Model 6100C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6100C Series: 6100CA, 6100CP, 6100CI, 6100CN), Model 6101C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6101C Series: 6101CA, 6101CP, 6101CI, 6101CN), Model 6102C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6102C Series: 6102CA, 6102CP, 6102CI, 6102CN), Model 8100AA/8100AP Reusable, Finger Clip Pulse Oximeter Sensor, Model 8101AA/8101AP Reusable, Finger Clip Pulse Oximeter Senso
K160865 · Nonin Medical, Inc. · DQA · Aug 5, 2016 · Cardiovascular
Device Facts
Record ID
K160865
Device Name
Model 6100C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6100C Series: 6100CA, 6100CP, 6100CI, 6100CN), Model 6101C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6101C Series: 6101CA, 6101CP, 6101CI, 6101CN), Model 6102C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6102C Series: 6102CA, 6102CP, 6102CI, 6102CN), Model 8100AA/8100AP Reusable, Finger Clip Pulse Oximeter Sensor, Model 8101AA/8101AP Reusable, Finger Clip Pulse Oximeter Senso
Applicant
Nonin Medical, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Aug 5, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Nonin's Model 6100C Series single-patient use, disposable pulse oximeter sensors are indicated for non-invasive spotchecking and/or continuous monitoring of patients (6100CA: adults > 30 kg / 66 Ib; 6100CP: pediatrics > 10 kg / 22 1b ; 6100CI: infants > 2 kg / 4 1b; and 6100CN: neonates < 2 kg / 4 1b and adults > 30 kg / 66 lb), who are well or poorly perfused, during both motion and non-motion conditions. It is intended for use hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healtheare environments, and mobile environments. Nonin's Model 6101C Series single-patient use, disposable pulse oximeter sensors are indicated for non-invasive spotchecking and/or continuous monitoring of patients (6101CA: adults > 30 kg / 66 lb; 6101CP: pediatrics > 10 kg / 22 1b ; 6101CI: infants > 2 kg / 4 1b; and 6101CN: neonates < 2 kg / 4 lb and adults > 30 kg / 66 lb), who are well or poorly perfused, during both motion and non-motion conditions. It is intended for use hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments. Nonin's Model 6102C Series single-patient use, disposable pulse oximeter sensors are indicated for non-invasive spot-checking and/or continuous monitoring of patients (6102CA: adults > 30 kg / 66 lb; 6102CP: pediatrics > 10 kg / 22 lb.; 6102Cl: infants > 2 kg / 4 lb; and 6102CN: neonates < 2 kg / 4 lb and adults > 30 kg / 66 Ib), who are well or poorly perfused, during both motion and nonmotion conditions. It is intended for use hospitals, medical facilities. Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments. The Model 8100AA reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of adult and pediatric patients (> 30 kg / 66 lb) who are well or poorly perfused, during both motion and non-motion conditions. The Model 8100AP reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of pediatric patients (8 – 60 kg / 18 – 132 lb) who are well or poorly perfused, during both motion and non-motion conditions. It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments. The Model 8101AA reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of adult and pediatric patients (> 30 kg / 66 lb) who are well or poorly perfused, during both motion and non-motion conditions. The Model 8101AP reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of pediatric patients (8 – 60 kg / 18 – 132 lb) who are well or poorly perfused, during both motion and non-motion conditions. It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments. The Model 8102AA reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of adult and pediatric patients (> 30 kg / 66 lb) who are well or poorly perfused, during both motion and non-motion conditions. The Model 8102AP reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of pediatric patients (8 - 60 kg / 18 - 132 lb) who are well or poorly perfused, during both motion and non-motion conditions. It is intended for use in hospitals, medical facilities. Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments. The Model 8100Q2 reusable, ear clip sensor is indicated for non-invasive, spot-checking and/or continuous monitoring of adult and pediatric patients (> 40 kg / 88 1b) who are well or poorly perfused, during non-motion conditions. It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, and mobile environments. The recommended application site is the earlobe. The Model 8101Q2 reusable, ear clip sensor is indicated for non-invasive, spot-checking and/or continuous monitoring of adult and pediatric patients (> 40 kg / 88 lb) who are well or poorly perfused, during non-motion conditions. It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments. The recommended application site is the earlobe. The Model 8102Q2 reusable, ear clip sensor is indicated for non-invasive, spot-checking and/or continuous monitoring of adult and pediatric patients (> 40 kg / 88 1b) who are well or poorly perfused, during non-motion conditions. It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, and mobile environments. The recommended application site is the earlobe.
