with varying skin pigmentation. SpO2 measurements (70-100% range) compared against laboratory CO-oximeter arterial blood samples. Accuracy determined via Arms (accuracy root mean square). Results demonstrated accuracy of ±2% for SpO2 and ±3 bpm for pulse rate under no motion and low perfusion conditions, consistent with predicate specifications.
Technological Characteristics
Transflectance pulse oximetry sensor; single-use; non-sterile. Materials: polycarbonate lens, polyurethane light barrier, acrylic adhesive, PVC cable jacket (all ISO-10993 compliant). Dimensions: 2.5 x 1.5 in. Connectivity: DB9, M15, or RD sensor connectors for Masimo SET/Rainbow SET technology boards. Sensing principle: spectrophotometry (red/IR light absorption) and photoplethysmography.
Indications for Use
Indicated for continuous noninvasive monitoring of SpO2 and pulse rate in adult and pediatric patients weighing >10 kg, including well or poorly perfused patients in healthcare environments.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an emblem consisting of three stylized human profiles facing to the right, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 23, 2016
Masimo Corporation Marguerite Thomlinson Senior Director, Regulatory Affairs 52 Discovery Irvine, California 92618
Re: K160940
Trade/Device Name: Masimo Disposable Transflectance Forehead Sensor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: October 24, 2016 Received: October 27, 2016
Dear Marguerite Thomlinson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image contains a signature and a name. The name is "Tina Kiang-S". The signature is to the left of the name and is a series of loops and curves.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K160940
Device Name
Masimo Disposable Transflectance Forehead Sensors
#### Indications for Use (Describe)
The Masimo Disposable Transflectance Forehead Sensors are indicated for continuous nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Masimo Disposable Transflectance Forehead Sensors are intended for use with adult and pediatric patients, weighing greater than 10 kg, who are well or poorly perfused in healthcare environments.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red symbol on the left, followed by the word "Masimo" in black, bold letters. To the right of the word "Masimo" is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618" also in gray.
| Submitter and Address of<br>Manufacturing Facility: | Masimo Corporation<br>52 Discovery<br>Irvine, CA 92618<br>Phone: (949) 297-7000<br>FAX: (949) 297-7592 |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Date: | November 21, 2016 |
| Contact: | Marguerite Thomlinson<br>Senior Director, Regulatory Affairs |
| Trade Name: | Masimo Disposable Transflectance Forehead Sensors |
| Common Name: | Oximeter |
| Classification Regulation/<br>Product Code: | 21 CFR 870.2700, Class II/DQA |
| Establishment Registration<br>Number: | 2031172 |
| Reason for Premarket<br>Notification: | New Device |
| Predicate Device: | K032551, Masimo LNOP TF-1 Reusable Transflectance<br>Sensor |
| Performance Standards | No performance standards for the above device have been<br>promulgated pursuant to Section 514. |
### Device Description
The subject device, Masimo Disposable Transflectance Forehead Sensor (Forehead Sensor), is a single-use sensor and is supplied non-sterile. The sensor is available in two configurations, consisting of the LNCS TFA-1 with the DB9 sensor connector and the M-LNCS TFA-1 with the M15 sensor connector. The different sensor connectors allow for sensor connection to instruments with corresponding mating DB9 or M15 connectors. The subject device shall also be available with RD sensor connector. The subject device is intended to be used with instruments that include Masimo SET technology for the monitoring of:
- Functional Oxygen Saturation of Arterial Hemoglobin (SpO2): The amount of ● oxyhemoglobin expressed a percentage of the hemoglobin that is available
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the logo is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".
- Pulse Rate (PR): Measured in beats per minute (BPM) is based on the optical detection of peripheral flow pulse
- . Perfusion Index (PI): The ratio of the pulsatile blood flow to the non-pulsatile or static blood in peripheral tissue
- Pleth Variability Index (PVI): Measure of dynamic changes in perfusion index that occur ● during the respiratory cycle.
The Masimo SET technology is provided by the Masimo SET or Masimo Rainbow SET technology board, which is included in Masimo pulse oximeter instruments, Masimo pulse CO-oximeter instruments and third-party multi-parameter instruments.
| Forehead Sensor Specifications | |
|--------------------------------|----------------------------------------------------|
| Features | Specifications |
| Patient population | Adults and pediatrics > 10 kg |
| Measurement site | Forehead |
| Type of use | Single-use |
| Sterility | Supplied non-sterile |
| Accuracy* | |
| SpO2, no motion | 70-100 ± 2 % |
| SpO2, low perfusion | 70-100 ± 2 % |
| Pulse Rate, no motion | 25-240 ± 3 bpm |
| Pulse Rate, low perfusion | 25-240 ± 3 bpm |
| Mechanical | |
| Sensor dimensions/weight | Approximately 2.5 x 1.5 in/ 1.0 oz (without cable) |
| Sensor cable length | 36 in, nominal |
| Patient-contacting materials | ISO-10993 compliant |
| Environmental | |
| Storage temperature | -40°C to +70°C |
| Operating temperature | +5°C to +40°C |
| Humidity | 10-95% humidity, non-condensing |
The specifications of the Forehead Sensor are listed in the table below.
