The Fingertip Pulse Oximeters, MD300C series, are portable, non-invasive devices intended for spot checking of arterial hemoglobin oxygen saturation (SpO2) and pulse rate of adult and pediatric patient at hospital (including clinical use in internist/surgery, Anesthesia, and intensive care units).
Device Story
Fingertip pulse oximeter series (MD300C) measures SpO2 and pulse rate via non-invasive optical sensor. Device uses dual-wavelength LEDs (660 nm red, 940 nm infrared) and photodetector to measure light absorption in pulsating arteriolar vascular bed. CPU calculates SpO2 based on systolic/diastolic light absorption ratios. Operated by clinicians in hospital settings; device clips onto finger. Features include LED/OLED display, low battery indicator, and automatic power-off after 8 seconds of no signal. Select models (MD300CF3, MD300CF61) include pulse rate tone modulation. Provides real-time physiological data to assist clinicians in patient monitoring. Powered by 2 AAA batteries.
Clinical Evidence
Clinical study conducted per ISO 9919:2005, Annex EE.2. Sample size: 12 healthy adult volunteers (ages 21-31). Evaluated SpO2 accuracy under steady-state, non-motion conditions across 70-100% SaO2 range compared to reference CO-Oximetry. No adverse events reported. Results confirmed compliance with manufacturer accuracy specifications.
Technological Characteristics
Materials: ABS plastic enclosure, medical silicon gel fingertip cushion. Sensing: Dual-wavelength LED (660nm/940nm) and photodetector. Power: 2 AAA alkaline batteries. Form factor: Fingertip-mounted. Connectivity: Standalone. Software: Embedded firmware. Standards: IEC 60601-1, IEC 60601-1-2, ISO 9919:2005, ISO 10993.
Indications for Use
Indicated for spot-checking arterial hemoglobin oxygen saturation (SpO2) and pulse rate in adult and pediatric patients in hospital settings (internist, surgery, anesthesia, intensive care).
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Model 9550 Onyx® II Finger Pulse Oximeter (K051107)
Submission Summary (Full Text)
{0}------------------------------------------------
# Section 3 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
There is no prior submission for the devices.
The assigned 510(K) number is K130947.
### 3.1 Submitter Information
### Manufacturer Name and Address:
Establishment Registration Number: 3005569927 Beijing Choice Electronic Technology Co., Ltd. Room 320, West Building 4, No.83 Fuxing Road, Beijing 100039, P.R.China
### Contact Person:
Mr. Lei Chen Beijing Choice Electronic Technology Co., Ltd. North Building 3F, No.9 Shuangyuan Road, Badachu Hi -tech Zone, Shijingshan District, Beijing, P.R. China, 100041 Phone: +86-10-88798300 Ext 6020 Fax: 215-4052545 Email: cc@choicemmed.com Date Prepared: April 27, 2013
### 3.2 Proposed Device Information
Device Common Name: Pulse Oximeter Device Trade/Proprietary Name: Fingertip Pulse Oximeter Model: MD300C1/MD300C2/MD300CF3/MD300CF61/MD300C61/MD300C63 Classification Name: Oximeter Regulation Number: 870.2700 Product Code: DQA Class: II Panel: Anesthesiology
### 3.3 Predicate Device
Model 9550 Onyx® II Finger Pulse Oximeter K-number: K051107 Product Code: DQA
Indications for Use: The Nonin® Onyx® II Model 9550 Finger Pulse Oximeter is a small, lightweight portable device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. It is
OCT 03 2013
{1}------------------------------------------------
intended for spotchecking of adult and pediatric patients on fingers (other than the thumb) between 0.3 - 1.0 inch (0.8 - 2.5 cm) thick. The index finger is the recommended site. Manufactured by:
Nonin Medical, Inc. 13700 Ist Ave. N. Plymouth, MN 55441-5443
### 3.4 Device Description
The proposed devices of Fingertip Pulse Oximeters MD300C series are fingertip devices, which can display SpO2 and pulse rate value.
The proposed devices consist of detector and emitter LED, signal amplify unit, CPU, display unit and power unit.
The pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The wavelength of one light source is 660 nm, which is red light; the other is 940 nm, which is infrared light.
Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photodetector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.
The power sources of the proposed devices are 2 AAA alkaline batteries. All of the proposed devices have low battery voltage indicator function and all of the proposed devices will automatically power off when there is no signal for longer than 8 seconds.
