K082113 · Nox Medical · MNR · Nov 7, 2008 · Anesthesiology
Device Facts
Record ID
K082113
Device Name
NOX T3, NOXTURNAL (PC APPLICATION)
Applicant
Nox Medical
Product Code
MNR · Anesthesiology
Decision Date
Nov 7, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2375
Device Class
Class 2
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K082113 · Nov 7, 2008
NOX T3, NOXTURNAL (PC APPLICATION)
Nox Medical
Retrospective clinical sleep study recordings (Embla N7000)
The sponsor used a retrospective cohort of 1,057 clinical sleep recordings to validate the performance of the Noxturnal application's automatic scoring algorithm (AHI and ODI) by comparing it to manual scoring performed by certified technicians and physicians.
Manual scoring of full PSG data by RPSGT certified technicians and physicians
Apnea Hypopnea Index (AHI) and Oxygen Desaturation Index (ODI) values
Indications for Use
The Nox T3 device is intended for ambulatory recording of physiological signals during sleep. The recorded signals are then downloaded to a PC where the signals can be viewed and analyzed by use of the Nox T3 application (Noxturnal). The Nox T3 system is indicated for use in patients greater than 2 years of age. The Nox T3 system is NOT intended for any patient monitoring or automatic diagnosis. The intended environments are hospitals, institutions, sleep clinics, or other test environments, including patient's home.
Device Story
Nox T3 is an ambulatory sleep recorder; records physiological signals during sleep for later analysis. Inputs: respiratory effort (abdominal/thoracic), oximetry (via wireless Bluetooth), nasal/mask pressure, snoring, body position/activity (3D accelerometer), audio (microphone), and two auxiliary channels (ECG, EMG, EEG, or EOG). Device is pocket-sized, battery-powered, worn on chest via respiratory effort belt. Data stored on device; downloaded via USB to PC running Noxturnal application. Used in hospitals, clinics, or home; operated by clinicians or patients (hook-up). Noxturnal software provides visualization, manual analysis, and automatic scoring of AHI/ODI. Output assists physicians in diagnosing sleep disorders. Benefits: enables ambulatory sleep studies, reducing need for in-lab polysomnography.
Clinical Evidence
No clinical trials; performance validated via bench testing and comparative analysis. Comparison of Nox T3 signals to Embla N7000 confirmed reliability. Validation of Noxturnal automatic scoring (AHI and ODI) performed using 1,057 historical Embla N7000 recordings previously scored by RPSGT-certified technicians and reviewed by physicians. Results demonstrated automatic scoring is substantially equivalent to manual scoring of full PSG data.
Technological Characteristics
ABS plastic housing; 79x63x21mm; 65g. Sensors: RIP respiratory effort, solid-state pressure, 3D accelerometer, microphone. Connectivity: USB 2.0 (PC), Bluetooth 2.0 (oximeter). Power: 1 AA battery. Software: Noxturnal application for PC (Windows 2000/XP). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-25/26/40, FCC Part 15.
Indications for Use
Indicated for patients > 2 years of age suspected of Sleep Disordered Breathing (SDB) or Periodic Limb Movement Disorder (PLMD). Not for patient monitoring or automatic diagnosis.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains a sequence of characters that appear to be handwritten. The characters are 'K0821113'. The characters are written in a simple, somewhat stylized manner, with varying stroke thicknesses. The overall impression is that of a handwritten code or identifier.
# Nox Medic
Nox Medical • Impra Keldnaholti • It 112 Reykjavik • Iceland • Tel.: • 354 570 7111 • www.noxmedical.com
### 510(k) Summary
#### Submitter
NOV - 7 2008
Nox Medical Keldnaholti IS-112 Reykjavik Iceland
Tel: 011 354 522 9000 Fax: 011 354 522 9111
Registration Number: Nox Medical will register following FDA clearance
#### Contact person
Kolbrun E Ottosdottir Keldnaholti IS-112 Reykjavik Iceland Email: keo@noxmedical.com Direct tel: 011 354 522 9140
#### Preparation Date
July 25, 2008
#### Device
| Trade Name: | Nox T3 |
|-----------------------|-----------------------------|
| Common Name: | Sleep Recorder |
| Classification Name: | Ventilatory Effort Recorder |
| Regulation Number: | 868.2375 |
| Product Code: | MNR |
| Device Class: | Class II |
| Classification Panel: | Anesthesiology |
#### Predicate Devices
Embla N7000 from Medcare Flaga Product Code: MNR 510(k) Number: K024322
Compass M10 from Medcare Flaga Product Code: MNR 510(k) Number: K041724
4100 Patient Oximeter Module from Nonin Medical Inc. Product Code: DQA 510(k) Number: K043359
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a handwritten string of characters. The string appears to be "K082113". The characters are written in black ink on a white background. The handwriting is somewhat stylized, with some characters slightly overlapping or connected.
