K061256 · Respironics, Inc. · NPF · Aug 2, 2006 · Anesthesiology
Device Facts
Record ID
K061256
Device Name
SMARTMONITOR 2PS
Applicant
Respironics, Inc.
Product Code
NPF · Anesthesiology
Decision Date
Aug 2, 2006
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 868.2377
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The SmartMonitor 2PS is intended for use in the continuous monitoring of respiration, heart rate and SpO2 levels of infant, pediatric, and adult patients. It detects and alarms for periods of high or low heart rate, high or low breath rate, and high or low saturation. When used as an infant monitor it is intended for use in a home or hospital environment. For infants only, it monitors and alarms for central apneas. When used as a pediatric or adult monitor, it is intended for use in a hospital environment. The SmartMonitor 2PSL is intended for use in the continuous monitoring of respiration, and heart rate of infant, pediatric, and adult patients. It detects and alarms for periods of high or low heart rate, and high or low breath rate. When used as an infant monitor it is intended for use in a home or hospital environment. For infants only, it monitors and alarms for central apneas. When used as a pediatric or adult monitor, it is intended for use in a hospital environment.
Device Story
Microprocessor-based monitor; acquires physiologic signals (ECG, respiration via transthoracic impedance, SpO2, pulse rate) via patient-attached sensors; analyzes signals to detect physiologic events (central apnea, bradycardia, abnormal breath/heart rates, SpO2 levels); generates visual/audible alarms; portable design for home/hospital use; healthcare provider reviews stored or real-time data via Respironics Synergy-E Host PC software; output facilitates clinical decision-making regarding patient status; benefits include continuous monitoring and timely intervention for respiratory/cardiac events.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and functionality verified through design verification testing and compliance with IEC 60601-1 and IEC 60601-1-2 standards.
Technological Characteristics
Microprocessor-based; plastic injection molded enclosure; transthoracic impedance for respiration; ECG and oximetry sensors; portable form factor (7.4" x 10" x 2.5", ~2 lbs); serial connectivity for data download via Synergy-E Host PC software; software-controlled; no hardware/firmware modifications from predicate.
Indications for Use
Indicated for continuous monitoring of respiration, heart rate, and SpO2 (2PS only) in infant, pediatric, and adult patients. Infant use in home or hospital; pediatric/adult use in hospital only. Alarms for high/low heart rate, breath rate, and saturation; central apnea monitoring for infants only.
Regulatory Classification
Identification
An apnea monitor is a complete system intended to alarm primarily upon the cessation of breathing timed from the last detected breath. The apnea monitor also includes indirect methods of apnea detection such as monitoring of heart rate and other physiological parameters linked to the presence or absence of adequate respiration.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Apnea Monitors; Guidance for Industry and FDA.”
CAS Medical Systems, Inc. 9303 Neonatal / Adult Vital Signs Monitor (K982776)
Submission Summary (Full Text)
{0}------------------------------------------------
Abbreviated 510(k) Tab 5 – 510(k) summary
# TAB 5
SmartMonitor 2PS
K061256
AUG 0 2 2006
# 510(K) SUMMARY
| Date of Submission | 01 May 2006 |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact / Address of Manufacturing facility | Zita A. Yurko Manager, Regulatory Affairs Respironics Inc. 1001 Murry Ridge Lane Murrysville, PA 15668 Phone: 724-387-4120 Fax: 724-387-4216 Zita.Yurko@Respironics.com |
| Proprietary Name | SmartMonitor 2PS |
| Common/Usual Name | Apnea Monitor / Vital Signs Monitor |
| Device Classification Name | Monitor, Apnea, Home Use / Monitor, Breathing Frequency / |
| | Oximeter |
| Classification Reference | 21 CFR 868.2377, 21 CFR 868.2375, and 21 CFR 870.2700 |
| Classification | Class II |
| Appropriate Classification Panel | Anesthesiology |
| Product Code | NPF, BZQ, and DQA |
| Predicate Devices | Respironics, Inc. SmartMonitor 2PS (K032403) CAS Medical Systems, Inc. 9303 Neonatal / Adult Vital Signs Monitor (K982776) |
| Reason for submission | Modified Indications for use and design |
.. .. ........ .. ...
{1}------------------------------------------------
## Substantial Equivalence
This premarket notification submission demonstrates that the SmartMonitor 2PS with expanded claims is substantially equivalent to the Respironics SmartMonitor 2PS (K032403) cleared for the infant patient population and CAS Medical Systems 9303 Neonatal / Adult Vital Signs Monitor (K982776) cleared for the adult, pediatric, and neonatal patient population.
The functionality of the design of the monitor was verified through the use of design verification testing. The safety of the design was assured by the completion of IEC 60601-1 and IEC 60601-1-2 testing. The Risk Traceability Matrix provided in the Risk Analysis assures that all hazards identified by the risk analysis are successfully mitigated.
This submission is seeking to extend the existing claims of the SmartMonitor 2PS to include adult and pediatric use.
#### Intended Use
The SmartMonitor 2PS is intended for use in the continuous monitoring of respiration, heart rate and SpO2 levels of infant, pediatric, and adult patients. It detects and alarms for periods of high or low heart rate, high or low breath rate, and high or low saturation. When used as an infant monitor it is intended for use in a home or hospital environment. For infants only, it monitors and alarms for central apneas. When used as a pediatric or adult monitor, it is intended for use in a hospital environment.
