K083787 · Smiths Medical Pm, Inc. · DQA · Apr 15, 2009 · Cardiovascular
Device Facts
Record ID
K083787
Device Name
BCI WW1000 PULSE OXIMETER, MODEL WW1000
Applicant
Smiths Medical Pm, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Apr 15, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The BCI® model WW1000 handheld pulse oximeter is intended for spot-checking applications (non-continuous use). It monitors and displays a patient's functional oxygen saturation (%SpO2), pulse rate, pulse signal strength, and pulse amplitude index readings. It may be used by physicians, respiratory therapists, nurses, certified nurse assistants, emergency medical technicians, sleep technicians, clinicians, and home users. The intended patient population ranges from infant to adult. It can be used on patients with low perfusion. The WW1000 may be used in the hospital or clinical environment, during emergency land transport, and in the home.
Device Story
Handheld pulse oximeter for spot-check applications; monitors %SpO2, pulse rate, pulse amplitude index, and signal strength. Device utilizes Micro-power OEM daughter-board (31392B1) to process optical sensor inputs; board performs signal processing and calculates physiological parameters; transmits digital values to host device for display on red LED. Operated by clinicians or home users in hospitals, clinics, homes, or during emergency transport. No audible/visual alarms. Output provides immediate physiological status to assist clinical decision-making; enables non-invasive assessment of oxygenation and perfusion status in diverse patient populations.
Clinical Evidence
Three clinical desaturation studies (Desat 24, 27, 40) conducted on the 31392B1 oximeter daughter-board. Desat 24 and 27 evaluated baseline performance; Desat 40 verified performance after hardware/software enhancements for low perfusion. Comparison testing across three host systems (Evaluation Kit 3375, OxiLink-I, and WW1000) confirmed consistent readings within ±1 digit under identical simulator settings. Results demonstrate safety and effectiveness for intended use.
Technological Characteristics
Handheld pulse oximeter; utilizes Micro-power OEM daughter-board (31392B1) for signal processing. Features red LED display; powered by AA batteries or optional AC charger. Connectivity includes patient-isolated USB. Biocompatible materials used for patient-contacting sensors. Bench testing performed for EMC, electrical, mechanical durability, and environmental safety.
Indications for Use
Indicated for spot-check monitoring of functional oxygen saturation (%SpO2), pulse rate, pulse signal strength, and pulse amplitude index in patients ranging from infant to adult, including those with low perfusion. Suitable for use by clinicians and home users in hospital, clinical, home, and emergency land transport settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
K05-3787
# 510(k) Summary of Safety and Effectiveness
Manufacturer: Address:
Telephone Number:
Fax Number:
Contact Person:
Date Prepared:
Proprietary Name:
Common/Classification Name: Product Code:
Regulation Number: Predicate Devices:
Smiths Medical N7 W22025 Johnson Drive Waukesha, WI 53186
(262) 542-3100
(262) 542-3325
Donald Alexander Director Regulatory Affairs December 1, 2008
BCI® model WW1000 Pulse Oximeter
Pulse Oximeter DQA
870.2700 (Class II) Masimo SET Radical pulse oximeter (K040214) BCI® Autocorr model 3304 pulse oximeter (K070732) BCI® MiniCorr model 3402 pulse oximeter (K991410)
### Device Description:
The aim of this submission is to obtain market clearance for the BCI model WW1000 pulse oximeter.
The BCI® model WW1000 handheld pulse oximeter is intended for spot-check applications. The WW1000 monitors and displays patient functional oxygen saturation (%SpO2), pulse rate, pulse amplitude index, and pulse signal strength information. It does not have audible or visual patient alarms. The user interface includes a red LED display and an ON/OFF button.
The WW1000 comes with disposable AA batteries, the 3044S reusable pulse oximetry sensor and relevant manuals. Optional accessories include other oximetry sensors, docking station, thermal printer (attaches to dock), rechargeable battery pack, universal mains AC charger, patient isolated USB cable, universal mounting bracket and protective
#### Indications for Use
The BCI® model WW1000 handheld pulse oximeter is intended for spot-checking applications (non-continuous use). It monitors and displays a patient's functional oxygen saturation (%SpO2), pulse rate, pulse signal strength, and pulse amplitude index readings. It may be used by physicians, respiratory therapists, nurses, certified nurse assistants, emergency medical technicians, sleep technicians, clinicians, and home users. The intended patient population ranges from infant to adult. It can be used on patients with low perfusion. The WW1000 may be used in the hospital or clinical environment, during emergency land transport, and in the home.
