K223675 · Itamar Medical , Ltd. · MNR · Jan 6, 2023 · Anesthesiology
Device Facts
Record ID
K223675
Device Name
WatchPAT ONE (WP1)
Applicant
Itamar Medical , Ltd.
Product Code
MNR · Anesthesiology
Decision Date
Jan 6, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.2375
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The WatchPAT ONE (WP1) device is a noninvasive home care device for use with patients suspected to have sleep related breathing disorders. The WP1 is a diagnostic aid for the detection of sleep related breathing disorders, sleep staging (Rapid Eye Movement (REM) Sleep, Light Sleep, Deep Sleep and Wake), snoring level and body position. The WP1 generates a peripheral arterial tonometry ("PAT") Respiratory Disturbance Index ("PRDI"), Apnea- Hypopnea index ("PAHI"), Central Apnea-Hypopnea index ("PAHIC"), PAT sleep staging identification (PSTAGES) and optional snoring level and body position discrete states from an external integrated snoring and body position sensor. The WP1's PSTAGES and snoring level and body position provide supplemental information to its PRD/PAHI/PAHIc. The WP1's PSTAGES and snoring level and body position are not intended to be used as the sole or primary basis for diagnosing any sleep related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted. PAHIc is indicated for use in patients 17 years and older. All other parameters are indicated for 12 years and older.
Device Story
WatchPAT ONE (WP1) is a non-invasive, patient-worn home diagnostic aid for sleep-related breathing disorders. Inputs include PAT signals and blood oxygen saturation from a finger-mounted uPAT probe, wrist motion from an embedded actigraph, and optional chest movement/snoring data. A microprocessor records these signals; offline software analyzes data to generate respiratory indices (PRDI, PAHI, PAHIc), sleep staging (PSTAGES), snoring levels, and body position. The device now includes an arrhythmia flagging feature (inherited from predicate WP300) that identifies suspected Atrial Fibrillation and premature beats using PPG signal analysis. The arrhythmia output is for informational use only, intended to assist physicians in determining if further cardiac investigation is required; it is not a diagnostic tool for arrhythmia. The device is used at home by the patient; data is analyzed by a physician to aid clinical decision-making regarding sleep disorders and potential cardiac follow-up.
Clinical Evidence
Bench testing only. Verification testing confirmed that the WP1 acquisition system (PAT and actigraphy) generates input signals equivalent to the predicate WP300, validating the performance of the integrated arrhythmia software algorithm.
Technological Characteristics
Ventilatory effort recorder utilizing Peripheral Arterial Tone (PAT) signal. Components: wrist-worn control unit with embedded actigraphy, uPAT finger probe (optical plethysmography), and optional chest sensor. Power: 1.5V Alkaline AAA battery. Connectivity: offline software analysis. Software: includes arrhythmia flagging algorithm (PPG-based) for Atrial Fibrillation and premature beats.
Indications for Use
Indicated for patients 12 years and older suspected of sleep-related breathing disorders; PAHIc parameter indicated for patients 17 years and older.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 6, 2023
Itamar Medical, LTD % Jonathan Kahan, Partner Hogan Lovells US LLP 555 Thirteenth Street. NW Washington, District of Columbia 20004-1109
Re: K223675
Trade/Device Name: WatchPAT ONE (WP1) Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: MNR Dated: December 7, 2022 Received: December 7, 2022
Dear Jonathan Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachana Visaria -S
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223675
Device Name WatchPAT ONE (WP1)
#### Indications for Use (Describe)
The WatchPAT ONE (WP1) device is a noninvasive home care device for use with patients suspected to have sleep related breathing disorders. The WP1 is a diagnostic aid for the detection of sleep related breathing disorders, sleep staging (Rapid Eye Movement (REM) Sleep, Light Sleep, Deep Sleep and Wake), snoring level and body position. The WP1 generates a peripheral arterial tonometry ("PAT") Respiratory Disturbance Index ("PRDI"), Apnea- Hypopnea index ("PAHI"), Central Apnea-Hypopnea index ("PAHIC"), PAT sleep staging identification (PSTAGES) and optional snoring level and body position discrete states from an external integrated snoring and body position sensor. The WP1's PSTAGES and snoring level and body position provide supplemental information to its PRD/PAHI/PAHIc. The WP1's PSTAGES and snoring level and body position are not intended to be used as the sole or primary basis for diagnosing any sleep related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted.