Device Story
Pulse oximeter sensors (disposable and reusable finger/ear clips) capture optical signals via PureLight technology; input processed by Nonin X-100 SenSmart Universal Oximetry System; system uses smart chip authentication via Hirose connector; output provides SpO2 and pulse rate data; used in hospitals, EMS, home, and mobile environments by clinicians; aids in patient oxygenation monitoring; benefits include non-invasive, real-time physiological assessment.
Clinical Evidence
Clinical SpO2 accuracy testing conducted on healthy subjects (adults) and neonates (up to 30 days). Compared sensor SpO2 to co-oximetry arterial blood samples across 70-100% SaO2 range. Testing performed in both motion and non-motion conditions (where applicable). All models met acceptance criteria for accuracy (ARMS).
Technological Characteristics
Sensors utilize PureLight optical technology; disposable and reusable form factors (finger/ear clips); Hirose connector with integrated smart chip for system authentication; biocompatible materials per ISO 10993; compliant with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and ISO 80601-2-61 (performance).
Indications for Use
Indicated for non-invasive spot-checking and/or continuous monitoring of patients (neonates, infants, pediatrics, adults) based on specific model weight ranges; well or poorly perfused; motion and non-motion conditions (varies by model).
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Nonin disposable cloth SpO2 Model 6000CX Series (K09853)
Nonin Model 8100S(X) Series Soft Pulse Oximetry Sensors (K132402)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to create a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 5, 2016
Nonin Medical, Inc. Ms. Nancy DeAngelo Senior Regulatory Affairs Specialist 13700 1st Ave. North Plymouth, Minnesota 55441
Re: K160865
Trade/Device Name: Model 6100C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6100C Series: 6100CA, 6100CP, 6100CL, 6100CN) Model 6101C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6101C Series: 6101CA, 6101CP, 6101CL, 6101CN) Model 6102C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6102C Series: 6102CA, 6102CP, 6102CL, 6102CN) Model 8100AA/8100AP Reusable, Finger Clip Pulse Oximeter Sensor Model 8101AA/8101AP Reusable, Finger Clip Pulse Oximeter Sensor Model 8102AA/8102AP Reusable, Finger Clip Pulse Oximeter Sensor Model 8100Q2 Reusable, Ear Clip Pulse Oximeter Sensor Model 8101Q2 Reusable, Ear Clip Pulse Oximeter Sensor Model 810202 Reusable, Ear Clip Pulse Oximeter Sensor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA, DPZ Dated: July 5, 2016 Received: July 6. 2016
Dear Ms. DeAngelo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known) K160865
#### Device Name
Model 6100C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6100C Series: 6100CA, 6100CN) Model 6101C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6101C Series: 6101CA, 6101CP, 6101CN)
#### Indications for Use (Describe)
Model 6100C Series Single-Patient Use, Disposable Pulse Oximeter Sensors
Nonin's Model 6100C Series single-patient use, disposable pulse oximeter sensors are indicated for non-invasive spotchecking and/or continuous monitoring of patients (6100CA: adults > 30 kg / 66 Ib; 6100CP: pediatrics > 10 kg / 22 1b ; 6100CI: infants > 2 kg / 4 1b; and 6100CN: neonates < 2 kg / 4 1b and adults > 30 kg / 66 lb), who are well or poorly perfused, during both motion and non-motion conditions.
It is intended for use hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healtheare environments, and mobile environments.
Model 6101C Series Single-Patient Use, Disposable Pulse Oximeter Sensors
Nonin's Model 6101C Series single-patient use, disposable pulse oximeter sensors are indicated for non-invasive spotchecking and/or continuous monitoring of patients (6101CA: adults > 30 kg / 66 lb; 6101CP: pediatrics > 10 kg / 22 1b ; 6101CI: infants > 2 kg / 4 1b; and 6101CN: neonates < 2 kg / 4 lb and adults > 30 kg / 66 lb), who are well or poorly perfused, during both motion and non-motion conditions.
