The Pulse Oximeter is intended for continuous monitoring, spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of single adult, adolescent, child and infant patients in hospitals and clinics.
Device Story
Pulse oximeter (MD300M, MD300K2) monitors SpO2 and pulse rate; uses photoelectric oxyhemoglobin inspection technology; inputs light signals (660nm red, 905nm near-infrared) via finger-clamping sensor; processes signals via microprocessor using Lambert-Beer Law; outputs SpO2%, pulse rate, pulse amplitude index, and alarm status to TFT display; used in hospitals/clinics by clinicians; provides physiological/technical alarms; aids clinical assessment of patient oxygenation status.
Clinical Evidence
Clinical study conducted on 12 healthy adult volunteers (diverse pigmentation, ages 21-43) per ISO 80601-2-61. Evaluated SpO2 accuracy (70-100% range) for three probe types. Results showed Arms of 1.75 (M-50E012CS09), 1.48 (M-50B008CS09), and 1.75 (M-50J033CS045).
Technological Characteristics
Photoelectric sensor (660nm/905nm); ABS plastic case; medical silicone/microfoam finger cushions; 3 AA batteries or 5V adapter; TFT color display; microprocessor-based signal processing; software-driven; complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, ISO 80601-2-61, and ISO 10993.
Indications for Use
Indicated for continuous monitoring and spot-checking of functional SpO2 and pulse rate in adult, adolescent, child, and infant patients in hospital and clinical settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three heads, representing the department's commitment to health, human services, and science. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 18, 2016
Beijing Choice Electronic Technology Co., Ltd. Lei Chen Quality Director North Building 3F, No.9 Shuangyuan Road Badachu Hi-tech Zone, Shijingshan District Beijing, China 100041
Re: K152563
Trade/Device Name: Pulse Oximeter (MD300M, MD300K2) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: March 7, 2016 Received: March 11, 2016
Dear Mr. Lei Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K152563
Device Name Pulse Oximeter
Indications for Use (Describe)
The Pulse Oximeter is intended for continuous monitoring, spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of single adult, adolescent, child and infant patients in hospitals and clinics.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section III 510(k) Summary
This summary of 510(k) is being submitted in accordance with the requirements of 21 CFR 807.92.
There is no prior submission for the device.
#### 3.1Submitter Information
#### ● Manufacturer Name:
Establishment Registration Number: 3005569927 Beijing Choice Electronic Technology Co., Ltd. Room 4104,No.A12 Yuquan Road Haidian District 100143 Beijing, P.R.China
#### . Contact Person:
Mr. Lei Chen Beijing Choice Electronic Technology Co., Ltd. North Building 3F, No. 9 Shuangyuan Road, Badachu Hi-tech Zone, Shijingshan District Beijing China 100041 Phone: +86-10-88798300 Ext 6020 Fax:215-4052545 Email: cc@choicemmed.com
- Date prepared: March 17, 2016
#### 3.2Proposed Device Information
Device Common Name: Pulse Oximeter Device Trade/Proprietary Name: Pulse Oximeter Model: MD300M, MD300K2 Classification Name: Oximeter Regulation Number: 870.2700 Product Code: DQA
Class: II
Panel: Anesthesiology
### 3.3Predicate Device
510(k) Number: K090599
{4}------------------------------------------------
Premarket Notification 510(k) Submission-Section III 510(k) Summary Common Name: Pulse Oximeter Device Trade/Proprietary Name: Pulse Oximeter Model: MD300K1 Classification Name: Oximeter Product Code: DQA Regulation Number: 870.2700
Device Class:
Panel: Anesthesiology
Manufacturer: Beijing Choice Electronic Technology Co., Ltd.
Intended Use: MD300K1 Pulse Oximeter is intended for continuous monitoring, spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of single adult and pediatric patients in hospitals and home care.
