OXIMAX SOFTCARE SENSORS, MODELS SC-A, SC-PR, SC-NEO, SC-N

K030930 · Nellcor Puritan Bennett, Inc. · DQA · May 23, 2003 · Cardiovascular

Device Facts

Record IDK030930
Device NameOXIMAX SOFTCARE SENSORS, MODELS SC-A, SC-PR, SC-NEO, SC-N
ApplicantNellcor Puritan Bennett, Inc.
Product CodeDQA · Cardiovascular
Decision DateMay 23, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2
AttributesPediatric

Indications for Use

The OxiMax SoftCare Sensors are indicated for single-patient use when continuous noninvasive arterial oxygen saturation and pulse rate monitoring are required.

Device Story

OxiMax SoftCare sensors are sterile, non-adhesive, single-patient use pulse oximetry sensors. They function as part of the OxiMax pulse oximetry system, connecting to OxiMax-compatible monitors. Sensors contain an integrated memory chip storing sensor-specific data, including Advanced Signal Evaluation parameters, lot codes, and model information, which the monitor uses for operation. Designed with soft materials to eliminate the need for adhesive attachment. Used in clinical settings for continuous monitoring of arterial oxygen saturation and pulse rate. Healthcare providers use the monitor's output to assess patient oxygenation status and guide clinical interventions. Benefits include noninvasive, continuous monitoring for various patient populations ranging from preterm infants to adults.

Clinical Evidence

Human oxygenation evaluations were conducted to confirm conformance to accuracy and precision specifications. Bench testing was also performed. No specific statistical metrics (e.g., AUC, p-values) were provided in the summary.

Technological Characteristics

Sterile, latex-free, non-adhesive pulse oximetry sensors. Features an integrated memory chip for sensor data storage. Compatible with OxiMax monitors. Four models (SC-A, SC-PR, SC-NEO, SC-N) sized for specific patient weight ranges. Employs electro-optical sensing for pulse oximetry.

Indications for Use

Indicated for continuous noninvasive arterial oxygen saturation and pulse rate monitoring. Models: SC-A for adults >40 kg; SC-PR for preterm infants <1.5 kg; SC-NEO for neonates 1.5-5 kg; SC-N for neonates <3 kg. For single-patient use.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and head. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Public Health Service MAY 2 3 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Gina To Senior Regulatory Affairs Project Manager Nellcor Puritan Bennett Incorporated 4280 Hacienda Drive Pleasanton, California 94588 Re: K030930 Trade/Device Name: OxiMax SoftCare Sensor Models SC-A, SC-PR, SC-NEO, SC-N Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: March 24, 2003 Received: March 25, 2003 Dear Ms. To: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ ### Page 2 - Ms. To Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a " promatics nothleated." The 127 I midnig of backanial of your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Susan Ruasr Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Page of 510(k) Number (if known): KO30 930 Nellcor® OxiMax™ SoftCare™ Sensors Device Name: Indications For Use: The Nellcor OxiMax SoftCare adult oxygen sensor, model SC-A, is indicated for single patient use when continuous noninvasive arterial oxygen saturation and pulse rate monitoring are required for adults weighing more than 40 kg. The Nellcor OxiMax SoftCare preterm infant oxygen sensor, model SC-PR, is indicated for single-patient use when continuous noninvasive arterial oxygen saturation and pulse rate monitoring are required for neonates weighing less than 1.5 kg. The Nellcor OxiMax SoftCare neonatal oxygen sensor, model SC-NEO, is indicated for single-patient use when continuous noninvasive arterial oxygen saturation and pulse rate monitoring are required for neonates weighing 1.5 kg to 5 kg. The Nellcor OxiMax SoftCare neonatal oxygen sensor, model SC-N, is indicated for single-patient use when continuous noninvasive arterial oxygen saturation and pulse rate monitoring are required for neonates weighing less than 3 kg. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) G.H. Watson eneral Hospital. 510(k) Number: K030930 Prescription Use V OR Over-The-Counter Use Optional Format (1-2-96) {3}------------------------------------------------ K030930 4280 Hacienda Drive Pleasanton, CA 94588 Tele: 925 463-4000 Fax: 925 463-4020 52 12 tuco Healthcare Nellcor ## 510(k) Summary Submitted by: Nellcor Puritan Bennett, Inc. 4280 Hacienda Drive Pleasanton, CA 94588 Company Contact: Gina To Senior Regulatory Affairs Project Manager (925) 463-4427 (925) 463-4020 - FAX March 24, 2003 Date Summary Prepared: Trade Name: Common/Usual Name: Oxygen Sensor Oximeter (DQA) per 21 CFR §870.2700 Classification Name: Nellcor Puritan Bennett, Inc., OxiMax Pulse Oximetry Legally Marketed System with N-595 Pulse Oximeter and OxiMax Sensors, (Unmodified) Device: K012891 OxiMax™ SoftCare™ Sensors ### DEVICE DESCRIPTION The OxiMax SoftCare sensors are sterile, latex-free, single patient use, non-adhesive The Oxiviax SoftCare Schools are Sterne, Intell 1900, estem as a means of attachment. Sensors. These sensors have a noon and soft and soft as a more is eliminated. These sensors contain a memory chip carrying information about the sensor which the oximeter needs for correct operation, including in-sensor data, Advanced Signal Evaluation, lot code and data set revision, and sensor model. The OxiMax SoftCare sensors are compatible with OxiMAX monitors. #### INTENDED USE The OxiMax SoftCare Sensors are indicated for single-patient use when continuous noninvasive arterial oxygen saturation and pulse rate monitoring are required. {4}------------------------------------------------ # SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE The OxiMax SoftCare Sensors have the same technological characteristics as the OxiMax MAX-N. The differences relate to patient contact surface, attachment means, electro-optical specifications, and labeling. The dimensions of the MAX-N have been modified to result in four distinct SoftCare sensors, to provide different sized sensors and graphics for various patient weight ranges within the adult and neonatal populations. ## TESTS PERFORMED TO SUPPORT DETERMINATION OF SUBSTANTIAL EQUIVALENCE Human and bench tests were performed to support the determination of substantial equivalence. Human oxygenation evaluations were conducted to confirm conformance to accuracy and precision specifications. #### CONCLUSIONS The technological characteristics of the OxiMax SoftCare Sensors and the results of testing do not raise new questions of safety or effectiveness when compared to the legally marketed predicate device. 25 13
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...