Tetragraph Neuromuscular Transmission Monitor

K190795 · Senzime AB · KOI · Oct 18, 2019 · Anesthesiology

Device Facts

Record IDK190795
Device NameTetragraph Neuromuscular Transmission Monitor
ApplicantSenzime AB
Product CodeKOI · Anesthesiology
Decision DateOct 18, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.2775
Device ClassClass 2

Indications for Use

The TetraGraph Neuromuscular Transmission (NMT) Monitor is indicated for monitoring the relaxation of the patient when neuromuscular blockade is administered.

Device Story

Portable, battery-operated EMG-based monitor; used perioperatively, in recovery, and critical care. Inputs: electrical stimulation pulses applied to peripheral nerves (e.g., ulnar nerve) via single-use electrode array. Device measures evoked Muscle Action Potential (MAP) response via EMG electrodes on hand. Outputs: quantitative, automatic measurement of muscle response to stimulus. Operated by clinicians in hospitals. Supplements clinical data (SpO2, ETCO2) to assess neuromuscular block adequacy, aiding in management of mechanical ventilation and surgical conditions. Benefits: provides objective, quantitative monitoring of muscle relaxation to guide NMBA administration.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing, including software performance, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2, IEC 60601-2-40), biocompatibility (ISO 10993-1), mechanical strength, and electrode performance (ANSI/AAMI EC12:2000).

Technological Characteristics

Portable EMG-based monitor; rechargeable lithium-polymer battery; touch screen interface. Single-use electrode array (4 electrodes). Stimulation: 10-60 mA constant current, 200/300 μs pulse width. Connectivity: USB port for charging and data transfer to PC. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-40, ISO 10993-1, ANSI/AAMI EC12:2000.

Indications for Use

Indicated for monitoring patient muscle relaxation during neuromuscular blockade administration in hospital settings; prescription use only.