#### 3.4Device Description
The Pulse Oximeter MD300M/MD300K2 can display the SpO2%, pulse rate, Pulse Amplitude Index and other indication parameters, such as time, ID number, pulse amplitude bar and battery power status, alarm limits and the connections of sensors. It is used for adult, adolescent, child and infant patients. The device has physiological alarm and technology alarm function. There are two-level alarm priorities in oximeter. High priority: "Di-Di-Di-----------Di-Di" indicates the patient is in the very dangerous situation. Low priority: "Di" indicates the technical alarm caused by the device itself. The device also has visual alarm function to indicate users by lamp and information on the device. The power supply of the applicant device is 3 AA alkaline batteries, rechargeable batteries or adapter.
The proposed device consists of photo detector, display screen, signal amplify unit, CPU, display unit and power supply unit.
Principle of the oximeter is as follows: A mathematic formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin (RHb) and Oxyhemoglobin (HbO2) in red and near-infrared zones. Operation principle of the instrument: Photoeletric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm red and 905nm near infrared light) can be focused on a human nail tip through a clamping finger-type sensor. A measured signal obtained by a photosensitive element, will be shown on the oximeter's display through process in electronic circuits and microprocessor.
The proposed device is not for life-supporting or life-sustaining, not for implant.
The device is not sterile and the transducers are reusable and do not need sterilization and re-sterilization.
The device is for prescription.
{5}------------------------------------------------
The device does not contain drug or biological products.
The device is software-driven and the software validation is provided in software.
{6}------------------------------------------------
## 3.5 Comparison list of the technological characteristics
| Comparison Elements | Proposed Device | Predicate Device | | |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Product Name | Pulse Oximeter | Pulse Oximeter | | |
| Model | MD300M | MD300K1 | | |
| Regulation No. | 21 CFR 870.2700 | 21 CFR 870.2700 | | |
| Classification | II | II | | |
| Classification Name | Oximeter | Oximeter | | |
| Product Code | DQA | DQA | | |
| Indications for Use | The MD300M pulse oximeter is intended for continuous monitoring, spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of single adult, adolescent, child and infant patients in hospitals and clinics. | MD300K1 Pulse Oximeter is intended for continuous monitoring, spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of single adult and pediatric patients in hospitals and home care. | | |
| Comparison Statement | The proposed device and the predicated device have the similar intended use and classification. The only difference is that the proposed device cannot be used in the home care environment while the predicated device can be used in the home care environment. | | | |
| Components | Photo detector, display screen, signal amplify unit, CPU, display unit and power unit. | Photo detector, display screen, signal amplify unit, CPU, display unit and power unit. | | |
| Design Principle | Principle of the oximeter is as follows: A mathematic formula is established making use | The MD300K1 Pulse Oximeter works by applying a sensor to a pulsating arteriolan | | |
| | | | | |
| | | of Lambert Beer Law according to Spectrum | vascular bed. The sensor contains a dual light | |
| | | Characteristics<br>Absorption<br>of<br>Reductive | source and photo detector. The one wavelength | |
| | | (RHb) and<br>Oxyhemoglobin<br>hemoglobin | of light source is 660nm, which is red light; the | |
| | | red and near-infrared zones.<br>(HbO2) in | other is 940nm, which is ultra red light. Skin, | |
| | | principle<br>Operation<br>of<br>the<br>instrument: | bone, tissue, and venous normally absorb a | |
| | | Oxyhemoglobin<br>Photoeletric<br>Inspection | constant amount of light during systole and | |
