← Product Code [MNR](/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR) · K151361

# Nox RIP Belts & Cables (K151361)

_Nox Medical Ehf · MNR · Nov 6, 2015 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR/K151361

## Device Facts

- **Applicant:** Nox Medical Ehf
- **Product Code:** [MNR](/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR.md)
- **Decision Date:** Nov 6, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.2375
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Pediatric

## Indications for Use

The Nox RIP Belts are intended for measuring of respiratory effort signals. They function as accessories for sleep/polysomnography (PSG) systems. The Nox RIP Belts are indicated for use on patients greater than 2 years of age. The intended environments are hospitals, institutions, sleep clinics, or other test environments, including the patient's home. The RIP Belt Cables are intended to interconnect Nox RIP belts (respiratory effort sensors) and Nox sleep devices, to allow measuring of respiratory effort signals. The RIP Belt Cables are indicated for use on patients greater than 2 years of age. The intended environments are hospitals, institutions, sleep clinics, or other test environments, including the patient's home. The Third Party RIP Belt Cables are intended to allow measuring of respiratory effort signals by interconnecting Nox RIP belts (respiratory effort sensors) and sleep devices with oscillation circuitry capable of measuring inductance between 1 and 5 µH. The Third Party RIP Belt Cables are indicated for use on patients greater than 2 years of age. The intended environments are hospitals, institutions, sleep clinics, or other test environments, including the patient's home.

## Device Story

Nox RIP Belts and Cables are passive accessories for sleep/polysomnography (PSG) systems; they measure respiratory effort via Respiratory Inductance Plethysmography (RIP) technology. Belts are placed around patient thorax and abdomen; cables interconnect these sensors to sleep recorders. Third-party cables interface with recorders having oscillation circuitry (1-5 µH inductance). Used in hospitals, clinics, or homes; operated by healthcare professionals. Devices provide raw respiratory effort signals to the recorder; clinicians use these signals to assess respiratory effort, aiding in sleep disorder diagnosis. Benefits include non-invasive monitoring of respiratory effort during sleep.

## Clinical Evidence

Bench testing only. Verification and validation performed per product requirements. Testing included signal integrity (SNR, range, bandwidth, linearity), EMC (IEC 60601-1-2), and electrical safety (AAMI/ANSI/ES 60601-1). Usability testing conducted per AAMI/ANSI/IEC 62366. Signal integrity compared to predicate demonstrated clinical equivalence.

## Technological Characteristics

Passive respiratory effort sensors using RIP technology. Materials: Polyester/Dorlastan elastic, tin-plated copper wire, ABS/PC plastic, TPE/TPE-ET strain relief, gold-plated stainless steel snaps. Dimensions/form factor: wearable belts and cables. Connectivity: wired connection to sleep recorders. Standards: AAMI/ANSI/ES 60601-1, IEC 60601-1-2, ISO 14971, ISO 15223-1, AAMI/ANSI/IEC 62366, 21 CFR 898.

## Regulatory Identification

A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.

## Predicate Devices

- QDC-PRO AND NOX-RIP ([K124062](/device/K124062.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 6, 2015

Nox Medical ehf C/O Kolbrun E Ottosdottir Director of Quality & Regulatory Katrinartuni 2 IS-105 Reykjavik ICELAND

Re: K151361

Trade/Device Name: Nox RIP Belts and Nox RIP Belt Cables Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing frequency monitor Regulatory Class: Class II Product Code: MNR Dated: October 2, 2015 Received: October 5, 2015

Dear Mr. Ottosdottir:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

*Tejashri Purohit-Sheth, M.D.*

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indication for Use

510(k) Number (if known): K151361

Device Name: Nox RIP Belts

Indications for Use (Describe):

The Nox RIP Belts are intended for measuring of respiratory effort signals. They function as accessories for sleep/polysomnography (PSG) systems.

The Nox RIP Belts are indicated for use on patients greater than 2 years of age.

The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments, including the patient 's home.

Type of Use ( Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

# Indication for Use

510(k) Number (if known): K151361

Device Name: RIP Belt Cables

Indications for Use (Describe):

The RIP Belt Cables are intended to interconnect Nox RIP belts (respiratory effort sensors) and Nox sleep devices, to allow measuring of respiratory effort signals.

The RIP Belt Cables are indicated for use on patients greater than 2 years of age.

