ASSURANCE ALAR SENSOR MODEL 10078; AB-N ADAPTER CABLE MODEL 10085
Applicant
Xhale
Product Code
DQA · Cardiovascular
Decision Date
Apr 10, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Assurance Alar Sensor is indicated for single patient use for continuously noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate from the nasal ala of adult and pediatric patients (weighing > 30kg), who are well or poorly perfused. The sensor can be used in a variety of healthcare environments where compatible pulse oximetry monitors are indicated for use, under professional supervision.
Device Story
Assurance Alar Sensor is a disposable, single-patient-use pulse oximetry sensor. It attaches to the nasal alar region using soft silicone rubber cushions for retention. The device utilizes red (660 nm) and infrared (890 nm) LED light sources and a silicon photodiode detector to measure light absorption changes in blood during each pulse. It connects via a DB-9 cable to compatible pulse oximetry monitors (e.g., Nellcor N-395). The monitor processes these signals to calculate and display SpO2 and pulse rate. Used in clinical settings under professional supervision, the device provides continuous monitoring to assist clinicians in assessing patient oxygenation status, potentially benefiting patients by enabling timely intervention for desaturation events.
Clinical Evidence
Controlled desaturation study performed per ISO/IEC 80601-2-61 on 12 healthy volunteers (varying age, gender, ethnicity, skin tone). Sensors connected to Nellcor N-395 monitors. Reference values obtained via indwelling arterial catheter and Co-oximeter. Analysis of 552 data pairs yielded an accuracy (ARMS) of 1.84% over the 70-100% SaO2 range.
Technological Characteristics
Transmissive pulse oximetry sensor; red (660 nm) and IR (890 nm) LEDs; silicon photodiode detector; silicone rubber skin contact material; DB-9 connector; disposable; single-patient use; compatible with Nellcor OxiSensor II technology monitors; tested per ISO 10993-1 (biocompatibility), IEC 60601-1-2 (EMC), and ISO 80601-2-61 (safety/performance).
Indications for Use
Indicated for continuous noninvasive monitoring of functional arterial oxygen saturation and pulse rate in adult and pediatric patients weighing >30kg. Suitable for well or poorly perfused patients in healthcare environments under professional supervision.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Ki2296
Page 1 of 6
Image /page/0/Picture/2 description: The image shows the logo for Xhale, a company that focuses on breathing new life into medicine. The logo features a stylized "X" symbol on the left, followed by the company name "Xhale" in bold, black letters. Below the name is the tagline "Breathing New Life into Medicine" in a smaller font.
# 510(k) SUMMARY
APR 1 0 2013
| Submitter | Xhale, Inc.<br>(via Assurance Biosense, Inc - a wholly owned subsidiary)<br>701 Hebron Avenue<br>Glastonbury, CT 06033<br>Phone: (860) 616-1610<br>Fax: (352) 375-3133 |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | David Rich<br>Sr. Vice President<br>Phone: (860) 616-1610 |
| Date Prepared | Feb 4, 2013 |
| Trade Name | Assurance™ Alar Sensor |
| Regulation (Common) Name | Pulse Oximeter Sensor |
| Classification | Class II, 21CFR 870.2700, 21CFR 870.2900 |
| Product Code | 74 DQA, 74 DSA |
| Predicate Device(s) | Nellcor N-395 Pulse Oximeter System - K991823<br>Nellcor OxiMax-Fast Adhesive Forehead Sensor - K021089<br>Masimo E1 Ear Sensor - K121912 |
### Device Description
The Assurance™ Alar Sensor is a disposable, single patient use Pulse Oximetry sensor designed to attach to the patient's nasal alar region - the fleshy region at the side of the nose. Skin contact and adhesive free sensor retention is via soft silicone rubber cushions encapsulating the optical components. The Assurance™ Alar Sensor with its associated patient cable, terminates in a DB-9 connector compatible with monitors employing Nellcor OxiSensor II SpO2 technology such as the Nellcor N-395.
The sensor utilizes red and IR LED light sources of 660 nm and 890 nm respectively along with a silicon photodiode detector to detect changes in oxygen saturation in the blood. Since oxygen saturated blood absorbs different amounts of light at each wavelength (red and infrared) as compared with unsaturated blood, the amount of light absorbed at each wavelength by the blood in each pulse can be used to calculate oxygen saturation.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Xhale. The logo consists of a stylized symbol to the left of the word "Xhale" in a bold, sans-serif font. Below the word "Xhale" is the tagline "Breathing New Life into Medicine" in a smaller font. The symbol to the left of the word "Xhale" is a geometric shape that resembles a stylized "X" or a four-way intersection.
# Indications for Use
The Assurance™ Alar Sensor is indicated for single patient use for continuous noninvasive monitoring of functional arterial oxygen saturation and pulse rate from the nasal alar of adult and pediatric patients, (weighing >30kg). The sensor can be used in a variety of healthcare environments where compatible pulse oximetry monitors are indicated for use, under professional supervision.