* Accuracy was determined in terms of Arms (accuracy root mean square) per ISO-80601-2-61.
#### Intended Use/Indications for Use
The Masimo Disposable Transflectance Forehead Sensors are indicated for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Masimo Disposable Transflectance Forehead Sensors are intended for use with adult and pediatric patients, weighing greater than 10 kg, who are well or poorly perfused in healthcare environments.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a stylized "V" symbol to the left of the word. To the right of the word "Masimo" is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".
### Technological Characteristics
#### Principle of Operation
The subject device, Forehead Sensor, is a pulse oximetry sensor which is governed by the following principles:
- Oxyhemoglobin (oxygentated blood) and deoxyhemoglobin (non-oxygenated blood) differ in their absorption of red and infrared light (spectrophotometry).
- The amount of arterial blood in tissue changes with a person's pulse (photoplethysymography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.
### Mechanism of Action for Achieving the Intended Effect
The Forehead Sensor is placed on the patient forehead and secured with a headband. The other end of the sensor is connected to a patient cable, which in turn is connected to a pulse oximeter. The instrument is turned on when use begins. Once use is complete, the user turns the instrument power "off" and removes the sensor from the patient.
### Summary of Technological Characteristics of Subject Device Compared to Predicate Device
#### Similarities and Differences between Subject Device and Predicate Device (K032551)
The subject device, Forehead Sensor, and the predicate (K032551) have the following key similarities:
- . Both devices have the same intended use of pulse oximetry measurements under no motion and low perfusion conditions;
- Both devices have the same forehead measurement site;
- Both devices are supplied non-sterile; and ●
- Both devices have the same accuracy specifications ●
The subject and predicate devices are mainly different in that:
- the subject device is single-use device whereas the predicate is reusable; and ●
- the subject device is intended for patients > 10 kg whereas the predicate is intended for ● patients >30 kg.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a red circle containing a white checkmark to the left of the word. To the right of the word "Masimo" is the text "MASIMO CORPORATION" and the address "52 Discovery, Irvine, CA 92618".
## Substantial Equivalent
| Table 12.5 Substantial Equivalent Discussion | | |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Masimo Disposable Forehead Sensor<br>Subject Device, Pending | Masimo Reusable Forehead Sensors, LNOP TF-1<br>Predicate, K032551 |
| 510(k) No.<br>General | | |
| Indications for use | The Masimo Disposable Transflectance<br>Forehead Sensors are indicated for<br>continuous noninvasive monitoring of<br>functional oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate. The<br>Masimo Disposable Transflectance<br>Forehead Sensors are intended for use<br>with adult and pediatric patients,<br>weighing greater than 10 kg, who are well<br>or poorly perfused in healthcare<br>environments. | The following additional Masimo Sensor is<br>indicated for the continuous noninvasive functional<br>oxygen saturation of arterial hemoglobin (SpO2) and<br>pulse rate:<br><br>Reusable oximetry transflectance sensor intended<br>for adults and pediatrics greater than 30 kg in<br>hospitals, hospital-type facilities, mobile, and home<br>environments. |
| Classification regulation/<br>product code | Same as K032551 | 21 CFR 870.2700, Class II / DQA |
| Principle of operation | Same as K032551 | Pulse oximetry is governed by the principles of a)<br>Oxyhemoglobin (oxygenated blood) and b)<br>deoxyhemoglobin (non-oxygenated blood) differ in<br>their absorption of red and infrared light<br>(spectrophotometry). The amount of arterial blood in<br>tissue changes with your pulse<br>(photoplethysymography). Therefore, the amount of<br>light absorbed by the varying quantities of arterial<br>blood changes as well. |
| Mechanism of actions for<br>achieving the intended use | Same as K032551 | The Forehead Sensor is placed on the patient<br>forehead and secured with a headband. The other<br>end of the sensor is connected to a patient cable,<br>which in turn is connected to a pulse oximeter. The<br>instrument is turned on when use begins. Once use<br>is complete, the user turns the instrument power<br>"off" and removes the sensor from the patient. |
| Patient population | Adults and pediatrics >10 kg | Adults and pediatrics >30 kg |
| Measurement site | Same as K032551 | Forehead |
| Type of use | Single-use | Reusable |
| Sterility | Same as K032551 | Supplied non-sterile |
| Compatibility | Instruments with Masimo SET<br>technology provided by Masimo SET or<br>Masimo Rainbow SET technology<br>boards, including the MS, MS-2K and<br>MX series boards | Instruments with Masimo SET technology provided<br>by Masimo SET technology boards, including the<br>MS series boards |