The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments, including the patient´s home.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{4}------------------------------------------------

# Indication for Use

510(k) Number (if known): K151361

Device Name: Third Party RIP Belt Cables

Indications for Use (Describe):

The Third Party RIP Belt Cables are intended to allow measuring of respiratory effort signals by interconnecting Nox RIP belts (respiratory effort sensors) and sleep devices with oscillation circuitry capable of measuring inductance between 1 and 5 µH.

The Third Party RIP Belt Cables are indicated for use on patients greater than 2 years of age.

The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments, including the patient's home.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{5}------------------------------------------------

## 510(k) Summary

### Submitter

Nox Medical ehf Katrinartuni 2 IS-105 Reykjavik Iceland Tel: +354 570 7170

Establishment Registration Number: 3007389703

#### Contact person

Kolbrun E Ottosdottir Katrinartuni 2 IS-105 Reykjavik Iceland Email: keo@noxmedical.com Direct tel: +354 570 7173

## Preparation Date

06 November 2015

## Device Details

| Trade Name:              | Nox RIP Belts                  | Nox RIP Belt Cables         |
|--------------------------|--------------------------------|-----------------------------|
| Product Groups           |                                |                             |
|                          | RIP Belt Cables                | Third Party RIP Belt Cables |
| Common Name:             | Respiratory Effort<br>Sensor   | Respiratory Effort Sensor   |
| Classification<br>Name:  | Ventilatory Effort<br>Recorder | Ventilatory Effort Recorder |
| Regulation<br>Number:    | 868.2375                       | 868.2375                    |
| Product Code:            | MNR                            | MNR                         |
| Device Class             | Class II                       | Class II                    |
| Classification<br>Panel: | Anesthesiology                 | Anesthesiology              |

{6}------------------------------------------------

#### Predicate Device

QDC-PRO AND NOX-RIP from Nox Medical ehf Product Code: MNR 510(k) Number: K124062

This predicate has not been subject to a design-related recall.

Note1: Please note that the Nox RIP Belts will be regarded as a subsystem of the predicate, QDC-PRO AND NOX-RIP. The QDC-PRO AND NOX-RIP includes a recorder and several accessories but only the RIP belts will be addressed in the comparison between the new device and the predicate.

Note2: Please note that the Nox RIP Belt Cables will be regarded as a subsystem of the predicate, QDC-PRO AND NOX-RIP. The QDC-PRO AND NOX-RIP includes a recorder and several accessories but only the RIP belt cable will be addressed in the comparison between the new device and the predicate.

{7}------------------------------------------------

### Device Description

The Nox RIP Belts are respiratory effort sensors that are intended to function as an accessory with sleep/polysomnography (PSG) systems. The RIP Belts measure respiratory effort signals based on Respiratory Inductance Plethysmography (RIP) technology, which is the gold standard technology for respiratory effort belts.

Two RIP belts are used to measure the respiratory effort of the patient. One belt is placed around the patient's abdomen and the other around the patient's thorax. Both abdomen and thorax belts are identical.

The Nox RIP Belt Cables are used to connect between the respiratory effort sensor (RIP belts) and the applicable sleep recorder/polysomnography (PSG) system.

There are two product groups for the Nox RIP Belt Cables; RIP Belt Cables and Third Party RIP Belt Cables.

The RIP Belt Cables are designed for use with Nox recorders only. Those are abdomen cables only because the thorax belt is attached directly to the Nox recorders.

The Third Party RIP Belt Cables are designed for use with third party recorders. The Third Party RIP Belt Cables come in pairs for abdomen and thorax.

{8}------------------------------------------------

## Intended Use

#### Nox RIP Belts

The Nox RIP Belts are intended for measuring of respiratory effort signals. They function as accessories for sleep/polysomnography (PSG) systems.

The Nox RIP Belts are indicated for use on patients greater than 2 years of age.

The intended environments are hospitals, institutions, sleep clinics, or other test environments, including the patient's home.

#### RIP Belt Cables

The RIP Belt Cables are intended to interconnect Nox RIP belts (respiratory effort sensors) and Nox sleep devices, to allow measuring of respiratory effort signals.

The RIP Belt Cables are indicated for use on patients greater than 2 years of age.

The intended environments are hospitals, institutions, sleep clinics, or other test environments, including the patient's home.

