K191095 · Neurovirtual USA, Inc. · MNR · Jan 16, 2020 · Anesthesiology
Device Facts
Record ID
K191095
Device Name
Maxxi Snore Sensor
Applicant
Neurovirtual USA, Inc.
Product Code
MNR · Anesthesiology
Decision Date
Jan 16, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2375
Device Class
Class 2
Indications for Use
The Maxxi Snore Sensor is a device intended to acquire snore bursts. It responds to snoring and other sounds in the audio range picked up through the skin and convert them to a small analog voltage that provides an indication of the presence of these sound/vibration bursts. This sensor is intended to be used with polysomnography devices and adult patients. Intended to be use in a sleep laboratory, clinics or hospitals.
Device Story
Piezoelectric sensor captures mechanical vibrations/sounds from patient neck surface during sleep; converts vibrations into small analog voltage signals; output transmitted via DIN 42-802 connector to polysomnography (PSG) recording device. Used in sleep labs, clinics, or hospitals by clinicians; output displayed as snore bursts on PSG monitor; assists physicians in qualifying/quantifying snoring events to diagnose sleep disorders. Benefits include non-invasive monitoring of respiratory sounds to support clinical decision-making.
Clinical Evidence
Bench testing only. Comparative performance testing against the predicate device evaluated dimensions, cable length, visual conditions, and signal output (frequency and amplitude). Results demonstrated that the Maxxi Snore Sensor provides equivalent signals within acceptable ranges (±15% for frequency/amplitude) compared to the predicate.
Technological Characteristics
Piezoelectric sensing element; TPC-ET (thermoplastic elastomer) housing; insulated Teflon wire; DIN 42-802 touch-proof connector. Signal output: AC signal, 0-500 Hz frequency range, 0-500uV amplitude sensitivity. Dimensions: 18mm sensor diameter, 3ft/7ft cable lengths. Complies with ISO 13485:2003, ISO 10993 (biocompatibility), IEC 60601-1.
Indications for Use
Indicated for adult patients undergoing polysomnography in sleep laboratories, clinics, or hospitals to acquire snore bursts for the assessment of sleep-disordered breathing.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 16, 2020
Neurovirtual USA, INC. Eduardo Faria CEO 3303 W Commercial Blvd #100 Fort Lauderdale, Florida 33309
Re: K191095
Trade/Device Name: Maxxi Snore Sensor Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: MNR Dated: December 27, 2019 Received: December 31, 2019
Dear Eduardo Faria:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Michael Ryan Division Director DHT1C: Division of ENT. Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191095
Device Name Maxxi Snore Sensor
#### Indications for Use (Describe)
The Maxxi Snore Sensor is a device intended to acquire snore bursts. It responds to snoring and other sounds in the audio range picked up through the skin and convert them to a small analog voltage that provides an indication of the presence of these sound/vibration bursts. This sensor is intended to be used with polysomnography devices and adult patients. Intended to be use in a sleep laboratory, clinics or hospitals.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------|--|
| <span style="white-space: nowrap;"> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |
| <span style="white-space: nowrap;"> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Neurovirtual. The logo consists of a stylized waveform graphic in shades of orange, followed by the word "NEUROVIRTUAL" in blue. The "O" in Neurovirtual is replaced with a solid orange circle. The waveform graphic is on the left side of the logo, and the text is on the right.
### Section 5 510(k) SUMMARY
A) Submitter's Name: Neurovirtual USA, Inc.
Owner / Operator Registration Number: 9091724 Manufacture Registration Number: 3006125362
- B) Address: 3303 W Commercial Blvd #100 Fort Lauderdale, FL 33309 USA
- C) Phone and Fax Numbers Phone: (786) 693-8200 Fax: (305) 393-8429
- D) Contact Person: Eduardo J. Faria
- E) Preparation Date: December 27, 2019
- F) Classification Name: Common / Usual Name: Breathing frequency monitor Proprietary Name: Maxxi Snore Sensor Product Code: MNR Class: Class II Regulation: 21 CFR 868.2375
- G) Device Description:
MAXXI SNORE SENSOR piezo electric based sensors detect sound and vibration of the snore through skin contact. The sensor is placed on the patient neck surface where it can easily detect the snoring bursts during the sleep study.
