Smart Cable NMT Module and Accessories, Life Scope BSM 3000 Series and Life Scope BSM 6000 Series
K201949 · Nihon Kohden Corporation · KOI · May 2, 2021 · Anesthesiology
Device Facts
Record ID
K201949
Device Name
Smart Cable NMT Module and Accessories, Life Scope BSM 3000 Series and Life Scope BSM 6000 Series
Applicant
Nihon Kohden Corporation
Product Code
KOI · Anesthesiology
Decision Date
May 2, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2775
Device Class
Class 2
Indications for Use
The Smart Cable NMT Module and Accessories are indicated for monitoring the relaxation of the patient when neuromuscular blockades are administered. The Smart Cable NMT Module and Accessories are comprised of: - AF-201P NMT Module with Smart Cable - Disposable Electrodes - Main cable - Holder (optional) The Smart Cable NMT Module and Accessories are intended to be used as a system that requires Nihon Kohden compatible electrodes and bedside monitoring systems. The Smart Cable NMT Module and Accessories are intended for use by medical personnel in clinical settings and are available by prescription only.
Device Story
Smart Cable NMT Module (AF-201P) is an accessory for Nihon Kohden bedside monitors (BSM-3000, BSM-6000, CSM-1901). System uses EMG technology to monitor neuromuscular transmission. Input: electrical stimulation pulses applied to ulnar nerve via disposable electrode array; output: EMG signals representing muscle action potential. Operation: medical personnel apply electrodes to patient hand/wrist; module delivers stimulation and captures EMG response; data transmitted to bedside monitor for display. Provides quantitative measurement of muscle relaxation to guide dosing of muscle relaxants and regional anesthetics. Benefits: objective monitoring of neuromuscular blockade to ensure patient safety during anesthesia and mechanical ventilation.
Clinical Evidence
Bench testing only. No clinical data submitted. Testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2, IEC 60601-2-40), biocompatibility (ISO 10993-1), mechanical strength, environmental, usability, and shelf-life testing. Electrode performance verified per ANSI/AAMI EC12:2000.
Technological Characteristics
EMG-based neuromuscular transmission monitor. Components: AF-201P module, main cable, disposable 5-electrode array (2 stimulating, 2 recording, 1 ground). Stimulation: 10-60 mA constant current, 200/300µs pulse width. Patterns: Single, TOF, PTC, TET, DBS. Connectivity: Proprietary Smart Cable interface to Nihon Kohden bedside monitors. Software: Integrated into host monitor firmware.
Indications for Use
Indicated for monitoring patient relaxation during administration of neuromuscular blockades. For use by medical personnel in clinical settings (OR, recovery room, ICU).
Regulatory Classification
Identification
An electrical peripheral nerve stimulator (neuromuscular blockade monitor) is a device used to apply an electrical current to a patient to test the level of pharmacological effect of anesthetic drugs and gases.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Nihon Kohden Corporation % Sunita Teekasingh Sr. Regulatory Affairs Consultant - GSA2 Group LLC Nihon Kohden America, Inc. 15353 Barranca Parkway Irvine, California 92618
Re: K201949
Trade/Device Name: Smart Cable NMT Module and Accessories, Life Scope BSM 3000 Series Bedside Monitors, Life Scope BSM 6000 Series Bedside Monitors, and Nihon Kohden CSM-1901 Bedside Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX, KOI Dated: March 30, 2021 Received: April 2, 2021
Dear Sunita Teekasingh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation -emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Todd Courtnev Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K201949
### Device Name
Smart Cable NMT Module and Accessories; Life Scope BSM-3000 Series Bedside Monitors; Life Scope BSM-6000 Series Bedside Monitors; and Nihon Kohden CSM-1901 Bedside Monitor
### Indications for Use (Describe)
The Smart Cable NMT Module and Accessories are indicated for monitoring the relaxation of the patient when neuromuscular blockades are administered.