PAHIc is indicated for use in patients 17 years and older. All other parameters are indicated for 12 years and older.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
| Applicant's Name: | Itamar Medical Itd. |
|-------------------|--------------------------|
| | 9 Halamish st. |
| | Caesarea 3088900, Israel |
| | Tel: +972 4 617 7000 |
| | Fax: +972 4 627 5598 |
- Contact Person: Jonathan Kahan, Esq. Hogan Lovells US LLP Columbia Square 555 Thirteenth Street, NW Washington, DC 20004-1109 Tel: (202)637-5794 Fax: (202)637-5910 Email: jonathan.kahan@hoganlovells.com
- November 30th, 2022 Date Prepared:
Trade Name: WatchPAT ONE (WP1)
Common or Usual Name: Ventilatory Effort Recorder
Classification Name: Breathing Frequency Monitor
Medical Specialty: Anesthesiology
Product Code: Ventilatory Effort Recorder, MNR
Device Class: Class II
Regulation Number: 868.2375
Panel: Anesthesiology
Predicate Device: WatchPAT300 (WP300) (Itamar Medical Ltd), cleared under K222331; product code MNR
Reference Device: WatchPAT ONE (WP1) (Itamar Medical Ltd.), cleared under K183559; product code MNR
### Purpose of the Special 510(k) notice.
The subject WP1 combines the software arrhythmia feature of the predicate device WP300 (K222331) into the reference device WP1 (K183559).
{4}------------------------------------------------
### Intended Use
The WatchPAT ONE is a non-invasive home care device for use with patients suspected to have sleep related breathing disorders. The WP1 is a diagnostic aid for the detection of sleep related breathing disorders, sleep staging (Rapid Eye Movement (REM) Sleep, Light Sleep, Deep Sleep and Wake), snoring level and body position. The WP1 generates a peripheral arterial tonometry ("PAT") Respiratory Disturbance Index ("PRDI"). Apnea-Hypopnea index ("PAHI"). Central Apnea-Hypopnea index ("PAHIc"). PAT sleep staging identification (PSTAGES) and optional snoring level and body position discrete states from an external integrated snoring and body position sensor. The WP1's PSTAGES and snoring level and body position provide supplemental information to its PRDI/PAHI/PAHIc. The WP1's PSTAGES and snoring level and body position are not intended to be used as the sole or primary basis for diagnosing any sleep related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted.
PAHIc is indicated for use in patients 17 years and older. All other parameters are indicated for 12 years and older.
### Device Description
The subject WatchPAT ONE (WP1) has the same intended use and indications for use as the cleared predicate WP300 (K222331) and the cleared reference WP1 (K183559). The subject WP1 combines the software arrhythmia feature of the WP300 cleared predicate (K222331) into the WP1 cleared reference (K183559).
The subject WP1 consists of: A wrist worn Control Unit, Actigraphy (in the Control Unit), Wrist Strap, Power Supply (battery), uPAT probe, Chest sensor (optional), Management and Analysis Software (SW) and mobile application.
The subject WP1, like the predicate WP300 and reference WP1, uses the same Management and Analysis Software (SW). The arrhythmia feature is to be used for informational use only, to flaq patients suspected of having arrhythmias, thereby aiding the physician to decide if further arrhythmia investigation is needed. The device is not intended to be used as a diagnostic device for cardiac arrhythmia and it is not intended to replace traditional methods of diagnosis. The arrhythmia detection is not intended for use in life supporting or sustaining systems or monitor and alarm devices. It provides the sleep physician with additional information to that of the WP1 cleared capabilities of detecting sleep disorders, to be considered in conjunction with the physician's knowledge of patient background, clinical history, symptoms, and other diagnostic information.
