K160231 · Nonin Medical, Inc. · DQA · Dec 23, 2016 · Cardiovascular
Device Facts
Record ID
K160231
Device Name
Model X-100C CO-Met™ Oximetry System
Applicant
Nonin Medical, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Dec 23, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Nonin's Model X-100C CO-Met™ Oximetry System is indicated for noninvasive measuring of functional oxygen saturation of arterial hemoglobin (%SpO2), carboxyhemoglobin saturation (%COHb), methemoglobin saturation (%MetHb), and pulse rate of adult and pediatric patients (>66 lbs / 30 kg) using the Model 8300AA sensor. The measurements may be multiple spot-checks to observe change and/or continuous monitoring. The X-100C system is indicated for use by trained personnel in clinical settings, including Emergency Medical Service (EMS), hospitals, medical facilities, and mobile environments. This device is not meant for sole use in clinical decision making; it must be used in conjunction with additional methods of assessing clinical signs and symptoms.
Device Story
Model X-100C CO-Met Oximetry System uses Model 8300AA reusable finger clip sensor to measure arterial hemoglobin saturation (%SpO2), carboxyhemoglobin (%COHb), methemoglobin (%MetHb), and pulse rate. Device processes signals from sensor; displays numeric values, pleth waveform, and alarm status on backlit LCD. Used by trained healthcare professionals in hospitals, EMS, and mobile environments. Provides real-time data and alarm management; supports AC or battery power. Output interfaces include Bluetooth, RS-232, and Nurse Call/Analog. Device intended for use alongside other clinical assessment methods to aid patient monitoring; does not provide diagnostic conclusions independently.
Clinical Evidence
Clinical accuracy testing conducted on healthy adult subjects (18+ years) of varying skin tones. COHb and MetHb accuracy evaluated against co-oximetry (0-15% range). SpO2 accuracy evaluated during induced hypoxia (70-100% range) and in the presence of COHb/MetHb (80-100% SpO2, 0-15% COHb/MetHb). All testing performed per ISO 80601-2-61. Results demonstrated performance within acceptance criteria.
Technological Characteristics
Noninvasive pulse oximetry using reusable finger clip sensor (Model 8300AA). Backlit LCD display. Powered by 7.2V Li-ion battery or 100-240 VAC. Connectivity: Bluetooth, RS-232, Nurse Call/Analog. Biocompatible materials (ISO 10993). IP32 ingress protection. Complies with IEC 60601-1, IEC 60601-1-2, and ISO 80601-2-61.
Indications for Use
Indicated for noninvasive measurement of %SpO2, %COHb, %MetHb, and pulse rate in adult and pediatric patients (>66 lbs/30 kg). Used for spot-checks or continuous monitoring by trained personnel in clinical, EMS, hospital, and mobile environments. Not for sole clinical decision-making.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Masimo Rainbow SET Radical 7 Pulse CO-Oximeter (K061204)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three abstract human figures connected by flowing lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 23, 2016
Nonin Medical, Inc. Kim Aves Principal Regulatory Affairs Specialist 13700 1st Avenue North Plymouth, Minnesota 55441
Re: K160231
Trade/Device Name: Model X-100C CO-Met™ Oximetry System Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: November 28, 2016 Received: December 1, 2016
Dear Kim Aves:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K160231
#### Device Name
Model X-100C CO-Met™ Oximetry System
# Indications for Use (Describe)
Model X-100C
Nonin's Model X-100C CO-Met™ Oximetry System is indicated for noninvasive measuring of functional oxygen saturation of arterial hemoglobin (%SpO2), carboxyhemoglobin saturation (%COHb), methemoglobin saturation (%MetHb), and pulse rate of adult and pediatric patients (>66 lbs / 30 kg) using the Model 8300AA sensor. The measurements may be multiple spot-checks to observe change and/or continuous monitoring. The X-100C system is indicated for use by trained personnel in clinical settings, including Emergency Medical Service (EMS). hospitals, medical facilities, and mobile environments. This device is not meant for sole use in clinical decision making; it must be used in conjunction with additional methods of assessing clinical signs and symptoms.