The Smart Cable NMT Module and Accessories are comprised of:
- · AF-201P NMT Module with Smart Cable
- Disposable Electrodes
- Main cable
- · Holder (optional)
The Smart Cable NMT Module and Accessories are intended to be used as a system that requires Nihon Kohden compatible electrodes and bedside monitoring systems. The Smart Cable NMT Module and Accessories are intended for use by medical personnel in clinical settings and are available by prescription only.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
The Life Scope BSM-3000 Series Bedside Monitor is intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to produce a visual record of the electrical signals produced by the heart and monitor the electrocardiogram to generate audible and/or visible alarms when an arrhythmia exists.
The device is also intended to monitor heart rate, blood oxygen saturation (SpO2), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), body temperature,BIS, cardiac output (CO), oxygen concentration (O2), carbon dioxide concentration (CO2), EtCO2, respiratory rate, and inspired and expired anesthetic gases including N20. Halothane. Isoflurane. Enflurane. Sevoflurane, and Desflurane.
The device may generate an audible and/or visual alarm when a measured rate falls outside preset limits. The device may also be used to condition and transmit physiological signals via radio frequency. The device can interface to external equipment to display numerical and waveform data and alarms from the external devices. Supported external devices include CO2 monitors, BIS monitors, Anesthetic agents/gases detection system, Anesthesia machine, Ventilators, CCO monitors, TOF monitors, CCO/SvO2 Monitors, EEG monitoring device, tcPO2/tcPCO2 monitors, rSO2 monitors and external devices which output analog voltage signal.
The device will be available for use by medical personnel on patients within a medical facility on all patient populations.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
The Life Scope BSM-6000 Series Bedside Monitor is intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to produce a visual record of the electrical signals produced by the heart and monitor the electrocardiogram to generate audible and/or visible alarms when an arrhythmia exists.
{4}------------------------------------------------
The device is also intended to monitor heart rate, plood oxygen saturation (SpO2), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), body temperature,BIS, cardiac output (CO), oxygen concentration (FiO2), carbon dioxide concentration (CO2). EtCO2. respiratory rate, and inspired and expired anesthetic agents and anesthetic gases including N20, Halothane, Isoflurane, Enflurane, Sevoflurane, and Desflurane.
The device may generate an audible and/or visual alarm when a measured rate falls outside preset limits. The device may also be used to condition and transmit physiological signals via radio frequency. The device can interface to external equipment to display numerical and waveform data and alarms from the external devices. Supported external devices include CO2 monitors, BIS monitors, Anesthetic agents/gases detection system, Anesthesia machine, Ventilators, CCO monitors, TOF monitors, CCO/SvO2 Monitors, EEG monitoring device, tcPO2/tcPCO2 monitors, rSO2 monitors and external devices which output analog voltage signal.
The device will be available for use by medical personnel on patients within a medical facility on all patient populations.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
The Nihon Kohden CSM-1901 Bedside Monitor is intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. The device is intended to produce a visual record of the electrical signal produced by the heart and monitor the electrocardiogram to generate visible alarms when an arrhythmia exists. The device is also intended to monitor heart rate, blood oxygen saturation (SpO2), noninvasive blood pressure (NIBP), invasive blood pressure (IBP), body temperature, BIS, cardiac output (CO), oxygen concentration (FiO2), carbon dioxide concentration (CO2), EtCO2, respiratory rate, inspired and expired anesthetic agents and anesthetic gases including N20, halothane, isoflurane, enflurane and desflurane. The device also displays patient data from external devices such as ventilators, TOF monitors, and EEG measuring unit.
The device may generate and audible and/or visual alarm when a measured rate falls outside preset limits.