It is intended for use hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|-----------------------------------------------------------------|
| <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{3}------------------------------------------------
510(k) Number (if known) K160865
#### Device Name
Model 6102C Series Single-Patient Use, Disposable Pulse Oximeter Sensors (6102C Series: 6102CA, 6102CN) Model 8100AA/8100AP Reusable, Finger Clip Pulse Oximeter Sensor
#### Indications for Use (Describe)
Model 6102C Series Single-Patient Use, Disposable Pulse Oximeter Sensors
Nonin's Model 6102C Series single-patient use, disposable pulse oximeter sensors are indicated for non-invasive spotchecking and/or continuous monitoring of patients (6102CA: adults > 30 kg / 66 Ib; 6102CP: pediatrics > 10 kg / 22 1b ; 6102C1: infants > 2 kg / 4 1b; and 6102CN: neonates < 2 kg / 4 1b and adults > 30 kg / 66 lb), who are well or poorly perfused, during both motion and non-motion conditions.
It is intended for use hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healtheare environments, and mobile environments.
### Model 8100AA/8100AP Reusable, Finger Clip Pulse Oximeter Sensor
The Model 8100AA reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of adult and pediatric patients (> 30 kg / 66 lb) who are well or poorly perfused, during both motion and non-motion conditions.
The Model 8100AP reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of pediatric patients (8 – 60 kg / 18 – 132 lb) who are well or poorly perfused, during both motion and non-motion conditions.
It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------|
| <div> <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{4}------------------------------------------------
510(k) Number (if known) K160865
#### Device Name
Model 8101AA/8101AP Reusable, Finger Clip Pulse Oximeter Sensor Model 8102AA/8102AP Reusable, Finger Clip Pulse Oximeter Sensor
#### Indications for Use (Describe)
Model 8101AA/8101AP Reusable, Finger Clip Pulse Oximeter Sensor
The Model 8101AA reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of adult and pediatric patients (> 30 kg / 66 lb) who are well or poorly perfused, during both motion and non-motion conditions.
The Model 8101AP reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of pediatric patients (8 – 60 kg / 18 – 132 lb) who are well or poorly perfused, during both motion and non-motion conditions.
It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healtheare environments, and mobile environments.
Model 8102AA/8102AP Reusable, Finger Clip Pulse Oximeter Sensor
The Model 8102AA reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of adult and pediatric patients (> 30 kg / 66 lb) who are well or poorly perfused, during both motion and non-motion conditions.
The Model 8102AP reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of pediatric patients (8 - 60 kg / 18 - 132 lb) who are well or poorly perfused, during both motion and non-motion conditions.
It is intended for use in hospitals, medical facilities. Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|------------------------------------------------------------------------------------|
| <span style="font-size:10pt">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{5}------------------------------------------------
510(k) Number (if known) K160865
#### Device Name
Model 8100Q2 Reusable, Ear Clip Pulse Oximeter Sensor Model 8101Q2 Reusable, Ear Clip Pulse Oximeter Sensor Model 8102Q2 Reusable, Ear Clip Pulse Oximeter Sensor
#### Indications for Use (Describe)
Model 8100Q2 Reusable, Ear Clip Pulse Oximeter Sensor
The Model 8100Q2 reusable, ear clip sensor is indicated for non-invasive, spot-checking and/or continuous monitoring of adult and pediatric patients (> 40 kg / 88 1b) who are well or poorly perfused, during non-motion conditions. It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, and mobile environments.
The recommended application site is the earlobe.
Model 8101Q2 Reusable, Ear Clip Pulse Oximeter Sensor
The Model 8101Q2 reusable, ear clip sensor is indicated for non-invasive, spot-checking and/or continuous monitoring of adult and pediatric patients (> 40 kg / 88 lb) who are well or poorly perfused, during non-motion conditions. It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.
The recommended application site is the earlobe.
