The ReDe Mask is to be used as a breathing frequency monitor by recording nasal and/or oral airflow while supplying oxygen. The ReDe Mask itself performs no diagnostic functions and only supports the recording of airflow for use as an adjunct with pulse oximetry. It is intended for adult prescription and single use only in healthcare facility procedural areas and recovery rooms.
Device Story
ReDe Mask is a single-use, prescription-only breathing frequency monitor for sedated adult patients. It consists of a thermistor-based temperature-sensing circuit integrated into a standard face mask. The device continuously measures temperature changes in the nasal/oral airflow path; exhalation causes warming, inhalation causes cooling. Firmware analyzes the slope of temperature changes to detect breath events (inflection points). Output is provided via colored LEDs: green for intervals <7.5s, yellow for 7.5-20s, and flashing red for ≥20s. Used by healthcare professionals in procedural/recovery areas as an adjunct to pulse oximetry. It provides real-time visual feedback to assist clinicians in identifying respiratory depression or airway obstruction, potentially improving patient safety during sedation.
Clinical Evidence
Clinical evidence includes human factors/usability validation (n=30 anesthesia providers; n=spectrum of providers) confirming ease of use and correct interpretation of LED indicators. Performance accuracy was validated in two studies (n=38 and n=50) comparing the ReDe Mask against capnography (gold standard) and bioimpedance (ExSpiron). Results showed equivalence in detecting exhalation and respiratory rates, with no false positives or negatives reported in the study cohorts.
Technological Characteristics
Thermistor-based temperature sensing; 3V CR2032 non-replaceable battery; microcontroller-based firmware analysis; integrated into a standard face mask with elastic strap; dimensions 43.43mm H x 30.48mm W x 14.99mm D; weight 40g; operating temperature 16°C to 31°C; biocompatibility per ISO 10993 (cytotoxicity, sensitization, intracutaneous, systemic toxicity); electromagnetic compatibility per IEC 60601-1-2; electrical safety per IEC 60601-1.
Indications for Use
Indicated for adult patients sedated for diagnostic or therapeutic procedures in healthcare facility procedural areas and recovery rooms. Used as an adjunct to pulse oximetry to monitor breathing frequency when supplemental oxygen is provided via the mask.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized image of an eagle or bird in flight, composed of three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 17, 2017
TereoPneuma, Inc. Warren Young President and COO 13223 Black Mountain Rd., Suite 1-224 San Diego, California 92129
Re: K161953
Trade/Device Name: ReDe Mask Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: PRK Dated: January 3. 2017 Received: January 9, 2017
Dear Mr. Young:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Tina Kiang
-s
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K161953
Device Name ReDe Mask
Indications for Use (Describe)
The ReDe Mask is indicated for use by healthcare facility procedural areas and recovery rooms as an adjunct to monitor breathing in adult patients who are sedated for a diagnostic or therapeutic procedure. The ReDe Mask measures the time period between the current and previous exhalation and illuminates a colored light during the exhalation that reflects the interval of time between breaths. If the interval is less than 7.5 seconds, the green light illuminates during exhalation; if the interval is greater than 7.5 seconds, the yellow light illuminates during exhalation; and if the interval between breaths is 20 seconds or longer, the red light flashes continuously. The ReDe Mask is only to be used when supplemental oxygen is provided by the facemask. The ReDe Mask is not a standalone device and is only to be used as an adjunct to pulse oximetry.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for TereoPneuma. The logo consists of a stylized graphic on the left and the company name on the right. The graphic is a circular shape with a green center, surrounded by yellow and red curved shapes. The company name is written in green, with the "TM" symbol in a smaller font size.