#### Third Party RIP Belt Cables

The Third Party RIP Belt Cables are intended to allow measuring of respiratory effort signals by interconnecting Nox RIP belts (respiratory effort sensors) and sleep devices with oscillation circuitry capable of measuring inductance between 1 and 5 µH.

The Third Party RIP Belt Cables are indicated for use on patients greater than 2 years of age.

The intended environments are hospitals, institutions, sleep clinics, or other test environments, including the patient's home.

## Technological Characteristics and Comparison with Predicate

#### Nox RIP Belts

The comparison table below is provided as a summary of the intended use and most relevant characteristics of the Nox RIP Belts relative to the predicate device.

Please note that the Nox RIP Belts will be regarded as a subsystem of the predicate. ODC-PRO AND NOX-RIP. The ODC-PRO AND NOX-RIP includes a recorder and several accessories but only the RIP belts will be addressed in the comparison between the new device and the predicate.

{9}------------------------------------------------

| Characteristic                       | QDC - PRO AND NOX-RIP (K124062)                                                                                                                                                                                                                                                                    | Nox RIP Belts                                                                                                                                     | Comparison Result              |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| General                              |                                                                                                                                                                                                                                                                                                    |                                                                                                                                                   |                                |
| Intended Use                         | The QDC-PRO device is a sensor unit intended for measuring of physiological signals during sleep. The signals measured are processed in the QDC-PRO device and the resulting signals made available at the output connector for acquisition by a third party Polysomnography (PSG)/sleep recorder. | The Nox RIP Belts are intended for measuring of respiratory effort signals. They function as accessories for sleep/polysomnography (PSG) systems. | Equivalent<br>See Note 1 below |
| Intended Environments                | The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments.                                                                                                                                                                                   | The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments, including the patient´s home.    | Equivalent<br>See Note 2 below |
| Patient Population                   | The QDC-PRO AND NOX-RIP are indicated for use in patients greater than 2 years of age.                                                                                                                                                                                                             | The Nox RIP Belts are indicated for use on patients greater than 2 years of age.                                                                  | Identical                      |
| Prescription Use                     | Yes                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                               | Identical                      |
| Type of Modules                      | RIP belts are disposable                                                                                                                                                                                                                                                                           | RIP belts are disposable                                                                                                                          | Identical                      |
| Various Belt Sizes                   | Yes                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                               | Identical                      |
| Physical/Material                    |                                                                                                                                                                                                                                                                                                    |                                                                                                                                                   |                                |
| RIP Belts Material                   | Elastic: Polyester/Dorlastan (not made with natural rubber latex), tin plated copper wire<br>Connector: ABS plastic                                                                                                                                                                                | Elastic: Polyester/Dorlastan (not made with natural rubber latex), tin plated copper wire<br>Connector: ABS plastic                               | Identical                      |
| Disposable Components                | Disposable respiratory effort sensor/RIP belts.                                                                                                                                                                                                                                                    | Disposable respiratory effort sensor/RIP belts.                                                                                                   | Identical                      |
| Functional                           |                                                                                                                                                                                                                                                                                                    |                                                                                                                                                   |                                |
| Connection to Patient                | RIP Belts are attached to patient                                                                                                                                                                                                                                                                  | RIP Belts are attached to patient                                                                                                                 | Identical                      |
| Type of Equipment to Be Connected to | Sleep recorder and RIP Belt Cable                                                                                                                                                                                                                                                                  | Sleep recorder and RIP Belt Cable                                                                                                                 | Identical                      |
| Connector Type                       | RIP belts have connector for to connect to Nox proprietary RIP snaps                                                                                                                                                                                                                               | RIP belts have connector for to connect to Nox proprietary RIP snaps                                                                              | Identical                      |
| Signals and Sensors                  |                                                                                                                                                                                                                                                                                                    |                                                                                                                                                   |                                |
| Signals Measured                     | Respiratory Effort (Abdomen and Thorax)                                                                                                                                                                                                                                                            | Respiratory Effort (Abdomen and<br>Thorax)                                                                                                        | Identical                      |
| Respiratory Effort<br>Technology     | RIP (Respiratory Inductive Plethysmography)<br>technology                                                                                                                                                                                                                                          | RIP (Respiratory Inductive<br>Plethysmography) technology                                                                                         | Identical                      |
| Signal Processing                    | No signal processing is done in RIP Belts                                                                                                                                                                                                                                                          | No signal processing is done in<br>RIP Belts                                                                                                      | Identical                      |
| Power and Isolation                  |                                                                                                                                                                                                                                                                                                    |                                                                                                                                                   |                                |
| Power Source                         | No power source. RIP Belts contain only<br>passive components                                                                                                                                                                                                                                      | No power source. RIP Belts<br>contain only passive components                                                                                     | Identical                      |
| Isolation                            | No component included in the RIP Belts that<br>are relied on as means of protection                                                                                                                                                                                                                | No component included in the<br>RIP Belts that are relied on as<br>means of protection                                                            | Identical                      |
| Standards                            |                                                                                                                                                                                                                                                                                                    |                                                                                                                                                   |                                |
| AAMI/ANSI/ES<br>60601-1              | Yes                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                               | Identical                      |
| IEC 60601-1-2                        | Yes                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                               | Identical                      |