#### Model 8300AA
The Model 8300AA reusable finger clip sensor is intended for noninvasive measuring of function of arterial hemoglobin (%SpO2), carboxyhemoglobin saturation (%COHb), methemoglobin saturation (%MetHb), and pulse rate of adult and pediatric patients (> 66 lbs / 30 kg). The measurements may be multiple spot-checks to observe change and/or continuous monitoring. It is intended for use in Emergency Medical Service (EMS), hospitals, medical facilities, and mobile environments. This device is not meant for sole use in clinical decision making: it must be used in conjunction with additional methods of assessing clinical signs and symptoms.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="white-space: nowrap;"> <div style="display: inline-block;"> <input checked="true" type="checkbox"/> </div> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="white-space: nowrap;"> <div style="display: inline-block;"> <input type="checkbox"/> </div> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 5. "510(k) Summary" as required by section 807.92(c)
| Submitter: | Nonin Medical, Inc.<br>13700 1st Ave. North<br>Plymouth, MN 55441-5443 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Phone: | 763 577 3196 |
| Fax: | 763 553 7807 |
| Contact Person: | Kim E. Aves<br>Principal Regulatory Affairs Specialist |
| Date Prepared: | January 29, 2016; Revised December 22, 2016 |
| Trade Names: | Model X-100C CO-Met™ Oximetry System |
| Common Name: | Oximeter |
| Classification Name: | Oximeter |
| Regulation Number: | Class II, 21 CFR 870.2700 (Oximeter) |
| Product Code, Panel: | DQA, Anesthesiology |
| Predicate Device(s): | Masimo Rainbow SET Radical 7 Pulse CO-Oximeter, K061204 |
| | Device Description: The Masimo Rainbow SET® Radical 7 Pulse CO-<br>Oximeter (Radical 7) with Rainbow technology allows for noninvasive<br>monitoring of arterial oxygen saturation (%SpO2), pulse rate,<br>carboxyhemoglobin saturation (%SpCO), and/or methemoglobin<br>saturation (%SpMet). The Radical 7 features an LCD display that<br>continuously displays numeric values for %SPO2 and pulse rate. Other<br>information displayed by the Radical 7 include: %SpCO and/or %SpMet,<br>Low Signal IQ (Low SIQ), Perfusion Index (PI), alarm status, alarm<br>silence, battery life, sensor status, trends, and pleth waveform. The<br>Radical 7 has output interfaces including SatShare connection to multi- |
The Radical 7 is intended to be used with Masimo LNOP series of oximetry sensors, LNCS series of oximetry sensors, and patient cables. The Radical 7 is also intended to be used with Masimo Rainbow (SpCO/SpMet) sensors and Rainbow patient cables, cleared under K061204 (July 27, 2006.)
parameter monitors, Nurse Call analog output, and RS-232 serial output.
#### Radical 7 Indications for Use (as cleared under K061204)
The Masimo SET® Radical 7 Pulse CO-Oximeter and accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SP02), pulse rate (measured by an SP0 2 sensor), carboxyhemoglobin (measured by an SpCO/SpMet sensor), and/or methemoglobin saturation (measured by an SpCO/SpMet sensor). The Masimo SET® Radical 7 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both
{4}------------------------------------------------
no motion and motion conditions, and for patients who are well or poorly perfused in hospitals. hospital-tvpe facilities, mobile, and home environments. In addition, the Masimo SET® Radical 7 Pulse CO-Oximeter and accessories is indicated to provide the continuous noninvasive monitoring data obtained from the Masimo SET® Radical 7 Pulse CO-Oximeter and accessories of functional oxygen saturation of arterial hemoglobin (Sp0-) and pulse rate (measured by an Sp0- sensor) to multiparameter devices for the display of those devices.
Device Description: The Model X-100C CO-Met Oximetry System displays (COHb), methemoglobin (MetHb), pulse oxygen saturation (SpO2), and pulse rate (PR) data transmitted from one channel of data through a direct connection of the signal processor to the monitor. The monitor receives, displays, provide alarm management, and storage of COHb, MetHb, SpO2, and PR. The device is capable of running on battery power or on AC power via an external power supply. The device is equipped with patient alarms to alert the user to abnormal conditions and shall provide real time data output of COHb, MetHb, SpO2, and PR.
#### Intended Use: Model X-100C
Nonin's Model X-100C CO-Met™ Oximetry System is indicated for noninvasive measuring of functional oxygen saturation of arterial hemoglobin (%SpO2), carboxyhemoglobin saturation (%COHb), methemoglobin saturation (%MetHb), and pulse rate of adult and pediatric patients (>66 lbs / 30 kg) using the Model 8300AA sensor. The measurements may be multiple spot-checks to observe change and/or continuous monitoring. The X-100C system is indicated for use by trained personnel in clinical and non-clinical settings, including Emergency Medical Service (EMS), hospitals, medical facilities, and mobile environments. This device is not meant for sole use in clinical decision making; it must be used in conjunction with additional methods of assessing clinical signs and symptoms.