| Perfusion index (PI) | | |
| PI Function | Same as K032551 | PI is the measure of the ratio of the pulsatile blood<br>flow to the non-pulsatile or static blood in peripheral<br>tissue |
| Table 12.5 Substantial Equivalent Discussion | | |
| Feature | Masimo Disposable Forehead Sensor | Masimo Reusable Forehead Sensors, LNOP TF-1 |
| 510(k) No. | Subject Device, Pending | Predicate, K032551 |
| PI Range | Same as K032551 | 0.02% to 20% |
| Pleth variability index (PVI) | Same as K032551 | PVI is the measure of dynamic changes in perfusion index that occur during the respiratory cycle. PVI is available based on whether the feature is enabled in the instrument with Masimo SET technology |
| PVI Range | Same as K032551 | 0% to 100% |
| Accuracy, ARMS | | |
| SpO2, no motion | 70-100 ± 2 % | 70-100 ± 2 % |
| SpO2, low perfusion | 70-100 ± 2 % | 70-100 ± 2 % |
| Pulse rate, no motion | 25-240 ± 3 bpm | 25-240 ± 3 bpm |
| Pulse rate, low perfusion | 25-240 ± 3 bpm | 25-240 ± 3 bpm |
| Biocompatibility | | |
| Patient-contacting materials | Disposable adhesive sensor with<br>biocompatible patient contact materials:<br>Lens: optically clear material for light<br>transmission, polycarbonate<br>Light barrier: polyurethane<br>Sensor face: acrylic adhesive<br>Cable jacket: PVC medical grade | Reusable sensor with biocompatible patient contact<br>materials:<br>Lens: optically clear material for light transmission,<br>silicon rubber<br>Light barrier: PVC<br>Sensor face: PVC<br>Cable Jacket: PVC medical grade |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a red circle containing a white checkmark to the left of the word. To the right of the word "Masimo" are the words "MASIMO CORPORATION" in a smaller, sans-serif font, followed by the address "52 Discovery, Irvine, CA 92618".
The difference in the indications for use (IFU) between the subject and predicate devices does not raise different questions of safety and effectiveness because both devices have the same intended use as forehead sensors for pulse oximetry, same principle of operation and same accuracy specifications. Although the patient-contacting materials for the subject device and predicate are slightly different, the materials are the same in that they are compliant to ISO-10993-1. Additionally, since the patient population is the same for the subject device and the predicate, there is no expected difference in the subject device's performance with the slight extended weight range for the subject device. Please note the reference device in K051271, where the same patient population of greater than 10 kg does not require additional testing. Furthermore, the subject device has been validated in regards to human factors and usability.
In summary, the accompanied verification and validation data for the subject device, including testing for any differences, demonstrated that the subject device is as safe and effective as the predicate and therefore substantially equivalent to the predicate.
#### Non-clinical Testing
The subject devices were subjected to bench testing. The following non-clinical testing, as applicable, was performed in accordance with Masimo design control requirements and quality
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a stylized red checkmark to the left of the word. To the right of the logo is the text "MASIMO CORPORATION" in a smaller, sans-serif font, followed by the address "52 Discovery, Irvine, CA 92618".
system to demonstrate substantial equivalence of the subject device with its predicates:
- Electrical safety testing per IEC60601-1
- . EMC testing per IEC60601-1-2
- Pulse oximetry testing, including simulator testing of pulse rate and low perfusion, per ISO 80601-2-61,
- Biocompatibility testing per ISO-10993-1, ISO-10993-5 and ISO-10993-10 ●
- Usability testing per FDA Human Factors and Usability Guidance
- Software verification per FDA Software Guidance
- Mechanical and environmental testing
#### Clinical Testing
Clinical performance testing was completed in accordance with ISO 80601-2-61 to evaluate the accuracy of functional oxygen saturation (SpO2) measurements. The Forehead Sensors, connected to MX technology boards with Masimo SET technology, were applied to healthy adult volunteers with light to dark skin pigmentation. SpO2 measurements were compared against arterial blood samples analyzed by a laboratory CO-oximeter, in the range of 70 to 100%, to determine accuracy specifications in terms of accuracy root mean square (Arms). The variation in the accuracy specifications equals plus or minus one standard deviation, which encompassed approximately 68% of the tested population. Arms calculation incorporates error due to bias and imprecision (standard deviation).
#### Conclusion
The non-clinical and clinical testing provided in this 510(k) submission demonstrates that the subject devices. Forehead Sensors are as safe and effective as the predicate and therefore substantially equivalent to the predicate in K032551.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.