K221816 · Wesper, Inc. · MNR · Mar 9, 2023 · Anesthesiology
Device Facts
Record ID
K221816
Device Name
Wesper Lab
Applicant
Wesper, Inc.
Product Code
MNR · Anesthesiology
Decision Date
Mar 9, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2375
Device Class
Class 2
Indications for Use
The Wesper Lab is a digital recording device designed to be used under the direction of a physician or trained technician but may be applied by a layperson. Wesper Lab records multiple physiological parameters from a sleeping patient for the purpose of simultaneous or subsequent display of the parameters. The displayed data assists in the identification of sleep apnea by trained personnel. Wesper Lab is intended to be used for adult sleep studies at home or clinical environment. The device does not actively monitor or diagnose the patient and does not issue any alarms.
Device Story
Wesper Lab is a wearable digital recording system for adult sleep studies. It consists of two flexible, thin, wireless patches (thoracic and abdominal) and a mobile application. Patches use accelerometry and optical sensing to measure sleep position, respiratory effort, and airflow/pressure. The mobile app receives data via Bluetooth Low-Energy (BLE) from the patches and an external FDA-cleared pulse oximeter (for heart rate and SpO2). Data is uploaded to a secure cloud, then downloaded to a local PC where Wesper staff use a Study Output Module (SOM) to save data for physician interpretation. The device is applied by laypeople under physician/technician direction. It does not provide real-time monitoring, diagnosis, or alarms. The system benefits patients by enabling sleep data collection in home environments.
Clinical Evidence
Bench testing only. Performance testing included electrical safety (IEC 60601 series), electromagnetic compatibility, wireless coexistence (IEEE/ANSI C63.27), enclosure protection (IEC 60529), and battery safety (IEC 62133). Usability was validated per IEC 62366 and FDA human factors guidance, confirming no critical tasks.
Technological Characteristics
Wearable sensor system; thoracic/abdominal patches with optical sensors and accelerometers. Materials: biocompatible silicone-based adhesive (ISO 10993). Power: internal Li-ion battery, rechargeable via charging pad. Connectivity: BLE to mobile app, cloud-based storage. Dimensions: 3.8" L x 2.1" W x 0.23" D. Non-sterile. Software: Study Output Module (SOM) for data processing.
Indications for Use
Indicated for adult patients undergoing sleep studies in home or clinical environments to record physiological parameters for sleep apnea identification. Not for active monitoring or diagnosis; no alarms.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 9, 2023
Wesper Inc. Amir Reuveny CEO 234 5th Ave New York, New York 10001
Re: K221816
Trade/Device Name: Wesper Lab Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: MNR Dated: February 8, 2023 Received: February 8, 2023
Dear Amir Reuveny:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Rachana Visaria -S
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "WESPER" in bold, black letters. To the left of the word is a gray circle that is partially filled in, resembling a crescent moon. The font is simple and sans-serif, giving the logo a clean and modern look. The overall composition is straightforward and emphasizes the text.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K221816
Device Name Wesper Lab
Indications for Use (Describe)
The Wesper Lab is a digital recording device designed to be used under the direction of a physician or trained technician but may be applied by a layperson. Wesper Lab records multiple physiological parameters from a sleeping patient for the purpose of simultaneous or subsequent display of the parameters. The displayed data assists in the identification of sleep apnea by trained personnel. Wesper Lab is intended to be used for adult sleep studies at home or clinical environment. The device does not actively monitor or diagnose the patient and does not issue any alarms.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}------------------------------------------------
## WESPER
# 510(k) Summary
### 1. SUBMITTER
| Company & Address: | 234 5th Avenue<br>New York, NY 10001<br>516-654-4166 |
|--------------------|------------------------------------------------------|
| Contact Person: | Amir Reuveny |
| Date Prepared: | 11/18/2022 |
### 2. DEVICE
| Name of Device: | Wesper Lab |
|-----------------------|------------------------------|
| Common or Usual Name: | Breathing frequency monitor. |
| Classification Name: | Breathing frequency monitor. |
| Regulation: | 21 CFR 868.2375 |
| Regulatory Class: | II |
| Product Code: | MNR |
### 3. PREDICATE DEVICE
- Wesper Lab (K203343)
### 4. DEVICE DESCRIPTION
Wesper Lab ("the device") is a digital recording device designed to be used under the direction of a physician or trained technician but may be applied by a layperson. Wesper Lab records multiple physiological parameters from a sleeping patient for the purpose of simultaneous or subsequent display of the parameters. The displayed data assists in the identification of sleep apnea by trained personnel.
