K060576 · Nellcor Puritan Bennett, Inc. · DQA · Aug 14, 2006 · Cardiovascular
Device Facts
Record ID
K060576
Device Name
NELLCOR OXIMAX N-600X PULSE OXIMETER
Applicant
Nellcor Puritan Bennett, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Aug 14, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Nellcor OxiMax N-600x Pulse Oximeter is indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Nellcor OxiMax N-600x Pulse Oximeter is intended for use with neonatal, pediatric, and adult patients, and for patients who are well or poorly perfused, in hospitals, hospital-type facilities, intra-hospital transport, and home environments. For prescription use only
Device Story
Device monitors SpO2 and pulse rate via non-invasive optical sensors; uses red/infrared LEDs and photodiode to detect pulsatile blood flow changes; processes signals via OxiMax digital signal processing algorithm to output real-time SpO2, pulse rate, and plethysmographic waveform/blip bar. Used in hospitals, transport, and home environments by clinicians or patients. Features SatSeconds alarm management, audible/visual alarms, and serial data output for multiparameter monitors. OxiMax sensors store patient history/alarm data, enabling data transfer between care points. Benefits include continuous monitoring, motion/low-perfusion artifact filtering, and improved patient assessment during transitions.
Clinical Evidence
Clinical and non-clinical testing performed per 21 CFR Parts 812, 50, and 56. No specific performance metrics (e.g., sensitivity/specificity) provided in summary; testing confirmed performance equivalent to predicate devices.
Technological Characteristics
Non-invasive optical sensing using red/infrared LEDs and photodiode. Powered by AC (115V/230V) or internal 6V rechargeable lead-acid battery. Connectivity via serial port (EIA-232/RS-422). Includes OxiMax sensor memory chip for data storage. Software includes digital signal processing, interference filtering, and alarm management.
Indications for Use
Indicated for continuous non-invasive monitoring of SpO2 and pulse rate in neonatal, pediatric, and adult patients, including those with good or poor perfusion, in clinical and home settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Nellcor OxiMax Pulse Oximetry System with N-595 Pulse Oximeter (K012891)
{0}------------------------------------------------
Kobos76
AUG 1 4 2006
# 510(k) Summary
| Submitted by: | Nellcor Puritan Bennett Incorporated<br>4280 Hacienda Drive<br>Pleasanton, CA 94588 |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact: | Patrick Garvey<br>Regulatory Affairs Manager<br>Phone: (925) 463-4479<br>Fax: (925) 463-4020 |
| Date Summary Prepared: | October 10, 2005 |
| Trade Name: | OxiMax Pulse Oximetry System with N-600x Pulse<br>Oximeter and OxiMax Sensors and Cables (aka<br>"Accessories") |
| Common/Usual Name: | Pulse Oximeter, Sensors and Cables |
| Classification Name: | Oximeter (74DQA) (per 21 CFR §870.2700)<br>Patient Transducer and Electrode Cable (including<br>connector) (74DSA) (per 21 CFR §870.2900) |
| Substantially Equivalent<br>Devices: | 1. Nellcor OxiMax Pulse Oximetry System with<br>N-595 Pulse Oximeter, 510(k) #K012891<br>2. Nellcor N-200 Pulse Oximeter, 510(k)<br>#K863784 & K022819 |
## I. Device Description
The OxiMax N-600x Pulse Oximeter is a modification OxiMax Pulse Oximetry System with N-595 Pulse Oximeter and OxiMax Sensors and Cables. The N-600x is designed for continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate using OxiMax pulse oximetry sensors and the DOC-10 cable. The N-600x Pulse Oximeter displays digital values of SpO2 and Pulse Rate. Pulse Amplitude is displayed by means of a "blip bar" presentation or plethysmographic waveform. The N-600x can be powered by an internal power supply operating on AC from a standard electrical utility receptacle (manually switchable from 115V to 230V) or alternatively by an internal 6V rechargeable sealed lead-acid battery. The OxiMax N-600x Pulse Oximeter is intended for prescription use with adult, pediatric and neonatal patients in hospitals, hospital-type facilities, intra-hospital transport, and home environments.
510(k) Summary - Page 1
{1}------------------------------------------------
Audible and visual alarms for high/low saturation, pulse rate and pulse search are provided. The N-600x Pulse Oximeter also includes adjustable alarm silence duration and other configurable power-on settings. The N-600x provides a back-up piezo alarm that sounds if a high-level audible alarm has not been responded to within 45 seconds. The piezo alarm will also activate in the event of a primary alarm failure. The N-600x provides an audible low battery warning to alert the user of impending loss of power and consequent loss of monitoring capability. The N-600x Pulse Oximeter has visual indicators for pulse search, interference, power mode (i.e., battery or AC), alarm silence and alarm features. There is also a serial port (EIA-232 and RS-422 interface) that provides ASCII output of real-time data. Via the serial port, the N-600x interfaces with selected multiparameter monitors.
