LIFESCREEN APNEA

K042745 · Del Mar Reynolds Medical, Inc. · MNR · Jan 19, 2005 · Anesthesiology

Device Facts

Record IDK042745
Device NameLIFESCREEN APNEA
ApplicantDel Mar Reynolds Medical, Inc.
Product CodeMNR · Anesthesiology
Decision DateJan 19, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.2375
Device ClassClass 2

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Sleep Disordered Breathing (SDB) detectionAlgorithm deriving RR interval and ECG-derived respiratory information—Accuracy 89.1%, Sensitivity 87.0%, Specificity 90.3%, Positive predictivity 84.9%Diagnostic criteria established using a separate training group—Clinical trial—

Indications for Use

Lifescreen Apnea is indicated for screening adults for the probability that they suffer from obstructive sleep apnea, mixed apnea or hypopnea, in order to evaluate the necessity for a polysomonographic examination. The ECG recording may be obtained at any location specified by a physician including home, hospital or clinic. Subjects screened for sleep apnea should have periods of sleep of at least 4 hours duration during which the ECG is predominantly sinus rhythm in nature.

Device Story

Lifescreen Apnea is a software-only module for Del Mar Reynolds Holter ECG systems. It processes recorded ECG data to identify Sleep Disordered Breathing (SDB). The algorithm derives RR intervals and ECG-Derived Respiratory (EDR) signals from the ECG to detect apneic events and estimate an Apnea-Hypopnea Index (AHI). Used in home, hospital, or clinic settings; operated by clinicians. Output provides probability of SDB, assisting physicians in determining the necessity for formal polysomnographic evaluation. Benefits include identifying patients requiring further diagnostic sleep studies.

Clinical Evidence

Clinical trial conducted with a separate training group to establish diagnostic criteria. Performance metrics (per-minute statistics): Accuracy 89.1% (95% CI 88.58-89.52), Sensitivity 87.0% (95% CI 86.19-87.82), Specificity 90.3% (95% CI 89.74-90.88), and Positive Predictivity 84.9%.

Technological Characteristics

Software-only module for existing Holter ECG hardware. Utilizes ECG signal processing to derive RR intervals and ECG-Derived Respiratory (EDR) information. Operates on standard Holter scanning system platforms.

Indications for Use

Indicated for adult patients to screen for probability of obstructive sleep apnea, mixed apnea, or hypopnea to determine need for polysomnography. Requires minimum 4 hours sleep duration with predominantly sinus rhythm ECG.

Regulatory Classification

Identification

A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread. JAN 1 9 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Del Mar Reynolds Medical, Incorporated C/O Dr. George Myers Official Correspondent Medsys, Incorporated 377 Route 17S Hasbrouck Heights, New Jersey 07604 Re: K042745 Trade/Device Name: Lifescreen Apnea Regulation Number: 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: MNR Dated: December 13, 2004 Received: December 16, 2004 Dear Dr. Myers: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ ### Page 2 - Dr. Myers Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known): K042745 Device Name: Lifescreen Apnea Indications for Use: The device is intended for use on adult patients only as a screening device to determine the need for clinical diagnosis and evaluation by polysomnography based on the patient's score. The ECG recording may be obtained at any location specified by a physician including home, hospital or clinic. Subjects screened for sleep apnea should have periods of sleep of at least 4 hours duration during which the ECG is predominantly sinus rhythm in nature. × Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Aeen Suliom Page 1 of (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices 510(k) Number: K042745 \ {4}------------------------------------------------ K042745 Lifescreen Apnea Del Mar Reynolds Inc. 510(k) Submission Lifescreen Apnea September 7, 2004 510(k) Summary #### Submitter Information 1. Name: Del Mar Reynolds Medical Inc. Address: Del Mar Reynolds Inc. 13 Whatney Irvine, CA 92618 Telephone Number: 949-699-3300 Fax Number: 949-699-3380 Contact Person: Dr. George Myers (Official Correspondent) Medsys Inc. 377 Route 17 S Hasbrouck Heights, NJ 07604 Telephone 201-727-1703 Fax 201-727-1708 Date Prepared: September 7, 2004 #### 2. Name of Device Trade Name: Lifescreen Apnea Common Name: Apnea Examination System Classification name: Ventilatory Effort Recorder #### 3. Equivalent legally-marketed devices. Silent Night V, K000253, manufactured by Sleep Solutions #### 4. Description Lifescreen Apnea is a software option for the LifeScreen ECG Holter scanning software. The Lifescreen Apnea option is a software addition only - no hardware changes are involved. {5}------------------------------------------------ ## K042745 Lifescreen Apnea ### Page A 8 operation, electrocardiogram (ECG) data is collected by a standard Del Mar Reynolds Holter system and is operation, clock out a graming system with the Lifescreen Apnea option added. By recognizing episodes ment and from thing (SDB), the program indicates the probability that apneic events have occurred or breep Disodered Dreaming onea (AHI) Index. To detect SDB, the algorithm derives both RR Interval and ECG-Derived Respiratory (EDR) information from the ECG. #### Intended Use ડ. "Lifescreen Apnea is indicated for screening adults for the probability that they suffer from obstructive sleep apnea, mixed apnea or hypopnea, in order to evaluate the necessity for a polysomonographic examination. The ECG recording may be obtained at any location specified by a physician including home, hospital or clinic. Subjects screened for sleep apnea should have periods of sleep of at least 4 hours duration during which the ECG is predominantly sinus rhythm in nature." #### 6. Performance Data - Non-clinical tests (a) - 4. Tests with input data from data bases - 5. Validation tests - 6. Tests to check the operation of the algorithms - '५) Clinical tests Lifescreen Apnea has been extensively tested in a clinical trial. The diagnostic criteria have also been established with a separate training group. Diagnostic criteria were established using a separate training group. The results of the tests were: ### Per-minute statistics | Algorithm | | Expert | | |-----------|-----------------------|--------|-------------| | | | N | P | | | n | 9449 | 845 | | | p | 1012 | 5669 | | | | mean | 95% C.I. | | | Accuracy | 89.1 | 88.58-89.52 | | | Sensitivity | 87.0 | 86.19-87.82 | | | Specificity | 90.3 | 89.74-90.88 | | | Positive predictivity | 84.9 | | ### Where: N or n = negative (non-SDB) upper case: expert classification {6}------------------------------------------------ # K042745 Lifescreen Apnea ## Page A 9 P or p = positive (SDB detected) lower case: algorithm classification | Accuracy | $[(Nn + Pp) / Total] * 100 %$ | |-----------------------|-------------------------------| | Sensitivity | $[Pp / (Pn + Pp)] * 100 %$ | | Specificity | $[Nn / (Nn + Np)] * 100 %$ | | Positive Predictivity | $[Pp / (Pp + Np)] * 100 %$ | #### (c) Conclusions Lifescreen Apnea is equivalent in safety and efficacy to the legally marketed predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...