The device consists of an abdominal patch, a thoracic patch and a mobile application. The patches are multi-use wearable, flexible, thin, and wireless, and are designed to record sleep data in adult patients. Both patches are the same and differ only in their anatomical designation, which is determined by the order in which the user applies them at test setup time. The flexible fabric allows the patch to contract and expand as the patient breathes and moves during sleep.
The mobile application ("the app") resides on the patient's personal mobile device, relaying sleep data wirelessly to a secure remote storage location ("the cloud") for subsequent analysis by a healthcare professional. The app has 3 Bluetooth Low-Energy (BLE) ports, each of which receives multiple physiological channels from the patient.
The device's BLE ports connect to the following:
- 1. Abdominal patch
- 2. Thoracic patch
- 3. Pulse oximeter
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "WESPER" in black font. To the left of the word is a gray circle that is partially shaded. The shading gives the circle a crescent moon appearance. The font is sans-serif and appears to be a bold typeface.
The patches collect multiple physiological parameters related to sleep to be used by a healthcare professional. Specifically, the patches measure sleep position, respiratory effort, and airflow / air pressure. The patches leverage accelerometry and optical sensing to provide these measurements.
Data from the patches is transmitted via BLE throughout the night to the app, which uploads the data to the cloud. A third BLE port on the app connects to an FDA cleared pulse oximeter, which provides heart rate and blood oxygen saturation measurements.
The data recorded by the patches is relayed to a remote secure storage, where it will be downloaded to a local PC. Then, Wesper staff will execute the Study Output Module (SOM) and save the data locally on the desktop computer. The data is then ready for interpretation by a healthcare provider.
### 5. INDICATIONS FOR USE
The Wesper Lab is a digital recording device designed to be used under the direction of a physician or trained technician but may be applied by a layperson. Wesper Lab records multiple physiological parameters from a sleeping patient for the purpose of simultaneous or subsequent display of the parameters. The displayed data assists in the identification of sleep apnea by trained personnel. Wesper Lab is intended to be used for adult sleep studies at home or clinical environment. The device does not actively monitor or diagnose the patient and does not issue any alarms.
### 6. COMPARISON OF INTENDED USE
Both the Wesper Lab and the predicate device have the same intended use, specifically, for the recording of physiological parameters from a sleeping patient.
### 7. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject device is substantially equivalent to the predicate device (Wesper Lab, K203343). Both devices have the same intended use and similar indications, technological characteristics and principles of operation, with two design changes introduced in the subject device:
- . A charging element was added to the patch's electronic assembly, which will receive energy from a corresponding charging pad ("Wesper Charger").
- . The biocompatible adhesive layer at the bottom of the patch has been made replaceable, intended to be discarded after each nightly application. The adhesive material is identical to the one used in the predicate.
A substantial equivalence chart comparing the similarities and differences between the subject and the predicate device is provided in Table 1. Any minor differences in the technological characteristics do not raise different questions of safety or effectiveness.
{6}------------------------------------------------
| | Wesper Lab<br>(Subject) - K221816 | Wesper Lab<br>(Predicate) - K203343 | Similarities and<br>Differences |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification<br>regulation | 21 CFR 868.2375 | | Same |
| Product code | MNR | MNR | Same |
| Intended Use | For the recording of physiological parameters<br>from a sleeping patient. | | Same |
| Indications for Use | The Wesper Lab is a<br>digital recording device<br>designed to be used<br>under the direction of a<br>physician or trained<br>technician but may be<br>applied by a layperson.<br>Wesper Lab records<br>multiple physiological<br>parameters from a<br>sleeping patient for the<br>purpose of<br>simultaneous or<br>subsequent display of<br>the parameters. The<br>displayed data assists<br>in the identification of<br>sleep apnea by trained<br>personnel. Wesper<br>Lab is intended to be<br>used for adult sleep<br>studies at home or<br>clinical environment.<br>The device does not<br>actively monitor or<br>diagnose the patient<br>and does not issue<br>any alarms. | The Wesper Lab is a<br>digital recording device<br>designed to be used<br>under the direction of a<br>physician or trained<br>technician but may be<br>applied by a layperson.<br>Wesper Lab records<br>multiple physiological<br>parameters from a<br>sleeping patient for the<br>purpose of<br>simultaneous or<br>subsequent display of<br>the parameters. The<br>displayed data assists<br>in the identification of<br>sleep apnea by trained<br>personnel. Wesper Lab<br>is intended to be used<br>for adult sleep studies<br>at home or clinical<br>environment. The body-<br>worn component of the<br>system is single-use, to<br>be discarded after its<br>dedicated nightly<br>application. The device<br>does not monitor or<br>diagnose the patient<br>and does not issue any<br>alarms. | Similar<br>The indications for<br>use for the subject<br>device have been<br>modified from the<br>predicate to<br>remove references<br>to single use<br>limitations. This<br>change does not<br>raise any different<br>questions of safety<br>or effectiveness. |