| | | Technology is adopted in accordance with | diastole, as blood volume increases and | |
| | | Capacity Pulse Scanning and Recording | decreases. The ratio of light absorbed at systole | |
| | | Technology, so that two beams of different | and diastole is translated into an oxygen | |
| | | wavelength of lights (660nm red and 905nm | saturation measurement. This measurement is | |
| | | near infrared light) can be focused on a human | referred to as SpO2. | |
| | | nail tip through a clamping finger-type sensor. | | |
| | | signal<br>measured<br>obtained<br>A<br>by<br>ದ | | |
| | | photosensitive element, will be shown on the | | |
| | | display through process<br>oximeter's<br>in | | |
| | | electronic circuits and microprocessor. | | |
| Measurement | Red | 660±3nm | 660 nanometers | |
| Wavelength | Infrared | 905 ± 8nm | 940 nanometers | |
| Comparison Statement | | The proposed device and the predicate device have the same design principle and similar<br>measurement wavelength. | | |
| | | | | |
| | Display Type | TFT (colorized) | OLED | |
| | Battery | 3*AA alkaline batteries, rechargeable | 2*AA alkaline batteries | |
| | | batteries or adapter | | |
| Power adapter | 5V 2A | Not available | | |
| SpO2 display range | 0%~100% | 0~100% | | |
| SpO2 measurement<br>range | 0~100% | 0~100% | | |
| SpO2 accuracy | 70%~100%, $ \u00b1 $ 2%;<br>0~69% no definition | 70%~100%, $ \u00b1 $ 3%;<br>0~69% no definition | | |
| PR display range | 0~250bpm | 0~235bpm | | |
| PR measurement<br>range | 30~250bpm | 30~235bpm | | |
| PR accuracy | 30bpm~99bpm, $ \u00b1 $ 2bpm<br>100bpm~250bpm, $ \u00b1 $ 2% | 30bpm~99bpm, $ \u00b1 $ 2bpm<br>100bpm~235bpm, $ \u00b1 $ 2% | | |
| PR Resolution | 1% | 1% | | |
| Pulse amplitude<br>index measurement<br>range | 0.1%~20% | --- | | |
| Operating<br>temperature | 0°C~40°C | 5°C~40°C | | |
| Relative humidity | $ \le $ 80% no condensation in operation;<br>$ \le $ 93% no condensation in storage | $ \le $ 80% no condensation in operation;<br>$ \le $ 93% no condensation in storage | | |
| Atmosphere pressure | 86kPa~106kpa | 86kPa~106kpa | | |
| Alarm type | | Audio type, visual alarm and information | Audio type, visual alarm and information | |
| Alarm limit range | | SpO2 70%~100%<br>PR 30bpm~250bpm | SpO2 70%~100%<br>PR 30bpm~235bpm | |
| Default<br>Limit | SpO2 | High 100%, Low 90% | High 99%, Low 90% | |
| | PR | High 100bpm, Low 60bpm | High 100bpm, Low 60bpm | |
| Comparison Statement | | | The proposed device has similar product specification as predicate device. | |
| Contacting<br>Material | Battery Cover | ABS | ABS | |
| | Plastic Case Cover | | | |
| | Fingertip Cushion | Medical Silicone Gel (Finger probe M-50E012CS09, M-50B008CS09) | Medical Silicone Gel (Finger probe M-50 series) | |
| | Fingertip Cushion | Microfoam (Finger probe M-50J033CS045) | N/A | |
| Comparison Statement | | | The contacting materials of the proposed device are similar to those of the predicate device. The only difference is that the material of MD300M supporting finger probe M-50J033CS045 is microfoam while which is not available to the predicate device. | |
| Performance Testing | Laboratory Testing | The laboratory tests include SpO2 and PR accuracy Test, Weak Perfusion Test, High and Low Temperature and Humidity Test, Performance Test After Cleaning and ISO80601-2-61 | Meet the requirements of FDA Guidance | |
| | | | | |
| | Clinical Testing | Conformed to ISO9919&ISO 80601-2-61 | Conformed to ISO9919 | |
| | | | | |
| | | Clinical Test for the device accuracy is conducted in Yue Bei people's Hospital. The<br>clinical test report and protocol are provided<br>in Performance Testing-Clinical Test Report | | |
| Electrical | Electrical Safety | Conformed to IEC60601-1, IEC 60601-1-8 | | Conformed to IEC60601-1 |
| | Electromagnetic Compatibility | Conformed to IEC60601-1-2 | | Conformed to IEC60601-1-2 |
| | EMC<br>and<br>Safety | Software | Moderate level of concern | |
| | | | Compliance with FDA Guidance for the<br>content of Premarket Submissions for<br>Software Contained in Medical Devices | |
