NELLCOR PORTABLE SPO2 PATIENT MONITORING SYSTEM ( NPSPMS )
Applicant
Covidien, LLC
Product Code
DQA · Cardiovascular
Decision Date
Oct 2, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Nellcor Portable SpO2 Patient Monitoring System is indicated for prescription use only for spot check or continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpOz) and pulse rate. It is intended for use with neonatal, pediatric, and adult patients during both no motion conditions and for patients who are either well or poorly perfused, in hospitals, hospital-type facilities, mobile, and home environments.
Device Story
Portable pulse oximeter; monitors functional oxygen saturation (SpO2) and pulse rate. Inputs: optical signals from Nellcor OxiMAX sensors. Processing: OxiMAX technology PCBA; calculates SpO2 and pulse rate. Outputs: digital SpO2/pulse rate values; pulse amplitude via blip bar or plethysmographic waveform. Used in hospitals, mobile, and home environments; operated by clinicians or patients. Provides real-time physiological data to support clinical decision-making; enables non-invasive patient monitoring.
Clinical Evidence
No new clinical data; relies on clinical data previously submitted for predicates K060576 and K123581. Performance established via non-clinical bench testing, including ISO 80601-2-61:2011 and IEC 60601-1:2005 compliance, pulse oximetry performance verification, and human factors summative usability validation. Motion testing validated pulse rate accuracy (20-250 bpm) using a functional tester.
Technological Characteristics
Hand-held portable pulse oximeter. Powered by four AA batteries. Uses Nellcor OxiMAX sensors. PCBA and software identical to N-600X predicate. Complies with ISO 80601-2-61:2011 and IEC 60601-1:2005. Connectivity via sensor cable.
Indications for Use
Indicated for prescription use for spot check or continuous non-invasive monitoring of SpO2 and pulse rate in neonatal, pediatric, and adult patients. Suitable for well or poorly perfused patients under motion or no-motion conditions in hospital, mobile, or home settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of a human figure in profile, with three faces overlapping to represent the department's focus on health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 2, 2014
Covidien Ms. Mia Ware Senior Regulatory Affairs Product Specialist 6135 Gunbarrel Ave. Boulder, CO 80301
Re: K141542
Trade/Device Name: Nellcor Portable SpO2 Patient Monitoring System Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: August 28, 2014 Received: September 4, 2014
Dear Ms. Ware:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Ware
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows the name "Tejashri Purohit-Sheth, M.D." in a serif font. The name is written in black ink on a white background. The text appears to be a professional title, possibly indicating a medical doctor.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K141542
Device Name
NellcorTM Portable SpO2 Patient Monitoring System
#### Indications for Use (Describe)
The Nellcor Portable SpO2 Patient Monitoring System is indicated for prescription use only for spot check or continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpOz) and pulse rate. It is intended for use with neonatal, pediatric, and adult patients during both no motion conditions and for patients who are either well or poorly perfused, in hospitals, hospital-type facilities, mobile, and home environments.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
| Applicant Name and Address: | Covidien LP.<br>6135 Gunbarrel Ave<br>Boulder, CO 80301<br>Phone: (303) 305-2750<br>Fax: (303) 305-2212 |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration Number: | 2936999 |
| Device Name(s): | Nellcor™ Portable SpO2 Patient Monitoring System |
| Classification: | Class II |
| Classification Name: | Oximeter (74DQA) (per 21 CFR §870.2700) |
| Product Code: | DQA |
| Date Prepared: | 06/09/2014 |
| 510(k) Contact Person and Phone Number: | Mia M. Ware<br>Sr. Regulatory Affairs Specialist<br>Covidien - Respiratory and Monitoring Solutions<br>6135 Gunbarrel Ave.<br>Boulder, CO 80301<br>Phone: (303)305-2750<br>Fax: (303) 305-2212 |
| Name and Address of Manufacturing Site(s) | |
| Establishment Registration Number: | 30003591740 |
| Registered Establishment Name: | Mediana Co. LTD |
| Address: | Wonju Medical Industry Park,<br>1650-1, Donghwa-ri, Munmak-<br>eup, Wonju-si, Gangwon-do, Korea |
## Predicate Devices:
The predicate device(s) to which the Nellcor™ Portable SpO2 Patient Monitoring System is claiming substantial equivalence are as follows:
| Trade Name: | OxiMAX NPB-40 Pulse Oximeter | N-600X Pulse Oximeter |
|----------------|-------------------------------------------------------------------------------------------|--------------------------------------------------------|
| 510(k) Number: | K051352 (cleared on 8/11/05) | K123581 (cleared on 05/09/2013) |
| Applicant: | Nellcor Puritan Bennett, Incorporated<br>4280 Hacienda Drive<br>Pleasanton, CA 94588-8604 | Covidien LP<br>6135 Gunbarrel Ave<br>Boulder, CO 80301 |
{4}------------------------------------------------
# Purpose of this 510(k):
This 510(k) submission is to obtain market clearance for the Nellcor Portable Sp02 Patient Monitoring System, a line extension of the Nellcor pulse oximeters with OxiMAX technology.