## 510(k) Summary
#### Date summary prepared Jan 3, 2017
### 510(k) Submitter/Holder
TereoPneuma 13223 Black Mountain Rd. Ste 1-224 San Diego, CA 92129-4400
#### Contact
Warren G. Young, Ph.D. President and COO Telephone: 858-842-5036 Email: wyoung@tereopneuma.com
#### Name of Device
| Trade Name | ReDe Mask |
|---------------------|--------------------------------------------------------------------|
| Common Name | Breathing Monitor |
| Classification Name | monitor, breathing frequency<br>(21 CFR § 868.2375, class II, PRK) |
#### Predicate Device
| Trade Name: | Oral/Nasal Thermal Airflow Sensor |
|--------------------|-----------------------------------|
| Device Common Name | Airflow Sensor |
| 510(k) Number: | K080922 (cleared 7/2008) |
| Manufacturer: | Salter Labs |
#### Reference Devices
Two reference devices are used in the clinical evaluations:
| Trade Name: | MicroCap/NPB-75 |
|--------------------|---------------------------------------------------|
| Device Common Name | Combination Oximeter/ Carbon Dioxide Gas Analyzer |
| 510(k) Number: | K964239 |
| Manufacturer: | Spegas Industries Ltd. |
| Trade Name: | ExSpi |
|--------------------|-------|
| Device Common Name | Respi |
| 510(k) Number: | K1200 |
| Manufacturer: | Respi |
iron ratory Monitor 087 ratory Motion, Inc.
#### Indications for Use
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for TereoPneuma. The logo consists of a stylized image of a flame in yellow and red with a green circle in the center on the left. To the right of the flame image is the company name TereoPneuma in green font with a trademark symbol.
The ReDe Mask is indicated for use by healthcare professionals in healthcare facility procedural areas and recovery rooms as an adjunct to monitor breathing in adult patients who are sedated for a diagnostic or therapeutic procedure. The ReDe Mask measures the time period between the current and previous exhalation and illuminates a colored light during the exhalation that reflects the interval of time between breaths. If the interval is less than 7.5 seconds, the green light illuminates during exhalation; if the interval is greater than 7.5 seconds but up to 20 seconds, the yellow light illuminates during exhalation; and if the interval between breaths is 20 seconds or longer, the red light flashes continuously. The ReDe Mask is only to be used when supplemental oxygen is provided by the facemask. The ReDe Mask is not a standalone device and is only to be used as an adjunct to pulse oximetry.
## Intended Use
The ReDe Mask is to be used as a breathing frequency monitor by recording nasal and/or oral airflow while supplying oxygen. The ReDe Mask itself performs no diagnostic functions and only supports the recording of airflow for use as an adjunct with pulse oximetry. It is intended for adult prescription and single use only in healthcare facility procedural areas and recovery rooms.
Because sedation produces respiratory depression in every patient, healthcare providers must monitor ventilation by counting the respiratory rate during and after procedures in which sedatives are administered. Counting the respiratory rate may be inaccurate because it is difficult to gauge the rise and fall of a chest in a patient who is sedated, both because the size of the breaths are diminished by the sedatives and because the sedatives may cause airway obstruction. When the airway is obstructed, the chest continues to rise and fall as the diaphragm moves up and down, but no breath is going in and out of the patient due to the airway obstruction.
#### Device Description
The ReDe Mask is a breathing frequency monitor that provides a visual signal at each breath event (i.e., an inhalation followed by an exhalation). The proposed device consists of a temperature-sensing circuit attached externally to a standard face mask. This circuit continuously analyzes the temperature inside the face mask and determines when a breath event has occurred.
The ReDe Mask is designed to detect breathing events (cycles of inhalation followed by exhalation) by measuring temperature changes in the immediate vicinity of a patient's nose and/or mouth. Exhalations produce a temperature warming as expired air exits the mouth and nose, while inhalations result in a temperature cooling as ambient air and supplied oxygen enter the mask. The overall pattern is thus one of repeating periods of a warming and a cooling phase with every breath. The rate of warming and cooling, that is, the slope of the temperature change (degrees C per unit time) depends on the vigor with which the patient is breathing, which can range from very shallow breathing (small slope values) to vigorous breaths (large slope values). The ReDe Mask's breath detection algorithms are based on continuously measuring the warming and cooling slopes coupled with real-time analysis to determine the changeover point from negative slope (inhalation phase) to
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for "TereoPneuma". The logo consists of a stylized graphic to the left of the company name. The graphic has three curved lines in the colors yellow, green, and red. The company name is in green font, and the letters are connected. There is a trademark symbol to the right of the company name.
positive slope (exhalation phase). It is the detection of inflection points in the slope (negative to positive) that vield the elapsed time period between successive breaths. With each new breath the elapsed time between inflection points is used to determine which LED to illuminate.
