← Product Code [MNR](/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR) · K042253

# SLEEPSENSE SLEEP SENSORS (K042253)

_S.L.P. , Ltd. · MNR · Oct 6, 2004 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR/K042253

## Device Facts

- **Applicant:** S.L.P. , Ltd.
- **Product Code:** [MNR](/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR.md)
- **Decision Date:** Oct 6, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.2375
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Intended Use

SleepSense sensors provide a qualitative measure of a patient's physiological parameters for recording onto an FDA-cleared data acquisition system. Their target population: Children and adult patients who are screened during sleep disorder studies. Their environment of use is usually at a sleep laboratory or sometimes at the patient's home.

## Device Story

SleepSense sensors convert physiological parameters (respiration movement, limb movement, snoring sounds, airflow, body position) into electrical signals; sensing elements include piezo-crystals (mechanical force/vibration), thermocouples (temperature), or gravity switches (position). Sensors are housed in small enclosures, applied to patient, and connected via cable to an FDA-cleared data acquisition system. No electrical contact between sensor and patient. Device performs no monitoring, provides no patient condition indication, and has no alarms/thresholds. Used in sleep labs or home settings. A sleep technician (scorer) reviews recorded signals alongside other channels (EKG, EEG) to diagnose sleep disorders like apnea or insomnia. Device provides qualitative data to assist clinical decision-making.

## Clinical Evidence

No clinical data. Bench testing only; device is a passive sensor component intended for use with FDA-cleared data acquisition systems compliant with IEC 60601-1.

## Technological Characteristics

Passive physiological sensors; sensing elements include piezo-crystals, thermocouples, and gravity switches. Enclosed in patient-friendly housings with flexible cables. No electrical contact with patient. Standalone sensors; no internal processing or monitoring. Compliant with IEC 60601-1 electrical safety standards.

## Regulatory Identification

A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.

## Predicate Devices

- snoring sensor ([K941759](/device/K941759.md))
- moving images(tm) ([K935518](/device/K935518.md))
- the tracker ([K923033](/device/K923033.md))
- easyflow(tm) ([K922112](/device/K922112.md))
- resp-ez(tm) respirat ([K903300](/device/K903300.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 510(K) SUMMARY

# SleepSense Sleep Sensors 510(k) Number K_ O42253

| Applicant's Name: | S.L.P Ltd.             |
|-------------------|------------------------|
|                   | 18 Hazfira Street      |
|                   | Tel-Aviv 67779, Israel |
|                   | Tel: 972-3-537-1281    |
|                   | Fax: 972-3-537-1282    |

| Contact Person: | Noam Hadas             |
|-----------------|------------------------|
|                 | S.L.P Ltd.             |
|                 | 18 Hazfira Street      |
|                 | Tel-Aviv 67779, Israel |
|                 | Tel: 972-3-537-1281    |
|                 | Fax: 972-3-537-1282    |

#### Trade Name:

SleepSense sleep disorder sensors

#### Classification Name:

MNR - Ventilatory effort recorder

#### Classification:

The FDA has classified sensors for sleep disorder sensor device as class II devices. EPM development systems have classified the predicate devices as product code BZQ (monitor, respiration frequency). SLP selected to classify these products as product code MNR (Respiration effort recorder) under the same regulation #. 868.2375. This classification is based on the fact that these products do not monitor, but only output signals which are recorded in a system.

{1}------------------------------------------------

No monitoring takes place at all, since no indication of the patient's condition is presented to any caregiver, there is no need or intention to treat the patient, and there are no alarms or thresholds of any kind.

#### Predicate Device:

| Device               | Applicant               | 510(k)  | Decision Date |
|----------------------|-------------------------|---------|---------------|
| snoring sensor       | EPM development systems | K941759 | 11/10/1994    |
| moving images(tm)    | EPM information systems | K935518 | 05/17/1994    |
| the tracker          | EPM information systems | K923033 | 06/01/1993    |
| easyflow(tm)         | EPM information systems | K922112 | 12/07/1992    |
| resp-ez(tm) respirat | EPM information systems | K903300 | 12/28/1990    |

### Performance Standards:

No performance standards are specified for physiological sensors for sleep disorder testing. The SleepSense sensors are designed, and are specified to be used, only with FDA approved sleep recording systems, which are required to comply with all electrical safety requirements as specified in IEC 60601-1.

### Intended Use:

SleepSense sensors provide a qualitative measure of a patient's physiological parameters for recording onto an FDA-cleared data acquisition system. Their target population: Children and adult patients who are screened during sleep disorder studies. Their environment of use is usually at a sleep laboratory or sometimes at the patient's home.

### Device Description:

Monitoring various physiological parameters is standard practice in sleep disorder testing. Standard overnight recordings show, among others, traces of parameters like respiration movement, leg and arm movement, snoring sounds, respiration airflow and body position during sleep.

In order to record tracings showing these parameters, sensors are needed to convert the physiological parameter into an electrical signal. These sensors are very simple sensing 

{2}------------------------------------------------

elements like piezo-crystals that convert mechanical force or vibrations to an electrical signal. Other sensing elements may be thermocouples which generate a signal proportional to temperature, or gravity switches, that switch and electrical circuit on and off depending on their position.

In practice, these sensing elements are packaged in small, patient-friendly enclosures which are applied to the patient, and connected to the recording system via a long and flexible cable. There is no electrical contact of any kind between the sensors and the patient.

All signals received from the sensors are qualitative, and are only used to record the dynamic nature or existence of the physiological parameter recorded. A specially trained sleep technician called "scorer" reviews the overall recording in the morning following the study. The signals recorded, together with additional channels like EKG or EEG, are analyzed to arrive at a diagnosis of a sleep disorder like sleep apnea or insomnia.

## Substantial Equivalence:

S.L.P Ltd. claims that SleepSense sensors are the same products as the predicate devices, since S.L.P is the designer and OEM manufacturer of the sensors distributed by EPM since 1990. It is therefore obvious that they are substantially equivalent to these predicate devices cited above, without raising new safety and/or effectiveness issues.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three horizontal lines representing its wings and a stylized wave pattern below.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 6 - 2004

S.L.P. Limited C/O Ms. Patricia Murphy Responsible Third Party Official KEMA Quality BV 4377 County Line Road Chalfont, Pennsylvania 18914

Re: K042253

Trade/Device Name: SleepSense Sleep Sensors Regulation Number: 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: MNR Dated: October 4, 2004 Received: October 4, 2004

Dear Ms. Murphy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Murphy

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Chiu Lin, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### INDICATIONS FOR USE STATEMENT

510(k) Number: ____

Device Name: SleepSense sensors for sleep disorder testing

Indications for Use:

SleepSense sleep sensors provide a qualitative measure of some physiological parameters for recording onto an FDA-cleared data acquisition system. Their target population: Children and adult patients who are screened during sleep disorder studies. Their environment of use is usually at a sleep laboratory or sometimes at the patient's home.

Que Subcom

(Division Sign-Off) General Hospital, Division of Anesthesiology, C Infection Control, Dental f 510(k) Number

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

510(k) Number

× Prescription Use

OR

Over the Counter Use

(Per 21 CFR 801.109)

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR/K042253](https://fda.innolitics.com/submissions/AN/subpart-c%E2%80%94monitoring-devices/MNR/K042253)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
