← Product Code [MNR](/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR) · K180001

# Disposable BreathSensor Adult, Disposable BreathSensor Small Adult / Child, Disposable BreathSensor Preemie, Disposable BreathSensor Infant (K180001)

_Embla Systems · MNR · Sep 27, 2019 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR/K180001

## Device Facts

- **Applicant:** Embla Systems
- **Product Code:** [MNR](/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR.md)
- **Decision Date:** Sep 27, 2019
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 868.2375
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Pediatric

## Indications for Use

The Embla BreathSensor provides a qualitative air flow signal by oral and/or nasal temperature sensitive resistive components for recording onto a recording system in support of airflow analysis and sleep studies. The Embla BreathSensor does not provide any diagnostic conclusion about the patient's condition to the user. The Embla BreathSensor user is a qualified medical practitioner in a hospital environment who will exercise their professional judgment in using this information.

## Device Story

Embla BreathSensors are disposable, non-sterile thermistor-based sensors used in sleep studies (polysomnography). Sensors consist of carbon ink temperature-sensitive elements on a polyester substrate, with silver ink interfaces. Applied between the upper lip and nostrils, they detect temperature changes between ambient and exhaled air to provide a qualitative airflow signal. An interface cable connects the sensor to standard PSG recording equipment; active cable versions perform signal filtering and convert thermistor resistance changes to thermocouple-like signals. Used by medical practitioners in hospital settings to support airflow analysis; the device provides no diagnostic conclusions. Benefits include non-invasive, qualitative monitoring of patient respiration during sleep.

## Clinical Evidence

Bench testing only. Evaluations included verification of dimensions, material specifications, biocompatibility (cytotoxicity, sensitization, irritation), base resistance, temperature sensitivity, qualitative airflow response, and electromagnetic compatibility (IEC 60601-1-2:2014).

## Technological Characteristics

Thermistor-based airflow sensor. Materials: carbon ink (sensing element), silver ink (conductive interface), polyester substrate, and adhesive tape (Micro-Pore/Micro-Foam). Dimensions vary by model (Adult, Small Adult/Child). Connectivity: wired via interface cable to PSG amplifier. Non-sterile, single-patient use. Active interface cables provide signal filtering and thermistor-to-thermocouple signal conversion using internal battery power.

## Regulatory Identification

A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.

## Predicate Devices

- EdenTrace Airflow 3171, Sleep Lab Airflow 3170 ([K913749](/device/K913749.md))
- SleepSense Sleep Disorder Sensors ([K042253](/device/K042253.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

September 27, 2019

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Embla Systems (Embla) Sanjay Mehta Director, Quality Assurance & Regulatory Affairs 1 Hines Road, Suite 202 Kanata, ON, CA K2K 3C7

Re: K180001

Trade/Device Name: Embla BreathSensors Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing frequency monitor Regulatory Class: Class II Product Code: MNR Dated: August 26, 2019 Received: August 29, 2019

Dear Sanjay Mehta:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Michael Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K180001

Device Name Embla BreathSensors

#### Indications for Use (Describe)

The Embla BreathSensor provides a qualitative air flow signal by oral and/or nasal temperature sensitive resistive components for recording onto a recording system in support of airflow analysis and sleep studies.

The Embla BreathSensor does not provide any diagnostic conclusion about the patient's condition to the user.

The Embla BreathSensor user is a qualified medical practitioner in a hospital environment only who will exercise their professional judgment in using this information.

Type of Use (Select one or both, as applicable)

| [X] Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|--------------------------------------------------|-----------------------------------------------|
|--------------------------------------------------|-----------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services  

 Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA) Staff  

 PRAStaff@fda.hhs.gov
 

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*

| FORM FDA 3881 (7/17) | Page 1 of 1 | PSC Publishing Services (301)443-6740 EF |
|----------------------|-------------|------------------------------------------|
|----------------------|-------------|------------------------------------------|

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Natus Neurology. The word "natus" is written in a teal sans-serif font, with a registered trademark symbol next to the "s". Below "natus" is the word "neurology" in a smaller, italicized, black sans-serif font. The logo is simple and clean, with a focus on the company name.