# nox Medical
Nox Medical・Impra Keldnaholti・IS 112 Reykjavík・Iceland・Tel.: +354 570 7100・Fax: +354 570 711 ・www.noxmedical.com
#### Device Description
The Nox T3 is an ambulatory recording system. It includes a recording device, respiratory effort sensors, clip straps, filter tube connector and an USB cable for data download and the Nox T3 application (Noxturnal).
The Nox T3 device is a pocket size battery powered digital recorder that incorporates electronics to record and store up to three nights of physiological parameters. The Nox T3 device is worn on the patient's chest by snapping it to the thoracic respiratory effort sensor belt and securing its position with the clip straps. It has a display for status indication, signal integrity and preliminary results, and buttons for control.
The Nox T3 device records signals from five external sensors and three built-in sensors. The external sensors that can be used with the device are abdominal and thoracic respiratory effort sensors, oximeter (via wireless transmission), and two leads of the following: ECG, EMG, EEG or EOG. The built-in sensors include a pressure transducer allowing either recording of nasal pressure (via nasal cannula) or mask pressure and measuring of snoring, a three dimensional acceleration sensor for measure of patient's position and activity, and a microphone for true audio recording capabilities.
The Nox T3 device includes a class II Bluetooth transmitter/receiver to allow for wireless transmission of data from Nonin´s Model 4100 Patient Oximeter Module.
The Nox T3 application (Noxturnal) is used to configure the device for recording, and downloading, viewing and analyzing of recorded data on a PC.
#### Intended Use
The Nox T3 device is intended for ambulatory recording of physiological signals during sleep. The recorded signals are then downloaded to a PC where the signals can be viewed and analyzed by use of the Nox T3 application (Noxturnal). The Nox T3 system is indicated for use in patients greater than 2 years of age.
The Nox T3 system is NOT intended for any patient monitoring or automatic diagnosis.
The intended environments are hospitals, institutions, sleep clinics, or other test environments, including patient's home.
The Nox T3 system is used for patients suspected of suffering from Sleep Disordered Breathing (SDB) or Periodic Limb Movement Disorder (PLMD).
#### Technological Characteristics
The comparison table below is provided as a summary of the most relevant characteristics of the Nox T3 system relative to the predicate devices. The comparison table demonstrates that the Nox T3 system has no significant differences from the predicate devices for the intended use that would adversely affect product safety and effectiveness.
{2}------------------------------------------------
■ ПОХ 【 П Ре С | C | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
able 1 Substantial Equivalence Summ
| Characteristic | Embla N7000<br>(K024322) | Compass M10 System<br>(K962865) | 4100 Patient Oximeter Module<br>(K043359) | Nox T3 System |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Embla N7000 is intended for use<br>by a physician or trained technician<br>for the acquisition of EEG and<br>polysomnography (PSG) signals and<br>transmission of these signals to a PC<br>during neurophysiologic or sleep<br>examinations. | General<br>The intended use of the Compass<br>M10 system is to record physiological<br>signals during sleep, scan the signals<br>for abnormalities and represent the<br>count of abnormal events in a form<br>of a summary report. The results of<br>the scan may be manually<br>overwritten or corrected by the<br>physician. The device is intended for<br>use as a screening device to<br>determine the need for clinical<br>diagnosis and evaluation by<br>polysomnography based on the<br>patient's count of abnormal events.<br>It is not intended for any diagnosis.<br>It is not intended to be a monitor. | The Nonin's Bluetooth® enabled<br>Model 4100 Patient Oximeter Module<br>is indicated for use in measuring and<br>transmitting functional oxygen<br>saturation of arterial hemoglobin<br>(SpO2), pulse rate, and<br>plethysmographic data to a<br>compatible Bluetooth® enabled<br>device. | The Nox T3 device is intended for<br>ambulatory recording of physiological<br>signals during sleep. The recorded<br>signals are then downloaded to a PC<br>where the signals can be viewed and<br>analyzed by use of the Nox T3<br>application (Noxturnal).<br><br>The Nox T3 system is NOT intended<br>for any patient monitoring or<br>automatic diagnosis. |