## Predicate Device Comparison
Similarities in the intended use and indications for use between the ReDe Mask and predicate are:
- Both are to be used to record nasal and/or oral airflow while supplying oxygen.
- . Both do not perform diagnostic functions and only support the recording of airflow for use as an adjunct with other clinical devices.
- . Both are to be used as a breathing frequency monitor, with the proposed device conveying this information with colored indicator lights and the proposed predicate with a waveform display on a computer monitor.
- Both are non-invasive medical devices. .
Differences in the intended use and indications for use between the ReDe Mask and predicate are:
- . The ReDe Mask is intended to be used to measure the time period between the current and previous exhalation. The predicate averages the breathing rate over a period of time.
- . The ReDe Mask uses colored light indicators to reflect the interval of time between breaths. The predicate uses waveforms on a computer monitor and digital readouts to reflect the averaged breathing rate.
- . The ReDe Mask is indicated for use by healthcare professionals in healthcare facility procedural areas and recovery rooms as an adjunct to monitor breathing in adult patients who are sedated for a diagnostic or therapeutic procedure. The predicate is used in sleep laboratory settings to monitor a patient's breathing rate while sleeping.
These differences do not change the intended use for the ReDe Mask and do not raise different questions of safety and effectiveness per 807.92(a)(5).
#### Location of Use
The ReDe Mask is used in healthcare facility procedural areas and recovery rooms as an adjunct to monitor breathing in adult patients while the ThermiSense predicate is used in sleep laboratory settings. In both cases, the patients are not awake. The bench and clinical performance testing show that the device performs as specified within its operating ranges and effectively detects breathing in the patients in the healthcare facility procedural areas and recovery rooms. Both the ThermiSense and the ReDe Mask are designed to detect breathing rate by measuring breathing pattern waveforms. The waveforms in both devices are the electrical signals produced by each devices' thermistor circuitry in response to breathing-caused temperature changes. These waveform patterns are analyzed to determine a breathing rate. The ThermiSense transfers its waveform data to external
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for TereoPneuma. The logo consists of a stylized image of three curved lines in the colors yellow, green, and red. To the right of the image is the company name, TereoPneuma, in green font. The letters 'TM' are in superscript to the right of the image and the company name.
equipment for analysis. The ReDe Mask retains its waveform data and performs the waveform analysis internal to the device itself. However, regardless of where the analyses take place, both devices are measuring and analyzing the same phenomenon: breathing pattern temperature waveforms. These waveforms are necessarily fundamental to the operation of both devices. Both devices measure temperature changes using thermistors, produce breathing waveforms corresponding to measured temperature patterns, and analyze those waveform patterns.
Based on these fundamental characteristics, which we view as equally inherent in both devices, we believe a claim of substantial equivalence is supported
## Technological and Performance Characteristics
#### Technology Similarities
The ReDe Mask and the predicate were compared in the following areas and found to have similar technological characteristics and to be equivalent:
| Characteristic | ReDe Mask | ThermiSense (K080922) |
|------------------------------------------------|-----------|-----------------------|
| Similar power source<br>using a battery | Yes | Yes |
| Similar permanent, non-<br>replaceable battery | Yes | Yes |
| Similar technology using<br>a thermistor | Yes | Yes |
| Similar electronic<br>components | Yes | Yes |
| Similar device function | Yes | Yes |
| Similar oxygen supply<br>system | Yes | Yes |
| Similar sampling position | Yes | Yes |
- A. Both devices measure respiratory rate with the same technology:
- Both devices use a thermistor sensor to detect changes in temperature in the . nasal and/or oral airflow pathways reflecting exhalation and inhalation (i.e., a breath).