### 510(K): K180001, EMBLA BREATHSENSORS

# K180001 510(k) Summary

510(k) Summary This summary is provided to support the 510(k) pre-market notification for the Embla BreathSensors.

| Company Name:          | Embla Systems (Embla)<br>1 Hines Road, Suite 202<br>Kanata, ON, CA, K2K 3C7                                                                                  |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact:       | Mr. Sanjay Mehta<br>Director, Quality Assurance & Regulatory Affairs<br>Phone: (905) 287-5055<br>Email: Sanjay.Mehta@natus.com                               |
| Date Summary Prepared: | September 27, 2019                                                                                                                                           |
| Trade Name:            | Embla BreathSensors                                                                                                                                          |
| Common Name:           | Thermistor for Air Flow                                                                                                                                      |
| Classification Name:   | Breathing Frequency Monitor<br>21 CFR 868.2375<br>Product Code: MNR<br>Class II                                                                              |
| Predicate Device:      | K913749 EdenTrace Airflow 3171, Sleep Lab Airflow 3170<br>EdenTec Corporation (Primary predicate)<br>K042253 SleepSense Sleep Disorder Sensors<br>S.L.P. Ltd |

#### Product Description

The Embla BreathSensors (Model 970 Series) are airflow thermistors used as an accessory to a polysomnogram (PSG) or sleep study recording system. The BreathSensor family is comprised of two different variations: Models 971 Adult and Model 974 Small Adult/Child.

The BreathSensors are for use with any standard PSG amplifier or recorder. An interface cable connects the BreathSensor to the polysomnography equipment.

Embla BreathSensors are designed to provide a qualitative measure of respirations (flow) from the mouth and/or nose. The user selects the appropriate size of BreathSensor based on their clinical judgment.

The nasal and/or oral temperature sensor elements are carbon ink placed on a substrate. The carbon ink is the temperature sensitive material. Silver ink interfaces with the carbon ink temperature sensing elements to provide a connection to the interface cable.

The Embla BreathSensors are non-sterile, single patient use, disposable devices.

#### Intended Use of the Device

The Embla BreathSensor provides a qualitative air flow signal by oral and/or nasal temperature resistive resistive resistive components for recording onto a recording system in support of airflow analysis and sleep studies. The Embla BreathSensor does not provide any diagnostic conclusion about the patient's condition to the user. The Embla BreathSensor user is a qualified medical practitioner in a hospital environment who will exercise their professional judgment in using this information.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Natus Neurology. The word "natus" is in a teal sans-serif font, and the word "neurology" is in a black italicized sans-serif font. The logo is simple and modern, and it is likely used to represent the company's brand.

# Summary of Technological Characteristics

The BreathSensors are single patient use, disposable sensors applied to a patient below the nose, (above the mouth), to support a qualitative indication of airflow from the nostrils and/or mouth. The BreathSensors are a form of thermistor, indicating airflow by temperature change. The accessory interface cables connect between the BreathSensor and user equipment for display and recording by the user's equipment.

Table 1 provides a side-by-side comparison of the BreathSensors to the predicate device applied to support this premarket notification.

