WATCH PAT 100S

K042916 · Itamar Medical · MNR · Dec 29, 2004 · Anesthesiology

Device Facts

Record IDK042916
Device NameWATCH PAT 100S
ApplicantItamar Medical
Product CodeMNR · Anesthesiology
Decision DateDec 29, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.2375
Device ClassClass 2

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Sleep related breathing disorders———————
REM sleep stage identification———————

Indications for Use

The Watch-PAT100S (WP100S) device is a non-invasive home care device for use with patients suspected to have sleep related breathing disorders. The WP100S is a diagnostic aid for the detection of sleep related breathing disorders and rapid eye movement ("REM") sleep stages. The WP100S generates a peripheral arterial tonometry ("PAT") respiratory disturbance index ("PRDI") and PAT REM sleep stage identification ("PREM"). The WP100S's PREM provides supplemental information to its PRDI. The WP100S's PREM is not intended to be used as the sole or primary basis for diagnosing any sleep related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted. The WP100S is not indicated for children less than 17 years old. The WP100S is contraindicated for patients with latex allergy.

Device Story

Home-use diagnostic device for sleep-related breathing disorders; utilizes peripheral arterial tonometry (PAT) via finger-mounted sensor; inputs include PAT signals, oximetry, and actigraphy data; device processes signals to calculate respiratory disturbance index (PRDI) and identify REM sleep stages (PREM); intended for use by patients in home setting; healthcare providers review output to assist in diagnosis; PREM serves as supplemental information to PRDI; device aids in identifying sleep disorders; not for pediatric use (<17 years).

Clinical Evidence

Clinical data compared WP100S performance against manual scoring of polysomnography (PSG) and automated scoring of BioSleep for REM sleep stage detection. Results demonstrated WP100S is as safe and effective as manual PSG and BioSleep for REM detection, and as safe and effective as the WP100 for sleep apnea detection.

Technological Characteristics

Peripheral arterial tonometry (PAT) sensing; includes oximeter, actigraph, battery, charger, and computer analysis software. Non-invasive, home-use form factor.

Indications for Use

Indicated for patients aged 17+ suspected of sleep-related breathing disorders. Contraindicated for patients with latex allergy.

Regulatory Classification

Identification

A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ DEC 29 2004 K042916 # 510(k) SUMMARY #### Itamar's Watch PAT 100S # Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Itamar Medical Ltd. 2 Ha'eshel Street, P.O. Box 3579 Caesarea 38900 Israel 011-972-4-617-7000 Phone: 011-972-4-627-5598 Facsimile: Koby Sheffy, Ph.D. Contact Person: September 27, 2004 · Date Prepared: #### Name of Device and Name/Address of Sponsor Watch-PAT 100S Itamar Medical Ltd. 2 Ha'eshel Street, P.O. Box 3579 Caesarea 38900 Israel # Common or Usual Name Ventilatory Effort Recorder ### Classification Name Breathing Frequency Monitor (21 C.F.R. § 868.2375) #### Predicate Devices Itamar's Watch PAT 100 ("WP100") Oxford BioSignal's BioSleep ("BioSleep") {2}------------------------------------------------ ### Intended Use / Indications for Use The Watch-PAT100S (WP100S) device is a non-invasive home care device for use with patients suspected to have sleep related breathing disorders. The WP100S is a diagnostic aid for the detection of sleep related breathing disorders and rapid eye movement ("REM") sleep stages. The WP100S generates a peripheral arterial tonometry ("PAT") respiratory disturbance index ("PRDI") and PAT REM sleep stage identification ("PREM"). The WP100S's PREM provides supplemental information to its PRDI. The WP100S's PREM is not intended to be used as the sole or primary basis for diagnosing any sleep related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted. The WP100S is not indicated for children less than 17 years old. The WP100S is contraindicated for patients with latex allergy. #### Technological Characteristics The WP100S uses peripheral arterial tonometry to generate apnea indices ("PRDIs") and identify REM sleep stages ("PREMs"). The device's primary components include a cleared oximeter and actigraph, an associated computer analysis program, a battery and charger, and a carrying case. ### Performance Data Clinical data demonstrates that the device is as safe and effective as manual scoring of polysomnography ("PSG") and automated scoring of the BioSleep in detection REM sleep stages. Clinical data also demonstrated that the WP100S is at least as safe and effective as the Watch-PAT100 in detecting sleep apnea. ### Substantial Equivalence The WP100S is substantially equivalent to manual scoring of PSG data and a combination of the WP100 and the BioSleep. The WP100S has the same intended use as the WP100, very similar indications as a combination of the WP100 and the BioSleep, and the same hardware and principles of operation as the WP100. The minor technological differences between the WP100S and its predicate devices, namely the WP100S's software modifications, raise no new issues of safety or effectiveness. Clinical data demonstrate that the WP100S is as safe and effective as its predicates. Thus, the WP100S is substantially equivalent. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized caduceus or a representation of human figures. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 2 9 2004 Itamar Medical Limited C/O Mr. Jonathan S. Kahn Regulatory Counsel Hogan & Hartson, L.L.P 555 Thirteenth Street, NW Washington, DC 20004 Re: K042916 Trade/Device Name: Watch-PAT 100S Regulation Number: 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: MNR Dated: October 21, 2004 Received: October 21, 2004 Dear Mr. Kahn: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Kahn Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Charles Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### Indications for Use Statement 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: Watch-PAT 100S Indications for Use: The Watch-PAT 100S ("WP100S") device is a non-invasive home care device for use with patients suspected to have sleep related breathing disorders. The WP100S is a diagnostic aid for the detection of sleep related breathing disorders and rapid eye movement ("REM") sleep stages. The WP100S generates a peripheral arterial tonometry ("PAT") respiratory disturbance index ("PRDI") and PAT REM sleep stage identification ("PREM"). The WP100S's PREM provides supplemental information to its PRDI. The WP100S's PREM is not intended to be used as the sole or primary basis for diagnosing any sleep related breathing disorder, prescribing treatment, or determining whether additional diagnostic assessment is warranted. The WP100S is not indicated for children less than 17 years old. The WP100S is contraindicated for patients with latex allergy. Prescription Use _ X (Part 21 C.F.R. 801 Subpart D) AND/OR Over-The-Counter Use (21 C.F.R. 807 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | |-----------------------------------------------| | Division of Anesthesiology, General Hospital, | | Infection Control, Dental Devices | | 510(k) Number: | K042916 | |----------------|---------| |----------------|---------| \\\DC - 87424/0001 - 2034236 v2
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...