- . Both devices use a low-power 3V coin battery (CR2032) as the electrical power source.
- . Both devices use a battery that is permanent and non-replaceable.
- B. Both devices have components that perform data analysis:
- . The ReDe Mask uses a thermistor connected to a microcontroller and other electronic components on a printed circuit board that is sealed in a small plastic housing mounted on the face mask.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for TereoPneuma. The logo consists of a stylized image of a flame in yellow and red, with a green circle in the center. To the right of the flame is the word "TereoPneuma" in green, with the letters "TM" in a smaller font size in the upper right corner.
- . The predicate uses a thermistor housed in a nasal cannula that is connected to other electronic components housed in a separate box that is located next to the patient,
- . Both devices use firmware that performs algorithm analysis of temperature changes during exhalation and inhalation and correlates it to a breath.
- C. Both devices include an oxygen supply system.
- . Uses a standard oxygen supply tube.
- D. Both have similar sampling position under the nares and in the airflow path of the mouth.
- The ReDe Mask uses a face mask and elastic strap to hold in place on the face. .
- . The predicate uses a molded plastic nasal cannula.
#### Technology Differences
The ReDe Mask and the predicate were compared in the following areas and found to have different technological characteristics but do not impact safety or performance:
| Characteristic | ReDe Mask | ThermiSense (K080922) |
|------------------------------------|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Requires additional<br>connections | No | Yes |
| Data Reported | Breathing rate by<br>way of colored<br>indicator lights | Airflow and breathing rate by<br>readout of waveform on computer<br>monitor |
| Stored Data | Accumulated data<br>stored internal to<br>the device. | Dependent upon recording device<br>selected |
| Mounting Design | Face mask and<br>electronics housing<br>is placed on<br>patient's face. | Nasal cannula and is placed on<br>patient's face and connected to<br>external box of electronics. |
| Ambient Operating<br>Temperature | 16° to 31° C | 5° to 32° C |
| Weight | 40 grams | 30 grams |
| Dimensions (control unit) | 43.43 mm H | 17.2 mm H |
| | 30.48 mm W | 34.8 mm W |
| | 14.99 mm D | 57.5 mm D |
A. Different methods are used to report the breathing rate:
- The ReDe Mask reports the breathing rate by way of colored indicator lights. ●
- . The predicate reports the breathing rate by way of a readout of a waveform on a computer monitor.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for TereoPneuma. The logo consists of a stylized graphic on the left and the company name "TereoPneuma" on the right. The graphic features a yellow and red swoosh design with a green circle in the center. The company name is written in green, with the letters "TM" in a smaller font size as a superscript to the right of the name.
The use of green, yellow and red lights to convey breathing rate information is simpler and easier than looking a waveform and counting the number of peaks on a moving signal on the computer monitor. This method of signaling the breathing rate does not impact safety and effectiveness of the ReDe Mask.
- B. Different methods are used for the applied parts:
- The ReDe Mask uses a face mask, a housing and a thermistor as applied parts . to attach to the face of the patient.
- . The predicate uses nasal cannula as applied parts to attach to the face of the patient.
The use of a standard face mask is as safe and accepted as the use of a nasal cannula. The face mask applied part does not impact the safety and effectiveness of the ReDe Mask.
- C. The ambient operating temperature of the ReDe Mask is specified in a range within the ambient operating temperature range of the predicate device.
Because the ambient operating temperature range of the ReDe Mask falls within the ambient operating temperature range of the predicate device, there is no impact on its safety and effectiveness.
- D. The weight of the ReDe Mask is on 10 grams heavier than the proposed predicate.
The 10 grams heavier weight of the ReDe Mask does not impose any extra burden on the patient that would impact safety and effectiveness of the ReDe Mask.
- E. The overall dimensions of the ReDe Mask is smaller than the predicate when comparing the volumes. The ReDe Mask is 19.843 cm³ while the predicate is 34.417 cm³.