| Table 1: Substantial Equivalence Technical Characteristics |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature                                                    | BreathSensors<br>Under Review                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | EdenTrace Airflow<br>3171, Sleep Lab<br>Airflow 3170<br>(Primary Predicate<br>K913749)                                                                                                                                                                                                          | S.L.P. Ltd<br>SleepSense Sleep<br>Disorder Sensors<br>(Predicate K042253)                                                                                                                                                                                                                                                                                                                                       | Equivalence<br>Comments                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Intended Use                                               | Disposable temperature<br>sensitive sensor to<br>support recording<br>airflow.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Disposable temperature<br>sensitive sensor to<br>support recording<br>airflow.                                                                                                                                                                                                                  | Disposable temperature<br>sensitive sensor to<br>support recording<br>airflow.                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Indications for<br>Use                                     | The Embla<br>BreathSensor provides a<br>qualitative air flow<br>signal by oral and/or<br>nasal temperature<br>sensitive resistive<br>components for<br>recording onto a<br>recording system in<br>support of airflow<br>analysis and sleep<br>studies.<br>The Embla<br>BreathSensor does not<br>provide any diagnostic<br>conclusion about the<br>patient's condition to<br>the user.<br>The Embla<br>BreathSensor user is a<br>qualified medical<br>practitioner who will<br>exercise their<br>professional judgment<br>in using this<br>information.<br>Rx only | The airflow signal is<br>provided by oral and/ or<br>nasal temperature<br>sensitive components.<br>Accessory single use<br>airflow sensors are used<br>to provide resistance<br>changes as affected by<br>temperature changes in<br>airflow representing<br>inhaled and exhaled air.<br>Rx only | SleepSense sensors<br>provide a qualitative<br>measure of a patient's<br>physiological<br>parameters for<br>recording onto an FDA-<br>cleared data acquisition<br>system. Their target<br>population: Children<br>and adult patients who<br>are screened during<br>sleep disorder studies.<br>Their environment of<br>use is usually at a sleep<br>laboratory or sometimes<br>at the patient's home.<br>Rx Only | Both the Embla<br>BreathSensors under<br>review and primary<br>predicate Airflow Sensors<br>provide oral and/or nasal<br>temperature sensitive<br>components.<br>The population of use of<br>the Embla BreathSensors<br>under review and primary<br>predicate EdenTec Airflow<br>Sensors are similar.<br>The variations of the<br>primary predicate Airflow<br>Sensors and Embla<br>BreathSensors under<br>review are identified by a<br>Model variation name<br>(Adult, Small Adult /<br>Child).<br>All three devices apply<br>interface cables as<br>applicable to condition the<br>signal for medical<br>professional user's<br>recording equipment.<br>All Rx Only<br>The Embla BreathSensors<br>under review provide<br>additional detail with<br>regard to the indications. |
| Contraindication                                           | Contraindicated for use<br>on patients who exhibit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | None                                                                                                                                                                                                                                                                                            | None                                                                                                                                                                                                                                                                                                                                                                                                            | The allergic reaction to<br>tape adds clarification on                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Table 1: Substantial Equivalence Technical Characteristics |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Feature                                                    | BreathSensors<br>Under Review                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | EdenTrace Airflow<br>3171, Sleep Lab<br>Airflow 3170<br>(Primary Predicate<br>K913749)                                                                                                                                                                                                          | S.L.P. Ltd<br>SleepSense Sleep<br>Disorder Sensors<br>(Predicate K042253)                                                                                                                                                                                                                                                                                                                                       | Equivalence<br>Comments                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                                            | allergic reactions to<br>adhesive tape.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                 | use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Environment of<br>Use                                      | Hospitals                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Hospitals, clinics and<br>home.                                                                                                                                                                                                                                                                 | Hospitals, clinics (sleep<br>labs) and home.                                                                                                                                                                                                                                                                                                                                                                    | The Embla Breath Sensors<br>have a hospital<br>environment of use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Sensor<br>variations                                       | BreathSensor<br>Adult, Small Adult /<br>Child                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Airflow Sensor<br>Adult, Small Adult,<br>Child, Infant, Preemie                                                                                                                                                                                                                                 | Disposable Thermal<br>Flow Sensor<br>Adult, Child, Extra<br>Small                                                                                                                                                                                                                                                                                                                                               | Similar<br>The medical professional<br>selects the size of<br>BreathSensor that best fits<br>their patient with regard to<br>placement of nasal and oral<br>BreathSensor elements.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Temperature<br>Sensing<br>Elements                         | Breath Sensor<br>Nasal and Oral: Adult,<br>Small Adult / Child,                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Airflow Sensor<br>Nasal and Oral: Adult,<br>Small Adult, Child,<br>Infant<br>Oral only: Preemie                                                                                                                                                                                                 | Disposable Flow Sensor<br>Nasal and Oral, Oral<br>only options: Adult,<br>Child, Extra Small                                                                                                                                                                                                                                                                                                                    | Similar<br>The primary predicate<br>variation names of Small<br>Adult and Child are<br>renamed to one variation.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| BreathSensor:<br>Single patient<br>use, disposable         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Qualitative<br>temperature<br>output                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                             | Same<br>The signal output from the<br>BreathSensor is a<br>qualitative signal based<br>upon the difference in<br>ambient and exhaled air.<br>The medical professional<br>will adjust signal amplitude<br>on the recording<br>equipment.