### Model 8102Q2 Reusable, Ear Clip Pulse Oximeter Sensor
The Model 8102Q2 reusable, ear clip sensor is indicated for non-invasive, spot-checking and/or ing of adult and pediatric patients (> 40 kg / 88 1b) who are well or poorly perfused, during non-motion conditions. It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, and mobile environments.
The recommended application site is the earlobe.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span></span> </div> |
|----------------------------------------------|----------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span></span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{6}------------------------------------------------
# "510(k) Summary" as required by section 807.92(c)
| Submitter: | Nonin Medical, Inc. |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Phone:<br>Fax: | 13700 1st Ave. North<br>Plymouth, MN 55441-5443<br>763-553-9968<br>763-553-7807 |
| Contact Person: | Nancy DeAngelo<br>Senior Regulatory Affairs Specialist |
| Date Prepared: | March 24, 2016 |
| Trade Names: | Model 6100C Series Single-Patient Use, Disposable Pulse<br>Oximeter Sensors (6100C Series: 6100CA, 6100CP, 6100CI,<br>6100CN)<br>Model 6101C Series Single-Patient Use, Disposable Pulse<br>Oximeter Sensors (6101C Series: 6101CA, 6101CP, 6101CI,<br>6101CN)<br>Model 6102C Series Single-Patient Use, Disposable Pulse<br>Oximeter Sensors (6102C Series: 6102CA, 6102CP, 6102CI,<br>6102CN)<br>Model 8100AA/8100AP Reusable, Finger Clip Pulse Oximeter<br>Sensor<br>Model 8101AA/8101AP Reusable, Finger Clip Pulse Oximeter<br>Sensor<br>Model 8102AA/8102AP Reusable, Finger Clip Pulse Oximeter<br>Sensor<br>Model 8100Q2 Reusable, Ear Clip Pulse Oximeter Sensor<br>Model 8101Q2 Reusable, Ear Clip Pulse Oximeter Sensor<br>Model 8102Q2 Reusable, Ear Clip Pulse Oximeter Sensor |
| Common Name: | Pulse oximeter sensor |
| Classification Name: | Oximeter |
| Regulation Number: | Class II, 21 CFR 870.2700 (Oximeter) |
| Product Code, Panel: | DQA, Anesthesiology<br>DPZ, Anesthesiology |
| Predicate Device(s): | The Nonin Model 6100C Series (Models 6100CA, 6100CP,<br>6100CI, 6100CN), Model 6101C Series (Models 6101CA,<br>6101CP, 6101CI, 6101CN), Model 6102C Series (Models<br>6102CA, 6102CP, 6102CI, 6102CN), Model 8100AA/8100AP, |
{7}------------------------------------------------
Model 8101AA/8101AP, Model 8102AA/8102AP, Model 8100Q2, Model 8101Q2 and 8102Q2 Pulse Oximeter Sensors are predicated on Nonin's disposable cloth SpO2 Model 6000CX Series sensors (Models 6000CA, 6000CP, 6000Cl and 6000CN), Nonin Model 8100S(X) Series Soft Pulse Oximetry Sensors (Models 8100SS, 8100SM and 8100SL) and Nonin Model 8000Q2 Ear Clip Sensor cleared under K09853 (June 4, 2010), K132402 (February 21, 2014) and K080255 (May 23, 2008), respectively. The proposed and predicate sensors utilize Nonin's PureLight® technology. Device Description: The Model 6100C Series, Model 6101C Series and Model 6102C Series Pulse Oximeter Sensors are single-patient use. non-sterile disposable pulse sensors intended for use with the Nonin Medical Model X-100 SenSmart Universal Oximetry System (Model X-100). The Model 8100AA/8100AP, Model 8101AA/8101AP, Model 8102AA/8102AP, Model 8100Q2, Model 8101Q2 and Model 8102Q2 Pulse Oximeter Sensors are reusable, non-sterile pulse sensors intended for use with the Nonin Medical Model X-100 SenSmart Universal Oximetry System (Model X-100). Intended Use: Model 6100C Series Single-Patient Use, Disposable Pulse Oximeter Sensors Nonin's Model 6100C Series single-patient use, disposable pulse oximeter sensors are indicated for non-invasive spot-checking and/or continuous monitoring of patients (6100CA: adults > 30 kq / 66 lb; 6100CP: pediatrics > 10 kg / 22 lb.; 6100Cl: infants > 2 kg / 4 lb; and 6100CN: neonates < 2 kg / 4 lb and adults > 30 kg / 66 Ib), who are well or poorly perfused, during both motion and nonmotion conditions. It is intended for use hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments. Model 6101C Series Single-Patient Use, Disposable Pulse Oximeter Sensors Nonin's Model 6101C Series single-patient use, disposable pulse oximeter sensors are indicated for non-invasive spot-checking and/or continuous monitoring of patients (6101CA: adults > 30 kg / 66 lb; 6101CP: pediatrics > 10 kg / 22 lb.; 6101Cl: infants > 2 kg / 4 lb; and 6101CN: neonates < 2 kg / 4 lb and adults > 30 kg / 66 Ib), who are well or poorly perfused, during both motion and nonmotion conditions.