The device will be available for use by trained medical facility on all patient populations. including adult, neonate, infant, child, and adolescent subgroups.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.25 3.375L5.75 10.875L2.75 7.875" stroke="#000000" stroke-linecap="round" stroke-linejoin="round" stroke-width="1.5"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <rect height="12" stroke="#000000" stroke-width="1.5" width="12" x="2" y="2"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
{5}------------------------------------------------
#### 510(k) Summary – General Provisions 5.0
| | Manufacturer Name: | Nihon Kohden Corporation |
|--------------------|---------------------|------------------------------------------------------------------|
| | Address: | 1-31-4 Nishiochiai, Shinjuku-Ku<br>Tokyo 161-8560, Japan |
| General Provisions | Submitter Name: | Nihon Kohden America |
| | Address: | 15353 Barranca Parkway<br>Irvine, California, USA 92618 |
| | Contact: | Sandra Gadeyne<br>Sr. Director of Quality and Regulatory Affairs |
| | Phone Number | 949-356-3401 |
| | Fax Number: | 949-580-1550 |
| | Email | Sandra_Gadeyne@nihonkohden.com |
| | Date of Preparation | 30Apr2021 |
#### 5.1 Smart Cable NMT Module and Accessories
| Subject Device | Trade Name: | Smart Cable NMT Module and Accessories |
|------------------|-------------------------|---------------------------------------------------------------------------------------------------------------------|
| | Marketing Names | NMT Smart Cable TOF Pod; Smart Cable<br>NMT Pod, with EMG Support; NMT Pod<br>TOF Pod; TOF Smart Pod; NMT Smart Pod |
| | Common Name: | Electrical peripheral nerve stimulator |
| | Classification Name | Stimulator, Nerve, Peripheral, Electric |
| | Classification: | Class II |
| | Product Code: | KOI |
| | Regulation Number: | 21 CFR 868.2775 |
| | 510(k) Number | K201949 |
| Predicate Device | Trade Name: | TetraGraph Neuromuscular Transmission<br>Monitor |
| | Classification Name: | Electrical peripheral nerve stimulator |
| | Premarket Notification | Stimulator, Nerve, Peripheral, Electric |
| | Product Code: | KOI |
| | Device Classification: | Class II |
| | Regulation Number: | 21 CFR 868.2775 |
| | Manufacturer: | Senzime AB |
| | 510K number: | K190795 |
| Reference Device | Trade Name: | ToFscan |
| | Classification Name: | Electrical Peripheral Nerve Stimulator |
| | Premarket Notification: | Stimulator, Nerve, Peripheral, Electric |
| | Product Code: | KOI |
| | Device Classification: | Class II |
| | Regulation Number: | 21 CFR 868.2775 |
| | Manufacturer: | Idmed |
| | 510(k): | K172690 |
{6}------------------------------------------------
| Subject Device<br>Description | The Nihon Kohden Smart Cable NMT Module (NMT Module) and<br>Accessories is an optional accessory for the Nihon Kohden bedside<br>monitoring systems. The Smart Cable NMT Module and Accessories TOF<br>(Train of Four) are based on EMG technology. With this system, the user<br>can apply electrical stimulation on the ulnar nerve to detect the muscle's<br>action potential. The reaction to the electrical impulse can be visualized<br>on the connected monitoring system. The Smart Cable NMT Module and<br>Accessories can assist medical personnel to quantitatively determine the<br>level of muscle relaxation. This information can be used to determine the<br>dose of muscle relaxants and regional anesthetics when performing<br>anesthesia in a clinical setting. It is intended for use by medical personnel<br>in the operating room, recovery room, or intensive care unit. |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The NMT Module is a system comprised of NMT Module, Main Cable,<br>Holder, and EMG Electrode. The NMT module is connected to an<br>electrode via Main Cable. The electrode is a single-use electrode array<br>and each array includes two stimulating electrodes, two recording<br>electrodes, and one ground electrode. The NMT module can transmit an<br>electrical stimulation pulse to the patient and can receive EMG signals via<br>the electrode array. The captured data from the disposable electrode is<br>sent to the monitoring system via the Smart Cable interface connector.<br>The various stimulation settings are also sent to the monitoring system to<br>display. |