The ReDe Mask is smaller and does not increase its impact on the safety and effectiveness.
#### Performance Data
The design of the ReDe Mask features the same operating principle, technology and manufacturing processes as the predicate device. The following performance tests were conducted to confirm the safety and effectiveness of the ReDe Mask.
#### Electromagnetic Compatibility Testing
The tests were performed according to following standard:
- . IEC 60601-1-2: 2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for TereoPneuma. The logo consists of a stylized image of a flame in yellow and red, with a green circle in the center. To the right of the flame is the word "TereoPneuma" in green font, with a small "TM" symbol in the upper right corner.
Remark: This device is ranged to the Group 1 Class B apparatus according to the standard of CISPR 11: 2010 clause 5.2.
# Electrical Safety Testing
Electrical Safety Testing was conducted in accordance with the following standards:
- IEC 60601-1:2005 + CORR.1:2006 + CORR.2:2007 + A1:2012 •
# Bench Testing:
| Test | ReDe Mask<br>proposed device | ThermiSense<br>predicate device | Comment |
|------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Battery Discharge<br>Sufficiency | Self contained coin<br>battery (CR2032)<br>with an 8 hour life to<br>enforce single use. | Uses the same<br>CR2032 coin battery | Proposed device is<br>as safe and effective<br>as the predicate. |
| Ambient<br>Temperature<br>Sensor | Operates properly in<br>range from 16° - 31°<br>C as specified. | Specified operating<br>range is 5° to 32° C. | Proposed device is<br>as safe and effective<br>as the predicate by<br>operating well within<br>the range specified<br>by the predicate. |
| Operational Time<br>Limits | Properly indicates<br>that device is about<br>to expire to enforce<br>single use. | Device has no<br>enforceable single use<br>method. | Proposed device is<br>as safe and effective<br>as the predicate. |
| Single Use<br>Enforcement | Properly expires to<br>enforce single use. | Device has no<br>enforceable single use<br>method. | Proposed device is<br>as safe and effective<br>as the predicate. |
| Worst-Case<br>Scenario | Operates as<br>specified with<br>breathing rate, tidal<br>volume and oxygen<br>flow parameters. | Parameters unknown. | Proposed device is<br>as safe and effective<br>as the predicate<br>given that the<br>predicate can perform<br>at least the same<br>parameters that are<br>the minimum<br>physiological/medical<br>requirements. |
| Aging | Ambient aging up to<br>1 year does not<br>produce loss of<br>performance or<br>physical integrity. | Thermistor unit is<br>warranted for 3<br>months and an<br>outboard interface box<br>is warranted for 1<br>year. | Proposed device is<br>as safe and effective<br>as the predicate and<br>will not expire for at<br>least the maximum<br>warranty time period<br>offered by the<br>predicate. |
| Transit | Transit conditions do<br>not produce loss of<br>performance or<br>physical integrity. | Unknown, but<br>presumably operates<br>after transit to<br>distributors and end-<br>users. | Proposed device is<br>as safe and effective<br>as the predicate by<br>continuing to perform<br>after transit<br>conditions, as does<br>the predicate. |
| Breath Detection | Operates as<br>specified with<br>breathing rate, tidal<br>volume and oxygen<br>flow parameters and<br>beyond the<br>specifications. | Parameters unknown. | Proposed device is<br>as safe and effective<br>as the predicate<br>given that the<br>predicate can perform<br>at least the same<br>parameters that are<br>the minimum<br>physiological/medical<br>requirements. |
| Breath Detection<br>at BPM, TV and<br>O² Limits | Operates as<br>specified with<br>breathing rate, tidal<br>volume and oxygen<br>flow parameters at<br>the limits of the<br>specifications. | Parameters unknown. | Proposed device is<br>as safe and effective<br>as the proposed<br>predicate given that<br>the proposed<br>predicate can perform<br>at least the same<br>parameters that are<br>the minimum<br>physiological/medical<br>requirements. |