| | | | Risk Management in Compliance with<br>ISO14971:2007 | |
| Biocompati | Microfoam | In Vitro Cytotoxicity | No cytotoxic potential | N/A |
| | | Animal skin irritation test | No evidence of significant irritation from the test extract to rabbits | |
| | | Skin Sensitization Test | No evidence of sensitization was observed. | |
| Comparison Statement | | | Compliance with ISO10993 | |
| Label and Labeling | | Compliance with the Guidance of pulse oximeter-premarket notification issued on March 4,2013 | Compliance with FDA guidance | |
Table 3-1 Performance Specification Comparison Table between the Proposed Device (MD300M) and Predicate Device
{7}------------------------------------------------
{8}------------------------------------------------
Performance Specification
{9}------------------------------------------------
Premarket Notification 510(k) Submission—Section III 510(k) Summary
{10}------------------------------------------------
{11}------------------------------------------------
### Table 3-2 Performance Specification Comparison Table between the Proposed Device (MD300K2) and Predicate Device
| Comparison Elements | Proposed Device | Predicate Device |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | Pulse Oximeter | Pulse Oximeter |
| Model | MD300K2 | MD300K1 |
| Regulation No. | 21 CFR 870.2700 | 21 CFR 870.2700 |
| Classification | II | II |
| Classification Name | Oximeter | Oximeter |
| Product Code | DQA | DQA |
| Indications for Use | The MD300K2 pulse oximeter is intended for<br>continuous monitoring, spot-checking of<br>functional pulse oxygen saturation (SpO2) and<br>pulse rate (PR) of single adult, adolescent,<br>child and infant patients in hospitals and<br>clinics. | MD300K1 Pulse Oximeter is intended for<br>continuous monitoring, spot-checking of<br>functional pulse oxygen saturation (SpO2)<br>and pulse rate (PR) of single adult and<br>pediatric patients in hospitals and home care. |
{12}------------------------------------------------
| | Comparison Statement | The proposed device and the predicated device have the similar intended use and<br>classification. The only difference is that the proposed device cannot be used in the home<br>care environment while the predicated device can be used in the home care environment. | | |
|--|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| | Components | Photo detector, display screen, signal amplify<br>unit, CPU, display unit and power unit. | Photo detector, display screen, signal amplify<br>unit, CPU, display unit and power unit. | |
| | Design Principle | Principle of the oximeter is as follows: A<br>mathematic formula is established making use<br>of Lambert Beer Law according to Spectrum<br>Absorption Characteristics of Reductive<br>hemoglobin (RHb) and Oxyhemoglobin<br>(HbO2) in red and near-infrared zones.<br>Operation principle of the instrument:<br>Photoeletric Oxyhemoglobin Inspection<br>Technology is adopted in accordance with<br>Capacity Pulse Scanning and Recording<br>Technology, so that two beams of different<br>wavelength of lights (660nm red and 905nm<br>near infrared light) can be focused on a human<br>nail tip through a clamping finger-type sensor.<br>A measured signal obtained by a photosensitive<br>element, will be shown on the oximeter's<br>display through process in electronic circuits<br>and microprocessor. | The MD300K1 Pulse Oximeter works by<br>applying a sensor to a pulsating arteriolar<br>vascular bed. The sensor contains a dual light<br>source and photo detector. The one<br>wavelength of light source is 660nm, which<br>is red light; the other is 940nm, which is ultra<br>red light. Skin, bone, tissue, and venous<br>normally absorb a constant amount of light<br>during systole and diastole, as blood volume<br>increases and decreases. The ratio of light<br>absorbed at systole and diastole is translated<br>into an oxygen saturation measurement. This<br>measurement is referred to as SpO2. | |
| | Measurement<br>Wavelength | Red | 660±3nm…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.