| | The AF-201P NMT Module is used to control the electrical stimulation<br>and to measure the response. The operational setting is controlled via<br>buttons on the module or a touch screen. |
| Indications for<br>Use and<br>Intended Use | The Smart Cable NMT Module and Accessories are indicated for monitoring<br>the relaxation of the patient when neuromuscular blockades are<br>administered. |
| | The Smart Cable NMT Module and Accessories are comprised of: |
| | • AF-201P NMT Module with Smart Cable<br>• Disposable Electrodes<br>• Main cable<br>• Holder (optional) |
| | The Smart Cable NMT Module and Accessories are intended to be used<br>as a system that requires Nihon Kohden compatible electrodes and<br>bedside monitoring systems. The Smart Cable NMT Module and<br>Accessories are intended for use by medical personnel in clinical settings<br>and are available by prescription only. |
{7}------------------------------------------------
| Characteristics | Subject Device:<br>Smart Cable NMT<br>Module and<br>Accessories | Predicate Device:<br>Tetragraph<br>Neuromuscular<br>Transmission Monitor<br>(K#190795) | Reference Device:<br>ToFscan Neuromuscular<br>Transmission Monitor<br>(K#172690) |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Panel | Anesthesiology | Anesthesiology | Anesthesiology |
| Classification Name | 21 CFR 868.2775<br>Electrical Peripheral<br>Nerve Stimulator | 21 CFR 868.2775<br>Electrical Peripheral Nerve<br>Stimulator | 21 CFR 868.2775<br>Electrical Peripheral Nerve<br>Stimulator |
| Regulatory Class | Class II | Class II | Class II |
| Product Code | KOI | KOI | KOI |
| Indications for Use | The Smart Cable NMT<br>Module and Accessories<br>are indicated for<br>monitoring the relaxation<br>of the patient when<br>neuromuscular<br>blockades are<br>administered.<br><br>The Smart Cable<br>Module and Accessories<br>are comprised of:<br>AF-201P NMT Module<br>with Smart Cable<br>Disposable<br>ElectrodesMain cableHolder (optional)<br>The Smart Cable NMT<br>Module and Accessories<br>are intended to be used<br>as a system that<br>requires Nihon Kohden<br>compatible electrodes<br>and bedside monitoring<br>systems. The Smart<br>Cable NMT Module and<br>Accessories are<br>intended for use by<br>medical personnel in<br>clinical settings and are<br>available by prescription<br>only. | The TetraGraph<br>Neuromuscular<br>Transmission (NMT)<br>Monitor is indicated for<br>monitoring the relaxation<br>of the patient when<br>neuromuscular blockade is<br>administered. | ToFscan is a<br>neuromuscular<br>transmission monitor for<br>monitoring the<br>neuromuscular block of a<br>patient in the operating<br>theatre, recovery room or<br>intensive care unit. The<br>effect of neuromuscular<br>blocking agents (NMBA) is<br>monitored by measuring the<br>acceleration of the muscle<br>movement<br>(acceleromyography) or by<br>visually observing muscle<br>contractions consequent to<br>electrical stimulation. The<br>ToFscan has a three-<br>dimensional acceleration<br>sensor (accelerometer) to<br>detect and quantify a<br>patient's thumb movement<br>(contracting adductor<br>pollicis). The sensor is<br>directly integrated into the<br>finger's splint, making it<br>possible to obtain its optimal<br>and reproducible positioning |