1
{1}------------------------------------------------
## Risk Mitigation Table
Below is a summary of risks common to pulse oximeters and how this submission addresses those risks.
| Identified Risk | Mitigation Measures |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Inadequate Device Performance | Accuracy Testing (Section 11)<br>Benchtop Performance Testing (Section 9)<br>Software Documentation and Testing<br>(Section 8) |
| Electrical or Mechanical Failure | Electrical, Mechanical, and Environmental<br>Testing (Section 9) |
| Electromagnetic Interference | Electromagnetic Compatibility Testing<br>(Section 9) |
| Adverse Tissue Reactivity | Biocompatibility Information (Section 10) |
| Cross-contamination or Infection | Cleaning and Disinfection Testing (Section<br>9) |
| Improper Use | Labeling (Section 7) |
An in-depth risk management analysis, including mitigation measures, was performed on the BCI WW1000 system. The results are provided under Section 5.
#### Performance Testing
The BCI WW1000 pulse oximeter passed all performance bench top testing including EMC, electrical, mechanical durability, safety (operator and patient), and temperature/humidity. Test reports, including test protocols, pass/fail criteria, results and conclusions, are provided under Section 9.
#### Clinical Testing
The WW1000 pulse oximeter utilizes an existing oximeter daughter-board (Micro-power OEM board 31392B1) that provides the complete measurement of SpO2 and Pulse Rate (PR) and digitally transmits the calculated values to a host device. The 31392B2 oximeter daughter-board underwent three clinical desaturation studies; two trials (Desat 24 and Desat 27) used the Micro-power OEM Board Evaluation kit (3375) with additional optoisolation of the communication signals, and one (Desat 40) used the OxiLink-I host device which provides signal isolation.
Comparison testing among the Evaluation Kit (3375). OxiLink-I and WW1000 host devices was performed to demonstrate that all three systems produce the same readings under identical simulator settings. Results indicate that the Micro-power OEM board produced the same readings (within ±1 digit) in all three host devices.
2
{2}------------------------------------------------
The 31392B1 oximeter board underwent clinical desaturation study Desat 24 in July 2002 and Desat 27 in September 2003. In 2007/2008, hardware and software enhancements were made to the 31392B1 board to extend its operation over a wider range of perfusion index values and improve lower perfusion performance. Desat 40 was performed with the intent of verifying that these enhancements did not adversely affect SpO2 accuracy.
Results of the three individual clinical desaturation trials indicate that the WW1000 pulse oximeter performs as intended and is safe and effective for use. Please find complete desaturation test reports for Desat 24, 27 and 40 and applicable project file memos in Section 11.
## Conclusion
Supporting information per this premarket submission confirms that the BC1® WW1000 pulse oximeter is substantially equivalent to its predicate devices.
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# APR 1 5 2009
Mr. Donald Alexander Director, Regulatory Affairs Smiths Medical PM, Incorporated N7 W22025 Johnson Drive Waukesha, Wisconsin 53186-1856
Re: K083787
Trade/Device Name: BCI® Model WW1000 Pulse Oximeter Regulation Number: 21 CFR 880.2700 Regulatory Class: II Product Code: DQA, DPZ Dated: March 26, 2009 Received: March 30, 2009
Dear Mr. Alexander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2- Mr. Alexander
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susan Runne
S. Susan Runne, DDS, MA
-Susan Runner, D.D.S., MA Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: BCI® Model WW1000 Pulse Oximeter
Indications for Use: .. .
The BCI® model WW1000 handheld pulse oximeter is intended for spot-checking applications (non-continuous use). It monitors and displays a patient's functional oxygen saturation (%SpO2), pulse rate, pulse signal strength, and pulse amplitude index readings. It may be used by physicians, respiratory therapists, nurses, certified nurse assistants, emergency medical technicians, sleep technicians, clinicians, and home users. The intended patient population ranges from infant to adult. It can be used on patients with low perfusion. The WW1000 may be used in the hospital or clinical environment, during emergency land transport, and in the home.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Susan Rums
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: KCS
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.