Table 1 Comparison Summary of the New Device (Nox RIP Belts) and Predicate Device

{10}------------------------------------------------

Note 1: Nox RIP Belts are intended for measuring of respiratory effort signals and function as accessories for sleep systems. The predicate, ODC-PRO AND NOX RIP, is intended for measuring of physiological signals during sleep. Respiratory effort is one example of physiological signals and both the new device and the predicate measure the signals during sleep. Furthermore, the predicate RIP belts are intended for measuring respiratory effort signals as an accessory to the ODC-PRO recorder. Therefore the indication for use can be reqarded the same.

Note 2: The intended environment depends on the classification of the sleep recorder to be used with the Nox RIP Belts. The predicate device, ODC-PRO AND NOX-RIP, is intended to be used with a third party polysomnography (PSG) recorder. PSG studies are rarely performed in home environment and therefore it is not listed in the intended environments for the predicate. Simpler sleep recorders like polygraph (PG) recorders are commonly used at patients' home. The Nox RIP Belts may be used with any type of sleep recorders/PSG recorders including PG recorders. Therefore, the home environment was added for the Nox RIP Belts. If a sleep recorder is intended for home use, the type of RIP Belts does not change or influence the intended environment of the recorder. The home inclusion to the environment of use does therefore not pose any new safety and effectiveness concerns for the Nox RIP Belts. Additionally, risk analysis, verification and validation results demonstrate that the use of Nox RIP Belts in the home environment does not raise any new concerns regarding the intended use, safety or efficiency. Furthermore, these are prescription devices and are always used under supervision of a licensed medical practitioner. The intended environments are therefore deemed equivalent.

The result of the above comparison is that the indication for use, patient population and intended environments for the Nox RIP Belts can be regarded the same as that of the predicate QDC-PRO AND NOX-RIP device.

The comparison above demonstrates that all major technological characteristics, including features, materials, and principles of operation of the Nox RIP Belts are identical to the predicate QDC-PRO AND NOX RIP.

{11}------------------------------------------------

#### Nox RIP Belt Cables

The comparison table below is provided as a summary of the intended use and most relevant characteristics of the Nox RIP Belt Cables relative to the predicate device.

Please note that the Nox RIP Belt Cables will be regarded as a subsystem of the predicate, QDC-PRO AND NOX-RIP. The QDC-PRO AND NOX-RIP includes a recorder and several accessories but only the RIP belt cable will be addressed in the comparison between the new device and the predicate.