# Technological Characteristics
The subject WP1 has the same technological characteristics as the predicate WP300 and the reference WP1 to which it has been modified.
Identically to WP300 and WP1, the subject WP1 device is a ventilatory effort recorder that utilizes Peripheral Arterial Tone (PAT) signal. It is a patient-worn device used at home for aiding in the diagnosis of sleep related breathing disorders based on the PAT signal. In all 3 systems, the controller part of the device is worn on the wrist and records the PAT signal and blood oxygen saturation levels in the distal part of the finger-mounted probe based on an optical plethysmographic method, and the wrist motion from an embedded actigraphy.
{5}------------------------------------------------
The following report output information is provided by the cleared WP300, cleared WP1, and the subject WP1 using the same offline SW and algorithms that analyze recorded data to present respiratory disturbance index (pRDI), Apnea Hypopnea Index (pAHI), Central Apnea Hypopnea Index (pAHIc), sleep stages (pSTAGES - REM Sleep, Light Sleep, Deep Sleep, Wake), snoring level and body position discrete states. The new arrhythmia information in the subject WP1's report is identical to the one in the cleared predicate WP300.
The subject WP1, identically to both the predicate and the reference device, consists of the same components: (1) unified PAT probe (uPAT) is used to measure the PAT and oximetry signals: (2) an embedded actigraph which provides a signal that is used to determine periods of sleep/wake based on the motion of the wrist; (3) Electronics, which include a microprocessor that records the information supplied by the uPAT finger probe, actigraph and chest movement; and (4) chest sensor (SBP/RESBP sensors same as in the predicates).
The SW characteristics of the subject WP1 and the reference WP1 are the same except for the newly added SW algorithm that was cleared for use in the predicate WP300 (K222331) to identify or 'flag' the patient for further arrhythmia investigation (Atrial Fibrillation and Premature Beats). This additional SW algorithm is the proposed modification to the subject WP1.
The HW of the subject WP1 is the same as the HW of the reference WP1 (K183559). Further, the HW of the WP1 was designed to provide identical signals to that of the predicate WP300. The differences in HW between the WP1 and the WP300 devices have been thoroughly discussed and found to be substantially equivalent in K183559.
The cleared arrhythmia SW algorithm from the predicate WP300 added to the subject WP1 does not alter the technological characteristics of the device or its principles of operation.
# Performance Data
SW Design verification was performed on the subject WP1 device with the SW containing the additional arrhythmia feature. The testing and acceptance criteria are the same as those in the predicate WP300 and reference WP1 devices. Bench testing was previously conducted on the reference WP1 (K183559) to show that the acquisition system of the WP1 and WP300 generate equivalent input signals to the analysis algorithms. Specifically, the equivalence of the PAT and the actigraphy signals which are used in the newly added arrhythmia algorithm were tested. This testing provides evidence that the WP1 signals are equivalent to the WP300 signals.
# Substantial Equivalence
The subject WP1 has the same intended use as both the predicate WP300 and the reference WP1. The WP1's new SW arrhythmia feature is offered as supplemental information to its sleep information. The device is not intended to be used as a diagnostic device for any cardiac arrhythmia. All other information supplied is the same as the information supplied by the predicate device and the differences in HW between the predicate WP300 (in which the new SW feature for Arrhythmia detection was cleared) and the subject WP1 do not raise new questions of safety or effectiveness.