It is intended for use hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.
{8}------------------------------------------------
# Model 6102C Series Single-Patient Use, Disposable Pulse Oximeter Sensors
Nonin's Model 6102C Series single-patient use, disposable pulse oximeter sensors are indicated for non-invasive spot-checking and/or continuous monitoring of patients (6102CA: adults > 30 kg / 66 lb; 6102CP: pediatrics > 10 kg / 22 lb.; 6102Cl: infants > 2 kg / 4 lb; and 6102CN: neonates < 2 kg / 4 lb and adults > 30 kg / 66 Ib), who are well or poorly perfused, during both motion and nonmotion conditions.
lt is intended for use hospitals, medical facilities. Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.
# Model 8100AA/8100AP Reusable, Finger Clip Pulse Oximeter Sensor
The Model 8100AA reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of adult and pediatric patients (> 30 kg / 66 lb) who are well or poorly perfused, during both motion and non-motion conditions.
The Model 8100AP reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of pediatric patients (8 – 60 kg / 18 – 132 lb) who are well or poorly perfused, during both motion and non-motion conditions.
It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.
# Model 8101AA/8101AP Reusable, Finger Clip Pulse Oximeter Sensor
The Model 8101AA reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of adult and pediatric patients (> 30 kg / 66 lb) who are well or poorly perfused, during both motion and non-motion conditions.
The Model 8101AP reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of pediatric patients (8 - 60 kg / 18 - 132 lb) who are well or poorly perfused. during both motion and non-motion conditions.
It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.
{9}------------------------------------------------
# Model 8102AA/8102AP Reusable, Finger Clip Pulse Oximeter Sensor
The Model 8102AA reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of adult and pediatric patients (> 30 kg / 66 lb) who are well or poorly perfused, during both motion and non-motion conditions.
The Model 8102AP reusable, finger clip sensor is intended for non-invasive spot-checking and/or continuous monitoring of pediatric patients (8 – 60 kg / 18 – 132 lb) who are well or poorly perfused, during both motion and non-motion conditions.
It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.
## Model 8100Q2 Reusable, Ear Clip Pulse Oximeter Sensor
The Model 8100Q2 reusable, ear clip sensor is indicated for noninvasive, spot-checking and/or continuous monitoring of adult and pediatric patients (> 40 kg / 88 lb) who are well or poorly perfused, during non-motion conditions. It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.
The recommended application site is the earlobe.
# Model 8101Q2 Reusable, Ear Clip Pulse Oximeter Sensor
The Model 8101Q2 reusable, ear clip sensor is indicated for noninvasive, spot-checking and/or continuous monitoring of adult and pediatric patients (> 40 kq / 88 lb) who are well or poorly perfused. during non-motion conditions. It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.
The recommended application site is the earlobe.
# Model 8102Q2 Reusable, Ear Clip Pulse Oximeter Sensor
The Model 8102Q2 reusable, ear clip sensor is indicated for noninvasive, spot-checking and/or continuous monitoring of adult and pediatric patients (> 40 kg / 88 lb) who are well or poorly perfused, during non-motion conditions. It is intended for use in hospitals, medical facilities, Emergency Medical Service (EMS) environments, home healthcare environments, and mobile environments.