{12}------------------------------------------------

| Characteristic                                | QDC - PRO AND NOX-<br>RIP (K124062)                                                                                                                                                                                                                                                                                                    | Nox RIP Belt Cables                                                                                                                                                                       |                                                                                                                                                                                                                                                                                            |                                |
|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
|                                               |                                                                                                                                                                                                                                                                                                                                        | RIP Belt Cables                                                                                                                                                                           | Third Party RIP Belt<br>Cables                                                                                                                                                                                                                                                             | Comparison Results             |
|                                               |                                                                                                                                                                                                                                                                                                                                        | General                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                            |                                |
| Intended Use                                  | The QDC-PRO device is<br>a sensor unit intended<br>for measuring of<br>physiological signals<br>during sleep. The signals<br>measured are processed<br>in the QDC-PRO device<br>and the resulting signals<br>made available at the<br>output connector for<br>acquisition by a third<br>party Polysomnography<br>(PSG)/sleep recorder. | The RIP Belt Cables are<br>intended to interconnect<br>Nox RIP belts<br>(respiratory effort<br>sensors) and Nox sleep<br>devices, to allow<br>measuring of respiratory<br>effort signals. | The Third Party RIP Belt<br>Cables are intended to<br>allow measuring of<br>respiratory effort signals<br>by interconnecting Nox<br>RIP belts (respiratory<br>effort sensors) and sleep<br>devices with oscillation<br>circuitry capable of<br>measuring inductance<br>between 1 and 5 μH. | Equivalent<br>See Note 3 below |
| Intended<br>Environments                      | The intended<br>environments are<br>hospitals, institutions,<br>sleep centers, sleep<br>clinics, or other test<br>environments.                                                                                                                                                                                                        | The intended<br>environments are<br>hospitals, institutions,<br>sleep centers, sleep<br>clinics, or other test<br>environments, including<br>the patient's home.                          | The intended<br>environments are<br>hospitals, institutions,<br>sleep centers, sleep clinics,<br>or other test<br>environments, including<br>the patient's home.                                                                                                                           | Equivalent<br>See Note 4 below |
| Patient<br>Population                         | The QDC-PRO AND<br>NOX-RIP are indicated<br>for use in patients<br>greater than 2 years of<br>age.                                                                                                                                                                                                                                     | The RIP Belt Cables are<br>indicated for use on<br>patients greater than 2<br>years of age.                                                                                               | The Third Party RIP Belt<br>Cables are indicated for<br>use on patients greater<br>than 2 years of age.                                                                                                                                                                                    | Identical                      |
| Prescription<br>Use                           | Yes                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                        | Identical                      |
| Type of Modules                               | RIP Belt Cable for<br>connection to recorder<br>and RIP belts                                                                                                                                                                                                                                                                          | RIP Belt Cable for<br>connection to recorder<br>and RIP belts                                                                                                                             | RIP Belt Cables for<br>connection to recorder and<br>RIP belts                                                                                                                                                                                                                             | Identical                      |
| Physical / Material                           |                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                            |                                |
| RIP Cables<br>Material                        | ● PVC – wire jacket<br>● PC with 10% glass<br>fiber – belt end<br>● PC with 10% glass<br>fiber – device end<br>● Gold plated<br>stainless steel -<br>snaps<br>● TPE - strain relief                                                                                                                                                    | ● PVC – wire jacket<br>● ABS/PC - belt end<br>● ABS/PC - device<br>end<br>● Gold plated<br>stainless steel -<br>snaps<br>● TPE - strain relief                                            | ● PVC – wire jacket<br>● ABS/PC - belts end<br>● TPE-ET - device end<br>strain relief<br>● Gold plated stainless<br>steel - snaps<br>● TPE-ET - strain relief<br>Note: TPE-ET is also<br>known as Riteflex®                                                                                | Equivalent<br>See Note 5 below |
| Disposable                                    | RIP Belt Cable is                                                                                                                                                                                                                                                                                                                      | RIP Belt Cable is                                                                                                                                                                         | Third Party RIP Belt Cables                                                                                                                                                                                                                                                                | Identical                      |
| Components                                    | reusable                                                                                                                                                                                                                                                                                                                               | reusable                                                                                                                                                                                  | are reusable                                                                                                                                                                                                                                                                               |                                |
| Functional                                    |                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                            |                                |
| Connection to<br>Patient                      | RIP Belt Cable is<br>attached to patient via<br>connection to RIP belts                                                                                                                                                                                                                                                                | RIP Belt Cable is<br>attached to patient via<br>connection to RIP belts                                                                                                                   | Third Party RIP Belt Cables<br>are attached to patient via<br>connection to RIP belts                                                                                                                                                                                                      | Identical                      |
| Type of<br>Equipment to<br>Be Connected<br>to | Interconnect sleep<br>recorder and respiratory<br>effort sensor (RIP belt)                                                                                                                                                                                                                                                             | Interconnect sleep<br>recorder and respiratory<br>effort sensor (RIP belt)                                                                                                                | Interconnect sleep<br>recorder and respiratory<br>effort sensor (RIP belt)                                                                                                                                                                                                                 | Identical                      |
| Belt Cable<br>Connectors -<br>Recorder End    | Proprietary connector for<br>Nox RIP snaps                                                                                                                                                                                                                                                                                             | Proprietary connector for<br>Nox RIP snaps                                                                                                                                                | Touch proof 1 mm 2 pin<br>female connector                                                                                                                                                                                                                                                 | Equivalent<br>See Note 6 below |
| Belt Cable<br>Connectors -<br>Belt End        | Gold coated proprietary<br>metal snaps                                                                                                                                                                                                                                                                                                 | Gold coated proprietary<br>metal snaps                                                                                                                                                    | Gold coated proprietary<br>metal snaps                                                                                                                                                                                                                                                     | Identical                      |
| Signals and Sensors                           |                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                            |                                |
| Signal<br>Processing                          | No signal processing<br>done in RIP Belt Cable                                                                                                                                                                                                                                                                                         | No signal processing<br>done in RIP Belt Cable                                                                                                                                            | No signal processing done<br>in Third Party RIP Belt<br>Cables                                                                                                                                                                                                                             | Identical                      |
| Power and Isolation                           |                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                            |                                |
| Power Source                                  | No power source. RIP<br>Belt Cable contains only<br>passive components                                                                                                                                                                                                                                                                 | No power source. RIP<br>Belt Cable contains only<br>passive components                                                                                                                    | No power source. Third<br>Party RIP Belt Cables<br>contain only passive<br>components                                                                                                                                                                                                      | Identical                      |
| Isolation                                     | No component included<br>in the RIP Belt Cable<br>that is relied on as<br>means of protection                                                                                                                                                                                                                                          | No component included<br>in the RIP Belt Cable<br>that is relied on as<br>means of protection                                                                                             | No component included in<br>the Third Party RIP Belt<br>Cables that is relied on as<br>means of protection                                                                                                                                                                                 | Identical                      |
| Standards                                     |                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                            |                                |
| AAMI/ANSI/ES<br>60601-1                       | Yes                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                        | Identical                      |
| IEC 60601-1-2                                 | Yes                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                        | Identical                      |