Regulatory Classification

Identification

An electrical peripheral nerve stimulator (neuromuscular blockade monitor) is a device used to apply an electrical current to a patient to test the level of pharmacological effect of anesthetic drugs and gases.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 18, 2019 Senzime AB % Elisa Maldonado-Holmertz RA/OA Consultant Obleix Consulting LLC 12416 Fairfax Ridge Place Austin, Texas 78738 Re: K190795 Trade/Device Name: Tetragraph Neuromuscular Transmission Monitor Regulation Number: 21 CFR 868.2775 Regulation Name: Electrical Peripheral Nerve Stimulator Regulatory Class: Class II Product Code: KOI Dated: September 11, 2019 Received: September 12, 2019 Dear Elisa Maldonado-Holmertz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {2}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K190795 Device Name TetraGraph Neuromuscular Transmission Monitor Indications for Use (Describe) The TetraGraph Neuromuscular Transmission (NMT) Monitor is indicated for monitoring the relaxation of the patient when neuromuscular blockade is administered. Type of Use (Select one or both, as applicable) | <div> <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |-----------------------------------------------------------------------------------------------------------------| | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for SENZIME. The logo consists of a blue circle with a white waveform inside, followed by the word "SENZIME" in large, blue, sans-serif letters. In the upper right corner of the image, the text "Page 1 of 7" is visible. # Section 005 - 510(k) Summary # 1. Submission Sponsor Senzime AB Ulls väg 29B 756 51 Uppsala Sweden Johanna Faris, QA/RA Director Email: johanna.faris@senzime.com Tel number: +46 (0)705286330 # 2. Submission Correspondent Obelix Consulting, LLC 12416 Fairfax Ridge Place Austin, TX 78738 USA Elisa Maldonado-Holmertz, RA/QA Consultant Email: elisamh@obelixconsult.com Cellphone number: +1 512 431 6069 ## 3. Date Prepared 15 October 2019 ## 4. Device Identification | Type of 510(k) Submission: | Traditional | |----------------------------|---------------------------------------------------------------------| | Trade or Proprietary Name: | TetraGraph Neuromuscular Transmission (NMT)<br>Monitor (TetraGraph) | | Common or Usual Name: | TetraGraph | | Product Code: | KOI | | Regulation Nr: | 868.2775 | | Regulation Name: | Electrical peripheral nerve stimulator | | Class of Device: | Class II | | Panel: | Anesthesiology | | Reason for Submission: | New device | | Multiple Devices: | No; this is the only device in the submission | {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for SENZIME. The logo consists of a blue circle with a white sine wave inside, followed by the word "SENZIME" in large, blue, sans-serif letters. The text "Page 2 of 7" is located in the upper right corner of the image. #### 5. Legally Marketed Predicate and Reference Devices Predicate Device- K051635, Datex Ohmeda S/5 NeuroMuscular Transmission Module, E-NMT (from GE Healthcare) Reference Device - K992598, TOF-Watch SX (from Merck & Co, Inc. (Organon Teknika Corp.)) #### 6. Device Description The TetraGraph Neuromuscular Transmission (NMT) Monitor (TetraGraph) is a portable, battery-operated EMG-based neuromuscular transmission monitor for use perioperative and in recovery and critical care environments following or during the application of Neuromuscular block. Neuromuscular Transmission (NMT) is the transfer of an electrical impulse between a motor nerve and its associated muscle. The NMT is blocked by neuromuscular blocking agents ("NMBAs") which cause transient muscle paralysis preventing the patient from moving and breathing spontaneously. Muscle relaxation is used during general anesthesia to enable endotracheal intubation and mechanical ventilation and to provide optimal surgical conditions. Muscle relaxation may also be used in critical care during mechanical ventilation. In these circumstances, TetraGraph can be used as an objective monitor of neuromuscular transmission. TetraGraph undertakes this function by electrical stimulation of the peripheral nerve and directly measuring the evoked response of the muscles (Muscle Action Potential (MAP)), thus providing a quantitative and automatic measurement of muscle response to a stimulus using electromyography (EMG). The TetraGraph is a prescription-only medical device and is indicated for use in hospitals. TetraGraph supplements the use of clinical information/data obtained from other monitors, such as peripheral oxygen saturation (SpO2), end-tidal carbon dioxide (ETCO2), as well as clinical assessment, to determine the adequacy of ventilation. The level of neuromuscular block is routinely measured by stimulating a peripheral nerve, usually in the forearm and by evaluating the muscle response typically in the thumb or little finger. The TetraGraph controls the level of electrical stimulation applied to the nerve and monitors the muscle response by the use of Electromyography (EMG) detected by electrodes on the muscle. TetraGraph consists of the following main components: #### TetraGraph Monitor The TetraGraph Monitor is used to control the electrical stimulation and to measure the EMG-response. The Monitor is controlled via a touch screen and a power button. The {6}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the Senzime logo in blue. The logo consists of a stylized waveform graphic to the left of the company name, "SENZIME," which is written in a bold, sans-serif font. Above the logo, the text "Page 3 of 7" is visible, indicating that the image is likely a page from a larger document. TetraGraph Monitor is connected to the electrode via a cable (the TetraCord