| Intended Environments | The intended environments are<br>hospitals, institutions, sleep centers,<br>sleep clinics, or other test<br>environments. | Not defined | Not defined | The Intended environments are<br>hospitals, institutions, sleep centers,<br>sleep clinics, or other test<br>environments, including patient's<br>home. |
| Patient Population | Not defined | The Compass M10 system is intended<br>to be used for adult and pediatric<br>patients. | Not defined | The Nox T3 system is indicated for<br>use in patients greater than 2 years<br>of age. |
| Prescription Use | Yes | Yes | Yes | Yes |
| Standards/Listing/Registrations | | | | |
| IEC60601-1:1988 | Yes | Yes | Yes | Yes |
| IEC60601-1-2:2001 | Yes | Yes | Yes | Yes |
| IEC 60601-2-25:1993 | Yes | NA | NA | Yes |
| IEC 60601-2-26: 2002 | Yes (1994 version) | NA | NA | Yes |
| IEC 60601-2-40: 1998 | Yes | NA | NA | Yes |
| Characteristic | Embla N7000<br>(K024322) | Compass M10 System<br>(K962865) | 4100 Patient Oximeter Module<br>(K043359) | Nox T3 System |
| 47 CFR Part 15 | NA | NA | Yes | Yes |
| FCC registration | NA | NA | Yes | Yes |
| Bluetooth EPL | NA | NA | Yes | Yes |
| Case Material | ABS Plastic (Patient Unit) | ABS Plastic and Aluminum | NA | ABS Plastic |
| RIP Belts Material | Polyester, latex free | Polyester, latex free | NA | Polyester, latex free |
| Dimension | 80mm (2.5") W<br>111mm (4.9") H<br>18.5mm(0.8") D<br>(Patient Unit) | 65mm (2.5") W<br>124mm (4.9") H<br>20mm(0.8") D | NA | 79mm (3.11") W<br>63mm (2.48")H<br>21mm(0.83") D |
| Weight | 280g (Patient Unit) | 100g | NA | 65g |
| Disposable Components | Disposable respiratory effort<br>sensor/RIP belts.<br>Nasal cannula<br>Remaining components require<br>cleaning. | Disposable respiratory effort<br>sensor/RIP belts.<br>Remaining components require<br>cleaning. | NA | Disposable respiratory effort<br>sensor/RIP belts.<br>Filter tube connector<br>Remaining components require<br>cleaning. |
| Acquisition units | Three units | One unit | NA | One unit |
| Number of channels | Seven channels (Patient Unit).<br>40 channels (Bedside Unit).<br>Eight auxiliary channels<br>(Communication Unit). | Five channels. | NA | Eight channels. |
| Recording time | Unlimited | Up to 12 hours | NA | Up to 24 hours |
| Control | Data acquisition and data storage<br>microprocessor controlled<br>Acquisition parameters set from<br>application SW | Data acquisition and data storage<br>microprocessor controlled<br>Acquisition parameters set from<br>application SW | NA | Data acquisition and data storage<br>microprocessor controlled<br>Acquisition parameters set from<br>application SW |
| Device Data Storage | No | Yes | NA | Yes |
Section 5 - Page 3 of 8
510(k) Summary
{3}------------------------------------------------
Nox Medical.
■ NOX 【 】 LE di C 【 】 【 】 【 】 【 】 】 】 【 】 】 】 】 】 【 】 】 】 【 】 【 】 】 】 【 】 【 】 】 【 】 】 【 】 】 】 【 】 】 】 【 】 】 】 【 】 】 】 【 】 】 】 【 】 】 】 【 】 】 】 【 】 】 】 】 【 】 】 】 【 】 】 】 】 【 】
KO82113
510(k) Summary
{4}------------------------------------------------
| medicar | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | |
| ﻟﻠﻘﻀﺎﺀ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮ<br>X<br>ﻟ | |
| l | |
Nox Medical · Impra Keldnaholti · Is 112 Reykjavik · Iceland · Tel. +354 570 7111 · www.noxmedical.com
| Characteristic | Embla N7000<br>(K024322) | Compass M10 System<br>(K962865) | 4100 Patient Oximeter Module<br>(K043359) | Nox T3 System |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Interface (PC) | Ethernet | USB v1.1 | NA | USB v2.0 |
| Connection to Patient | RIP belts for respiratory effort<br>Probes or Flexi Wrap for oximetry<br>Plastic tubing and cannula for<br>pressure sensing<br>Touch proof electrode cables<br>Thermistor<br>Snore Sensor<br>Piezo belts for respiratory effort<br>Elastic cloth material for support of<br>device | RIP belts for respiratory effort<br>Probes or Flexi Wrap for oximetry<br>Elastic cloth material for support of<br>device | NA | RIP belts for respiratory effort<br>Probes or Flexi Wrap for oximetry<br>Plastic tubing and cannula for<br>pressure sensing<br>Touch proof electrode cables<br>RIP belts for attaching of device and<br>clip straps to secure position of<br>device |