Table 2 Comparison Summary of the New Device (Nox RIP Belt Cables) and Predicate Device

{13}------------------------------------------------

Note 3: Nox RIP Belt Cables are intended to allow measuring of respiratory effort signals and function as accessories for sleep systems. The predicate, QDC-PRO AND NOX RIP, is intended for measuring of physiological signals during sleep. Respiratory effort is one example of physiological signals and both the new device and the predicate allow measuring of the signals during sleep. Furthermore, the predicate RIP belt cable is intended to interconnect RIP belts and the QDC-PRO recorder to allow measuring of respiratory effort signals. Therefore the indication for use can be regarded the same.

{14}------------------------------------------------

Note 4: The intended environment depends on the classification of the sleep recorder to be used with the Nox RIP Belt Cables. The predicate device, ODC-PRO AND NOX-RIP, is intended to be used with a third party polysomnography (PSG) recorder. Simpler sleep recorders like polygraph (PG) recorders are commonly used at patients' home. The Nox RIP Belt Cables may be used with any type of sleep recorders/PSG recorders including PG recorders. Therefore, the home environment was added for the Nox RIP Belt Cables. If a sleep recorder is intended for home use, the type of RIP belt cables does not change or influence the intended environment of the recorder. The home inclusion to the environment of use does therefore not pose any new safety and effectiveness concerns for the Nox RIP Belt Cables. Additionally, risk analysis, verification and validation results demonstrate that the use of Nox RIP Belt Cables in the home environment does not raise any new concerns regarding the intended use, safety or efficiency. Furthermore, these are prescription devices and are always used under supervision of a licensed medical practitioner. The intended environments are therefore deemed equivalent.

Note 5: The RIP Belt Cables and Third Party RIP Belt Cables are compliant to standard AAMI/ANSI ES/60601-1. Furthermore, verification testing and risk analysis show that minor difference in material does not raise new questions about safety and effectiveness. The material of the new devices and the predicate can thus be regarded as equivalent.

Note 6: All cables, both new and predicate, are compliant with standard AAMI/ANSI ES/60601-1. Furthermore, verification testing, signal integrity comparison and risk analysis show that different connectors do not raise new questions about safety and effectiveness. The recorder end connectors can thus be regarded as equivalent.

