The Dolphin 2000™ Oximetry Sensors are indicated for use in continuous monitoring of arterial oxygen saturation and pulse rate.
Device Story
Dolphin 2000 Oximetry Sensors are disposable or re-usable pulse oximetry probes. Disposable versions use adhesive bandages to secure emitter/detector diodes to measurement sites (finger, toe, or neonatal foot). Re-usable versions use clothespin-style clips or Y-shaped configurations with bandages to maintain contact with the patient's finger, ear, or hand. Sensors connect to compatible pulse oximeter monitors. Operation relies on optical assessment of tissue oxygenation via LEDs and photodiodes; light passes through tissue to detectors. Monitors process these signals to calculate arterial oxygen saturation (SpO2) and pulse rate. Used in clinical settings by healthcare providers to monitor patient status. Output provides real-time physiological data to assist in clinical decision-making and patient management.
Clinical Evidence
Clinical validation performed via breathe-down protocols at the VA Hospital of Wisconsin-Milwaukee. Study involved human volunteers with SpO2 values ranging from 70-100% at rest. Accuracy was demonstrated by comparing Dolphin 2000 SpO2 values to functional SaO2 values. Data analyzed to determine ARMS (Accuracy Root Mean Square) for each probe.
Technological Characteristics
Optical sensing via LEDs and photodiodes. Disposable sensors use adhesive tape laminates; re-usable sensors use spring-loaded clips or Y-shaped housings. Compatible with major pulse oximeter monitors. Tested to EN 60601-1 (1990), EN 60601-1-1 (1993), EN 60601-1-2 (1993), and ASTM F1415-92. Non-sterile. Biocompatible materials for prolonged skin contact.
Indications for Use
Indicated for continuous monitoring of arterial oxygen saturation and pulse rate in adult, pediatric, infant, and neonatal patient populations.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
KG33876
# MAR 1 1 2004
## 14. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
## Dolphin 2000™ Pulse Oximetry Y Sensor 12/11/03
#### Submitter ( Consultant name and Address)
Bill Cuman 9433 S. Morning Glory Lane Highlands Ranch, CO 80130
Phone: 720-939-6482 Fax: 786-551-8221
#### Sponsor Company Name and Address and Contact Person
Dolphin Medical Inc. 12525 Chadron Avenue Hawthorne, CA 90250
Tammy Conway, QA Manager (310) 978-0516 phone: (310) 978-1816 fax:
### Manufacturing Facility Name and Address
Opto Sensors (M) Sdn. Bhd. No. 6 Jalan Angkasa Mas 1 Tabrau Industrial Estate II 81100 Johor Bahru, Malaysia
#### Common, Classification & Proprietary Names
Common Name: oximetry sensor Classification Name: oximeter Dolphin™ 2000 Oximetry Sensors Proprietary Name:
### Predicate Devices
| Sensor Model # | Predicate<br>Model # | Predicate<br>510(k) # |
|-----------------------------|----------------------|-----------------------|
| 2020, 2040, 2060, &<br>2070 | Model 2010 | K030952 |
| 2321 & 2341 | 2351 | K012989 |
| 2342 | 2352 | K012989 |
| 2323, & 2343 | 2353 | K012989 |
| 2344 | 2354 | K012989 |
| 2220 & 2260 | 2210 | K032947 |
| 2422, 2424, & 2426 | 2421 | K030952 |
| Extension Cables | | |
{1}------------------------------------------------
#### Device Description
Device Describition
The Dolphin 2000™ Oximetry Sensors are fully compatible disposable and re-usable The Dolphin 2000 ~ Oximelly Ochson's are raily of pulse oximeter monitors. They replacement sensors for use with major brands of pares while in the first and BCI Compatabile Sensors.
The disposable Dolphin 2000 Oximetry Sensors are constructed in a similar manner to The disposable Dolphin 2000 Oximelly Ochson are embedded in a laminate of tapes
predicate devices. The emitter and detector diodes are embedded in a lands a predicate devices. The childer and development an adhesive bandage that is connected to the cable assombly. "The consortient by wrapping it around a finger backing that allows the sensor to be upplied to the End (Colphin 2000 Oximetry Sensors are of toe (measurement site). Tour sizes of aloposes, infant and neonatal application available, which are indicated for doc for duall, positive, mon-sterile for single patient use.