The proposed devices of MD300CF3 and MD300CF61 have pulse rate tone modulation function. The function of pulse rate tone modulation is that the device can make a sound of Di-Di-Di to reflect pulse rate in real time.
The proposed devices are not for life-supporting or life-sustaining, not for implant. The devices or transducers are not sterile and the transducers are reusable and do not need sterilization or re-sterilization. The devices are for prescription. The devices do not contain drug or biological products.
The Fingertip Pulse Oximeters MD300C series share the same measurement principle and oximeter sensor and oxygen saturation module and power supply. The indented target population and use environment of the Fingertip Pulse Oximeters MD300C series are the same.
The devices are software-driven and the software validation is provided in Section of Software
{2}------------------------------------------------
et Notification 510(k) Submission - Section 3 510(K) Summary
Premarket Notification 510(k) Submission – Section 3 10(K) Summary
The differences between each model of the proposed devices are shown the table 3-
| | | MD300C1 | MD300C2 | MD300CF3 | MD300CF61 | MD300C61 | MD300C63 |
|----------------------------------------|------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|--------------------------|--------------------------|--------------------------|--------------------------|
| Display Range | SpO₂ | 0-99% | 0-100% | 0-100% | 0-100% | 0-99% | 0-100% |
| | PR | 0~254BPM | 0~254BPM | 0~254BPM | 0~254BPM | 0~254BPM | 0~254BPM |
| Measurement<br>Range | SpO₂ | 70%-99% | 70%-100% | 70%-100% | 70%-100% | 70%-99% | 70%-100% |
| | PR | 30~235BPM | 30~235BPM | 30~235BPM | 30~235BPM | 30~235BPM | 30~235BPM |
| Accuracy | SpO₂ | 70%-99%: ±2% | 70%-100%: ±2% | 70%-100%: ±2% | 70%-100%: ±2% | 70%-99%: ±2% | 70%-100%: ±2% |
| | SpO₂ | 0%-69%:<br>no definition | 0%-69%:<br>no definition | 0%-69%:<br>no definition | 0%-69%:<br>no definition | 0%-69%:<br>no definition | 0%-69%:<br>no definition |
| | PR | 30~99BPM:<br>±2BPM | 30~99BPM:<br>±2BPM | 30~99BPM:<br>±2BPM | 30~99BPM:<br>±2BPM | 30~99BPM:<br>±2BPM | 30~99BPM:<br>±2BPM |
| | PR | 100~235BPM:<br>±2% | 100~235BPM:<br>±2% | 100~235BPM:<br>±2% | 100~235BPM:<br>±2% | 100~235BPM:<br>±2% | 100~235BPM:<br>±2% |
| Resolution | SpO₂ | 1% | 1% | 1% | 1% | 1% | 1% |
| | PR | 1BPM | 1BPM | 1BPM | 1BPM | 1BPM | 1BPM |
| Display Screen | | LED | OLED | OLED | OLED | LED | OLED |
| Pulse rate tone<br>modulation function | | N | N | Y | Y | N | N |
| The shape and size of<br>shell | | Although the proposed devices have different appearances, their sizes are similar. Please see the pictures of each model of the proposed device for the differences in appearance. | | | | | |
{3}------------------------------------------------
remarket Notification 510(k) Submission - Section 3 510(K) Summar
# 3.5 Comparison list of the technological characteristics
| Comparison Elements | Proposed Device | Predicate Device |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Fingertip Pulse Oximeter MD300C series | Finger Pulse Oximeter (K051107) |
| Model | MD300C series: MD300C1/MD300C2/MD300CF3<br>/MD300CF61MD300C61/MD300C63 | Onyx® II Model 9550 |
| Regulation No. | 21 CFR 870.2700 | 21 CFR 870.2700 |
| Classification | II | II |
| Classification Name | Oximeter | Oximeter |
| Product Code | DQA | DQA |
| Indications for Use | The Fingertip Pulse Oximeters, MD300C series, are<br>portable, non-invasive devices intended for spot<br>checking of arterial hemoglobin oxygen saturation<br>(SpO2) and pulse rate of adult and pediatric patient at<br>hospital (including clinical use in internist/surgery,<br>Anesthesia, and intensive care units)." | The Nonin® Onyx® II Model 9550 Finger Pulse Oximeter is a<br>small, lightweight portable device indicated for use in<br>measuring and displaying functional oxygen saturation of<br>arterial hemoglobin (SpO2) and pulse rate. It is intended for<br>spotchecking of adult and pediatric patients on fingers (other<br>than the thumb) between 0.3 - 1.0 inch (0.8-2.5 cm) thick. The<br>index finger is the recommended site. |
| Comparison Statement | The proposed devices have the same indications for use and classification. | |
| Components | The applicant device consists of detector and emitter<br>LED, signal amplify unit, CPU, data display unit and<br>power unit | photo detector, LED, signal amplify unit, CPU, display unit<br>and power unit |
able 3-2 Comparison Table
{4}------------------------------------------------
. . . .