| Device Description | The Smart Cable<br>NMT Module and<br>Accessories are<br>comprised of:<br>• AF-201P NMT<br>Module with<br>Smart Cable<br>• Disposabl<br>e<br>Electrodes<br>• Main cable<br>• Holder (optional)<br><br>The Smart Cable NMT<br>Module and<br>Accessories are<br>intended to be used as<br>a system that requires<br>Nihon Kohden<br>compatible electrodes<br>and bedside monitoring<br>systems. The Smart<br>Cable NMT Module and<br>Accessories are<br>intended for use by<br>medical personnelin<br>clinical settings and are<br>available by<br>prescription only. | The TetraGraph<br>Neuromuscular<br>Transmission (NMT)<br>Monitor (TetraGraph) is a<br>portable, battery-operated<br>EMG-based neuromuscular<br>transmission monitor for<br>use perioperative and in<br>recovery and critical care<br>environments following or<br>during the application of<br>Neuromuscular block.<br><br>Neuromuscular<br>Transmission (NMT) is the<br>transfer of an electrical<br>impulse between a motor<br>nerve and its associated<br>muscle. The NMT is<br>blocked by neuromuscular<br>blocking agents ("NMBAs")<br>which cause transient<br>muscle paralysis preventing<br>the patient from moving and<br>breathing spontaneously.<br><br>Muscle relaxation is used<br>during general anesthesia<br>to enable endotracheal<br>intubation and mechanical<br>ventilation and to provide<br>optimal surgical<br>conditions. Muscle<br>relaxation may also be<br>used in critical care during<br>mechanical ventilation. In<br>these circumstances,<br>TetraGraph can be used<br>as an objective monitor of<br>neuromuscular<br>transmission. TetraGraph<br>undertakes this function by<br>electrical stimulation of the<br>peripheral nerve and<br>directly measuring the<br>evoked response of the<br>muscles (Muscle Action<br>Potential (MAP)), thus<br>providing a quantitative<br>and automatic<br>measurement of muscle<br>response to a stimulus<br>using electromyography<br>(EMG). The TetraGraph is<br>a prescription-only medical<br>device and is indicated for<br>use in hospitals.<br><br>TetraGraph supplements<br>the use of clinical<br>information/data obtained<br>from other monitors, such | The effect of<br>neuromuscular blocking<br>agents (NMBAs) is<br>monitored by measuring<br>the acceleration of the<br>muscle movement<br>(acceleromyography) or<br>by visually observing<br>muscle contractions<br>consequent to electrical<br>stimulation. The ToFscan<br>has a three-dimensional<br>acceleration sensor<br>(accelerometer) to detect<br>and qualify a patient's<br>thumb movement<br>(contracting adductor<br>pollicis). The sensor is<br>directly integrated into<br>the finger's splint, making<br>it possible to obtain it's<br>optimal and reproducible<br>positioning. |
{8}------------------------------------------------
{9}------------------------------------------------
| | saturation (SpO2), end-<br>tidal carbon dioxide<br>(ETCO2), as well as<br>clinical assessment, to<br>determine the adequacy of | |
|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | ventilation.<br>The level of<br>neuromuscular block is | |
| | routinely measured by<br>stimulating a peripheral<br>nerve, usually in the<br>forearm and by evaluating | |
| | the muscle response<br>typically in the thumb or<br>little finger. The<br>TetraGraph controls the | |
| | level of electrical<br>stimulation applied to the<br>nerve and monitors the<br>muscle response by the | |
| | use of Electromyography<br>(EMG) detected by<br>electrodes on the muscle. | |
| | TetraGraph consists of the<br>followingmain components | |
| | TetraGraph Monitor<br>The TetraGraph Monitor is<br>used to control the electrical<br>stimulation and to measure<br>the EMG- response. The<br>Monitor is controlled via a | |
| | touch screen and a power<br>button. The Senzime AB<br>Traditional 510(k)<br>Premarket Submission<br>TetraGraph Page 005- | |
| | 3TetraGraph Monitor is<br>connected to the electrode<br>via a cable (the TetraCord<br>Cable). The battery in the<br>TetraGraph Monitor is | |
| | charged via a<br>communication port<br>connected to a USB- supply<br>adapter. | |
| | TetraSens Electrode<br>The TetraSens Electrode is<br>a single-use electrode<br>array. Each array includes | |
| | two stimulating electrodes<br>(applied along the ulnar<br>nerve at the wrist) and two<br>recording electrodes | |
| | (applied on the hand). The<br>TetraGraph Monitor can<br>transmit stimulation pulses<br>to the patient and can | |