{6}------------------------------------------------
| | Subject WP1 | WP300 (K222331) | WP1 (K183559) | Comparison |
|---------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The WatchPAT ONE<br>(WP1) device is a non-<br>invasive home care device<br>for use with patients<br>suspected to have sleep<br>related breathing<br>disorders. The WP1 is a<br>diagnostic aid for the<br>detection of sleep related<br>breathing disorders, sleep<br>staging (Rapid Eye<br>Movement (REM) Sleep,<br>Light Sleep, Deep Sleep<br>and Wake), snoring level<br>and body position. The<br>WP1 generates a<br>peripheral arterial<br>tonometry ("PAT")<br>Respiratory Disturbance<br>Index ("PRDI"), Apnea-<br>Hypopnea index ("PAHI"),<br>Central Apnea-Hypopnea<br>index ("PAHIc"), PAT<br>sleep staging identification<br>(PSTAGES) and optional<br>snoring level and body<br>position discrete states<br>from an external integrated<br>snoring and body position<br>sensor. The WP1's<br>PSTAGES and snoring<br>level and body position<br>provide supplemental<br>information to its<br>PRDI/PAHI/PAHIc. The<br>WP1's PSTAGES and<br>snoring level and body<br>position are not intended<br>to be used as the sole or<br>primary basis for<br>diagnosing any sleep<br>related breathing disorder,<br>prescribing treatment, or<br>determining whether<br>additional diagnostic<br>assessment is warranted.<br>PAHIc is indicated for use<br>in patients 17 years and<br>older. All other parameters<br>are indicated for 12 years<br>and older | The Watch-PAT300<br>(WP300) device is a non-<br>invasive home care device<br>for use with patients<br>suspected to have sleep<br>related breathing<br>disorders. The WP300 is a<br>diagnostic aid for the<br>detection of sleep related<br>breathing disorders, sleep<br>staging (Rapid Eye<br>Movement (REM) Sleep,<br>Light Sleep, Deep Sleep<br>and Wake), snoring level<br>and body position. The<br>WP300 generates a<br>peripheral arterial<br>tonometry ("PAT")<br>Respiratory Disturbance<br>Index ("PRDI"), Apnea-<br>Hypopnea index ("PAHI"),<br>Central Apnea-Hypopnea<br>index ("PAHIc"), PAT<br>sleep staging identification<br>(PSTAGES) and optional<br>snoring level and body<br>position discrete states<br>from an external integrated<br>snoring and body position<br>sensor. The WP300's<br>PSTAGES and snoring<br>level and body position<br>provide supplemental<br>information to its<br>PRDI/PAHI/PAHIc. The<br>WP300's PSTAGES and<br>snoring level and body<br>position are not intended<br>to be used as the sole or<br>primary basis for<br>diagnosing any sleep<br>related breathing disorder,<br>prescribing treatment, or<br>determining whether<br>additional diagnostic<br>assessment is warranted.<br>PAHIc is indicated for use<br>in patients 17 years and<br>older. All other parameters<br>are indicated for 12 years<br>and older | The WatchPAT ONE<br>(WP1) device is a non-<br>invasive home care device<br>for use with patients<br>suspected to have sleep<br>related breathing<br>disorders. The WP1 is a<br>diagnostic aid for the<br>detection of sleep related<br>breathing disorders, sleep<br>staging (Rapid Eye<br>Movement (REM) Sleep,<br>Light Sleep, Deep Sleep<br>and Wake), snoring level<br>and body position. The<br>WP1 generates a<br>peripheral arterial<br>tonometry ("PAT")<br>Respiratory Disturbance<br>Index ("PRDI"), Apnea-<br>Hypopnea index ("PAHI"),<br>Central Apnea-Hypopnea<br>index ("PAHIc"), PAT<br>sleep staging identification<br>(PSTAGES) and snoring<br>level and body position<br>discrete states from an<br>external integrated snoring<br>and body position sensor.<br>The WP1's PSTAGES and<br>snoring level and body<br>position provide<br>supplemental information<br>to its PRDI/PAHI/PAHIc.