3-5
.
{5}------------------------------------------------
| | | emarket Notification 510(k) Submission - Section 3 510(K) Summary | |
|------------------------|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| | SpO2 resolution | 1% | 1% |
| | PR Display Range | 0-254bpm | 18-321bpm |
| | PR Accuracy | ±2bpm (30-99bpm) and 2% (100-235bpm) | ±3% (20-250 BPM) |
| | PR resolution | 1bpm | 1 bpm |
| | Operating<br>temperature | 5℃~40℃ | 0°C~40 °C |
| | Relative humidity | ≤80% (operating)<br>≤93% (storage) | 10%~90% (operating)<br>10%~95% (storage) |
| | Atmosphere<br>pressure | 86kPa~106kPa | Up to 40,000 feet |
| | pulse rate tone<br>modulation function | Available: MD300CF3、MD300CF61<br>Not Available: MD300C1、MD300C2、MD300C61、MD300C63 | Not Available |
| | Comparison Statement | The applicant device has similar device specifications as the predicate device. | |
| | Battery<br>cover | ABS | ABS |
| Contacting<br>Material | Fingertip<br>Cushion | Medical Silicon gel | Medical Silicon gel |
| | Enclosure | ABS | ABS |
| | Comparison Statement | The contacting materials of applicant device are as same as that of the predicate device. | |
| | Performance<br>Testing | The bench tests include Test for Pulse signal strength<br>Indicator Bar、Pulse rate and SpO2 accuracy test for 3000<br>cycles Cleaning. Test for Random vibration, wide band and<br>Test according to ISO9919. All the bench test results are<br>provided in Performance Testing-Bench | |
| | Bench Test | Meet the requirements of FDA Guidance | |
| | Clinical Test | Conformed to ISO 9919 | |
| | | | Conformed to ISO 9919 |
ket Notification 510(k) Submission - Section 3 510(K) Summary
.
{6}------------------------------------------------
is and the same of the second of the states of the states of the country of the country of the country of the country of the country of the country of the country of the coun
.
.
3-7
{7}------------------------------------------------
| rket Notification 510(k) Submission - Section 3 510(K) Summary | | Label and Labeling |
|----------------------------------------------------------------|------------------------------------------------------------------------------|------------------------------|
| Animal skin irritation test | No evidence of significant<br>irritation from the test extract<br>to rabbits | Compliance with FDA guidance |
| In Vitro Cytotoxicity | No cytotoxic potential | |
| Skin Irritation Test | No evidence of causing<br>sensitization | |
| Animal skin irritation test | No evidence of significant<br>irritation from the test extract<br>to rabbits | Compliance with FDA guidance |
| ABS plastic | | |
. .
・
{8}------------------------------------------------
### 3.6 Intended Use
The Fingertip Pulse Oximeters, MD300C series, are portable, non-invasive devices intended for spot checking of arterial hemoglobin oxygen saturation (SpO2) and pulse rate of adult and pediatric patient at hospital (including clinical use in internist/surgery, Anesthesia, and intensive care units).
### 3.7 Test
### Non-clinical Test
The Fingertip Pulse Oximeters MD300C series are designed and tested and will be manufactured in accordance with both mandatory and voluntary standards, including:
- IEC 60601 1 Medical Electrical Equipment Partl : General requirements for safety. ●
- IEC 60601-1-2:2007 Medical Electrical Equipment Part 1: General requirements 0 for safety-2, Collateral Standard: Electromagnetic compatibility - Requirements and tests.
- � ISO 9919:2005, Medical electrical equipment - Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use.
The Software Validation is in compliance with FDA Guidance for the Content of PreMarket Submissions for Software Contained in Medical Devices.
