K183559 · Itamar Medical , Ltd. · MNR · Jun 5, 2019 · Anesthesiology
Device Facts
Record ID
K183559
Device Name
WatchPAT ONE
Applicant
Itamar Medical , Ltd.
Product Code
MNR · Anesthesiology
Decision Date
Jun 5, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2375
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The WatchPAT™ONE (WP1) device is a non-invasive home care device for use with patients suspected to have sleep related breathing disorders. The WP1 is a diagnostic aid for the detection of sleep related breathing disorders, sleep staging (Rapid Eye Movement (REM) Sleep, Light Sleep, Deep Sleep and Wake), snoring level and body position. The WP1 generates a peripheral arterial tonometry ("PAT") Respiratory Disturbance Index ("PRDI"), Apnea-Hypopnea index ("PAHI"), Central Apnea-Hypopnea index ("PAHIc"), PAT sleep staging identification (PSTAGES) and snoring level and body position discrete states from an external integrated snoring and body position sensor. The WP1's PSTAGES and snoring level and body position provide supplemental information to its PRDI/PAHI/PAHIc. The WP1's PSTAGES and snoring level and body position are not intended to be used as the sole or primary basis for diagnosing any sleep related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted. PAHIc is indicated for use in patients 17 years and older. All other parameters are indicated for 12 years and older.
Device Story
WatchPAT ONE (WP1) is a non-invasive home-use diagnostic aid for sleep-related breathing disorders. It records peripheral arterial tonometry (PAT), pulse rate, oximetry, actigraphy (wrist motion), snoring, body position, and chest movement. The device consists of a finger PAT probe, wrist-worn controller, and chest sensor. Data is transmitted via Bluetooth to a smartphone app, then to a web server for analysis. The zzzPAT software processes the data to generate PRDI, PAHI, PAHIc, sleep staging (PSTAGES), snoring levels, and body position. Physicians review the output to assist in diagnosing sleep disorders and determining if further assessment is needed. The device is intended for single-use, replacing the reusable predicate's on-device storage and display with cloud-based storage and a mobile app interface.
Clinical Evidence
No clinical data presented. Substantial equivalence is supported by bench testing, software verification and validation, electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and home healthcare environment testing (IEC 60601-1-11).
Technological Characteristics
Ventilatory effort recorder using PAT technology. Components: finger PAT probe, actigraph (3D accelerometer), chest sensor (RESBP), Nordic ARM microprocessor, TI AFE4404 analog front end. Connectivity: Bluetooth (BLE) to mobile app, sFTP to web server. Power: 1.5V AAA battery. Software: zzzPAT algorithm. Biocompatibility: Cytotoxicity, irritation, and sensitization testing performed on housing and wrist strap materials.
Indications for Use
Indicated for patients 12 years and older suspected of sleep-related breathing disorders. PAHIc parameter indicated for patients 17 years and older.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 5, 2019
Itamar Medical, Ltd. % Jonathan Kahan Partner Hogan Lovells US LLP 555 Thirteenth Street, NW Washington, District of Columbia 20004-1109
Re: K183559
Trade/Device Name: WatchPATTM ONE (WP1) Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: MNR Dated: May 3, 2019 Received: May 3, 2019
Dear Jonathan Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Michael Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183559
Device Name WatchPAT™ONE (WP1)
#### Indications for Use (Describe)
The WatchPAT™ONE (WP1) device is a non-invasive home care device for use with patients suspected to have sleep related breathing disorders. The WP1 is a diagnostic aid for the detection of sleep related breathing disorders, sleep staging (Rapid Eye Movement (REM) Sleep, Light Sleep, Deep Sleep and Wake), snoring level and body position. The WP1 generates a peripheral arterial tonometry ("PAT") Respiratory Disturbance Index ("PRDI"), Apnea-Hypopnea index ("PAHI"), Central Apnea-Hypopnea index ("PAHIC"), PAT sleep staging identification (PSTAGES) and snoring level and body position discrete states from an external integrated snoring and body position sensor. The WP1's PSTAGES and snoring level and body position provide supplemental information to its PRD/PAH/PAHIC. The WP1's PSTAGES and snoring level and body position are not intended to be used as the sole or primary basis for diagnosing any sleep related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted. PAHIc is indicated for use in patients 17 years and older. All other parameters are indicated for 12 years and older.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
× | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
- 5.1 Applicant's Name: Itamar Medical Itd. 9 Halamish st. Caesarea 3088900, Israel Tel: +972 4 617 7000 Fax: +972 4 627 5598
- 5.2 Contact Person: Jonathan Kahan, Esq. Hogan Lovells US LLP Columbia Square 555 Thirteenth Street, NW Washington, DC 20004-1109 Tel: (202) 637-5794 Fax: (202) 637-5910 Email: jonathan.kahan@hoganlovells.com
- December 20, 2018 5.3 Date Prepared:
- WatchPAT™ONE (WP1) 5.4 Trade Name:
- 5.5 Common or Usual Name: Ventilatory Effort Recorder
- 5.6 Classification Name: Breathing Frequency Monitor
- 5.7 Medical Specialty: Anesthesiology
- 5.8 Product Code: Ventilatory Effort Recorder, MNR
- 5.9 Device Class: Class II
- 5.10 Regulation Number: 868.2375
- 5.11 Panel: Anesthesiology
#### 5.12 Predicate Device:
Watch-PAT 300 ("WP300") (Itamar Ltd), cleared under K180775; product code MNR (ventilatory effort recorder)
#### 5.13 Intended Use / Indication for Use:
The WatchPAT™ONE (WP1) device is a non-invasive home care device for use with patients suspected to have sleep related breathing disorders. The WP1 is a diagnostic aid for the detection of sleep related breathing disorders, sleep staging (Rapid Eye Movement (REM) Sleep, Light Sleep, Deep Sleep and Wake), snoring level and body position. The WP1 generates a peripheral arterial tonometry ("PAT") Respiratory Disturbance Index ("PRDI"), Apnea-Hypopnea index ("PAHI"), Central Apnea-Hypopnea index ("PAHIc"), PAT sleep staging identification (PSTAGES) and snoring level and body position discrete states from an external integrated snoring and body position sensor. The WP1's PSTAGES and snoring level and body position provide
{5}------------------------------------------------
supplemental information to its PRDI/PAHI/PAHIc. The WP1's PSTAGES and snoring level and body position are not intended to be used as the sole or primary basis for diagnosing any sleep related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted.
PAHIc is indicated for use in patients 17 years and older. All other parameters are indicated for 12 years and older.
## 5.14 Device Description:
The WatchPAT™ONE (WP1) is a non-invasive home care device intended for use with patients suspected to have sleep related breathing disorders. The WP1 generates a PAT respiratory disturbance index (PRDI), Apnea Hypopnea Index (PAHI), Central Apnea Hypopnea Index (PAHIc), sleep stages (PSTAGES - REM Sleep, Light Sleep, Deep Sleep and Wake) and snoring level and body position discrete states.
The WP1 device consists of the following: (1) same unified finger PAT probe that was used in WP300 to measure the PAT and oximeter signals; (2) same Actigraph which provides a signal that is used to determine periods of sleep/wake based on the motion of the wrist; (3) same electronics, except for an antenna added in the WP1's PCB to support BLE communication; (4) same chest sensor (RESBP) to measure snoring level, body position and chest movements; and (5) same zzzPAT Software and algorithm to analyze the data.
The subject WP1 is a modified version of the predicate WP300 device (K180775). The proposed modifications mainly include: transition from a reusable device to a single use device; shifting the display and user interface from the device to a mobile application; and shifting of the data storage from main device to a web server. None of these changes alter the fundamental technology or its principles of operation. Moreover, the WP1 maintains the same capabilities of the cleared WP300 (K180775) and provides the user with the same output information.
## 5.15 Substantial Equivalence:
## Intended Use/Indications for Use
The subject device WP1 is identical to the cleared WP300 in its intended use and indications for use, except for the minor modification of omitting the word 'optional' from the use of snoring and body position sensor (RESBP sensor).
## Comparison of Technological Characteristics and principle of operation
The technological characteristics and principles of operation of the subject device are similar to the predicate device. The subject WP1, like its predicate, is a ventilatory effort recorder that utilizes PAT technology. In both systems, the controller part of the device is worn on the wrist and records the PAT signal and blood oxygen saturation levels by a finger-mounted probe based on an optical plethysmographic method, the wrist motion from an embedded actigraphy and snoring level, body position states and chest movements from the chest sensor (RESBP).