| Light Indicator<br>Accuracy | Operates as<br>specified with<br>breathing rate, tidal<br>volume and oxygen<br>flow parameters and<br>beyond the<br>specifications and<br>light indicators<br>perform as designed. | Parameters unknown. | Proposed device is<br>as safe and effective<br>as the predicate<br>given that the<br>predicate can perform<br>at least the same<br>parameters that are<br>the minimum<br>physiological/medical<br>requirements. |
| Lifetime<br>Consistency and<br>Robustness | Operates as<br>specified over the<br>entire lifetime of 8<br>hours of operation<br>detecting all breaths<br>(consistency) and<br>shows that large<br>changes in breathing<br>rate do not affect its<br>accuracy<br>(robustness). | Parameters unknown. | Proposed device is<br>as safe and effective<br>as the predicate<br>given that the<br>predicate can perform<br>at least the same<br>parameters that are<br>the minimum<br>physiological/medical<br>requirements. |
| Temperature<br>Swings at Upper<br>Operating<br>Temperature | Operates as<br>specified at the<br>upper limits of the<br>operating<br>temperature of 31° C<br>where all the correct<br>lights illuminate as<br>expected in their<br>breathing rate<br>ranges.<br><br>Temperature swings<br>from a low of 0.17° C<br>to 1.40° C over a<br>tidal volume of 100<br>and 500 ml and<br>breathing rates from<br>2 to 45 BPM. The<br>very low swings<br>reflect the precision<br>and accuracy of the<br>device in detecting<br>breaths | Parameters unknown. | Proposed device is<br>as safe and effective<br>as the predicate<br>given that the<br>predicate can perform<br>at least the same<br>parameters that are<br>the minimum<br>physiological/medical<br>requirements. |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for "TereoPneuma™". The logo consists of a stylized graphic on the left and the company name on the right. The graphic features three curved shapes in yellow, green, and red, stacked on top of each other. The company name is written in green, with the "™" symbol in a smaller font size.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image is a logo for TereoPneuma. The logo consists of a stylized graphic on the left and the company name on the right. The graphic is a circular shape with three curved lines emanating from it in yellow, green, and red. The company name is written in green, with the letters "TereoPneuma" in a sans-serif font and the trademark symbol "TM" in superscript.
## Conclusion from Bench Performance Tests
Overall, the bench performance tests shows that the ReDe Mask passes all expectations and requirements. The conclusions from bench performance test data after performing nonclinical performance and safety studies is that the data shows that the ReDe Mask is as safe, as effective, and performs as well as the predicate device.
#### Biocompatibility Testing:
Four biocompatibility tests were conducted by NAMSA:
- . Cytotoxicity Study Using the ISO Elution Method (skin contact and external communication)
- ISO Guinea Pig Maximization Sensitization Test (skin contact and external communication)
- ISO Intracutaneous Study in Rabbits (skin contact and external communication)
- . ISO Systemic Toxicity Study in Mice (external communication)
#### Conclusion from Biocompatibility Tests
No effects of the applied parts were noted for cytotoxicity, skin irritation or skin sensitization and is as safe
#### Clinical Testing:
ReDe Mask 510(K) Application
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for TereoPneuma. The logo consists of a stylized image of a green circle with a yellow arc above it and a red arc below it. To the right of the image is the company name, "TereoPneuma," in green font. The letters "TM" are in superscript to the right of the image and the company name.
The Human Factors and Usability testing of the proposed device is established by the evaluation of the user interface, user interactions and instructions accompanying the device in accordance with FDA quidance documents and established standards: Human Factors and Usability Testing per FDA Guidance Document "Applying Human Factors and Usability Engineering to Medical Devices" and ISO 14971:2012. See "Section 6 - Risk Analysis" for more information on risk and usability factor identification.