The product is offered in 2 different length sizes, 3ft and 7ft.
The sensor is compatible with any recording device with the DIN 42-802 receptacle which is the gold standard for PSG recording machines.
## H) Substantial Equivalence:
The Maxxi Snore Sensor is equivalent with the following products:
| 510(k) Number | Model | Company |
|---------------|----------------|--------------------------------------------------------|
| K941759 | SNORING SENSOR | S.L.P. Ltd.<br>Former EPM INFORMATION<br>SYSTEMS, INC. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Neurovirtual. The logo consists of a stylized waveform graphic in shades of orange on the left, followed by the word "NEUROVIRTUAL" in blue. The "O" in "NEUROVIRTUAL" is replaced with a solid orange circle.
# 1. Indications for Use/Intentions for Use:
| Intention for Use Comparison | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Neurovirtual | S.L.P. Ltd. |
| Maxxi Snore Sensor | Snoring Sensor |
| The Maxxi Snore Sensor is a device intended to<br>acquire snore bursts. It responds to snoring and<br>other sounds in the audio range picked up<br>through the skin and convert them to a small<br>analog voltage that provides an indication of<br>the presence of these sound/vibration bursts.<br>This sensor is intended to be used with<br>polysomnography devices and adult patients. | The Snoring Sensor is a device intended to<br>acquire snore bursts. It responds to snoring and<br>other sounds in the audio range picked up<br>through the skin and convert them to a small<br>analog voltage that provides an indication of<br>the presence of these sound/vibration bursts.<br>This sensor is intended to be used with<br>polysomnography devices and adult patients. |
| Intended to be use in a sleep laboratory, clinics<br>or hospitals. | Intended to be use in a sleep laboratory, clinics<br>or hospitals. |
# 2. Technological Characteristics Comparison:
The predicate devices used to establish substantial equivalence for the Maxxi Snore Sensor are outlined below. This section of this submission will provide a comparison of design, materials, and technical specifications of the Maxxi Snore Sensor to each of the predicate devices stratified by functional modality.
| Technological Characteristics Comparison | | |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | Neurovirtual<br>Maxxi Snore Sensor | S.L.P. Ltd.<br>Snoring Sensor |
| 510(k) Number | K191095 | K941759 |
| Classification | BZQ | BZQ |
| Target population | Adults | Adults |
| Environment | sleep laboratory, clinics or hospitals | sleep laboratory, clinics or hospitals |
| Principle of operation | The mechanism of snoring is<br>vibration of anatomical structures in<br>the pharyngeal airway. The sensor is<br>placed on the side of the patient neck<br>in order to capture the movement or<br>vibration during the snore's events.<br><br>Crystalline materials produce small<br>amounts of electricity when a force is<br>applied that changes their shape in<br>some way. These are called<br>piezoelectric materials. When the<br>small amounts of vibration applied to<br>the Maxxi Snore crystal, a small<br>voltage is produced for a PSG device<br>amplify and display the signal in the<br>sleep study.<br><br>The signal output for snoring is called<br>snore bursts.<br><br>Based on the snore signals during the | The mechanism of snoring is vibration<br>of anatomical structures in the<br>pharyngeal airway. The sensor is<br>placed on the side of the patient neck<br>in order to capture the movement or<br>vibration during the snore's events.<br><br>Crystalline materials produce small<br>amounts of electricity when a force is<br>applied that changes their shape in<br>some way. These are called<br>piezoelectric materials. When the<br>small amounts of vibration applied to<br>the Maxxi Snore crystal, a small<br>voltage is produced for a PSG device<br>amplify and display the signal in the<br>sleep study.<br><br>The signal output for snoring is called<br>snore bursts.<br><br>Based on the snore signals during the |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Neurovirtual. The logo consists of a stylized waveform graphic in shades of orange, followed by the word "NEUROVIRTUAL" in dark blue. The "O" in "NEUROVIRTUAL" is replaced by a solid orange circle. The waveform graphic is positioned to the left of the text, with a horizontal line extending from the waveform to the beginning of the word "NEUROVIRTUAL".