| Intended<br>environment(s) | Home or clinical environment | | Same |
| Patient population | Adults | Same | |
| Rx or OTC | Rx | Same | |
| Device Type | Wearable Sensor | Same | |
| Main anatomical<br>Site | Torso | Same | |
| Airflow | Indirect measurement using respiratory effort<br>signals. | Same | |
| Respiratory Efforts | Thoracic and abdominal effort based on optical<br>sensor. | Same | |
| Body Position | Solid State Accelerometer | Same | |
| Body adherence<br>method | Biocompatible double-<br>sided silicone-based<br>adhesive, separable | Biocompatible double-<br>sided silicone-based<br>adhesive, non-<br>separable | Similar |
| Display type | Visual display including LEDs and device<br>specific visual indicators. | Same | |
| Power source | Internally powered using li-ion<br>battery | Same | |
| Charging | Multiple times in<br>patient's home or<br>clinical environment | Once at manufacturing | Similar<br>The difference in<br>the charging<br>mechanism has<br>been tested using<br>IEC 60601 and<br>IEC 62133 to<br>show that there is<br>no change in the<br>safety, EMC or<br>usability risk<br>profile of the<br>device. |
| Data storage | Internal memory and secured cloud storage | Same | |
| Communication<br>interface | Bluetooth low energy (BLE) | Same | |
| Access to recorded<br>data | Output file available to healthcare provider for<br>review | Same | |
| Device dimensions | 3.8 L x 2.1 W x 0.23 D (inches, max) | | Same |
| Biocompatibility | Biocompatible in accordance with ISO 10993 | | Same |
| Signal Recorded<br>Channels | 1. Thoracic effort<br>2. Abdominal effort<br>3. Body position<br>4. Airflow<br>5. Pressure | | Same |
| Internal memory | Flash memory card.<br>Recording time 40 hours. | | Same |
| Pulse oximeter | Coupled with an authorized FDA-cleared pulse<br>oximeter to measure pulse-rate and SpO2. | | Same |
| Sterility | Non-sterile | | Same |
| Electrical safety | IEC 60601, IEC 62133 | IEC 60601 | Similar |
| | | | The predicate<br>device's test<br>standards are<br>included in the<br>subject device's<br>safety test profile. |
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "WESPER" in a bold, sans-serif font. To the left of the word is a gray circle that is partially filled in, resembling a crescent moon. The text and the circle are the only elements in the image. The overall impression is clean and modern.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "WESPER" in all caps. To the left of the word is a gray circle that is partially filled in, resembling a moon. The font is sans-serif and the letters are bold.
### 8. PERFORMANCE DATA
Wesper has performed bench testing on its device battery to establish its reusability lifetime and has provided cleaning instructions to the operator to apply between uses.
In addition, Wesper conducted comprehensive performance testing according to the following standards and guidelines:
#### 8.1. Electrical Safety / EMC
- . IEC 60601 – Medical Electrical Equipment, Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances requirements and tests
- . IEC 60601 - Medical electrical equipment - Part 1-1: General requirements for safety -Collateral standard: Safety requirements for medical electrical systems
- IEEE/ANSI C63.27 - Evaluation of wireless coexistence
- . IEC 60529 ED. 2.2 B:2013 - Degrees of protection provided by enclosures (IP Code).
- . IEC 62133 - Secondary cells containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications – Part 2: Lithium systems (2017)
Wesper Lab met all acceptance criteria.
{9}------------------------------------------------
## OWESPER
### 8.2. Usability / Human Factors
Wesper Lab has been assessed to have no critical tasks as established by a Use-Related Risk Analysis, confirmed by a usability study. This assessment was conducted in accordance with IEC 62366 - Medical devices - Part 1: Application of usability engineering to medical devices and FDA Guidance "Applying Human Factors and Usability Engineering to Medical Devices", issued February 2016.
### CONCLUSION
The Wesper Lab (subject) and Wesper Lab (predicate, K203343) have the same intended use and similar indications, technological characteristics, and principles of operation. The technological differences between the subject and the predicate device, as described above, do not present different questions of safety or effectiveness. The data provided demonstrates that Wesper Lab is substantially equivalent to its predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.