K060708 · Somnomedics GmbH & Co. KG · MNR · Jan 24, 2007 · Anesthesiology
Device Facts
Record ID
K060708
Device Name
SOMNOSCREEN
Applicant
Somnomedics GmbH & Co. KG
Product Code
MNR · Anesthesiology
Decision Date
Jan 24, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2375
Device Class
Class 2
Indications for Use
The SOMNOscreen is indicated for use in the recording, displaying, monitoring, and storage of biophysical parameters for the purpose of assisting in the diagnosis of Neurological and Sleep Disorders.
Device Story
Portable physiological signal recording system; records, displays, monitors, and stores biophysical events. Inputs: up to 28 channels including AC (referential/differential), respiratory, AUX, and internal sensors (SpO2, pulse, plethysmogram, body position, light, patient marker, thorax/abdominal effort). Operation: data acquisition via sensors; storage on Compact Flash card; analysis via DOMINO software. Used in office, sleep lab, or patient home; operated by physician or under physician prescription. Output: stored biophysical data for diagnostic assistance. Benefits: aids diagnosis of sleep-related breathing and neurological disorders. Not for life support or patients with cardiac pacemakers.
Clinical Evidence
Validation study of 25 patients comparing SOMNOscreen, predicate SleepScreen (K021138), and manual scored polysomnography. Results demonstrated clinical performance equivalent to predicates for recording, displaying, monitoring, and storing biophysical parameters for diagnosing sleep-related respiratory disorders.
Technological Characteristics
Portable recording system with 28 channels (10 AC, 11 respiratory/AUX, 7 internal). Includes Compact Flash card, Li-ION batteries (2000mAh), and DOMINO software. Connectivity: data transfer via card reader. Power: battery-operated. Compliance with international electrical safety and electromagnetic compatibility standards.
Indications for Use
Indicated for adult patients suspected of having sleep-related breathing disorders. Non-life-supporting portable physiological signal recording device. Contraindicated for patients with cardiac pacemakers.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
K060708
## 510(k) Summary
| Applicant: | Somnomedics GmbH & Co.KG<br>Nonnengarten 8<br>Kist, Germany D-97270<br>Phone: +49 93 06 98 23 0<br>Facsimile: +49 93 06 98 23 20 | JAN 24 |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| US Contact: | M Squared Associates, Inc.<br>Cherita James<br>719 A Street, NE<br>Washington DC 20002<br>Phone: 202-546-1262 ext. 257<br>Facsimile: 202-546-3848<br>E-mail: CJames@msquaredassociates.com | |
| Date submitted: | January 11, 2007 | |
| Proprietary Name: | Somnomedics SOMNOscreen | |
| Common Name: | Ventilatory Effort Recorder | |
| Classification Status: | Class II per regulations §868.2375 | |
| Product Codes: | MNR | |
| Establishment Registration Number: | Will be obtained prior to US distribution of the device. | |
| Manufacturing/<br>Distribution Address: | Somnomedics GmbH & Co.KG<br>Nonnengarten 8<br>Kist, Germany D-97270 | |
| Predicate Devices: | SleepScreen/ApnoeScreen Cardio (K021138), Compumedics<br>Siesta System (K003175) | |
2007
.
{1}------------------------------------------------
#### Device Description:
The SOMNOscreen is a portable physiological signal recording system intended to be used to record, display, monitor, print and store biophysical events to aid in the diagnosis of neurologic and sleep disorders. The device is intended to be prescribed for use by a physician in the office, sleep laboratory or patient's home.
This device is NOT designed to be used in a Life Support situation. This device is not designed for use on patients with cardiac pacemakers.
Indication for Use: The SOMNOscreen is a non-life-supporting portable physiological signal recording device intended to be used for testing adult patients suspected of having sleep-related breathing disorders.
#### Intended Use:
The SOMNOscreen is indicated for use in the recording, displaying, monitoring, and storage of biophysical parameters for the purpose of assisting in the diagnosis of Neurological and Sleep Disorders.
#### Summary of Technological Characteristics:
The system provides up to 28 channels for data acquisition; 10 AC Channels, 8 Referential and 2 Differential, 11 Respiratory and AUX Channels, 7 Internal Channels (SPO2, Pulse Rate, Plethysmogram, Body Position, Light, Patient Marker, Thorax/Abdominal Respiratory Effort)
The SOMNOscreen is available in 6 different configurations. All configurations include a Compact Flash Card and Reader, Li ION Batteries, (2000mAh) with 1 x Battery Charger, a Carry Bag for housing the SOMNOscreen and Sensors, Instruction Manuals and the DOMINO software for Initialization, Data Transfer and Analysis.
#### Summary of Nonclinical Testing:
Performance testing was conducted to confirm compliance to device specifications; all functions were verified to operate as designed. Measured parameters met required ranges and accuracies. Testing to the international standards for electrical safety and electromagnetic compatibility were
{2}------------------------------------------------
performed. The SOMNOscreen was found to be compliant with the requirements of these standards.
### Summary of Clinical Testing:
A validation study of 25 patients evaluated with the SOMNOscreen, the predicate Sleepscreen from Viasys (K021138, originally Jaeger) and the manual scored Polysomnography demonstrates that the clinical performance of the SOMNOscreen when used as intended in the targeted patient population, is equivalent to the predicate Sleepscreen and the manual scored Polysmnography cleared for the evaluation of the recording, displaying, monitoring, printing, and storage of biophysical parameters for the purpose of assisting in the diagnosis of sleep-related respiratory disorders.
#### Conclusion
SOMNOMEDICS Somnoscreen has the same principles of operation and similar technological characteristics as the previously 510k cleared predicates. The differences do not present new issues of safety or effectiveness.
Based on extensive performance testing and a comparison to the predicate devices, the SOMNOMEDICS Somnoscreen is substantially equivalent to devices already on the market (cleared by the 510(k) process) and presents no new concerns about safety and effectiveness. Additionally, the device has identical indications to the predicate devices and the labeling of the device is consistent both with FDA's guidance as well as current medical practice.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the department's name written around the perimeter. In the center of the seal is an emblem that resembles an eagle or other bird-like figure, stylized with three curved lines representing the wings and body.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Somnomedics GmbH & Company KG C/O Ms. Cherita James Regulatory Consultant M Squared Associates, Incorporated 719 A Street. NE Washington, DC 20002
JAN 2 4 2007
Re: K060708
Trade/Device Name: SOMNOscreen Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: MNR Dated: January 11, 2007 Received: January 12, 2007
Dear Ms. James:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Ms. James
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number: K060708
Device Name: SOMNOscreen
Indications For Use: The SOMNOscreen is a non-life-supporting portable physiological signal recording device intended to be used for testing adult patients suspected of having sleep-related breathing disorders.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(21 CFR 807 Subpart C)
(Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Wilm Wul
-
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.