| | sleep study the physician will qualify,<br>quantify and along with the other<br>sleep parameters determine the<br>diagnose of the sleep disorder. | | sleep study the physician will qualify,<br>quantify and along with the other<br>sleep parameters determine the<br>diagnose of the sleep disorder. | |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|--|---------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Parent device | To work in conjunction with a<br>polysomnography device. | | To work in conjunction with a<br>polysomnography device. | |
| Mechanical application | Sensor is placed on patient's neck | | Sensor is placed on patient's neck | |
| Structure | Piezo-electric element covered by<br>TPC plastic injection with a wire to<br>connect to the PSG amplifier. | | Piezo-electric element covered by TPC<br>plastic injection with a wire to connect<br>to the PSG amplifier. | |
| Contact Material | Thermoplastic elastomers (TPC-ETs) | | Thermoplastic elastomers (TPC-ETs) | |
| Wire Material | Insulated Teflon wire | | Insulated Teflon wire | |
| Wire Colors | White | | White | |
| Connector | DIN 42-802<br>touch proof | | DIN 42-802<br>touch proof | |
| Signal output | AC signal | | AC signal | |
| Signal Frequency range | 0-500 HZ | | 0-500 HZ | |
| Amplitude sensitivity<br>range | 0-500uV | | 0-500uV | |
| Sensor diameter | 18mm | | 18mm | |
| Cable length | 3ft and 7ft | | 3ft and 7ft | |
| Package | Plastic bag | | Plastic bag | |
| Image | Image: sensor | | Image: sensor | |
Discussion: The Maxxi Snore and the predicate device Snoring Sensor manufactured by S.L.P Ltd., are substantially equivalent in technology, function, and intended use: both sensors are indicated to provide snoring signal for sleep disordered breathing; both devices use the same method of action; both devices use equivalent material and both devices provide equivalent output signals.
# I) Applied Standards:
In order to reach high quality and effectiveness the Maxxi Snore Sensor is produced in compliance with the quality management standard ISO 13485:2003, "Medical Devices, Quality Management Systems: Requirements for Requlatory Purposes" and FDA GMP "Good Manufacturing Practices".
ISO 10993 for Cytotoxicity, Skin Sensitization and Irritation Studies
IEC 60601-1, IEC 60601-1-1
## J) Performance Testing:
The test performance comparing the Maxxi Snore and the predicate device Snoring Sensor manufactured by S.L.P Ltd was performed and the summary results are show below.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Neurovirtual. The logo consists of a stylized waveform in shades of orange on the left, followed by the word "NEUROVIRTUAL" in blue. The "O" in "NEUROVIRTUAL" is replaced by a solid orange circle.
| Test | Description | Criteria | Results |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Parts<br>Dimensions | Verify if the dimensions of the<br>Piezo electric unit are within<br>the acceptable range and<br>equivalent to the predicate | Allowance: +-10% | The dimensions of the piezo<br>electric unit are within the<br>acceptable criteria as when<br>compared to the predicate<br>device. |
| Cable Length | Verify if the cable length is<br>within the acceptable range<br>and equivalent to the predicate | Expected: 3 ft and 7 ft<br>Allowance: +-5% | The cable length of the sensor is<br>within the acceptable criteria as<br>when compared to the predicate<br>device. |
| Visual Conditions | Verify the visual aspects of the<br>product and equivalency with<br>the predicate. Cable aspects,<br>labeling, flexibility, plastic<br>finishing, and connector<br>conditions. | All aspects must be<br>substantially<br>equivalent to the<br>predicate device. | Maxxi Snore sensor was<br>inspected, and the result is<br>equivalent to the predicate<br>device. |
| Output signals<br>Frequency and<br>Amplitude Tests | The sensor Maxxi Snore and<br>the predicate Snoring Sensor<br>were connected to the same<br>PSG recorder and the acquired<br>data was compared in<br>frequency and amplitude. | Allowance: +-15% | Both sensors acquired equivalent<br>signals within the acceptable<br>range. |
Discussion: The Maxxi Snore and the predicate device Snoring Sensor manufactured by S.L.P Ltd., are substantially equivalent in technology, function, and intended use: both sensors are indicated to provide respiratory signal for sleep disordered breathing; both devices use the same method of action; both devices use equivalent material and both devices provide equivalent output signals.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.