## Predicate Device(s) Comparison
Intended Use and Indications for Use
| Attribute | Assurance™ Alar Sensor | Predicate Device (K121912)<br>Masimo E1 Ear Sensor |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Continuous noninvasive monitoring of functional arterial oxygen saturation and pulse rate. | Same |
| Indications for Use | The Assurance™ Alar Sensor is indicated for single patient use for continuous noninvasive monitoring of functional arterial oxygen saturation and pulse rate from the nasal alar of adult and pediatric patients, (weighing >30kg). The sensor can be used in a variety of healthcare environments where compatible pulse oximetry monitors are indicated for use, under professional supervision. | The Masimo El Ear Sensor is indicated for single patient use for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for use with adult and pediatric patients, (weighing > 30kg), who are well or poorly perfused, in hospitals, hospital-type facilities, mobile, and home environments. |
| Patient Population | Adults, Peds >66 lbs. / 30 kg. | Adults, Peds >66 lbs. / 30 kg. |
| Duration of Use | Single patient, multi-use | Single patient, multi-use |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for Xhale, a company that is "Breathing New Life into Medicine". The logo consists of a stylized "X" shape on the left, followed by the word "Xhale" in bold, black letters. Below the company name is the tagline, "Breathing New Life into Medicine".
## Technology & Materials
| Attributes | Assurance™ Alar Sensor | Predicate Device (K121912)<br>Masimo E1 Ear Sensor |
|---------------------------|-------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Principle of<br>Operation | Spectrophotometric measurement of<br>functional arterial Oxygen saturation<br>by transmissive mode Pulse Oximetry | Spectrophotometric measurement of<br>functional arterial Oxygen saturation<br>by transmissive mode Pulse Oximetry |
| Reusable/Disposable | Disposable, Single Patient Use | Disposable, Single Patient Use |
| LEDs | Red (~660 nm) and IR (~880 nm) | Red (~660 nm) and IR (~880 nm) |
| LED drive | 2 wire, anti-parallel | 2 wire, anti-parallel |
| Detector | photodiode | photodiode |
| Connector | 9 pin DB-9 style | 9 pin DB-9 style |
| Skin Contact | Silicone rubber | Silicone rubber |
## Performance & Accuracy
| Device | 510(k) | Sensor<br>Location | SpO2 Accuracy<br>(ARMS) | Pulse Rate Accuracy |
|--------------------------------------------------------------------------------------|---------|-----------------------|--------------------------------------------------|---------------------|
| Assurance™ Alar Sensor | - | Nose | 70-100%: ± 2% | 30-250 bpm: ± 3 bpm |
| Nellcor N-395 System:<br>Oxisensor D-25<br>Oxisensor R-15<br>Dura Y & Ear clip D-YSE | K991823 | Finger<br>Nose<br>Ear | 70-100%: ±2%<br>80-100%: ± 3.5%<br>70-100%: ± 2% | 30-250 bpm: ± 3 bpm |
| Nellcor OxiMax Forehead | K021089 | Forehead | 70-100%: ±2% | 30-250 bpm: ± 3 bpm |
| Masimo E1 Ear Sensor | K121912 | Ear | 70-100%: ±2.5% | 30-250 bpm: ± 3 bpm |
The Assurance ™ Alar Sensor has been compared to the predicate device(s) and is viewed as substantially equivalent because:
- The intended use and indications for use are the same ●
- . The principles of operation, optical components and materials are the same
- . The performance and accuracy is the same or better
{3}------------------------------------------------
Page 4 of 6
Image /page/3/Picture/1 description: The image shows the logo for Xhale. The logo consists of a stylized "X" symbol to the left of the word "Xhale" in bold, black letters. Below the word "Xhale" is the tagline "Breathing New Life into Medicine" in a smaller font.
### Nonclinical Testing Summary
The following testing of the Assurance™ Alar Sensor was performed in accordance with the requirements of the design control regulations and established quality assurance procedures.
#### Biocompatibility of materials
Materials were evaluated per ISO 10993-1. The sensor is a surface device where materials have direct or incidental skin contact for less than 29 days.
- Surface Communicating, Skin, Prolonged duration .
- o. Cytotoxicity (tested per ISO 10993-5)
- Sensitization (tested per ISO 10093-10) o
- Irritation (tested per ISO 10993-10) o
- Pass / fail criteria was met for each respective ISO 10993 test .
Discussion: All materials were tested according to ISO 10093-1 for the intended use level of patient contact and duration and found to meet the applicable requirements.
#### Electromagnetic Compatibility
The device was tested per IEC 60601-1-2.
- . Emissions Testing
- o CISPR 11: Radiated Electromagnetic Emissions (2004)
- . Immunity Testing
- IEC 61000-4-2: Electrostatic Discharge Immunity (2001) O
- IEC 61000-4-3: Radiated Electromagnetic Field Immunity (2006) o
- o IEC 61000-4-6: Radio-Frequency Common Mode Immunity
Discussion: The device met the requirements of IEC 60601-1-2 relevant to the device. The tests conformed to the requirements set forth in ISO 80601-2-61, Particular Requirements for Basic Safety and Essential Performance of Pulse Oximeter Equipment.