<br>The WP1's PSTAGES and<br>snoring level and body<br>position are not intended<br>to be used as the sole or<br>primary basis for<br>diagnosing any sleep<br>related breathing disorder,<br>prescribing treatment, or<br>determining whether<br>additional diagnostic<br>assessment is warranted.<br>PAHIc is indicated for use<br>in patients 17 years and<br>older. All other parameters<br>are indicated for 12 years<br>and older. | Same as WP300,<br>almost the same<br>as WP1 reference<br>device, except for<br>addition of the<br>word "optional",<br>highlighted in<br><b>bold</b> . |
| User Population | Adult and adolescence<br>(from age 12) | Adult and adolescence<br>(from age 12) | Adult and adolescence<br>(from age 12) | Same |
| Intended Use<br>Environment | Home Use | Home Use | Home Use | Same |
| | Subject WP1 | WP300 (K222331) | WP1 (K183559) | Comparison |
| Channels | PAT, Pulse rate, Oximetry,<br>Actigraphy, Snoring, Body<br>Position, Chest Movement | PAT, Pulse rate, Oximetry,<br>Actigraphy, Snoring, Body<br>Position, Chest Movement | PAT, Pulse rate, Oximetry,<br>Actigraphy, Snoring, Body<br>Position, Chest Movement | Same |
| SW Analysis<br>output | • Respiratory indices<br>(pRDI, pAHI,pAHIc)<br>• Sleep stages (REM,<br>light, deep and wake)<br>• Snoring level<br>• Body position<br>discrete states<br>• Heart rate statistics<br>• Oximetry statistics | • Respiratory indices<br>(pRDI, pAHI,pAHIc)<br>• Sleep stages (REM,<br>light, deep and wake)<br>• Snoring level<br>• Body position<br>discrete states<br>• Heart rate statistics<br>• Oximetry statistics | • Respiratory indices<br>(pRDI, pAHI,pAHIc)<br>• Sleep stages (REM,<br>light, deep and wake)<br>• Snoring level<br>• Body position<br>discrete states<br>• Heart rate statistics<br>• Oximetry statistics | Same |
| | Arrhythmia flagging output:<br>• Suspected Atrial<br>Fibrillation (AFib):<br>○ Total duration in<br>sleep<br>○ Longest event<br>duration<br>• Premature beats:<br>○ Events per<br>minute | Arrhythmia flagging output:<br>• Suspected Atrial<br>Fibrillation (AFib):<br>○ Total duration in<br>sleep<br>○ Longest event<br>duration<br>• Premature beats:<br>○ Events per<br>minute | No Arrhythmia output | Same as WP300 |
| Input uPAT Probe | Itamar proprietary probe<br>only | Itamar proprietary probe<br>only | Itamar proprietary probe<br>only | Same |
| HW Components | • uPAT finger probe<br>• Actigraph<br>• Controller<br>• Chest sensor (optional) | • uPAT finger probe<br>• Actigraph<br>• Controller<br>• Chest sensor (optional)<br>• External Tamper-Proof<br>Bracelet (optional) | • uPAT finger probe<br>• Actigraph<br>• Controller<br>• Chest sensor | Same except<br>optional Tamper<br>Proof Bracelet not<br>available in WP1,<br>and chest sensor<br>was not optional in<br>predicate WP1. |
| Power Supply | One OTS 1.5V Alkaline<br>AAA battery | One OTS 1.5V Alkaline<br>AAA battery<br>OR<br>One rechargeable AAA<br>1.2V Nickel-metal hydride<br>(NiMH) battery | One OTS 1.5V Alkaline<br>AAA battery | Same as WP1,<br>single use battery<br>only |
| Sensors<br>Placement | Wrist, finger and chest<br>(optional) | Wrist, finger and chest<br>(optional) | Wrist, finger and chest | Same, except for<br>chest not being<br>optional in<br>predicate WP1 |
| Intended user of<br>the Arrhythmia<br>output | Physician | Physician | No Arrhythmia output | Same as WP300 |
| Arrhythmia<br>Monitoring period | During prescribed<br>sleep study | During prescribed<br>sleep study | No Arrhythmia output | Same as WP300 |
| Arrhythmia<br>output<br>technological<br>characteristic | PPG | PPG…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.