#### Model 8300AA
The Model 8300AA reusable finger clip sensor is intended for noninvasive measuring of functional oxygen saturation of arterial hemoglobin (%SpO2), carboxyhemoglobin saturation (%COHb), methemoglobin saturation (%MetHb), and pulse rate of adult and pediatric patients (> 66 lbs / 30 kg). The measurements may be multiple spot-checks to observe change and/or continuous monitoring. It is intended for use in Emergency Medical Service (EMS), hospitals, medical facilities, and mobile environments. This device is not meant for sole use in clinical decision making; it must be used in conjunction with additional methods of assessing clinical signs and symptoms.
{5}------------------------------------------------
# Comparison of Intended Use (Indications for Use)
| Radical 7 | X-100C | Comment |
|---------------------------------------------------------------------|------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Continuous monitoring | Multiple spot-checks to observe change and/or continuous monitoring | Multiple spot-checks to observe change is a form of monitoring. |
| Adult, pediatric and neonatal patients | Adult and pediatric | Nonin is not seeking clearance for X-100C in the neonate population. The Radical 7 applied part labeling indicates neonate use for SpO2 and pulse rate only. |
| Intended use does not specify patient size, weight or age | Nonin specifies patient population based on weight (>66 lbs./30kg) | The X-100C clarifies the use population regarding appropriate fit of the applied part, to ensure accuracy of measurements. |
| Does not specify user qualifications | For use by trained personnel | The X-100C is meant to be used by trained healthcare professionals to ensure proper use of the device and understanding of the display values. The X-100C is meant to be used in conjunction with additional methods of assessing clinical signs and symptoms. |
| Motion and no motion conditions | Nonin is not seeking motion claims for any parameters | The Radical 7 labeling indicates motion and no motion specific to SpO2 and pulse rate only. The X-100C is meant to be used in conjunction with additional methods of assessing clinical signs and symptoms. |
| Poorly perfused patients | Nonin is not seeking low perfusion claims | The Radical 7 labeling indicates poor perfusion specific to SpO2 and pulse rate only. The X-100C is meant to be used in conjunction with additional methods of assessing clinical signs and symptoms. |
| Hospitals, hospital-type facilities, mobile, and home environments. | Emergency Medical Service (EMS), hospitals, medical facilities, and mobile environments. | The X-100C is intended for use by trained healthcare professionals, not the lay or home user who are not trained in the proper use of the device and/or understanding of the display values. |
{6}------------------------------------------------
Testing: The Nonin Model X-100C and 8300AA sensors are supported by electrical and mechanical safety, electromagnetic compatibility, device performance, and clinical testing to ensure appropriate functionality and to demonstrate substantial equivalence to the predicate devices. The devices were tested with the Model X-100 system. Table 1 summarizes the technological characteristics of the proposed device and the predicate device.
| CATEGORY | Identical/Similar/<br>Different | Model X-100C System<br>(Subject) | Radical-7™ K061204<br>(Predicate) |
|-------------------------------------------------------------------------------|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Standalone<br>• AC Power<br>Requirements<br>• Power<br>consumption<br>• Fuses | Different | 100-240 VAC, 50-60 Hz | 100-240 VAC, 47-63 Hz<br>55 VA<br>1 Amp, Fast Acting, Metric,<br>(5x20mm), 250V |
| Batteries<br>• Type<br>• Capacity<br>• Charging Time | Different | 7.2 volt Li-ion battery pack, 2.4 Ah<br>when charged | NiMH (Handheld and docking<br>station) |
| | | 6 hours1 | 4 hours2 (Handheld)<br>10 hours3 (Docking station)<br>3 hours (Handheld)<br>6 hours (Docking station) |
| | | 2.5 hours | |
| Environmental<br>Operating<br>Temperature | Similar | 32 °F to 104 °F (0 °C to 40 °C) | 41 °F to 104 °F (5 °C to 40 °C) |