Cable). The battery in the TetraGraph Monitor is charged via a communication port connected to a USBsupply adapter. # TetraSens Electrode The TetraSens Electrode is a single-use electrode array includes two stimulating electrodes (applied along the ulnar nerve at the wrist) and two recording electrodes (applied on the hand). The TetraGraph Monitor can transmit stimulation pulses to the patient and can receive EMG signals via the electrode array. The electrodes are neither supplied sterile nor intended to be sterilized by the user. The TetraSens Electrodes are sold separately from the TetraGraph monitor and are available in boxes of 20 pcs. ## TetraCord Cable The TetraCord Cable is connected to the TetraGraph Monitor via a port and is connected to the TetraSens Electrode via the cable connector at the other end of the TetraCord Cable. The TetraCord Cable is supplied in together with the TetraGraph monitor as part of the kit and can also be sold separately as a spare part. ## TetraGraph Pole Clamp kit A Pole clamp kit is available for mounting the TetraGraph on a pole stand. The kit includes a mounting device and an attachment to the TetraGraph. The Pole clamp kit is supplied separately from the TetraGraph kit. ## Rechargeable lithium-polymer battery A re-chargeable lithium-polymer battery is included in the TetraGraph, and also available as spare part. The rechargeable battery is charged using the USB power supply adapter. ## USB power supply adapter and USB cable The rechargeable battery is charged using the USB power supply adapter is configured for local power outlets and connects to the TetraGraph using a USB cable. Data can also be transferred to a PC via the USB-port by using the USB-cable. # 7. Indication for Use Statement Indications for Use: The TetraGraph Neuromuscular Transmission (NMT) Monitor is indicated for monitoring the relaxation of the patient when neuromuscular blockade is administered. {7}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for SENZIME. The logo consists of a blue circle with a white sine wave inside, followed by the word "SENZIME" in large, bold, blue letters. In the upper right corner of the image, the text "Page 4 of 7" is visible. #### 8. Substantial Equivalence Discussion A device comparison is made for the TetraGraph compared to the predicate device Datex-Ohmeda S/5 NeuroMuscular Transmission Module, E-NMT from GE Healthcare ("Datex-Ohmeda"). Datex-Ohmeda's existing 510(k) clearance (K051635) means it is legally marketed. ## Intended use and indications for use, substantial equivalence discussion The intended use for the predicate device (Datex-Ohmeda S/5 NeuroMuscular Transmission Module, E-NMT) is the same since the Regulation 868.2775 applies for both: An electrical peripheral nerve stimulator (neuromuscular blockade monitor) is a device used to apply an electrical current to a patient to test the level of pharmacological effect of anesthetic drugs and gases. | | TetraGraph Neuromuscular Transmission<br>(NMT) Monitor | Datex-Ohmeda S/5TNI E-NMT module | |---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for use | The TetraGraph Neuromuscular Transmission<br>(NMT) Monitor is indicated for monitoring the<br>relaxation of the patient when neuromuscular<br>blockade is administered. | The Datex-Ohmeda S/5TNI E-NMT module<br>is indicated for monitoring the relaxation of<br>the patient and regional block stimulation<br>for nerve location. The device is indicated<br>for use by qualified medical personnel only. | Both devices are indicated for monitoring the muscle relaxation of the patient, which is TetraGraph's sole indication. Datex-Ohmeda also has a second indication: nerve location. However, TetraGraph has only one of the predicate device's two indications meaning that the subject device is compared to its predicate device with respect to their mutual indication only. The predicate device states that it is indicated for use by qualified medical professionals only in its indications. However, TetraGraph is a prescription device, and thus, it is also limited to use by medical professionals. TetraGraph's indications for use statement and its device's labeling identify it as a prescription device. Thus, both devices have the same population of intended users. Senzime believes that TetraGraph's Rx designation in its indications for use statement and on its labeling conveys that limitation and that it would be redundant to identify its intended user population in the device's indications. To summarize, the indications for use statement for the TetraGraph and its predicate device are not identical, but their mutual indication and common user population mean that they have the same intended use. {8}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for SENZIME. The logo consists of a blue circle with a white sine wave inside, followed by the word "SENZIME" in large, bold, blue letters. The text "Page 5 of 7" is located in the upper right corner of the image. #### Technological Characteristics Both TetraGraph and the predicate device measure muscle response using electromyography (EMG) and have reusable cables for the electrode connection. The electrode used for the stimulation are single-use. There are minor technological differences in number of electrodes and stimulation settings. Performance testing show that TetraGraph met all the acceptance criteria for this device, and which are the same or very similar to the predicate device's specifications. Thus, the test data demonstrate that TetraGraph is at least as safe and effective in monitoring the relaxation of the patient when neuromuscular blockade is administered as Datex-Ohmeda. ## Substantial Equivalence Table: The following Comparison of Characteristics Table