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Duration of use                                            | Duration of a sleep<br>study, anticipated to be<br>up to 12 hours                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Duration of a sleep<br>study                                                                                                                                                                                                                                                                    | Duration of a sleep<br>study                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Patient applied<br>location                                | Between upper lip and<br>nostrils                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Between upper lip and<br>nostrils                                                                                                                                                                                                                                                               | Between upper lip and<br>nostrils                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Sensor materials                                           | Carbon Ink<br>Silver Ink<br>Polyester<br>Tape                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Carbon Ink<br>Silver Ink<br>Polyester (Mylar)<br>Tape                                                                                                                                                                                                                                           | Unknown                                                                                                                                                                                                                                                                                                                                                                                                         | Carbon ink is the<br>temperature sensitive<br>material. The conductive<br>silver ink conducts the<br>signal to the interface cable<br>input connector. The<br>carbon and silver ink are<br>placed on a polyester<br>substrate. The polyester<br>substrate and ink are                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Table 1: Substantial Equivalence Technical Characteristics |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Feature                                                    | BreathSensors<br>Under Review                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | EdenTrace Airflow<br>3171, Sleep Lab<br>Airflow 3170<br>(Primary Predicate<br>K913749)                                                                                                                                                                                                          | S.L.P. Ltd<br>SleepSense Sleep<br>Disorder Sensors<br>(Predicate K042253)                                                                                                                                                                                                                                                                                                                                       | Equivalence<br>Comments                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                 | covered by tape.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Patient contact<br>material                                | Micro-Pore,<br>Micro-Foam Tape                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Micro-Pore Tape                                                                                                                                                                                                                                                                                 | Unknown                                                                                                                                                                                                                                                                                                                                                                                                         | Equivalent<br>Final finished form<br>BreathSensors have been<br>evaluated for cytotoxicity,<br>irritation and sensitization<br>with passing results.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Sterility                                                  | Supplied non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Supplied non-sterile                                                                                                                                                                                                                                                                            | Supplied non-sterile                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Accessory<br>Interface cables                              | Yes<br>Active                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes<br>Passive and active                                                                                                                                                                                                                                                                       | Yes<br>Passive and active                                                                                                                                                                                                                                                                                                                                                                                       | Similar<br>Active interface cable<br>variations are available.<br>The interface cables<br>connect to the medial<br>professional's amplifier,<br>recording equipment,<br>supporting transfer of the<br>BreathSensor thermistor<br>signal to their equipment.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Accessory<br>Interface cable:<br>Active Cable              | Yes<br>Battery powered,<br>filtering and conversion<br>of sensor thermistor to<br>thermocouple sensor<br>signal.                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Yes<br>Battery Powered,<br>filtering and conversion<br>of sensor thermistor to<br>thermocouple sensor<br>signal.                                                                                                                                                                                | Yes<br>Battery Powered,<br>filtering and conversion<br>of sensor thermistor to<br>thermocouple sensor<br>signal.                                                                                                                                                                                                                                                                                                | Equivalent<br>The active interface cables<br>have sensor and user<br>recording equipment<br>compatible connectors and<br>are battery powered<br>(internal).<br>The active interface cables<br>are battery-powered by an<br>internal non-user<br>replaceable battery. The<br>active interface cables<br>provide filtering and<br>conversion of the<br>thermistor sensor signal to<br>thermocouple sensor signal<br>for presentation to the<br>user's recording equipment                                                                                                                                                                                                                                                                                                           |
| Accessory<br>interface cable:<br>Reusable                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Connection to<br>user's<br>equipment                       | Yes<br>Accessory Interface<br>Cable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes<br>Accessory Interface<br>Cable                                                                                                                                                                                                                                                             | Yes<br>Accessory Interface<br>Cable                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Interface cable<br>variations                              | Active                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Active and passive                                                                                                                                                                                                                                                                              | Active and passive                                                                                                                                                                                                                                                                                                                                                                                              | Similar<br>The active interface cables<br>interface the BreathSensor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Table 1: Substantial Equivalence Technical Characteristics |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Feature                                                    | BreathSensors<br>Under Review                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | EdenTrace Airflow<br>3171, Sleep Lab<br>Airflow 3170<br>(Primary Predicate<br>K913749)                                                                                                                                                                                                          | S.L.P. Ltd<br>SleepSense Sleep<br>Disorder Sensors<br>(Predicate K042253)                                                                                                                                                                                                                                                                                                                                       | Equivalence<br>Comments                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                 | to recording equipment.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Natus Neurology. The word "natus" is in a teal color and in a larger font. Below that, the word "neurology" is in a smaller, black font. There is a registered trademark symbol to the right of the word "natus".