The compatibility of the skin-contact component material in the finished product meets the requirement of Biocompatibility. The Biological Evaluation Tests are in compliance with the standards of ISO 10993, "Biological Evaluation of Medical Devices".
| No. | Test Name |
|-----|-----------------------------------------------------------------------|
| 1 | Broad-band Vibration Test |
| 2 | Pulse rate and SpO2 Accuracy Test for 3000 cycles' cleaning |
| 3 | System Performance Test |
| 4 | Shelf life Test |
| 5 | Test for Pulse Signal Strength Indicator Bar |
| 6 | Performance Test according to ISO 9919 |
| 7 | Electromagnetic Compatibility Test According to IEC 60601-1-2 |
| 8 | Electrical Safety Test According to IEC 60601-1 |
| 9 | Irritation、Sensitization and Cytotoxicity Test according to ISO 10993 |
The list of non-clinical test performed on the proposed device.
The test results indicate that the safety and effectiveness of the proposed devices is identical to that of the predicate device.
{9}------------------------------------------------
### Clinical Test
The Fingertip Pulse Oximeters MD300C1 / MD300CF3/ MD300CF3/ MD300CF61 / MD300C61 / MD300C63 share the same pulse oximeter sensor and oxygen saturation module. So we considered a clinical test of one of the proposed devices could cover that of other devices. The clinical test of other proposed devices can be exempted. And we conducted clinical test for one of the proposed devices, and the model is MD300C63.
The Clinical Test was conducted according to ISO 9919:2005, Annex EE.2 in the lab of the first affiliated hospital, Sun Yat-sen University. The study protocol was subjected to ISO 9919:2005 Annex EE. Procedures of testing required in EE2 were adopted.
### Subjects:
After Institutional Review Board (IRB) approval, 12 healthy adult volunteer subjects (ages 21-31yr, 45-78kg, with light to dark pigmentation) were included in the study conducted to evaluate the SpO2 accuracy performance.
### Methods:
Each system was evaluated during steady state / non-motion conditions with various levels of induced hypoxia resulting in stable oxygen saturation levels between 100% and 70% SaO2. Arterial blood samples were drawn during simultaneous data collection from the test devices. The blood was immediately analyzed on Reference CO-Oximetry providing functional SaO2 for the basis of the SpO2 accuracy comparison.
Adverse events and complications:
There are no adverse events during the clinical test.
Conclusion:
The results of the study provide supporting evidence that the MD300C63 Fingertip Pulse Oximeter is compliance to the accuracy specification claimed by the manufacturer. The Fingertip Pulse Oximeter can be used under steady state / non-motion conditions for the range 70-100%.
### 3.8 Determination of substantial equivalence
The proposed devices of Finger Pulse Oximeters MD300C series have the same classification information, same indications and intended use, same design principle, similar product design and specifications, same performance effectiveness, performance safety as the predicate device. The differences only exist in display screen、 SpO2 and PR display range、appearance and audible/visual indicator and pulse rate tone modulation functions. These differences are slight and do not influence the effectiveness and safety of the device. According to the non-clinical and clinical test results, the proposed devices are as safe, as effective and perform as well as the predicate device. So the proposed devices are Substantially Equivalent (SE) to the predicate device which is US legally market device.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its wings, and a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" around the eagle. The text is in all capital letters and is arranged in a circular fashion to match the border.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 3, 2013
Beijing Choice Electric Technology Company, Limited Mr. Lei Chen Quality Director North Building 3F. No. 9 Shuangyuan Road, Badachu Hi-tech Zone. Shijingshan District Beijing China 100041
Re: K130947
Trade/Device Name: Fingertip Pulse Oximeter MD300C1, MD300CF3. MD300CF61, MD300C61, and MD300C63 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: August 30, 2013 Received: September 3, 2013
Dear Mr. Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRFI docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not mislcading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{11}------------------------------------------------
### Page 2 - Mr. Chen
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer. Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.goy/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDevices/Safety/ReportaProhlem/gletault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.
Sincerely yours.
Tejashri R
Clinical D
DAGRID
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director
FOR
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital,
Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{12}------------------------------------------------
### Section 2 Indications for Use Statement
## Indications for Use
510(k) Number (if known): Fingertip Pulse Oximeter MD300C1、MD300C2、MD300CF3、 Device Name: ___ MD300CF61、MD300C61、MD300C63
Indications for Use:
The Fingertip Pulse Oximeters, MD300C series, are portable, non-invasive devices intended for spot checking of arterial hemoglobin oxygen saturation (SpO2) and pulse rate of adult and pediatric patient at hospital (including clinical use in internist/surgery, Anesthesia, and intensive care units).
Prescription Use __ イ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nayan J. Patel -S
2013.10.02 16:54:45
-04'00'
Page 1 of _l
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.