| Storage/Transport<br>Temperature | Similar | -22 °F to 158 °F (-30 ºC to 70 ºC) | -40 °F to 158 °F (-40 °C to 70 °C) |
| Operating Humidity | Similar | 15% to 93%, noncondensing | 5% to 95%, noncondensing |
| Storage/Transport<br>Humidity | Different | Up to 93%, noncondensing | Not specified |
| Operating Altitude | Different | 0 to 4,000 meters (0 to 13,124 feet) | -304 to 5,486 meters (-1,000 to<br>18,000 feet) |
| Physical Characteristics<br>Dimensions | Different | X-100MM<br>• 305 mm W x 180 mm H x 130<br>mm D (12.0" W x 7.2" H x 5.0"<br>D) X-<br>100SPMM<br>• 21.4 mm H x 21.7 mm W x 72.7<br>mm L (including strain relief)<br>with 0.75 m cable (0.84" H x<br>0.85" W x 3.1" L (including<br>strain relief) with 2.5 ft cable) | Handheld<br>• 8.9" x 3.3' x 2.1" (22.6 cm x<br>8.4 cm x 5.3 cm)<br>Standalone<br>• 3.5" x 10.5" x 7.7" (8.9 cm x<br>26.7 cm x 19.6 cm) |
| CATEGORY | Identical/Similar/<br>Different | Model X-100C System<br>(Subiect) | Radical-7 ™ K061204<br>(Predicate) |
| Weight | Different | X-100MM<br>• Approximately 900 grams (2 pounds)<br>X-100SPMM<br>• 40 grams (1.4 ounces) | Handheld: 1.3 lbs. (0.59 kg)<br>Docking station (RDS-1, RDS-2, RDS-3):<br>• 2.5 lbs. (1.14 kg)<br>Docking station (RDS-1B)<br>• 4.11 lbs. (1.86 kg)<br>Standalone (RDS-1, RDS-2, RDS-3)<br>• 3.8 lbs. (1.73 kg)<br>Standalone (RDS-1B)<br>• 5.4 lbs. (2.45 kg) |
| Trending | Different | Not specified | 72 hours of trending at 2 second<br>resolution, up to 18 days of<br>trending at 10 second resolution,<br>output to serial printer or other<br>serial devices. |
| SpO2 Averaging | Similar | 6 seconds | 2, 4, 8, 10, 12, 14, or 16 seconds4 |
| Display/Indicators | Similar | %SpO₂, %COHb, %MetHb, pulse<br>rate, pleth waveform, alarm status,<br>status messages, , sensor fault, poor<br>signal indicator, signal processor<br>communication error, visual alarm<br>indicator, alarm silence, Bluetooth,<br>battery indicator and date and time. | %SpO2, %SpCO, %SpMet,<br>pulserate, pleth waveform,<br>alarm status, trends, status<br>messages,battery status, visual<br>alarm indicator, Signal IQ,<br>perfusion index, pleth variability<br>index, APOD and FastSat. |
| Display Update<br>Rate | Similar | 2-5 seconds, typically < 2 seconds | 1 second |
| Response Time | Similar | <10 second delay | < 20 second delay |
| Display color | Similar | Black | Blue |
| Display Type | Identical | Backlit LCD | Backlit LCD |
| Pixels | Similar | 800 x 600 | 480 x 160 dots |
| Output Interface | Similar | Bluetooth<br>Serial RS-232<br>Nurse Call/Analog Output | SatShare (RDS-1)<br>Serial RS-232 (RDS-1, RDS-1B,<br>RDS-3)<br>Nurse Call/Analog Output (RDS-1,<br>RDS-1B, RDS-3)<br>Phillips Vuelink, Spacelabs<br>Universal Flexport, RadNet,<br>RadLink (RDS-1, RDS-1B, RDS-3) |
| Compliance | | | |
| EMC Compliance | Identical | IEC 60601-1-2, Class B | EN60601-1-2, Class B |
| Equipment<br>Classification | Similar | IEC 60601-1 /CAN/CSA-C22.2 No.<br>601.1 / UL 60601-1, | IEC 60601-1 / UL 60601-1 |
| Type of Protection | Similar | Internally powered (on<br>batterypower). Class II with AC | Class 1 (on AC power), Internally<br>powered (on battery power) |
| Degree of<br>Protection | Similar | Patient Cable: Type BF-Applied Part | Patient Cable: Type BF-Applied<br>Part<br>SatShare Cable: Type BF-Applied<br>Part |
| Mode of Operation | Identical | Continuous | Continuous |
| Enclosure Degree<br>of Ingress<br>Protection | Different | IP32 | Not Specified |
## Table 1
4 Fully charged battery and screen at default brightness
2 Fully charged battery with the backlight at minimum and Power save mode on
3 Fully charged battery with the backlight at minimum and Power save mode on
{7}------------------------------------------------
4 With FastSat the averaging time is dependent on the 2 and 4 second settings the averaging time may range from 2-4 and 4-6 seconds respectively.
{8}------------------------------------------------
## Functional, Electrical and
Mechanical Safety Testing: The results of the testing demonstrate equivalency with the predicate 2 summarizes test results for the proposed devices, which met the relevant requirements of the applicable recognized standards.