| Signals Recorded | Respiratory Effort (Abdomen and<br>Thorax)<br>Body position<br>Activity<br>Oxygen Saturation<br>Pulse<br>Nasal/mask pressure<br>Airflow<br>Snore<br>Respiratory sound<br>EEG, EOG, EMG, ECG | Signals and Sensors<br>Respiratory Effort (Abdomen and<br>Thorax)<br>Body position<br>Activity<br>Oxygen Saturation<br>Pulse | NA | Respiratory Effort (Abdomen and<br>Thorax)<br>Body position<br>Activity<br>Oxygen Saturation<br>Pulse<br>Nasal/mask pressure<br>Airflow<br>Snore<br>Respiratory sound<br>EEG, EOG, EMG, ECG |
| Sensor Technology used<br>in/with the system | Solid state pressure sensor<br>Solid state position/activity sensor<br>Respiratory Effort Sensors (RIP<br>technology)<br>Oximetry | Solid state position/activity sensor<br>Respiratory Effort Sensors (RIP<br>technology)<br>Oximetry | NA | Solid state pressure sensor<br>Solid state position/activity sensor<br>Respiratory Effort Sensors (RIP<br>technology)<br>Oximetry |
| Characteristic | Embla N7000<br>(K024322) | Compass M10 System<br>(K962865) | 4100 Patient Oximeter Module<br>(K043359) | Nox T3 System |
| Microphone | Gold cup electrodes<br>Ag/AgCL electrodes<br>Piezobelts for respiratory effort<br>Thermistor<br>Piezo snoring sensor | | | Microphone<br>Gold cup electrodes<br>Ag/AgCL electrodes |
| Power Source | 115/230V AC | Power and Isolation<br>3V by 2 AA batteries (Data<br>Acquisition) | Two 1.5V AA batteries | 1.5V by 1 AA battery (Data<br>Acquisition) |
| Patient Isolation | Isolation between mains and applied part | Host PC (Data Transfer)<br>Device has no galvanic connections<br>to mains as it is a battery operated<br>device | Device has no galvanic connections<br>to mains as it is a battery operated<br>device | Host PC (Data Transfer)<br>Device has no galvanic connections<br>to mains as it is a battery operated<br>device |
| | | Not possible to connect auxiliary<br>devices to the device | Not possible to connect auxiliary<br>devices to the device | Not possible to connect auxiliary<br>devices to the device |
| | | Transmitter | | |
| RF Data Transfer | No | No | Bluetooth wireless technology | Bluetooth wireless technology |
| Bluetooth Specification | NA | NA | Version 1.1 | Version 2.0 |
| RF Operating frequency | NA | NA | 2.4-2.4835 GHz | 2.4-2.4835 GHz |
| RF Emissions CISPR 11 | NA | NA | Group 2 | Group 2 |
| | | | Class B | Class B |
| Output power | NA | NA | <1.1mW (Class II) | < 1.62mW (Class II) |
| Antenna Type | NA | NA | Internal | Internal |
| Modulation Type | NA | NA | Frequency Shift Keying | Frequency Shift Keying |
| | | | Frequency Hopping Spread Spectrum | Frequency Hopping Spread Spectrum |
| Network Topology | NA | NA | Point-to-Point: | Point-to-Point: |
| Bandwidth | NA | NA | 1 MHz | 1 MHz |
| | | User Interface | | |
| Characteristic | Embla N7000<br>(K024322) | Compass M10 System<br>(K962865) | 4100 Patient Oximeter Module<br>(K043359) | Nox T3 System |
| Data validation | On-line in the application SW (not a part of the system).<br>Status light on Bedside Unit | Visual verification of the respiratory signals by light indicators on device | Connection status LED | The device has a display to view signal integrity and operation of the device. |
| Operating System | No PC application | Microsoft Windows™ 2000 and XP | NA | Microsoft Windows™ 2000 and XP |
| Data review on PC | No PC application | Yes: Real-time waveforms | NA | Yes: Real-time waveforms |
| Generate/Print reports | No PC application | Yes | NA | Yes |
| Patient data entry | No PC application | Yes | NA | Yes |
| Analysis | No PC application | Automatic, result may be manipulated.<br>Manual analysis<br>Event marking (scoring) | NA | Automatic, result may be manipulated.<br>Manual analysis<br>Event marking (scoring) |
510(k) Summary
{5}------------------------------------------------
■ ПОХ ㅣ П Ре С Г С ( С ( С ) С ( С ) С ( С ) - 3 5 4 570 710 - Fax: +354 570 7111 - www.noxmedical.com
Nox Medical
Impra
Keldnaholti
IS 112 Reykjavik Iceland
Tel.: +354 570 7100
Fax: +354 570 7111
[www.noxmedical.com](http://www.noxmedical.com)
510(k) Summary
.