The re-usable Dolphin 2000 Finger Clip Oximetry Sensor is an adult-sized clothespin-The re-usable Dolphil 2005 Finger Olip Oximony Consor consists of the style clip that is placed on the one on anyon one on anyon maintained in mild
emitter and detector components mounted in opposing clip halves, mainter and consected emitter and delection components mounted in opposing only his house the optiolection in complession by a spring mings. The montain contact with the patient's finger. Clear components within contoured paids that the read to pass through the finger for the measurements.
The Re-usable Y sensor is for use on the ear, finger, hand, or neonatal foot and held in The Re-usable 1 Sensor is for use on the sensor can also be used on the adult ear with the place with a disposable bandage. The concer or enounted in a sealed pouch (same ear clip accessory." The emitter and other cructed in a Y shape. All Dolphin sensors are provided non-sterile.
#### Intended Use
Intended Use
The Dolphin 2000™ Oximetry Sensors are indicated for use in continuous monitoring of arterial oxygen saturation and pulse rate.
## Technological Characteristics Comparison
The Dolphin 2000™ Oximetry Sensors are substantially equivalent in intended use, The Dolphill 2000 - Oximony Conbore e. e commercially available oximetry sensors.
All of the Dolphin 2000™ oximetry sensors and the predicate devices operate on the All of the Dolphin 2000 - (Allinor) optical assessment of tissue oxygenation using emitters (LEDs) and detectors (photodiode).
The Dolphin 2000™ oximetry sensors are designed, configured, and manufactured for The Dulphill 2000 - Uximistry beneercially-available oximetry monitors. full companibility for use with the laboled, connections of equivalent specifications as used in the predicate devices.
The Dolphin 2000™ oximetry sensors, like the predicate devices are available in both The Dolphin 2000 - oxinstly Schools, ill.o the programs, in adult, pediatric, infant and neonatal populations.
The labeled accuracy of the Dolphin 2000 sensors is equivalent to those of the predicate devices.
{2}------------------------------------------------
### Performance Testing
#### Biocompatibility t
Biocompatibility tests, appropriate for skin-contacting devices for prolonged in Blocompaniblity tests, approphiats for onf the device components used in the assembly of the Dolphin 2000™ pulse oximetry sensors. Test results assembly of the Dolphin 2000 - paiso existed , initiant, and non-sensitizing.
#### Electrical Safety 피
Electrical Salety
The Dolphin 2000 Oximetry Sensors have been tested and found to comply with the applicable clauses of the following standards:
- Medical electrical equipment part 1: General EN 60601-1 (1990) requirements for safety
- EN 60601-1-1 (1993) Medical electrical equipment part 1: General 1 EN 60001-1-1 (1507) Modical Standard: Safety requirements for medical electrical systems
- EN 60601-1-2 (1993) Medical electrical equipment part 1: General – requirements for safety - 2. Collateral standard: Electromagnetic compatibility requirements and tests
- Standard Specification for Pulse Oximeters ASTM F1415-92 .
#### Clinical Testing
The sensors were validated in breathe-down protocols at the VA Hospital of Wisconsin - Milwaukee, (I)r. Phillip Clifford, MD.). Scientific accuracy was Wisconsin - Millindakoo, Companing Dolphin 2000 SpO2 values to functional SaO2 demorristiated by statistically oompanies as a the down protocol at rest (i.e. no motion) while values. Volunteers partionated in thing from 70-100%. Data was analyzed to determine the ARMS for each probe.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a stylized image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 1 2004
Dolphin Medical, Incorporated C/O Mr. Bill Curnan Regulatory Specialist Bill Curnan 9433 S. Morning Glory Lane Littleton, Colorado 80130
Rc: K033876
Trade/Device Name: Dolphin 2000 Oximetry Senors: Models 2020, 2040, 2060, 2070, 2220, 2321, 2323, 2341, 2342, 2343, and 2344 Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: February 10, 2004 Received: February 12, 2004
Dear Mr. Curnan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Curnan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Attachment l
## Indications for Use
510(k) Number (if known): K033876
Device Name: Dolphin 2000 Oximetry Sensors
Indications For Use: The Dolphin 2000 Oximetry Sensors are indicated for use in continuous monitoring of arterial oxygen saturation and pulse rate.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
fthh
(Division Sign-Off) ( Cholon Oigh-Oit)
Division of Anesthesiology, General Hospital, e Malon of Anesthesiology, Gene Infection Control, Dental Devices
Page 1 of
510(k) Number: K033876
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.