#### Electrical Safety Testing
Fluid Ingress / Spill Resistance
- The device was tested per ISO 80601-2-61 requirement of IEC 60601-1 Clause 11.6. . overflow, spillage, leakage, ingress of water or particulate matter requirements.
Discussion: The device mects IPX1 requirements.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Xhale. The logo consists of a square-shaped symbol on the left and the word "Xhale" in bold, black letters on the right. Below the word "Xhale" is the tagline "Breathing New Life into Medicine" in a smaller font. The square symbol is made up of curved lines that form an abstract shape.
## Surface Temperature
- The device was tested per ISO 80601-2-61 Clause 201.11 and ANNEX BB: skin . temperature requirements for protection against excessive temperatures.
| Test | Max Temp (Right) (°C) | Max Temp (Left) (°C) |
|------|-----------------------|----------------------|
| 1 | 30.7 | 29.2 |
| 2 | 37.6 | 36.7 |
| 3 | 27.9 | 27.2 |
| 4 | 36.3 | 37.6 |
| 5 | 36.9 | 35.2 |
Discussion: The device passed the test with the skin temperatures under the device not exceeding 38°C, below the max temperature limit of 41°C.
### Pulse Rate accuracy (low signal)
- . The device was tested for accuracy of pulse rate over the range 30-250 bpm using an SpO2 simulator at minimum perfusion settings under conditions of no motion.
Discussion: The device measured the pulse rate to within ± 3bpm over the range.
#### Inter-device Reliability and Accuracy
- . A population of devices was tested under simulated conditions to assess the Interdevice Reliability and Accuracy. The devices were tested over the ranges; 85-99% SpO2 and 40-140 bpm.
Discussion: The deviation between simulated and measured values across the population was ≤2% SpOz and ≤1 bpm. This performance is within the stated accuracies of ±2% SpO2 and ± 3 bpm accordingly.
{5}------------------------------------------------
Page 6 of 6
Image /page/5/Picture/1 description: The image shows the logo for Xhale, a company that is "Breathing New Life into Medicine". The logo consists of a stylized "X" symbol to the left of the company name. The "X" symbol is made up of thick black lines that form a square shape with curved lines extending from each corner.
## Mechanical and Environmental Testing
#### Drop Test
- The device was tested per ISO 80601-2-61 requirement of IEC 60601-1 clause . 15.3.4.1 drop test requirements for hand-held ME equipment.
Discussion: The device passed the drop test without damage and satisfied the test requirements.
Storage Temperature and Humidity
- The device was tested following storage at the extremes of temperature and humidity: t -40°C to +70°C, 15% to 95% RH (non-condensing).
Discussion: The device was not affected by the ranges of temperature and humidity indicated as storage conditions.
Operating Temperature and Humidity
- . The device function was evaluated over the range of temperature and humidity: -5°C to +40°C, 15% to 95% RH (non-condensing).
Discussion: The device function was not affected over the ranges of temperature and humidity indicated as operating conditions.
#### Clinical Testing Summary
Hypoxia Performance Testing
- Controlled desaturation testing was performed per ISO/IEC 80601-2-61 under . no-motion conditions with 12 healthy volunteer subjects of varying age, gender, ethnicity and skin tone.
- Sensors were connected to Nellcor N-395 monitors representative of "Nellcor . OxiSensor II SpO2 technology". SpO2 values were recorded from the monitors over the range of 70-100% SaO2
- . Reference blood samples were drawn from an indwelling arterial catheter and analyzed on a Co-oximeter.
Discussion: Statistical analysis of 552 evenly distributed data pairs, yielded a device accuracy (Arms) of 1.84% over the range 70-100% SaO2.
### Substantial Equivalence Conclusion
The sponsor has demonstrated through design and feature comparison, performance testing, and controlled desaturation testing, the Assurance™ Alar Sensor is substantially equivalent to the listed predicate devices.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three wings, representing health, hope, and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is black and white.
#### April 10, 2013
Mr. David Rich Senior Vice President Xhale, Incorporated 701 Hebron Avenue GLASTONBURY CT 06033
Re: K122996
Trade/Device Name: Assurance Alar Sensor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DOA, DSA Dated: March 22, 2013 Received: March 27, 2013
Dear Mr. Rich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
{7}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcestoYour/Industry/default.htm.
Sincerely yours.
Kwame O. Ulmer for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
# 510(k) Number (if known): וֹז וֹב רו ברי רב
## Device Name: Assurance Alar Sensor
Indications for use: The Assurance Alar Sensor is indicated for single patient use for continuously noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate from the nasal ala of adult and pediatric patients (weighing > 30kg), who are well or poorly perfused. The sensor can be used in a variety of healthcare environments where compatible pulse oximetry monitors are indicated for use, under professional supervision.
Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Lester W. Schultheis 2013.04.08 1
(Division Sign-Off) (Division of Anesthesiology, General H Infection Control, Dental Devices
510(k) Number
Page 1 of 1
4.1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.