{6}------------------------------------------------
The WP1 mainly differs from the WP300 by combining the key components into an undetachable monolith device. A mobile application also replaced the predicate WP300's user interface and communication. The WP1's monolithic package uses the same hardware as the predicate WP300 to measure patient data. In addition, the user interface of the mobile application was designed to provide similar display and input as the WP300. Although the patient data is transmitted and stored on a Web Server, the integrity of the patient data can be verified and validated. Furthermore, the cybersecurity risks were addressed according to FDA's cybersecurity guidance document'. As such, the differences in technological characteristics and principle of operation do not raise different questions of safety or effectiveness.
## Performance Testing
The following consensus standards were used to evaluate the predicate device:
- . Bench test was conducted to show that the data acquired on the WP1 is identical to the data downloaded by zzzPAT at the physician computer.
- The WP1 was tested to ensure that the device complies with the Federal ● Communications Commission (FCC) rules.
- Electrical safety testing per IEC 60601-1:2005 + CORR.1 (2006) + ● CORR.2 (2007) and AM1:2012
- Electromagnetic compatibility testing per IEC 60601-1-2:2014
- Home healthcare environment per 60601-1-11:2015
- Cytotoxicity, irritation and sensitization testing of the new housing and wrist strap materials.
- Software verification and validation testing was performed to demonstrate that the software in the subject device meets design specifications.
The testing above demonstrated that the WP1 has substantially equivalent performance as its predicate.
## Summary
Based on the comparison of intended use, indications for use, technological characteristics and performance testing, Itamar Ltd. believes that the WP1 System is substantially equivalent to its predicate.
A summary of comparison between the subject and predicate devices is provided below.
<sup>1</sup> FDA quidance document title, "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices", issued on October 2, 2014
{7}------------------------------------------------
| | Predicate Device: | Subject Device: | Comparison |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Itamar Medical's Watch-<br>PAT300 (K180775) | Itamar Medical's WP1 | |
| Intended Use | The Watch-PAT300 (WP300)<br>device is a non-invasive home<br>care device for use with patients<br>suspected to have sleep related<br>breathing disorders. The WP300<br>is a diagnostic aid for the<br>detection of sleep related<br>breathing disorders, sleep<br>staging (Rapid Eye Movement<br>(REM) Sleep, Light Sleep, Deep<br>Sleep and Wake), snoring level<br>and body position. The WP300<br>generates a peripheral arterial<br>tonometry ("PAT") Respiratory<br>Disturbance Index ("PRDI"),<br>Apnea-Hypopnea index ("PAHI"),<br>Central Apnea-Hypopnea index<br>("PAHIc"), PAT sleep staging<br>identification (PSTAGES) and<br>optional snoring level and body<br>position discrete states from an<br>external integrated snoring and<br>body position sensor. The<br>WP300's PSTAGES and snoring<br>level and body position provide<br>supplemental information to its<br>PRDI/PAHI/PAHIc. The WP300's<br>PSTAGES and snoring level and<br>body position are not intended to<br>be used as the sole or primary<br>basis for diagnosing any sleep<br>related breathing disorder,<br>prescribing treatment, or<br>determining whether additional<br>diagnostic assessment is<br>warranted.