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Natus Neurology. The word "natus" is in a teal color and in a larger font. Below that, the word "neurology" is in a smaller, black font. There is a registered trademark symbol to the right of the word "natus".

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Natus Neurology. The word "natus" is written in a teal, sans-serif font. Below that, the word "neurology" is written in a black, sans-serif font. The logo is simple and modern.

As noted, the differences do not affect substantial equivalence of safety and effectiveness:

- A contraindication is added for patients who may have allergic reactions to adhesive tape. .
- . The BreathSensors are supported by available accessory interface cable(s).

## Performance Tests to Demonstrate Substantial Equivalency

To establish the technical equivalency of the BreathSensors, evaluations were conducted to confirm compliance with performance requirements. Table 2 summarizes evaluations and results supporting substantial equivalence.

| Table 2: Summary of Evaluations Supporting Substantial Equivalence |                                                                                                                       |        |
|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|--------|
| Feature                                                            | Evaluation Summary                                                                                                    | Result |
| Dimensions                                                         | Verification of dimensions.                                                                                           | Pass   |
| Labels and Labeling                                                | Verification of label and labeling content for variations.                                                            | Pass   |
| Material                                                           | Verification of specified materials.                                                                                  | Pass   |
| Biocompatibility of patient<br>contact materials                   | Supported with final finished form cytotoxicity, sensitization and irritation<br>evaluations.                         | Pass   |
| Resistance                                                         | Verify BreathSensor base resistance.                                                                                  | Pass   |
| Temperature Sensitivity                                            | Verify minimum temperature sensitivity of resistance change per ℃.                                                    | Pass   |
| Performance Testing                                                | Comparison with predicate Disposable Flow Sensor, evaluation of qualitative airflow<br>response with Interface Cable. | Pass   |
| Electromagnetic<br>compatibility                                   | BreathSensor and Interface Cable compliance with IEC 60601-1-2:2014,<br>4th edition                                   | Pass   |

#### Conclusion

The Embla BreathSensors meet performance requirements. The intended use and technology of the Embla BreathSensors are similiar to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR/K180001](https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR/K180001)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