The Nellcor pulse oximetry cable, Model DOC-10, links individual OxiMax sensors to the N-600x Pulse Oximeter. The N-600x and DOC-10 are intended for use only with the OxiMax family of sensors. These OxiMax sensors each contain a memory chip carrying information about the sensor which the oximeter needs for correct operation, including in-sensor data, Advanced Signal Evaluation, lot code and data set revision, and sensor model. These OxiMax sensors include the following models: OxiMax MAX-A, MAX-AL, MAX-P, MAX-I, MAX-N, MAX-R, MAX-FAST, OxiCliq-A, OxiCliq-P, OxiCliq-I, OxiCliq-N, DS-100A, D-YS, D-YSE, D-YSPD, SC-A, SC-NEO, SC-PR, OXI-A/N, OXI-P/I, and NeoMax.
### II. Intended Use
The Nellcor OxiMax Pulse Oximeter N-600x is indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate.
### III. Technological Characteristics
The OxiMax Pulse Oximetry System measures functional oxygen saturation noninvasively via a light signal interacting with tissue, by utilizing the time-varying changes in tissue optical properties that occur with pulsatile blood flow. Red and infrared lightemitting diodes (LEDs) are utilized as light sources. A photodiode acting as a photodetector senses the signal strengths of the two wavelengths of light, which vary with the amount of light transmitted through the tissue. The N-600x Pulse Oximeter receives this electrical information from the sensor and processes the information by use of an algorithm to provide real time values of SpO2, pulse rate and pulse amplitude.
OxiMax technology encompasses:
- OxiMax Advanced Digital Signal Processing for reading through motion artifact and . low perfusion
- . Advanced Signal Evaluation providing the user information on sensor placement
- . SatSeconds Alarm Management
- . OxiMax Sensor Technology, including in-sensor data, allows the N-600x to write data to and read data from OxiMax adhesive sensors
The N-600x uses SpO2 and Pulse Rate software algorithms, interference-filtering software, and SatSeconds alarm management similar to the software in the legally
{2}------------------------------------------------
marketed predicate device cleared under K012891, with minor changes that do not raise new questions of safety or efficacy.
The Nellcor brand sensors containing OxiMax technology are compatible with the N-600x, other Nellcor oximeters and instruments containing Nellcor OxiMax oximetry, or with instruments licensed to use Nellcor sensors (Nellcor-compatible instruments). OxiMax Sensor Technology, including in-sensor data, allows the N-600x to write data to and read data from OxiMax adhesive sensors allowing patient history, namely SpO2 and pulse rate alarm events, to travel with the patient, and enabling quick patient assessment upon transfer to a new point of care. The N-600x provides on-screen viewing of insensor patient data including SpO2 and pulse rate trend of events.
## IV. Tests Performed to Support Determination of Substantial Equivalence
Clinical and non-clinical tests were performed to support the determination of substantial equivalence. Clinical studies were conducted following regulations under Title 21 of the Code of Federal Regulations (21 CFR), Part 812 - Investigational Device Exemptions, Part 50 - Protection of Human Subjects and Part 56 - Institutional Review Boards.
### V. Conclusions
The technological characteristics of the OxiMax N-600x Pulse Oximetry System and the results of non-clinical and clinical tests do not raise new questions of safety or effectiveness when compared to the legally marketed predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name encircling a stylized caduceus symbol. The caduceus is represented by three abstract human profiles facing right, with three parallel lines above them. The text reads "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Patrick Garvey Regulatory Affairs Manager Nellcor Puritan Bennett, Incorporated 4280 Hacienda Drive Pleasanton, California 94588-2719
Re: K060576
Trade/Device Name: The Nellcor® OxiMax® N-600x Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: August 7, 2006 Received: August 8, 2006
Dear Mr. Garvey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 - Mr. Garvey
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
## Enclosure
{5}------------------------------------------------
510(k) Number (if known):
The Nellcor® OxiMax® N-600x Pulse Oximeter Device Name:
# Indications For Use:
The Nellcor OxiMax N-600x Pulse Oximeter is indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Nellcor OxiMax N-600x Pulse Oximeter is intended for use with neonatal, pediatric, and adult patients, and for patients who are well or poorly perfused, in bospitals, hospital-type facilities, intra-hospital transport, and home environments. For prescription use only
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mitchell Nichols for Ann Graham
്:r)
schesiology, General Hospital,
(Optional Format 3-10-98)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.