← Product Code [MNR](/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR) · K071556

# SOMNOSCREEN EEG10-20 (K071556)

_Somnomedics GmbH · MNR · Jul 18, 2007 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR/K071556

## Device Facts

- **Applicant:** Somnomedics GmbH
- **Product Code:** [MNR](/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR.md)
- **Decision Date:** Jul 18, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 868.2375
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology

## Intended Use

The SOMNOscreen EEG10-20 is a portable physiological signal recording system intended to be used to record, display, monitor, print and store biophysical events to aid in the diagnosis of neurologic and sleep disorders. The device is intended to be prescribed for use by a physician in the office, sleep laboratory or patient's home. This device is NOT designed to be used in a Life Support situation. This device is not designed for use on patients with cardiac pacemakers.

## Device Story

Portable physiological signal recording system; records, displays, monitors, prints, and stores biophysical events. Inputs: up to 40 channels (22 AC, 20 referential, 11 respiratory/AUX, 7 internal including SpO2, pulse rate, plethysmogram, body position, light, patient marker, respiratory effort). Used in office, sleep lab, or home; operated by physician or under physician prescription. Data stored on Compact Flash card; DOMINO software used for initialization, transfer, and analysis. Output aids diagnosis of neurologic and sleep disorders; assists clinical decision-making by providing physiological data for review. Benefits patient by enabling diagnostic testing for sleep-related breathing disorders outside clinical settings.

## Clinical Evidence

Bench testing only. Performance testing verified device specifications, including channel accuracy and data acquisition functions. Compliance with international standards for electrical safety and electromagnetic compatibility was demonstrated.

## Technological Characteristics

Portable physiological recorder; 40-channel data acquisition (22 AC, 20 referential, 11 respiratory/AUX, 7 internal). Includes Compact Flash card, Li-ion batteries (2350mAh), and DOMINO software. Connectivity via data transfer from flash card. Electrical safety and EMC testing performed per international standards.

## Regulatory Identification

A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.

## Predicate Devices

- Somnomedics SOMNOscreen ([K060708](/device/K060708.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Ko71556

# Section 6: 510(k) Summary

### 510(k) Summary

JUL 1 8 2007

| Applicant:                              | SOMNOmedics GmbH<br>Am Sonnenstuhl 63<br>Randersacker, Germany D-97236<br>Phone: +49-931-359094-0<br>Facsimile: +49-931-359094-49                                                          |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| US Contact:                             | M Squared Associates, Inc.<br>Cherita James<br>719 A Street, NE<br>Washington DC 20002<br>Phone: 202-546-1262 ext. 257<br>Facsimile: 202-546-3848<br>E-mail: CJames@msquaredassociates.com |
| Manufacturing/<br>Distribution Address: | SOMNOmedics GmbH<br>Am Sonnenstuhl 63<br>Randersacker, Germany D-97236                                                                                                                     |

Establishment Registration Number: Active, awaiting assignment of registration number

| Date submitted:        | June 6, 2007                                                      |
|------------------------|-------------------------------------------------------------------|
| Proprietary Name:      | Somnomedics SOMNOscreen EEG10-20                                  |
| Common Name:           | Ventilatory Effort Recorder                                       |
| Classification Status: | Class II per regulations §868.2375 Breathing frequency<br>monitor |
| Product Codes:         | MNR                                                               |
| Predicate Device:      | Somnomedics SOMNOscreen (K060708) cleared January 24, 2007        |

Device Description: The SOMNOscreen EEG10-20 is a portable physiological signal recording
system intended to be used to record, display, monitor, print and store biophysical events to

{1}------------------------------------------------

aid in the diagnosis of neurologic and sleep disorders. The device is intended to be prescribed for use by a physician in the office, sleep laboratory or patient's home.

This device is NOT designed to be used in a Life Support situation. This device is not designed for use on patients with cardiac pacemakers.

Indication for Use: The SOMNOscreen EEG10-20 is a non-life-supporting portable physiological signal recording device intended to be used for testing adult patients suspected of having sleep-related breathing disorders.

Summary of Technological Characteristics: The system provides up to 40 channels for data acquisition: 22 AC Channels, 20 Referential, 11 Respiratory and AUX Channels, 7 Internal Channels (SPO2, Pulse Rate, Plethysmogram, Body Position, Light, Patient Marker, Thorax/Abdominal Respiratory Effort).

The SOMNOscreen EEGI0-20 includes a Compact Flash Card and Reader, Li ION Batteries, (2350mAh) with 1 x Battery Charger, a Carry Bag for housing the SOMNOscreen ERGIO-20 and Sensors, Instruction Manuals and the DOMINO software for Initialization, Data Transfer and Analysis.

Summary of Nonclinical Testing: The SOMNOscreen EEGIO-20 was subject to the same preclinical requirements as the previously cleared predicate device the Somnomedics SOMNOscreen (K060708).

Performance testing was conducted to confirm compliance to device specifications; all functions were verified to operate as designed. Measured parameters met required ranges and accuracies. Testing to the international standards for electrical safety and electromagnetic compatibility were performed. SOMNOscreen EEGI0-200 was found to be compliant with the requirements of these standards.

### Substantial Equivalence Discussion:

The SOMNOscreen (K060708) and the SOMNOscreen EEGI0-20 differ only in the number of AC-Channels. The predicate device provides up to 28 channels for data acquisition; 10 AC Channels, 8 Referential and 2 Differential, 11 Respiratory and AUX Channels, 7 Internal

{2}------------------------------------------------

Channels (SPO2, Pulse Rate, Plethysmogram, Body Position, Light, Patient Marker, Thorax/Abdominal Respiratory Effort) as compared to the SOMNOscreen EEGI0-20 providing an additional 12 data acquisition Channels, 12 AC Channels, and 12 Referential Channels (40 channels for data acquisition; 22 AC Channels, 20 Referential and 2 Differential, 11 Respiratory and AUX Channels, 7 Internal Channels).

#### Conclusion:

SOMNOMEDICS Somnoscreen EEGI0-20 is substantially equivalent to Somnomedics SOMNOscreen (K060708) already on the market and presents no new concerns about safety and effectiveness. Additionally, the device has identical indications to the predicate device and the labeling of the device is consistent both with FDA's guidance as well as current medical practice. SOMNOMEDICS Somnoscreen EEGIO has the same principles of operation and technological characteristics as the previously 510k cleared predicate.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of three human figures in profile, arranged in a row and facing to the right. The figures are rendered in black and have a flowing, abstract design. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 1 8 2007

SOMNOmedics GmbH C/O Ms. Cherita James Regulatory Consultant M. Squared Associates, Incorporated 719 A Street North East Washington, DC 20002

Re: K071556

Trade/Device Name: SOMNOscreen EEG10-20 Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: MNR Dated: July 9, 2007 Received: July 10, 2007

Dear Ms. James:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 -Ms. James

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known): K071556

Device Name: SOMNOscreen eeg10-20

Indications For Use: The SOMNOscreen secuen is a non-life-supporting portable physiological signal recording device intended to be used for testing adult patients suspected of having sleeprelated breathing disorders.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nel J. Rebol

(Division Sign-Ott) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of __ 1__

510(k) Number: k071556

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR/K071556](https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR/K071556)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