K082113
{6}------------------------------------------------
Nox Medical
■ ПОХ ↑ Medical • Medical • Iceland • Tel. +354 570 7100 • Fax: +354 570 7111 • www.noxmedical.com
Nox Medical • Impra Keldnaholt • IS 112 Reykjavik • Iceland • Tel. +354 57
Section 5 -- Page 7 of 8
{7}------------------------------------------------
K082113
# nox Medical
Nox Medical • Impra Keldnaholi • IS 112 Reykjavik • Iceland • Tel.: +354 570 7111 • www.noxmedical.com
#### Performance Testing Summary
The Nox T3 system has been tested and verified in various phases to include internal testing, verification and validation as well as external testing to assure product safety, effectiveness and reliability.
The design was verified and validated throughout the design process according to requirement specifications and intended use. Risk analysis was performed according to ISO 14971:2007, appropriate measures were implemented and their effectiveness verified and validated.
External test house was used to conduct testing needed to comply with applicable standards reqarding EMC and patient safety as well as additional RF testing to assure compliance to FCC requirements for R&TTE (The Radio and Telecommunications Terminal Equipment Directive) approval.
The general process from configuration of device to reading out the results and generating report was validated by having untrained person perform this actions. The results demonstrates that the Nox T3 system has meet its objective of being easy to operate, the Noxturnal interface guides the user appropriately, minimizing the likelihood of errors and lapses, and the design of the Nox T3 components and user instruction allows the hook-up to be performed by untrained people.
#### Performance Data
The signals recorded with the Nox T3 system were compared to signals recorded with the predicate device Embla N7000. The result demonstrates the reliability and usability of all signals recorded with the Nox T3 system.
Analysis comparison was performed to validate the quality of the automatic scoring performed by the Noxturnal application compared to manual scoring of full PSG data. The validation study focused on the comparison of the Apnea Hypopnea Index (AHI) values and the Oxygen Desaturation Index (ODI) values. The recordings used consisted of 1057 Embla N7000 recordings which had all gone through the process of being manually scored by Sleep Technicians with RPSGT certification, and then reviewed by a Physician.
The analysis comparison result demonstrates that the Noxturnal application scores AHI and ODI events in a substantially equivalent manner to the manual scoring on full PSG recordings obtained using Embla N7000 recorders.
#### Conclusion
Verification testing, validation, risk analysis, technological characteristic and performance data comparisons have been performed for the Nox T3 system. The results of this testing and comparisons do not raise new questions of safety and effectiveness and demonstrate the same effectiveness and safety as that of the predicates Compass M10, Embla N7000 and Model 4100 Patient Oximeter Module.
The Nox T3 system is therefore substantial equivalent to the predicates Compass M10 from Medcare Flaga (K041724) and Embla N7000 from Medcare Flaga (K024322).
The wireless function of the Nox T3 system is therefore substantial equivalent to that of the predicate Model 4100 Patient Oximeter Module from Nonin Medical (K043359).
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 7 2008
Ms. Kolbrun E. Ottosdottir Quality Manager Nox Medical Impra Keldnaholti IS 112 Reykjavik ICELAND
Re: K082113
Trade/Device Name: Nox T3 Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: MNR Dated: October 20, 2008 Received: October 23, 2008
Dear Ms. Ottosdottir:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
#### Page 2 - Ms. Ottosdottir
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sylitte Y. Mchue O MD
FOR DR CHILE LIN
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
# Indication for Use
510(k) Number (if known): KOBA\/ 3
Device Name: Nox T3
Indications For Use:
The Nox T3 device is intended for ambulatory recording of physiological signals during sleep. The recorded signals are then downloaded to a PC where the signals can be viewed and analyzed by use of the Nox T3 application (Noxturnal). The Nox T3 system is indicated for use in patients greater than 2 years of age.
The Nox T3 system is NOT intended for any patient monitoring or automatic diagnosis.
The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments, including patient's home.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Emmett J. Michael Dmd.
Division Sign-Off) Division of Anesthesiology, General Hospital infection Control, Dental Devices
4-082113 510(k) Number:
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.