compares the TetraGraph to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence. {9}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the logo for Senzime, a medical technology company. The logo features a stylized waveform graphic to the left of the company name, which is written in a bold, sans-serif font. In the upper right corner of the image, the text "Page 6 of 7" is visible, indicating that this image is part of a larger document. The logo and page number are the primary elements in the image. Senzime AB Traditional 510(k) Premarket Submission TetraGraph | Table – Comparison of Characteristics | | | | |-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Trade Name | TetraGraph<br>Neuromuscular<br>Transmission Monitor | Datex-Ohmeda S/5<br>NeuroMuscular<br>Transmission Module, E-NMT | TOF-Watch SX | | Device | SUBJECT | PREDICATE (K051635) | REFERENCE (K992598) | | Product Code | KOI | KOI | KOI | | Device Class | Class II | Class II | Class II | | Classification name | Electrical peripheral nerve<br>stimulator | Electrical peripheral<br>nerve stimulator | Electrical peripheral nerve<br>stimulator | | Regulation Number | 868.2775 | 868.2775 | 868.2775 | | Classification Panel | Anesthesiology | Anesthesiology | Anesthesiology | | Intended Use | The TetraGraph<br>Neuromuscular (NMT)<br>Monitor is an electrical<br>peripheral nerve<br>stimulator<br>(neuromuscular blockade<br>monitor) that is used to<br>apply an electrical current<br>to a patient to test the<br>level of pharmacological<br>effect of anesthetic drugs<br>and gases. | Datex-Ohmeda is an<br>electrical peripheral nerve<br>stimulator<br>(neuromuscular blockade<br>monitor) that is used to<br>apply an electrical current<br>to a patient to test the<br>level of pharmacological<br>effect of anesthetic drugs<br>and gases. | TOF-Watch is an electrical<br>peripheral nerve stimulator<br>(neuromuscular blockade<br>monitor) that is used to<br>apply an electrical current<br>to a patient to test the level<br>of pharmacological effect of<br>anesthetic drugs and gases. | | Indications for Use | The TetraGraph<br>Neuromuscular<br>Transmission (NMT)<br>Monitor is indicated for<br>monitoring the relaxation<br>of the patient when<br>neuromuscular blockade<br>is administered. | The Datex-Ohmeda<br>S/5TNI E-NMT module is<br>indicated for monitoring<br>the relaxation of the<br>patient and regional block<br>stimulation for nerve<br>location. The device is<br>indicated for use by<br>qualified medical<br>personnel only. | The TOF-Watch is an<br>instrument for monitoring<br>the neuromuscular<br>transmission during surgery<br>or in the intensive care unit<br>by means of<br>acceleromyography.<br>The TOF-Watch is only to be<br>operated by trained medical<br>personnel. | | Muscle movement<br>detection technology | Electromyography (EMG) | Electromyography EMG<br>Mechanomyography<br>(MMG) | Acceleromyography (AMG) | | Electrode Connection | Reusable Cable | Reusable Cable | Reusable Cable | | Electrode for<br>Stimulation | Single use electrode<br>array (4 electrodes on an<br>array) | Single use standard ECG<br>electrodes (5 electrodes) | Single use standard ECG<br>electrodes (3 electrodes) | | Stimulation Patterns | Single Twitch (ST), Train-<br>of-Four (TOF), Post-<br>tetanic Count (PTC) | ST, TOF, PTC, DBS | TOF, PTC, DBS, 1Hz, 0,1 Hz,<br>TET | | Stimulation Current<br>Range | 10-60 mA | 10 - 70 mA | 0 - 60 mA | | Stimulation Pulse Width | Square wave, constant<br>current: 200 μs or 300 μs | Square wave, constant<br>current: 100, 200 or 300<br>μs | 200 μs or 300 μs | #### Comparison of Characteristics Tahle {10}------------------------------------------------ Image /page/9/Picture/1 description: The image shows the logo for SENZIME. The logo consists of a blue icon on the left and the word "SENZIME" in blue, sans-serif font on the right. Above the logo, the text "Page 7 of 7" is visible. ## 9. Performance Testing Performance testing has been made for the TetraGraph which supports substantial equivalence between the TetraGraph and the predicate device. The performance testing made for the device is summarized below: - Software performance testing - Electrical safety (in accordance with IEC 60601-1) - Electromagnetic compatibility (in accordance with IEC 60601-1-2 and IEC 60601-2-40) - Biocompatibility testing (in accordance with ISO10993-1) - Thermal testing - Mechanical strength - Environmental testing - Usability testing - Shelf life testing ● - . Battery life testing - Performance testing of electrodes; including tensile testing as well as electrical ● performance (as specified in the ANSI/AAMI EC12:2000 standard) The performance testing show that TetraGraph met all of the acceptance criteria for this device, and which are the same or very similar to the predicate device's specifications. Thus, the test data demonstrate that TetraGraph is at least as safe and effective in monitoring the relaxation of the patient when neuromuscular blockade is administered as the predicate #### 10. Clinical Performance Data There was no clinical testing required to demonstrate that TetraGraph is substantial equivalent to the Datex-Ohmeda Device, as the predicate has similar technological characteristics. #### 11. Statement of Substantial Equivalence In summary, TetraGraph, compared to its predicate device, has the same intended use, one of the same indications for use, very similar technological characteristics, where the technological differences do not raise different questions of safety or effectiveness. Test data demonstrates that TetraGraph is at least as safe and effective as its predicate device for the same indications. Thus, TetraGraph is substantially equivalent to its predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...