The recommended application site is the earlobe.
{10}------------------------------------------------
| Testing: | The Nonin Model 6100C Series, Model 6101C Series, Model<br>6102C Series, Model 8100AA/8100AP, Model 8101AA/8101AP,<br>Model 8102AA/8102AP, Model 8100Q2, Model 8101Q2 and<br>Model 8102Q2 Series sensors are supported by electrical safety,<br>electromagnetic compatibility, device performance, and clinical<br>testing to ensure appropriate functionality and to demonstrate<br>substantial equivalence to the predicate devices. The devices<br>were tested with the Nonin's Medical Model X-100 SenSmart<br>Universal Oximetry System. |
|----------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
# Functional and Safety Testing:
The results of the testing demonstrate equivalency with the predicate devices and compliance to recognized standards. Table 1 summarizes test results for the proposed devices, which met the relevant requirements of the applicable recognized standards.
| Table 1 | | |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------|--------|
| Test | Reference | Result |
| Electrical Safety | IEC 60601-1 | Pass |
| Temperature and Humidity | IEC 60601-1<br>EN 1789 | Pass |
| Atmospheric Pressure (Altitude) | IEC 60601-1 | Pass |
| Electromagnetic Immunity and<br>Emissions | IEC 60601-1-2 | Pass |
| Performance | ISO 80601-2-61<br>IEC 60601-1<br>IEC 60601-1-6; IEC 60601-1-12<br>IEC 62304<br>ANSI/AAMI EC13<br>ISO 14155 | Pass |
| Ingress Protection | ISO 80601-2-61 | Pass |
| Diaphoretic related ingress | Internal performance<br>characterization | Pass |
| Mechanical Durability | IEC 60601-1<br>ISO 80601-2-61 | Pass |
| Biocompatibility | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | Pass |
Note: The results reflected in the table above are representative for all sensors covered within this submission.
# Clinical Testing: SpO2 Accuracy testing
# Models 6100CA, 6101CA, 6102CA, 6100CP, 6101CP, 6102CP, 6100Cl, 6101Cl and 6102Cl
SpO2 accuracy testing was conducted at an independent research laboratory on healthy, male and female, non-smoking, light to darkskinned subjects that were 18 years of age and older. The measured oxygen saturation value (SpO2) of the sensors was compared to simultaneous arterial blood samples as assessed by co-oximetry. The
{11}------------------------------------------------
accuracy of the sensors in comparison to the co-oximeter samples was measured over the SaO2 range of 70-100% in motion and non-motion. Accuracy data was calculated using the root-mean-squared (ARMs value) for all subjects.
## Models 6100CN, 6101CN and 6102CN
SpO2 accuracy testing was conducted at a Children's Hospital on male and female, light to dark-skinned subjects up to 30 days of age. The measured oxygen saturation value (SpO2) of the sensors was compared to simultaneous arterial blood samples as assessed by co-oximetry. The accuracy of the sensors in comparison to the co-oximeter samples was measured over the SaO2 range of 70-100% in non-motion. Accuracy data was calculated using the root-mean-squared (ARMs value) for all subjects.
### Models 8100AA. 8101AA. 8102AA. 8100AP. 8101AP and 8102AP
SpO2 accuracy testing was conducted at an independent research laboratory on healthy, male and female, non-smoking, light to darkskinned subjects that were 18 years of age and older. The measured oxygen saturation value (SpO2) of the sensors was compared to simultaneous arterial blood samples as assessed by co-oximetry. The accuracy of the sensors in comparison to the co-oximeter samples was measured over the SaO2 range of 70-100% in motion and non-motion. Accuracy data was calculated using the root-mean-squared (ARMs value) for all subjects.
### Models 8100Q2, 8101Q2 and 8102Q2
SpO2 accuracy testing was conducted at an independent research laboratory on healthy, male and female, non-smoking, light to darkskinned subjects that were 18 years of age and older. The measured oxygen saturation value (SpO2) of the sensors was compared to simultaneous arterial blood samples as assessed by co-oximetry. The accuracy of the sensors in comparison to the co-oximeter samples was measured over the SaO2 range of 70-100% in non-motion. Accuracy data was calculated using the root-mean-squared (ARMS value) for all subjects.