| | receive EMG signals via the<br>electrode array. The<br>electrodes are neither<br>supplied sterile nor<br>intended to be sterilized by | |
{10}------------------------------------------------
| Electrode for<br>Stimulation | Single-use<br>electrode array (5<br>electrodes on an | Single-use electrode array<br>(4 electrodes on an array) | Uses 3D Accelerometer<br>Sensor to measure thumb<br>movement instead of |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrode Connection | Reusable Cable | Reusable Cable | Reusable Cable |
| movement<br>detection<br>technology | (EMG) | | |
| Muscle | Electromyography | from the Tetragraph kit.<br>Rechargeable lithium-<br>polymer battery<br>A re-chargeable lithium-<br>polymer battery is included<br>in the TetraGraph, and also<br>available as spare part. The<br>rechargeable battery is<br>charged using the USB<br>power supply adapter.<br>USB power supply adapter<br>and USB cable<br>The rechargeable battery is<br>charged using the USB<br>power supply adapter. The<br>adapter is configured for<br>local power outlets and<br>connects to the TetraGraph<br>using a USB cable. Data<br>can also be transferred to a<br>PC via the USB-port by<br>using the<br>USB-cable.<br>Electromyography (EMG) | Acceleromyography |
| | | TetraGraph Pole Clamp kit<br>A Pole clamp kit is available<br>for mounting the<br>TetraGraph on a pole<br>stand. The kit includes a<br>mounting device and an<br>attachment to the<br>Tetragraph. The Pole clamp<br>kitis supplied separately | |
| | | TetraCord Cable<br>The TetraCord Cable is<br>connected to the<br>TetraGraph Monitor via a<br>portand is connected to the<br>TetraSens Electrode<br>via the cable connector at<br>the other end of the<br>TetraCord Cable. The<br>TetraCord Cable is supplied<br>in together with the<br>TetraGraph monitor as part<br>of the kit and can also be<br>sold separately as a spare<br>part. | |
| | | the user. The TetraSens<br>Electrodes are sold<br>separately from the<br>TetraGraph monitor and are<br>available in boxes of 20<br>pcs. | |
| | array) | | electrode |
| Electrode applied part | Total of 5:<br>Two stimulating<br>electrodes are applied<br>along the ulnar nerve at<br>the wrist and two<br>recording electrodes are<br>applied on the hand.<br>One grounding electrode<br>applied on the hand | Total of 4:<br>Two stimulating electrodes<br>are applied along the ulnar<br>nerve at the wrist and two<br>recording electrodes are<br>applied on the hand. | Not Applicable (Uses D<br>Accelerometer Sensor to<br>measure thumb<br>movement instead of the<br>electrode) |
| Stimulation Patterns | Single stimulation,<br>TOF (Train Of Four),<br>PTC (Post Tetanic<br>Count), TET (tetanic<br>stimulation), DBS<br>(Double Burst<br>Stimulation) | Single Twitch (ST),<br>Train-of-Four (TOF),<br>Post-tetanic Count<br>(PTC) | TOF (Train Of Four)<br>- PTC (Post<br>Tetanic Count)<br>- ATP (Automated<br>TOF PTC)<br>- DBS (Double<br>Burst) (3,3) (3,2)<br>- ST (Single Twitch)<br>0.1 Hz and 1 Hz<br>- TET (Tetanus) 50 Hz |
| Stimulating<br>Current Range | 10-60 mA | 10-60 mA | 0-60 mA |
| Stimulation Pulse Width | Square wave, constant<br>current: 200µs or<br>300µs | Square wave, constant<br>current: 200µs or 300µs | Constant current, 0-<br>60 mA, monophasic<br>200µsec<br>pulse width |
| Power source | Power is supplied from<br>the Monitoring<br>System. | Battery operated | AC Power Supply |
| Compatibility | Works only with<br>Nihon Kohden<br>Monitoring System:<br>BSM-6000, BSM-<br>3000 CSM-1901 | Tetragraph Neuromuscular<br>Transmission Monitor<br>consists of a monitor,<br>electrode, and other<br>components. | The ToFscan system is<br>composed of the monitor,<br>the hand sensor, and the<br>US AC power supply unit. |
| General Safety | IEC 60601-1, IEC-<br>60601-1-2, IEC 60601-2-<br>40 | IEC 60601-1, IEC-60601-<br>1-2, IEC 60601-2-40 | IEC 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-2-10 |
| Intended Use | There are differences in the indications for use statement, but the<br>differences do not result in new intended use.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.