<br>PAHIc is indicated for use in<br>patients 17 years and older. All<br>other parameters are indicated<br>for 12 years and older | The WatchPAT™ONE (WP1)<br>device is a non-invasive home<br>care device for use with patients<br>suspected to have sleep related<br>breathing disorders. The WP1 is<br>a diagnostic aid for the detection<br>of sleep related breathing<br>disorders, sleep staging (Rapid<br>Eye Movement (REM) Sleep,<br>Light Sleep, Deep Sleep and<br>Wake), snoring level and body<br>position. The WP1 generates a<br>peripheral arterial tonometry<br>("PAT") Respiratory Disturbance<br>Index ("PRDI"), Apnea-<br>Hypopnea index ("PAHI"),<br>Central Apnea-Hypopnea index<br>("PAHIc"), PAT sleep staging<br>identification (PSTAGES) and<br>snoring level and body position<br>discrete states from an external<br>integrated snoring and body<br>position sensor. The WP1's<br>PSTAGES and snoring level and<br>body position provide<br>supplemental information to its<br>PRDI/PAHI/PAHIc. The WP1's<br>PSTAGES and snoring level and<br>body position are not intended to<br>be used as the sole or primary<br>basis for diagnosing any sleep<br>related breathing disorder,<br>prescribing treatment, or<br>determining whether additional<br>diagnostic assessment is<br>warranted.<br>PAHIc is indicated for use in<br>patients 17 years and older. All<br>other parameters are indicated<br>for 12 years and older. | ldentical except for<br>device name and<br>omission of the<br>word 'optional' |
| Channels | PAT, Pulse rate, Oximetry,<br>Actigraphy, Snoring, Body<br>Position, Chest Movement | PAT, Pulse rate, Oximetry,<br>Actigraphy, Snoring, Body<br>Position, Chest Movement | Identical |
| Intended Use<br>Environment | Home Use | Home Use | Identical |
| Input uPAT<br>Probe | Designed to use Itamar<br>proprietary probe only | Designed to use Itamar<br>proprietary probe only | Identical |
| Analysis output | • pRDI<br>• pAHI<br>• pAHIc<br>· Sleep stages<br>· Snoring level<br>· Body position discrete states | · pRDI<br>• pAHI<br>· pAHIc<br>· Sleep stages<br>· Snoring level<br>· Body position discrete states | Identical |
| | Predicate Device:<br>Itamar Medical's Watch-<br>PAT300 (K180775) | Subject Device:<br>Itamar Medical's WP1 | Comparison |
| Components | • uPAT finger probe<br>• actigraph<br>• Controller<br>• Microphone<br>• Accelerometer<br>• ZzzPAT software<br>• Chest sensor (optional)<br>• External Tamper-Proof<br>Bracelet (optional) | • uPAT finger probe<br>• actigraph<br>• Controller<br>• Microphone<br>• Accelerometer<br>• ZzzPAT software<br>• Chest sensor | Chest sensor was<br>already cleared in<br>the predicate.<br>Optional tamper-<br>proof bracelet for<br>identification of a<br>patient by means of<br>electronic bracelet<br>is not available in<br>WP1. |
| Sensors<br>Placement | Wrist, finger and chest (optional) | Wrist, finger and chest | Identical |
| Analysis<br>Software | zzzPAT | zzzPAT | Identical |
| Power Supply | One OTS 1.5V Alkaline AAA<br>battery OR<br>One rechargeable AAA 1.2V<br>Nickel-metal hydride battery<br>rechargeable (NiMH) battery | One OTS 1.5V Alkaline AAA<br>battery | Single use battery<br>only |
| Data Access<br>Method | USB using FTDI | Internet access using sFTP<br>protocol | Same data<br>delivered to the<br>analysis software<br>zzzPAT |
| Data Collection | • Sensors connect direct to main<br>device.<br>• Study data is stored on the<br>main device and then<br>uploaded via USB cable upon<br>return of the device. | • Sensors connect direct to<br>main device.<br>• Study data is wirelessly<br>transferred (Bluetooth) from<br>main device to mobile phone<br>and from the mobile phone to<br>a storage on a web server,<br>over the Internet. | Same data<br>delivered to the<br>analysis software<br>zzzPAT |
| Data Storage | Flash | Web Server | Data is stored in<br>web server instead<br>of on-board flash<br>memory. |
| Recording time | Approx. 10 hours | Approx. 10 hours | Identical |
| Motion Probe | 1 axis (out of 3D accelerometer) | 1 axis (out of 3D accelerometer) | Identical |
| Analog Front<br>End | AFE4404 (by TI) | AFE4404 (by TI) | Identical |
| Microprocessor | Nordic ARM | Nordic ARM | Identical |
| Sample<br>Resolution | PAT, Actigraphy, Snore: 12 bits<br>Oximetry: 1%<br>Body Position: 5 discrete states<br>(supine, prone, right, left and sit)<br>Chest Movements – 12 bits x 3<br>axis | PAT, Actigraphy, Snore: 12 bits<br>Oximetry: 1%<br>Body Position: 5 discrete states<br>(supine, prone, right, left and sit)<br>Chest Movements - 12 bits x 3<br>axis | Identical |
| Operating<br>Voltage | 3.3 V | 3.3 V | Identical |
| User Interface<br>Display/<br>Indicators | Via Main device: OLED | Via Smartphone User Interface<br>Via Main device (LED) | Similar user display<br>and input |
{8}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.