Testing conclusion: the proposed Model 6100C Series, Model 6101C Series, Model 6102C Series, Model 8100AA/8100AP, Model 8101AA/8101AP, Model 8102AA/8102AP, Model 8100Q2, Model 8101Q2 and Model 8102Q2 Series sensors meet all acceptance criteria. Based on test results and comparison to the leqally marketed predicates, the Model the proposed sensors perform equivalently to the predicate sensors for their intended use.
# Summary of Substantial Equivalence:
The Model 6100C Series, Model 6101C Series, Model 6102C Series, Model 8100AA/8100AP, Model 8101AA/8101AP, Model 8102AA/8102AP, Model 8100Q2, Model 8101Q2 and Model 8102Q2 sensors have the following similarities to their respective predicate devices:
- . Identical indications for use
{12}------------------------------------------------
- Identical intended use environments ●
- ldentical patient population
- Same primary mode of operation .
- ldentical critical sensor optics technology (PureLight® Technology) ●
- Perform equivalently to the same specifications .
- . Similar construction and materials
The following lists the differences and the rationale for those differences between the proposed sensors and their respective predicate devices:
- . Model 6100C Series, Model 6101C Series and Model 6102C Series
- Contraindications the contraindication regarding the use of o device in MR environment was deemed more appropriate as a warning and the contraindication regarding defibrillation proof is no longer applicable due to verification testing according to IEC 60601-1:2005 when used with the X-100 System.
- SpO2 Motion Accuracy internal and clinical testing was O performed to verify the addition of accuracy claims in accordance with ISO 80601-2-61:2011.
- Pulse Rate Accuracy (Motion and Non-motion) internal and O clinical testing was performed to verify the change in motion claims and the addition of non-motion claims in accordance with ISO 80601-2-61:2011.
- New connector type the X-100 System, along with the о cleared SpO2 and rSO2 sensors for the system, utilize a Hirose connector with a Smart chip for authentication of the sensor to the system.
- Model 8100AA/8100AP, Model 8101AA/8101AP, and Model ● 8102AA/8102AP
- Contraindications the contraindication regarding the use of o device in MR environment was deemed more appropriate as a warning.
- Sensor Optic Housing Materials the direct patient contacting о materials used on the proposed sensors are the same as other cleared Nonin oximetry devices.
- Model 8100Q2. Model 8101Q2 and Model 8102Q2 ●
- Contraindications the contraindication reqarding the use of о device in MR environment was deemed more appropriate as a warning and the contraindication regarding defibrillation proof is no longer applicable due to verification testing according to IEC 60601-1:2005 when used with the X-100 System.
- SpO2 and Pulse Rate Low Perfusion internal and clinical O testing was performed to verify the addition of these claims in accordance with ISO 80601-2-61:2011.
- New connector type the X-100 System, along with the O cleared SpO2 and rSO2 sensors for the system, utilize a Hirose connector with a Smart chip for authentication of the sensor to the system.
{13}------------------------------------------------
### Conclusion: No new questions of safety and effectiveness were raised. Based on the results of the above referenced testing, the same critical optics technology, and risk management assessment, Nonin Medical has determined that the proposed Model 6100C Series Pulse Oximeter Sensors, Model 6101C Series Pulse Oximeter Sensors, Model 6102C Series Pulse Oximeter Sensors, Model 8100AA/8100AP Pulse Oximeter Sensors, Model 8101AA/8101AP Pulse Oximeter Sensors, Model 8102AA/8102AP Pulse Oximeter Sensors, the Model 8100Q2 Pulse Oximeter Sensor, the Model 8101Q2 Pulse Oximeter Sensor and the Model 8102Q2 Pulse Oximeter Sensor are substantially equivalent to the predicate Nonin Model 6000CX Series Pulse Oximeter Sensors (Models 6000CA, 6000CP, 6000Cl, and 6000CN); Model 8100S(X) Series Soft Pulse Oximetry Sensors (Models 8100SS, 8100SM and 8100SL), and the Model 8000Q2 Ear Clip Sensor.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.