### Table 2
| Test | Reference | Result |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Electrical Safety | IEC 60601-1 | Pass |
| Temperature and Humidity | EN 1789<br>IEC 60601-1 | Pass |
| Atmospheric Pressure (Altitude) | IEC 60601-1 | Pass |
| Electromagnetic Immunity and<br>Emissions | IEC 60601-1-2 | Pass |
| Performance | ISO 80601-2-61<br>IEC 60601-1<br>IEC 60601-1-6<br>IEC 60601-1-12<br>IEC 62304<br>ANSI/AAMI EC13<br>ISO 14155 | Pass |
| Ingress Protection | ISO 80601-2-61 | Pass |
| Diaphoretic related ingress | Internal performance<br>characterization | Pass |
| Mechanical Durability | IEC 60601-1<br>ISO 80601-2-61<br>ISTA 2A ASTM<br>D-4169 | Pass |
| Biocompatibility | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | Biocompatible<br>per Cytotoxicity<br>Sensitization and<br>Irritation |
#### Accuracy (clinical) testing:
#### COHb Accuracy Testing
COHb accuracy testing was conducted at an independent research laboratory on healthy, male and female, non-smoking, light to darkskinned subjects that were 18 years of age and older.
The measured carboxyhemoglobin value (%COHb) of the sensors was compared to simultaneous arterial blood samples as assessed by cooximetry. The accuracy of the sensors in comparison to the co-oximeter samples measured over the COHb range of 0-15% with 95 - 100% SaO2. Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61, Medical Electrical Equipment- Particular requirements for basic safety and essential performance of pulse oximeter equipment.
{9}------------------------------------------------
#### MetHb Accuracy Testing
MetHb accuracy testing was conducted at an independent research laboratory on healthy, male and female, non-smoking, light to darkskinned subjects that were 18 years of age and older.
The measured methemoglobin value (%MetHb) of the sensors was compared to simultaneous arterial blood samples as assessed by cooximetry. The accuracy of the sensors in comparison to the co-oximeter samples measured over the MetHb range of 0 – 15% with 95 – 100% SaO2.
Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61, Medical Electrical Equipment- Particular requirements for basic safety and essential performance of pulse oximeter equipment.
#### SpO2 Accuracy Testing
During no-motion conditions at an independent research laboratory, SpO2 accuracy testing was conducted during induced hypoxia studies on healthy, male and female, non-smoking, light- to dark-skinned subjects that were 18 years of age and older. The measured arterial hemoglobin saturation value (SpO2) of the sensors was compared to arterial hemoglobin oxygen (SaO2) value, determined from blood samples witha laboratory co-oximeter. The accuracy of the sensors in comparison tothe co-oximeter samples measured over the SpO2 range of 70 - 100%. Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61, Medical Electrical Equipment— Particular requirements for basic safety and essential performance of pulse oximeter equipment.
#### SpO2 Accuracy Testing in Presence of COHb and MetHb
During no-motion conditions at an independent research laboratory, SpO2 accuracy testing in the presence of COHb and MetHb was conducted during induced hypoxia studies on healthy, male and female, non-smoking, light- to dark-skinned subjects that were 18 years of age and older.
The measured arterial hemoglobin saturation value (SpO2) of the sensors was compared to arterial hemoglobin oxygen (SaO2) value, determined from blood samples with a laboratory co-oximeter.
The accuracy of the sensors in comparison to the co-oximeter samples measured over the SpO2 range of 80 - 100%, range 0 - 15% COHb and SpO2 range of 80 – 100%, range 0 – 15% MetHb.
Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61, Medical Electrical Equipment— Particular requirements for basic safety and essential performance of pulse oximeter equipment.
{10}------------------------------------------------
| | Testing conclusion: the proposed Model X-100C CO-Met Oximetry System<br>meets all acceptance criteria. Based on test results, analysis and comparison<br>to the legally marketed predicates, the Model X-100C CO- Met Oximetry<br>System performance is substantially equivalent to the predicate device for its<br>intended use. |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of<br>Substantial<br>Equivalence: | The Model X-100C CO-Met Oximetry System has the following similarities to<br>the predicate devices:<br>Similar indications for use Similar intended use environments Similar patient population Same primary mode of operation Perform equivalently to the same specifications |
| Conclusion: | No new questions of safety and effectiveness were raised. Based on the<br>results of clinical, environmental, safety, performance and usability testing,<br>and risk management assessment, Nonin Medical has determined that the<br>proposed Model X-100C CO-Met Oximetry System is substantially equivalent<br>to the predicate Masimo Radical 7 CO-Oximetry System for its intended use. |