# General Description:
The Nellcor Portable SpO2 Patient Monitoring System is a modification of the OxiMax NPB-40 and N-600X Pulse Oximetry Systems. It is designed for continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate using Nellcor pulse oximetry sensors with OxiMAX technology, and the oximetry sensor cable. The monitor displays digital values of SpO2 and Pulse Rate. Pulse Amplitude is displayed by means of a "blip bar" presentation or plethysmographic waveform. The Nellcor Portable Sp02 Patient Monitoring System is powered by four AA batteries.
# Proposed Nellcor Portable SpO2 Patient Monitoring System Indications for Use:
The Nellcor Portable SpO2 Patient Monitoring System is indicated for prescription use only for spot check or continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (Sp02) and pulse rate. It is intended for use with neonatal, pediatric, and adult patients during both no motion and motion conditions and for patients who are either well or poorly perfused, in hospitals, hospital-type facilities, mobile, and home environments.
# Summary of Technical Characteristics
The Nellcor Portable Sp02 Patient Monitoring System is a line extension of the Nell-1 family of pulse oximeters. It is technologically identical to the predicate devices. It has the same oximetry PCBA and software as N-600X and has the same hand-held portable form factor and performs within the same specifications as the NPB-40. The clinical performance when used with adult, pediatric and neonatal patients is the same as in the N-600X pulse oximeter and was established with that oximeter in K060576 and K123581. The Nellcor Portable SpO2 Patient Monitoring System is intended to be used with the same Nellcor Sp02 sensors that are commercially available and used with the predicate devices. Based on the results of the non-clinical validation studies, Covidien has established that the Nellcor Portable SpO2 Patient Monitoring System is substantially equivalent to the predicate devices.
## Non-clinical/bench-testing data
The performance testing section of this submission includes verification and validation reports for pulse oximetry performance in accordance with FDA Guidance document: "Pulse Oximeters - Premarket Notification Submissions [510(k)s] Guidance for Industry and Food and Drug Administration Staff". Non-clinical testing in this submission includes, but is not limited to, ISO 80601-2-61:2011 and IEC 60601-1:2005 test reports, Oximetry performance verification, Human Factors Summative Usability validation, and testing incorporating simulated motion performed to validate the pulse rate accuracy of in the range of 20 -250 beats per minute during motion using a functional tester.
## Discussion of clinical data
{5}------------------------------------------------
Clinical data submitted in K060576, and K123581 is applicable to the Nellcor Portable SpO2 Patient Monitoring System. Because there are no changes to the performance, technology, and intended use of the device, the clinical data submitted as part of the premarket notifications for the predicate devices also applies to the Nellcor Portable SpO2 Patient Monitoring System.
## Conclusions
The technological characteristics of the Nellcor Portable SpO2 Patient Monitoring System and the results of non-clinical and clinical tests do not raise new questions of safety or effectiveness when compared to the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.