The SmartMonitor 2PSL is intended for use in the continuous monitoring of respiration, and heart rate of infant, pediatric, and adult patients. It detects and alarms for periods of high or low heart rate, and high or low breath rate. When used as an infant monitor it is intended for use in a home or hospital environment. For infants only, it monitors and alarms for central apneas. When used as a pediatric or adult monitor, it is intended for use in a hospital environment.
{2}------------------------------------------------
### Device Description
The SmartMonitor 2PS is a microprocessor-based, software-controlled device intended for use as an infant apnea monitoring system or as an adult or pediatric vital signs monitoring system. The electromechanical design of the SmartMonitor 2PS is unchanged from the SmartMonitor 2PS (K032403) except that cleared ECG and oximetry sensor accessories have been selected for adult and pediatric use of the monitor.
#### Patient Usage:
The SmartMonitor 2PS is designed to analyze and record physiologic signals (ECG, respiration, SpO2, and pulse rate) acquired from patients during sleep. Its primary function is to analyze the physiologic signals and generate visual and audible alarm indications upon detection of physiologic events such as central apnea (for infants), low breath rate (for adults and pediatrics), bradycardia, and high or low SpOx. The portable design of the device facilitates its use in a home environment. The primary functions are unchanged from the existing design of the SmartMonitor 2PS (K032403). The only difference is that the implementation of these functions has been expanded to include the adult and pediatric patient populations.
#### System Description:
ECG and respiration signals are acquired via a single transducer set that is attached to the patient and directly connected to the monitor. The measurement method used to derive the respiration signal is transthoracic impedance. SpO2 and plethysmographic pulse rate are acquired via an oximeter finger or foot sensor. The acquired physiologic signals are classified and stored for use at a later time. These system inputs are unchanged from the SmartMonitor 2PS (K032403), except that these features now include the adult and pediatric populations. No modifications to the SmartMonitor 2PS predicate firmware or hardware were needed to support the expanded intended use.
All of the following features are unchanged from the SmartMonitor 2PS (K032403):
- The SmartMonitor 2PS is a compact, lightweight unit. Two front panel connectors are provided . for the patient sensor input. The sensor connectors and associated sensor plugs are individually keyed to prevent improper insertion.
{3}------------------------------------------------
- The SmartMonitor 2PS is approximately 7.4 inches wide, 10 inches deep and 2.5 inches high. It u weighs approximately 2 pounds.
- The enclosure for the monitor is constructed of plastic injection molded materials, which are unchanged from K032403. Components and assemblies are securely mounted inside. The enclosure design is resistant to the entrance of liquids and other foreign materials.
- Locations for serial number plate and necessary user notes are provided at the bottom of the . monitor.
- Accessories for the SmartMonitor 2PS include a patient cable, lead wires, ECG electrodes, a l sensor belt to secure the sensors, and an oximeter sensor.
- Respironics Synergy-E Host PC Software is used to download or direct serial connect with the l SmartMonitor 2PS for the purpose of downloading the monitor's previously stored data or retrieving the monitor's real time data.
- The functionality of the design of the monitor was verified through the use of design verification . testing.
- . The safety of the design was assured by IEC 60601-1 and IEC 60601-1-2 testing.
- . The Risk Traceability Matrix provided the Risk Analysis assures that all hazards identified by the risk analysis are successfully mitigated.
(End of Tab.)
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread. The eagle is composed of four curved lines that represent the four major goals of the department.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## AUG # 2 2006
Respironics, Incorporated C/O Ms. Zita A. Yurko Manager, Regulatory Affairs Sleep & Home Respiratory Group 1001 Murry Ridge Lane Murrysville, Pennsylvania 15668
Re: K061256
Trade/Device Name: Respironics SmartMonitor 2PS and 2PSL Regulation Number: 21 CFR 868.2377 Regulation Name: Apnea Monitor Regulatory Class: II Product Code: NPF Dated: July 5, 2006 Received: July 7, 2006
Dear Mr. Yurko:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent ffor the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Yurko
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
Page 1 of 1
510(k) Number (if known): K061256_________________________________________________________________________________________________________________________________________
Device Name(s): Respironics SmartMonitor 2PS and 2PSL
The SmartMonitor 2PS is intended for use in the continuous monitoring of respiration, heart rate and SpO2 levels of infant, pediatric, and adult patients. It detects and alarms for periods of high or low heart rate, high or low breath rate, and high or low saturation. When used as an infant monitor it is intended for use in a home or hospital environment. For infants only, it monitors and alarms for central apneas. When used as a pediatric or adult monitor, it is intended for use in a hospital environment.
The SmartMonitor 2PSL is intended for use in the continuous monitoring of respiration, and heart rate of infant, pediatic, and adult patients. It detects and alarms for periods of high or low heart rate, and high or low breath rate. When used as an infant monitor it is intended for use in a home or hospital environment. For infants only, it monitors and alarms for central apneas. When used as a pediatric or adult monitor, it is intended for use in a hospital environment.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use XXXXXX (Per 21 CFR 801.109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96)
lille Salim
Thesiology, Gene
ber: K041254
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.