Human Factors and Usability Validation: The purpose of the HF/U validation was to determine if the instructions for use and overall device design facilitate safe and effective use for the intended users, uses and use environment. The test designs were comprised of participants running through the order of operations when operating the device, including unpacking the device, reading the IFU, using the device and disposing of the device, while a study monitor observed and recorded any instances of subject difficulty, mishandling and/or misuse. The first study utilized 30 anesthesia care providers. The participants reported that the instructions for use were easy to follow and the device was easy to use. The second study utilized the complete spectrum of health care providers who may use the device and was conducted inside the intended environment of use. In both studies, all participants were able to correctly interpret the green, yellow and red illuminations, and all were able to see the flashing LEDs from 20 feet away. The study monitor observed no observed participant use errors, close calls or use problems, and received overwhelming positive feedback that the device is easy to understand and use. These results and participant feedback support the conclusion that the ReDe mask is as safe and as effective for use as the predicate.
Comparison of Accuracy With Capnograph: The purpose of the study was to compare the detection of exhalation and low respiratory rates in non-patient volunteers using the ReDe mask device with simultaneous end-tidal CO2 (capnograph) measurements of exhalation. A total of 38 individuals participated in this study. The data show conclusively that the proposed device is equivalent in all respects to a capnograph, considered the gold standard in monitor the breathing rate of a patient. No false positives or false negatives occurred with the use of the proposed device, ensuring that it can be used safely and without confusion in the detection of breath events.
Comparison of Accuracy With ExSpiron and Capnograph: The purpose of the study was to compare the detection of exhalation and low respiratory rate in non-patient volunteers using the ReDe mask device with simultaneous bioimpedance (ExSpiron) and end-tidal CO₂ (capnograph) measurements of exhalation. A total of 50 individuals participated in this study. The study looked at normal patient breathing, reduced patient breathing and cessation of patient breathing. The study also looked at the occurrence of false negatives and false positives. The data show conclusively that the ReDe Mask is equivalent in performance to an ExSpiron and a capnograph,
#### Animal Testing:
Animal testing was not required to demonstrate that the proposed device met its design requirements and therefore there are no animal data associated with this device.
Conclusions from clinical performance data
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for TereoPneuma. The logo consists of a stylized graphic on the left and the company name on the right. The graphic is composed of three curved shapes in yellow, green, and red, with a green circle in the center. The company name "TereoPneuma" is written in green, with "TM" as a superscript.
Human factor and usability studies showed that the device use and interpretation of instructions were easy to follow. Clinical performance data shows that the proposed device performs similarly to the capnograph, and to the ExSpiron, another reference device used for monitoring breathing.
### Conclusion
In establishing substantial equivalence of the ReDe Mask to the predicate device, TereoPneuma evaluated the intended use, indications for use, technological characteristics, reported adverse events and instrument risk profiles. The use of the ReDe Mask in patient monitoring environments does not raise any new types of questions of safety and effectiveness compared with the predicate device currently in use.
We believe that the ReDe Mask and the ThermiSense are substantially equivalent. Both the ThermiSense and the ReDe mask are designed to detect breathing by measuring breathing pattern waveforms. The waveforms in both devices are the electrical signals produced by each devices' thermistor circuitry in response to breathing-caused temperature changes. These waveform patterns are analyzed to determine a breathing event. The ThermiSense transfers its waveform data to external equipment for analysis. The ReDe mask retains its waveform data and performs the waveform analysis internal to the device itself. However, regardless of where the analyses take place, both devices are measuring and analyzing the same phenomenon: breathing pattern temperature waveforms. These waveforms are necessarily fundamental to the operation of both devices. Both devices measure temperature changes using thermistors, produce breathing waveforms corresponding to measured temperature patterns, and analyze those waveform patterns to arrive at a final output showing that the patient is breathing. Both devices use the same underlying technology, both devices measure and analyze waveform patterns to determine breathing. and both devices present their outputs, a display of breathing events, to clinicians in real time.
The design intent of both devices is the same; the underlying technology of both devices is the same: and the output information content of both devices is the same: breathing by the patient. Based on these fundamental characteristics, a claim of substantial equivalence is supported.
Based on the clinical and non-clinical testing performed, we believe that the ReDe mask is as safe and as effective as the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.