The result of the above comparison is that the indication for use, patient population and intended environments for the Nox RIP Belt Cables can be regarded the same as that of the predicate QDC-PRO AND NOX-RIP device.

Furthermore, the comparison above shows that there are no significant differences in major technological characteristics, including features, materials, and principles of operation of Nox RIP Belt Cables compared to the predicate. The minor differences between the new and predicate device do not raise any new questions regarding safety and effectiveness. It can therefore be determined that Nox RIP Belt Cables have the same major technological characteristics as the predicate device, QDC-PRO AND NOX RIP.

{15}------------------------------------------------

## Performance Testing Summary

Thorough internal testing has demonstrated that the Nox RIP Belts and Nox RIP Belt Cables are suitable for their intended use.

The design of the Nox RIP Belts and Nox RIP Belt Cables was tested, verified and validated throughout the design process according to product requirement specifications and intended use.

The Nox RIP Belts and Nox RIP Belt Cables comply with the following standards:

- ISO 14971 Second edition 2007, Medical devices Application of risk management to . medical devices
- . ISO 15223-1:2012, Medical devices – Symbols to be used with medical device labels, labelling and information to be supplied
- . AAMI/ANSI/ES 60601-1:2005, Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
- . IEC 60601-1-2 Edition 3: 2007-03, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests
- AAMI/ANSI/IEC 62366:2007/(R)2013, Medical devices Application of usability . engineering to medical devices

Additionally, the Third Party RIP Belt Cables comply with

- 21 CFR 898 FDA PERFORMANCE STANDARD FOR ELECTRODE LEAD WIRES AND . PATIENT CABLES:2014
#### Verification

Nox RIP Belts and Nox RIP Belt Cables were tested towards test protocols written to cover all product requirement specifications. The test results demonstrate that the design output conforms to design input and that the device fulfills all physical characteristics, performance criteria, functional requiremental requirements, interface requirements, packaging and labeling requirements as well as safety and reliability requirements defined.

#### Risk Analysis

Risk analysis was performed for RIP Belts and RIP Belt Cables according to the standard: ISO 14971 - Medical devices - Application of risk management to medical devices, appropriate measures have been implemented and their effectiveness verified and validated.

#### Standards Testing

Assessment has been conducted for RIP Belts and RIP Belt Cables towards the standard AAMI/ANSI/ES 60601-1:2005. The result of this assessment is that the RIP Belt Cables are in compliance with the standard AAMI/ANSI/ES 60601-1:2005.

#### EMC Testing

Assessment has been conducted for RIP Belts and RIP Belt Cables towards the electromagnetic compatibility standard IEC 60601-1-2 Edition 3: 2007-03 to ensure compatibility with requirements addressing electrostatic discharge (ESD), electromagnetic disturbances, magnetic fields, electrical fast transients and bursts, and surge. The result of this assessment is that the RIP Belts and RIP Belt Cables are in compliance with the standard IEC 60601-1-2 Edition 3: 2007-03.

{16}------------------------------------------------

#### Usability

The RIP Belts and RIP Belt Cables were designed to minimize use errors and use-associated risks according to the usability engineering standard AAMI/ANSI/IEC 62366:2007/(R)2013. Usability testing resulted in all usability goals passed.

#### Signal Quality and Comparison Testing

Signal integrity tests were conducted for the Nox RIP Belts and Nox RIP Belt Cables with focus on signal to noise ratio, signal range, bandwidth and linearity and the test results compared to the signal integrity test conducted for the predicate QDC-PRO AND NOX-RIP.

The results from these signal integrity tests and comparison demonstrate that the Nox RIP Belts and Nox RIP Belt Cables signals are clinically equivalent to that of the predicate QDC-PRO AND NOX RIP.

#### Conclusion

Based on the testing, risk analysis and verification activates described above and detailed comparison to the predicate device provided in Table 2 above, it is the conclusion of Nox Medical that the Nox RIP Belts and Nox RIP Belt Cables are substantially equivalent to device already on the market (cleared by the 510(k) process) and present no new concerns about safety and effectiveness.

The Nox RIP Belts and Nox RIP Belt Cables are therefore substantial equivalent to the predicate QDC-PRO AND NOX-RIP from Nox Medical (K124062).

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR/K151361](https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR/K151361)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
